K172765 · Great Lakes Orthodontics, Ltd. · NXC · Jun 1, 2018 · Dental
Device Facts
Record ID
K172765
Device Name
Smart Moves Complete
Applicant
Great Lakes Orthodontics, Ltd.
Product Code
NXC · Dental
Decision Date
Jun 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Smart Moves Complete is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e. all second molars). Smart Moves Complete positions teeth by way of continuous gentle force.
Device Story
System consists of series of custom-fabricated, clear, thermoformed plastic aligners; intended as alternative to traditional brackets/wires. Dentist/orthodontist takes PVS impressions or digital scans; data sent to lab. Software generates 3D model of final treated state and intermediate stages. Prescribing physician reviews/approves treatment plan via software. Lab fabricates aligners using thermoforming. Patient wears aligners sequentially (2 weeks per stage), removing only for eating/brushing. Continuous gentle force moves teeth to programmed positions. Device is prescription-only; used in dental clinic/home settings. Benefits include virtually invisible appliance and avoidance of fixed braces.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-1), software verification/validation, and mechanical property analysis (tensile strength, elongation, stress, modulus, load, water absorption, and durometer).
Technological Characteristics
Material: 0.03" Thermoplastic PETG. Sensing/Actuation: Passive mechanical force via thermoformed plastic. Form factor: Patient-specific removable aligners. Connectivity: Electronic submission of digital scans/prescriptions. Software: Dental tooth alignment design software. Biocompatibility: ISO 10993-1 compliant. Hardness: 80 ± 3 Shore D.
Indications for Use
Indicated for treatment of tooth malocclusion in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Great Lakes Orthodontics Ltd Mr. David Graver Director of Logistics 200 Cooper Ave Tonawanda, New York 14150
#### Re: K172765
Trade/Device Name: Smart Moves Complete Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: April 30, 2018 Received: May 2, 2018
Dear David Graver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
June1, 2018
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172765
Device Name Smart Moves Complete
Indications for Use (Describe)
Smart Moves Complete is indicated for the treatment of tooth malocclusion in patients with permanent dentition (1.e. all second molars). Smart Moves Complete positions teeth by way of continuous gentle force.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### Smart Moves Complete
#### K172675
## Sponsor
Great Lakes Orthodontics Ltd
200 Cooper Ave.
Tonawanda, NY 14150
Contact person:
Dave Graver
Director of Logistics
## Date Prepared: May 31, 2018
# Device Identification
Trade/Proprietary name: Smart Moves Complete Regulation Number: 21CFR 872.5470 Regulation Name: Aligner Sequential/ Orthodontic Plastic Bracket
Panel: Dental
Product Class: II
Product Code: NXC
#### Legally Marketed Predicate Device
K113618, ClearCorrect System, ClearCorrect Inc.
{4}------------------------------------------------
## Indications for Use
Smart Moves Complete is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e. all second molars). Smart Moves Complete positions teeth by way of continuous gentle force.
#### Background
Orthodontic treatment is conventionally performed using brackets and archwires. A bracket is bonded to each tooth and a wire is inserted with the goal of moving each tooth to a desired final treatment position. An alternative to traditional bracket treatment has been in existence for the past 20 years. This alternative treatment uses clear plastic aligners to gradually move the teeth. The aligners are designed in a series of incremental stages. A patient wears each aligner stage for a period of two weeks to accomplish the programmed tooth movement. Once tooth movement has been accomplished for a given stage the current aligner is removed and the next aligner in the sequence is worn. This method of orthodontic treatment is popular with patients because it provides a virtually invisible appliance and avoids the appearance of braces on the teeth.
#### Design and Use
- This is a prescription use device by order of a Dentist or Orthodontist ●
- Designed for single patient use (2 weeks/aligner)
- Aligners are worn except for eating and brushing teeth. They are removed by the patient for those tasks.
- It is not provide sterile
- It does not contain any drugs or biologic
- Software as part of the system is used by the Dental lab to design the aligners that the ● Dental professional reviews and approves prior to fabrication.
- No clinical studies are included in the submission
## Device Description
## Aligner System
A dental health professional (e.g. orthodontist or dentist) prescribes Smart Moves Complete based on an assessment of the patient's teeth, takes molds of the patient's teeth, determines a course of treatment to reposition the teeth via gentle, corrective forces in the upper and lower dental arch, and completes a prescription form. Digital information concerning the patient's teeth and the prescription form are then sent electronically to Great Lakes Orthodontics.
{5}------------------------------------------------
Great Lakes then designs a series of plastic thermoformed aligners in a sequential series intended to gradually realign the patient's teeth in accordance with the physician's prescription using standard dental tooth alignment software. The prescribing physician accesses the doctor-facing portion of the standard dental tooth alignment software to review and approve the model scheme before the thermoformed aligners are produced. Once approved, Great Lakes Orthodontics produces these aligners, which are formed of clear, thin, thermoformed plastic.
The aligners in a sequential series are sent back to the dental health care professional who then provides them to the patient, confirming fit and design. Over a period of time, each progression of aligners are provided sequentially to the patient by the dental health professional to gradually move the target teeth to the designed position. The dental care professional monitors treatment from the moment the first aligner is delivered to when the final aligner is delivered. The aligners are held in place by pressure and can be removed by the patient at any time.
| Manufacturer | Great Lakes<br>Orthodontics, Ltd. | ClearCorrect Inc. | Device Comparison |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Smart Moves Complete | ClearCorrect System | N/A |
| 510(k) Number | K172765 | K113618 | N/A |
| Product Code | NXC | NXC | Same |
| Regulation Number | 872.5470 | 872.5470 | Same |
| Regulation Name | Orthodontic Plastic<br>Bracket | Orthodontic Plastic<br>Bracket | Same |
| Indications for Use<br>Statement | Smart Moves Complete is<br>indicated for the<br>treatment of tooth<br>malocclusion in patients<br>with permanent<br>dentition (i.e. all second<br>molars). Smart Moves<br>Complete positions teeth<br>by way of continuous<br>gentle force. | The ClearCorrect System<br>is indicated for the<br>treatment of tooth<br>malocclusion in patients<br>with permanent<br>dentition (i.e. all second<br>molars). The<br>ClearCorrect System<br>positions teeth by way of<br>continuous gentle force. | Same |
| Device Description | Smart Moves Complete<br>consists of a series of<br>doctor-prescribed, thin,<br>clear plastic removable<br>orthodontic appliances<br>(aligners). The aligners<br>gently move the patient's<br>teeth in small increments<br>from their original state<br>to a more optimal,<br>treated state. Smart<br>Moves Complete is<br>intended as an<br>alternative to | The ClearCorrect device<br>is fabricated of clear thin<br>thermoformed<br>polyurethane plastic in a<br>sequential series to<br>progressively reposition<br>the teeth. Corrective<br>force to straighten the<br>teeth is delivered via<br>minor changes into a<br>position in each<br>subsequent aligner. | Same; the device<br>performs treatment of<br>tooth malocclusions in<br>the same way by<br>continuous gentle force. |
| | | | |
| | conventional bracket<br>technology and fixed<br>appliances for the<br>treatment of patients<br>with malocclusion. | | |
| Mode of Action | Orthodontic tooth<br>movement occurs<br>through forces applied by<br>the appliance to the<br>dentition as each tooth<br>follows the programmed<br>displacement based on a<br>doctor's prescription. | Orthodontic tooth<br>movement occurs<br>through forces applied by<br>the appliance to the<br>dentition as each tooth<br>follows the programmed<br>displacement based on a<br>doctor's prescription. | Same |
| Anatomy Location | Mouth; mucosal<br>membranes | Mouth; mucosal<br>membranes | Same |
| Size | Patient specific | Patient specific | Same |
| Manufacturing Method | Thermoforming | Thermoforming | Same |
| Material | 0.03" Thermoplastic<br>PETG | 0.03" Thermoplastic<br>Polyurethane | Similar; these are both<br>thermoplastic forming<br>materials that have same<br>durometer, similar<br>properties in modulus<br>that do not raise any<br>additional questions of<br>safety or efficacy. |
| Biocompatibility | Passed ISO 10993-1 and<br>series | Passed ISO 10993-1 and<br>series | Same |
| Hardness Durometer | $80 \pm 3$ Shore D | $80 \pm 3$ Shore D | Same |
| Material Tensile<br>Strength/Stress | Strength and stress of<br>material tensile<br>produced acceptable<br>results for base material | Information not available | Base materials of<br>thermoplastic used to<br>fabricate aligners uses<br>similar material used in<br>these applications gentle<br>tooth alignment. The<br>materials are<br>thermoformed plastics<br>that are similar in<br>composition to the<br>predicate device not<br>raising any additional<br>questions of safety or<br>efficacy. |
| Material Elongation | Elongation at yield and<br>break produced<br>acceptable results for<br>base material | Information not available | Base materials of<br>thermoplastic used to<br>fabricate aligners uses<br>similar material used in<br>these applications gentle<br>tooth alignment. The<br>materials are<br>thermoformed plastics<br>that are similar in<br>composition to the<br>predicate device not<br>raising any additional |
| | | | |
| Material Modulus | Modulus results<br>produced similar results<br>for testing of base<br>material | Information not available | Base materials of<br>thermoplastic used to<br>fabricate aligners uses<br>similar material used in<br>these applications gentle<br>tooth alignment. The<br>materials are<br>thermoformed plastics<br>that are similar in<br>composition to the<br>predicate device not<br>raising any additional<br>questions of safety or<br>efficacy. |
| Material Load | Load at yield and break<br>produced acceptable<br>results for base material | Information not available | Base materials of<br>thermoplastic used to<br>fabricate aligners uses<br>similar material used in<br>these applications gentle<br>tooth alignment. The<br>materials are<br>thermoformed plastics<br>that are similar in<br>composition to the<br>predicate device not<br>raising any additional<br>questions of safety or<br>efficacy. |
| Dental Tooth Alignment<br>Software Description | The Smart Moves<br>Complete 3-D software<br>uses a scan of a PVS<br>impression or a digital<br>scan to generate the<br>image of a final, treated<br>state and then interprets<br>a series of images that<br>represent intermediate<br>teeth states. Once the<br>dental practitioner<br>approves the treatment<br>plan, the software<br>converts the files to<br>produce the series of<br>patient specific aligners. | The standard dental<br>tooth alignment software<br>uses a scan of tooth<br>impression or a digital<br>scan to generate the<br>image of a final, treated<br>state and then interprets<br>a series of images that<br>represent intermediate<br>teeth states. Once the<br>dental practitioner<br>approves the treatment<br>plan, the software<br>converts the files to<br>produce the series of<br>patient specific aligners. | Same |
| Healthcare Professional<br>Review | The dental practitioner<br>reviews final aligner<br>series prior to fabrication<br>and has the option to<br>reject or request<br>modifications to the set-<br>up prior to approving it<br>for aligner fabrication. | The dental practitioner<br>reviews final aligner<br>series prior to fabrication<br>and has the option to<br>reject or request<br>modifications to the set-<br>up prior to approving it<br>for aligner fabrication. | Same |
#### Predicate Device Comparison
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### Non-Clinical Performance Testing
As part of demonstrating safety and effectiveness of Smart Moves Complete and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Great Lakes Orthodontics completed a number of non-clinical performance tests. The Smart Moves Complete meets all the requirements for overall design, biocompatibility, and performance results confirming that the design output meets the design inputs and specifications for the device.
The Smart Moves Complete passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
- Biocompatibility testing per ISO 10993-1 passed cytotoxicity, sensitization, and irritation .
- Software verification and validation testing using FDA's guidance document confirmed . acceptance to required specifications
- There is no defined specification or standard for tensile strength so a sample of base . materials used for fabrication of aligners were tested and analyzed producing acceptable results for tensile strength yield
- . There is no defined specification or standard for aligner elongation which a sample of based materials used for fabrication of aligners were tested and analyzed showing acceptable results in comparison of break point
- . There is no defined specification or standard for tensile stress so a sample of base materials used for fabrication of aligners were tested and analyzed producing acceptable results for tensile stress at break point
- . There is no defined specification or standard for modulus so this was tested for the base materials used for fabrication of aligners to show acceptable results between the samples tested
- There is no defined specification or standard for load of materials which found acceptable ● results when tested and analyzed for samples of base material tested used for fabrication of aligners
- There is no defined specification or standard for water absorption which found acceptable ● results when tested and analyzed for fabricated aligners
- Durometer testing of thermoplastic forming material 80 ± 3 Shore D: Acceptable .
{9}------------------------------------------------
- Process Flow validation was performed to ensure that the finished device matches the software output specifications. The output and work model and aligner were tested and compared. Aligners met the specifications of this testing.
#### Clinical Performance Data
There was no clinical testing performed to support the medical device as the indications for use is equivalent to the predicate device. The substantial equivalence is supported by non-clinical testing and the comparative table, so this is not required.
#### Substantial Equivalence Conclusion
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Based on the comparison and analysis above, the Smart Moves Complete is determined to be substantially equivalent to the referenced predicate device(s).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.