A retrospective study was conducted to evaluate the effect of high-frequency mechanical vibration on root resorption (measured by tooth length) in patients undergoing orthodontic treatment with clear aligners.
Effect of the Application of High-Frequency Mechanical Vibration on Tooth Concurrent with Orthodontic Treatment Using Clear Aligners: A Retrospective Study; Retrospective, controlled, multicenter; Follow-up/Duration: 12 months or completion of treatment, whichever was shorter
Patients with Class I malocclusion, good oral hygiene, complete permanent dentition (except third molars), and initial anterior crowding of 3-5 mm.; Sample Size: 30; Number of Sites: Multicenter
Control (no vibration device or sham)
Root resorption measured by tooth length
Indications for Use
The VPro5 is intended for use by the orthodonic patient during treatment with aligners to facilitate minor anterior tooth movement.
Device Story
VPro5 is a rechargeable, hand-held, powered, high-frequency vibration device used by orthodontic patients at home. It consists of an oscillator and a C-shaped mouthpiece. The patient manually applies the mouthpiece to the occlusal surface of the teeth for a single 5-minute session daily. The device ramps up to a 120Hz vibration frequency and automatically shuts off after 5 minutes. LED indicators show cycle progression. Usage data can be retrieved via USB connection to a computer. By applying high-frequency mechanical vibration, the device facilitates minor anterior tooth movement during aligner treatment. Clinical studies indicate faster tooth movement rates and improved tracking compared to control groups, with no significant increase in root resorption.
Clinical Evidence
Clinical evidence includes a prospective randomized controlled trial (n=63) and a retrospective study (n=30). The prospective trial compared VPro5 to control/sham groups; results showed statistically significant faster tooth movement rates and improved tracking in VPro5 arms. The retrospective study evaluated root resorption, finding no statistically significant change in tooth length for the VPro5 group compared to a control group. Bench testing confirmed electrical safety (IEC 60601-1 series), biocompatibility (ISO 10993), and mechanical performance.
Technological Characteristics
Rechargeable, hand-held, powered vibration device. Components: C-shaped mouthpiece (patient-contacting), oscillator, USB charging cable. Vibration frequency: 120Hz. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1, 60601-1-2, and 60601-1-11. Not supplied sterile; intended for cleaning with water. Software: moderate level of concern.
Indications for Use
Indicated for orthodontic patients undergoing treatment with aligners to facilitate minor anterior tooth movement.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
November 1, 2019
Propel Orthodontics LLC Bryce Way President and CEO 394 South Abbott Avenue Milpitas, California, 95035
Re: K183018
Trade/Device Name: VPro5 Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: OYH Dated: October 14, 2019 Received: October 15, 2019
Dear Bryce Way:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183018
Device Name
VPro5
Indications for Use (Describe)
The VPro5 is intended for use by the orthodonic patient during treatment with aligners to facilitate minor anterior tooth movement.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K183018
# 510(k) Summary
| Submitter Name: | Propel Orthodontics LLC |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 394 South Abbott Ave Milpitas, CA 95035 |
| Contact Person: | Bryce Way |
| Phone Number: | 408-394-5851 |
| Fax Number: | 914-560-2056 |
| Date Prepared: | October 31, 2019 |
| Device Trade Name: | VPro5® |
| Classification Name: | Orthodontic Vibratory Accessory |
| Product Code: | OYH |
| Classification: | Class II pursuant to 21 C.F.R. § 872.5470 |
| Predicate Device: | AcceleDent® Aura (K153048) |
| Device Description: | The VPro5® is a rechargeable, hand-held, powered, high-frequency vibration device. The device consists of the following components: the VPro5 Oscillator, the VPro5 Mouthpiece, a charging cable, and a wall charging adapter. The oscillator and mouthpiece are coupled together prior to use. |
| | During use, the C-shaped mouth piece adapter is manually applied to the occlusal surface of the teeth. The VPro5® is to be used 5 minutes per day in a single session in conjunction with aligner orthodontic treatment. |
| | The mouthpiece is connected to the oscillator and turned on by depressing a white on-switch located on the unit. When turned on the device will slowly ramp up to full vibration frequency of 120Hz. The vibrations are applied to the occlusal surface of the teeth. The device will automatically shut off after a duration of 5 minutes. The oscillator also contains LED light notifications which indicate status of the cycle progression. |
| | The mouthpiece is to be cleaned and stored in a dry place. The USB port can also be connected to a computer to display usage data. |
| Indications for Use: | The VPro5® is intended for use by the orthodontic patient during treatment with aligners to facilitate minor anterior tooth movement. |
{4}------------------------------------------------
## Performance Data:
Non-clinical and clinical performance testing completed with the VPro5® has demonstrated that the device performs safely and as intended.
## Non-Clinical Performance Testing:
The VPro5® was tested to ensure conformance with the following electrical safety and electromagnetic compatibility (EMC) standards:
- IEC 60601-1 Medical Electrical Equipment Part 1 General Requirements for Basic . Safety and Essential Performance
- IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2 General Requirements for . Basic Safety and Essential Performance
- . IEC 60601-1-11 (2nd Edition): 2015 Medical Electrical Equipment -- Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
The VPro5® is patient-contacting. The only patient-contacting component in the VPro5® is the C-shaped mouthpiece. A biocompatibility risk assessment was conducted for all components of the device per the FDA Guidance Document for Use of ISO 10993-1. Results of biocompatibility testing demonstrate that the VPro5® is biocompatible for use as intended. Cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and intracutaneous irritation (ISO 10993-5) biocompatibility testing was conducted on the final finished device.
Software documentation of moderate level of concern was provided per the FDA Guidance Document for the "Content of Premarket Submissions for Software Contained in Medical Devices."
Bench testing conducted on 40 units of the VPro5® demonstrated that all the devices tested met criteria by being within the product specifications including evaluation of the ramp-up, vibratory frequency, acceleration, and time functionality of the device.
A two-year shelf life validation study was conducted to evaluate the performance and integrity of real time 2-year shelf-life of the mouthpiece and vibration device including through mechanical wear and exposure to artificial saliva to simulate the usage environment in the patient mouth.
{5}------------------------------------------------
#### Clinical Performance Testing:
Two clinical studies were performed using the VPro5®, one prospective, pivotal clinical trial and one retrospective study examining the effect of high-frequency vibration on root resorption. The two clinical studies are summarized in Tables 1A and 1B, belov:
### Table 1A.
## Effect of VPro5 Therapy on Clear Aligner Tracking: A Prospective Randomized Controlled Clinical Trial
| Study Design | Randomized, controlled multicenter, prospective | |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Arms | Control: Subjects received clear aligners only and progressed to the next aligner every 14 days 7-Day Sham: Subjects received a sham device and progressed to the next aligner every 7 days 7-Day VPro5: Subjects received the VPro5 and progressed to the next aligner every 7 days 5-Day Sham: Subjects received a sham device and progressed to the next aligner every 5 days 5-Day VPro5: Subjects received a VPro5 and progressed to the next aligner every 5 days | |
| Inclusion<br>Criteria | Subjects must be 18-45 years of age Subject is willing and able to comply with all study procedures and sign informed consent/HIPAA forms Subject must have complete adult dentition (excluding third molars) Subjects must have class I malocclusion or mild class II/III malocclusions Subject is at least one month into aligner treatment Subject has history of and currently healthy oral hygiene (PD is <4 mm, GI<1, PI=1) | |
| Exclusion<br>Criteria | Subjects who have received periodontal treatment in the previous 6 months Subjects who are taking medication that could affect the level of inflammation, such as chronic antibiotics, phenytoin, cyclosporin, anti-inflammatory drugs, systemic corticosteroids, or calcium channel blockers Subjects with severe class II or class III malocclusion Subjects with skeletal class I but extreme dental malocclusion Severe crowding that requires extraction Subjects with more than 4 mm positive overjet or more than 2 mm negative overjet Subjects with extreme deep bite (more than 90%) Subjects with severe open bite (more than 2 mm) Pregnant women Subjects with any systemic diseases affecting bone metabolism Smoking | |
{6}------------------------------------------------
| | Subjects with active, untreated caries Subjects that require interproximal reduction or attachments during the study period |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Number of<br>Subjects | 63 |
| Mean Age<br>(years) | 29 |
| Primary<br>Outcome<br>Measure | Rate of tooth movement (mm/week) Total tooth movement (percent tracking) |
| Study Duration | 20-56 days (depending on study arm) |
| Results | Subjects in the 7-Day VPro5 and 5-Day VPro5 arms achieved a statistically significant faster rate of tooth movement compared to<br>subjects in the Control arm. In addition, subjects in the 5-Day VPro5 arm achieved statistically significantly greater tracking (i.e.,<br>percentage of prescribed study tooth movement achieved) as compared to subjects in the 7-Day Sham arm. Both VPro5 arms had<br>tracking comparable to the Control arm. (The 5-Day Sham arm was discontinued due to inadequate tracking and subject discomfort.)<br>No serious device related adverse events were observed. No differences were seen in VPro5 performance between male and female<br>subjects or based upon age. Appropriate statistical adjustments were made for the use of multiple comparisons. |
# Table 1B.
# Effect of the Application of High-Frequency Mechanical Vibration on Tooth Concurrent with Orthodontic Treatment Using Clear Aligners: A Retrospective Study
| Study Design | Retrospective, controlled, multicenter |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Arms | VPro5 Control (no vibration device or sham) |
| Inclusion Criteria | Class I malocclusion Good oral hygiene Complete permanent dentition except for third molars Initial anterior crowding ranging from 3 to 5 mm |
| Exclusion Criteria | External apical root shortening observed at the preoperative radiographic examination Proposed treatment included extraction of premolars and dental stripping |
| Number of Subjects | 30 |
| Mean Age (years) | 26 |
| Primary Outcome Measure | Root resorption measured by tooth length |
{7}------------------------------------------------
| Study Duration | 12 months or completion of treatment, whichever was shorter |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Results | The control group showed a statistically significant decrease in tooth lengths compared to the VPro5 group, which showed change<br>of tooth lengths that was not statistically significant. |
These studies demonstrate that the VPro5® is substantially equivalent to the predicate device for use by the orthodontic patient during treatment with aligners to facilitate minor anterior tooth movement.
{8}------------------------------------------------
# Substantial Equivalence:
The VPro5® is substantially equivalent to AcceleDent® Aura, cleared under K153048.
Table 2 below provides a comparison of the characteristics of the proposed device and legally marketed predicate device.
| Specification | VPro5® | AcceleDent® Aura | Comparison |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (TBD) | (K153048) | |
| Intended Use | Facilitation of minor<br>anterior tooth movement<br>during orthodontic<br>treatment with aligners. | Facilitation of minor<br>anterior tooth movement<br>during orthodontic<br>treatment with brackets<br>and wires or aligners. | Similar. The predicate<br>device is intended for use<br>with brackets and wires<br>or aligners, and the<br>proposed device is<br>intended for use only<br>with aligners. |
| Indications for Use | The VPro5® is intended<br>for use by the orthodontic<br>patient during treatment<br>with aligners to facilitate<br>minor anterior tooth<br>movement. | AcceleDent® Aura is<br>intended for use during<br>orthodontic treatment. It<br>is used in conjunction<br>with brackets and wires<br>or aligners and helps<br>facilitate minor anterior<br>tooth movement. | Similar. The predicate<br>device is intended for use<br>with brackets and wires<br>or aligners, and the<br>proposed device is<br>intended for use only<br>with aligners. |
| Classification Regulation | 21 C.F.R. § 872.5470 | 21 C.F.R. § 872.5470 | Same. |
| Product Code | OYH | OYH | Same. |
| Device Class | II | II | Same. |
| Duration of Use | 5 minutes per day during<br>orthodontic treatment | 20 minutes per day<br>during orthodontic<br>treatment | Different. The proposed<br>device has a shorter<br>duration of use than the<br>predicate device. |
| Technological<br>Characteristics | Intraoral applianceIntegrated,<br>rechargeable batteryVibration mechanism<br>of action | Intraoral applianceIntegrated,<br>rechargeable batteryVibration<br>mechanism of action | Same. |
| Vibration Frequency | High frequency (120 Hz) | Low frequency (30 Hz) | Different. The proposed<br>device operates at a<br>higher vibration<br>frequency than the<br>predicate device. The<br>substantial equivalence<br>of high frequency<br>vibration for facilitating<br>tooth movement is<br>supported by the clinical<br>data included in this<br>submission. |
| Specification | VPro5®<br>(TBD) | AcceleDent® Aura<br>(K153048) | Comparison |
| Components | Mouthpiece Battery-powered oscillator Charging cable (USB) Wall charging adapter | Mouthpiece Battery-powered oscillator Charging cable (USB) Wall charging adapter | Same. |
| Power Source | Rechargeable battery | Rechargeable battery | Same. |
| Sterilization | Not supplied sterile Intended to be rinsed with water after use | Not supplied sterile Intended to be rinsed with water after use | Same. |
| Rx Only? | Yes | Yes | Same. |
Table 2. Comparison of Proposed VPro5® Device and Predicate Device
{9}------------------------------------------------
The VPro5® has similar intended use and indications for use as the previously cleared AcceleDent® Aura, use by the orthodontic patient during treatment to facilitate minor anterior tooth movement. The AcceleDent® Aura is intended for use with aligners or brackets and wires, and the VPro5® is intended for use with aligners only. In addition, the VPro5® has similar technological characteristics and principles of operation as its predicate. Although there are technological differences between the VPro5® and its predicate device, these differences do not raise different questions of safety or efficacy as confirmed by the company's testing and validation activities described in this submission.
As demonstrated by the clinical testing, summarized above, the VPro5® is substantially equivalent for facilitating minor anterior tooth movement as the predicate device. The pivotal clinical study demonstrates that the VPro5® is substantially equivalent to the predicate and that the VPro5® does not raise different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.