K214008 · Shenzhen Soga Technology Co., Ltd. · NVK · May 10, 2022 · General, Plastic Surgery
Device Facts
Record ID
K214008
Device Name
Dental diode laser
Applicant
Shenzhen Soga Technology Co., Ltd.
Product Code
NVK · General, Plastic Surgery
Decision Date
May 10, 2022
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue; marginal and interdental gingival and epithelial lining of free gingiva; frenotomy; biopsy; operculectomy; Implant recovery; gingivectomy; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of deseased epithelial lining ; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty; excision of lesions ; exposure of unerupted/ partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth inability); pulpotomy; pulpotomy as adjunct to root canal therapy; Fibroma removal; Gingival incision; Treatment of canker sores; herpetic ulcers of the oral mucosa: Laser soft tissue curettage: Reduction of gingival hypertrophy:.
Device Story
Dental diode laser (ILaser I) uses semiconductor laser diode to generate laser energy; energy transmitted via optical fiber tip to treatment site. Operated by dental surgeons in hospitals, clinics, or medical offices. Device includes mainframe, power system, laser drive, optical path, LCD screen, and emergency stop. Laser radiation absorbed by soft tissue; converted to heat for cutting, coagulation, and vaporization. Used for intra/extra-oral soft tissue procedures; periodontal pocket debridement; gingival treatments. Provides hemostasis and tissue management; benefits include precise soft tissue removal and improved clinical indices (e.g., probe depth, gingival index).
Clinical Evidence
No clinical data. Bench testing only. Safety and effectiveness verified through electrical, electromagnetic, and laser performance standards (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22).
Technological Characteristics
Class IV semiconductor diode laser; 980nm ±20nm wavelength; 4.0W CW / 8.0W peak power; 1-100Hz frequency. Enclosure: ABS; Handpiece cover: Silica gel. Fiber optic tip delivery. Powered by internal supply. Dimensions: 18cm x 16cm x 26cm. Connectivity: MCU-based control. Sterilization: Fiber optic tips autoclavable at 121°C for 30 min.
Indications for Use
Indicated for soft tissue oral surgery, including incision, excision, hemostasis, coagulation, vaporization, and periodontal procedures (e.g., sulcular debridement, curettage) in all patient populations requiring dental soft tissue management.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
May 10, 2022
Shenzhen Soga Technology Co., Ltd. Tse Adrian Quality Manager D906, Yinxing Technology Building No. 1301, Sightseeing Road Xinlan Community, Guanlan Street, Longhua District Shenzhen, Guangdong 518110 CHINA
Re: K214008
Trade/Device Name: Dental diode laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: NVK, GEX Dated: March 11, 2022 Received: March 11, 2022
Dear Tse Adrian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K214008
Device Name Dental Diode Laser
#### Indications for Use (Describe)
Intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue; marginal and interdental gingival and epithelial lining of free gingiva; frenotomy; biopsy; operculectomy; Implant recovery; gingivectomy; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of deseased epithelial lining ; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty; excision of lesions ; exposure of unerupted/ partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth inability); pulpotomy; pulpotomy as adjunct to root canal therapy; Fibroma removal; Gingival incision; Treatment of canker sores; herpetic ulcers of the oral mucosa: Laser soft tissue curettage: Reduction of gingival hypertrophy:.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| <div><span style="font-size:10pt"></span> <span style="font-size:10pt"></span></div> <div><span style="font-size:10pt"></span></div> | <span style="font-size:10pt"><label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label></span> |
| | <span style="font-size:10pt"><label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label></span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for SOGA. The logo is in a dark gray color, with the letter A having a green triangle in the middle. Below the logo is the company name in Chinese.
## 005_510(k) Summary
### 5.1 Submitter Information
- Company: Shenzhen Soga technology Co., Ltd.
- Address: D906, Yinxing Technology Building No. 1301, Sightseeing Road, Xinlan Community, Guanlan Street, Longhua District, Shenzhen, Guangdong, China.
- Phone: +086-15915873605
- Contact: Tse Adrian, Quality Manager
- Mail box: 15915873605@Soga12.com ●
- Website: www.soga12.com
### 5.2 Device Information
- Trade/Device Name: Dental diode laser ●
- Model: ILaser I
- Common Name: Dental diode laser
- Classification regulation: Regulation number: 21 CFR 878.4810 Requlation Description: Manual surgical instrument for general use. Regulation Medical Specialty: General & Plastic Surgery
- Review Panel: General & Plastic Surgery
- Product Code: Primary product code: NVK; Secondary product code: GEX ●
- Regulation Number: 21 CFR 878.4810 ●
- Device Class: Class II
### 5.3 Predicate Device Information
Predicate Device: SIROLaser Advance (Model: FonaLaser)
Manufacturer: Sirona Dental Systems GmbH
510(k) number: K103753
Indication of use:
Intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue; marginal and interdental
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for SOGA. The logo is in a dark gray color, with the letter A in green. Below the logo is the company name in Chinese characters.
gingival and epithelial lining of free gingiva; frenectomy; frenotomy; biopsy; operculectomy; Implant recovery; gingivectomy; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of deseased epithelial lining ; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty; excision of lesions ; exposure of unerupted/partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; Laser removal of diseased, infected, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth inability); pulpotomy; pulpotomy as adjunct to root canal therapy; Fibroma removal; Gingival incision and excision; Treatment of canker sores; herpetic ulcers of the oral mucosa; Laser soft tissue curettage; Reduction of gingival hypertrophy;.
### 5.4 Device Description
Principle: The semiconductor laser diode is excited by the generation power supply to generate the laser, which is effectively transmitted to the treatment site through the optical fiber tip.
Intended operator: Dental surgeon.
Component part: It is composed of laser host (including internal power system, laser drive system, optical path system, LCD screen, emergency stop button, laser) and optic fiber tip.
Sterile: The device mainframe not supplied sterile and do not require sterilization prior to use. The consumable component (Fiber optic tip) not supplied sterile but require sterilization prior to use.
Sterilization condition: 121℃ (250°F), 30min
### 5.5 Indications for Use
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for SOGA. The logo is in a dark gray color, with the letter A having a green triangle in the middle. Below the logo is the company name in Chinese, which is 索感科技.
Intra- and extra-oral surgery including incision, excision, hemostasis, coagulation and vaporization of soft tissue; marginal and interdental gingival and epithelial lining of free gingiva; frenectomy; frenotomy; biopsy; operculectomy; Implant recovery; gingivectomy; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of deseased epithelial lining ; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty; excision of lesions ; exposure of unerupted/partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; Laser removal of diseased, infected, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth inability); pulpotomy; pulpotomy as adjunct to root canal therapy; Fibroma removal; Gingival incision and excision; Treatment of canker sores; herpetic ulcers of the oral mucosa; Laser soft tissue curettage; Reduction of gingival hypertrophy;.
| Comparison<br>Items | Subject Device:<br>SOGALaser<br>(Model: ILaser I) | Predicate Device:<br>SIROLaser Advance<br>Model: FONALaser<br>(K103753) |
|-------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Classification & Intended Use | | |
| Classification | GEX, NVK<br>Class II<br>21 CFR 878.4810 | GEX<br>Class II<br>21 CFR 878.4810 |
| Intended use | Intra- and extra-oral<br>surgery including incision,<br>excision, hemostasis, | Intra- and extra-oral<br>surgery including incision,<br>excision, hemostasis, |
## 5.6 Comparison of Technological Characteristics with the Predicate Device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the text "Shenzhen Soga technology Co. Ltd." in a clear, sans-serif font. The text is horizontally aligned and appears to be the name of a company. The background is plain white, providing a clean and professional look to the text.
Image /page/6/Picture/1 description: The image shows the logo for SOGA, a technology company. The logo is in a dark gray color, with the letter 'A' in the logo having a green triangle in the middle. Below the logo, there are Chinese characters, and a black line is present at the bottom of the image.
coagulation and vaporization of soft tissue; marginal and interdental gingival and epithelial lining of free gingiva; frenectomy; frenotomy; biopsy; operculectomy; Implant recovery; gingivectomy; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of deseased epithelial lining ; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty; excision of lesions ; exposure of unerupted/partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; Laser removal of diseased, infected, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of diseased,
coagulation and vaporization of soft tissue; marginal and interdental gingival and epithelial lining of free gingiva; frenectomy; frenotomy; biopsy; operculectomy; Implant recovery; gingivectomy; gingivoplasty; gingival troughing; crown lengthening; hemostasis of donor site; removal of granulation tissue; laser assisted flap surgery; debridement of deseased epithelial lining ; incisions and draining of abscesses; tissue retraction for impressions; papillectomy; vestibuloplasty; excision of lesions ; exposure of unerupted/partially erupted teeth; removal of hyperplastic tissues; treatment of aphthous ulcers; leukoplakia; Laser removal of diseased, infected, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the text "Shenzhen Soga technology Co. Ltd." in bold, black font. The text appears to be a company name. The background is plain white, providing a clear contrast for the text to stand out. There is a black line underneath the text.
Image /page/7/Picture/1 description: The image shows the logo for SOGA. The logo is in a dark gray color, with the letters "SOGA" in a stylized font. Below the letters is the company name in Chinese characters, which translates to "Suo Gan Technology". The "A" in SOGA has a green triangle in the middle.
| | infected, inflamed and | diseased, infected, |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | necrosed soft tissue in the<br>periodontal pocket to<br>improve clinical indices<br>including gingival index,<br>gingival bleeding index,<br>probe depth, attachment<br>loss and tooth inability);<br>pulpotomy; pulpotomy as<br>adjunct to root canal<br>therapy; Fibroma removal;<br>Gingival incision and<br>excision; Treatment of<br>canker sores; herpetic<br>ulcers of the oral mucosa;<br>Laser soft tissue curettage;<br>Reduction of gingival<br>hypertrophy;. | inflamed and necrosed<br>soft tissue in the<br>periodontal pocket to<br>improve clinical indices<br>including gingival index,<br>gingival bleeding index,<br>probe depth, attachment<br>loss and tooth inability);<br>pulpotomy; pulpotomy as<br>adjunct to root canal<br>therapy; Fibroma removal;<br>Gingival incision and<br>excision; Treatment of<br>canker sores; herpetic<br>ulcers of the oral mucosa;<br>Laser soft tissue<br>curettage; Reduction of<br>gingival hypertrophy;. |
| Patient Population | For use in all patients | For use in all patients |
| Environment of<br>Use | For Dental surgeon | Prescription Use - For<br>professional |
| Patient Population | No restriction on the<br>applicable population | All Population |
| Comparison<br>Statement | The subject device enjoys the same classification and<br>intended use with the predicate device. | |
| Technological Characteristics | | |
| Principle / Method<br>of Operation | The semiconductor laser<br>diode is excited by the<br>generation power supply to<br>generate the laser, which is<br>effectively transmitted to | It is generated by a laser<br>diode in the control unit<br>and is guided through a<br>quartz fiber to the<br>treatment area. The laser |
| | the treatment site through<br>the fiber optic tip. | radiation is absorbed by<br>the tissue and converted<br>into heat for cutting,<br>solidification, bacteria<br>reduction and<br>desensitization. |
| Environment<br>of<br>Use | Hospital, clinic, and medical<br>office setting | Hospital, clinic, and<br>medical office setting |
| Size | 18cm x 16cm x 26cm | 182 x 197 x 189 mm |
| Shipping list | The product is composed<br>of:<br>● Mainframe<br>● Fiber optic tip<br>● Charging adaptor<br>● Fiber optic cleaner<br>● Fiber bender<br>● Laser area symbol<br>● Protection glasses | ● FONALaser mainframe<br>including A control<br>unit including a<br>handpiece with a<br>manual switch<br>● Handpiece cover<br>● Fiber head pack<br>(including 1 x 200 μm<br>and 4 x 320 μm)<br>● Fiber cutter<br>● Disposable operation<br>head (including a<br>bending tool)<br>● Protection glasses<br>● Switch power supply |
| Software | MCU software | MCU software |
| Laser Class | IV (4) | IV (4) |
| Wavelength | 980nm ±20nm | 970 nm ± 15nm |
| Frequency | 1Hz to 100Hz | 1Hz to 100Hz |
| Output Power | 4.0 W Max CW / 8.0 W | 4.0 W Max CW / 7.0 W |
| | Peak Power (Pulse Mode) | Peak Power (Pulse Mode) |
| Pulse mode | Continuous | Continuous |
| Aiming Beam | Laser diode, 5 mW,<br>680nm±20nm, Class 1 | Laser diode, max 1 mW,<br>635-650nm, Class 1 |
| N.O.H.D. | 3.03 meters | 1.5 meters |
| Comparison<br>statement | The subject device has almost the same technological<br>characteristics with the predicate device, and their slight<br>difference in Size, Aiming Beam, Wavelength and<br>N.O.H.D. will not affect the core usage of the devices or<br>validated by relevant standard evaluation or not<br>bringing new safety and effectiveness concerns, thus<br>will not affecting the substantial equivalence<br>comparison. their differences are not affecting the core<br>usage of the devices<br>The difference in Wavelength had verified the safety<br>characteristics by IEC 60825-1 and IEC 60601-2-22, and<br>the efficiency of subject device is better than predicate<br>device. It could be found in literature referred.[1] | |
| Material Characteristics | | |
| Application part | Fiber optic tip | Fiber |
| Enclosure | ABS | Not know |
| Hand piece Cover | Silica gel | elasto-plastic |
| Safety & Effectiveness | | |
| Patient-Contacting<br>Materials | Fiber optic tip | Fiber |
| Electrical Safety | Verified according IEC<br>60601-1 | Not know |
| EMC | Verified according IEC<br>60601-1-2 | Not know |
| | | |
| Performance Safety | Verified according IEC 60825-1 and IEC 60601-2-22 | Not know |
| Comparison statement | The safety and essential effectiveness of the subject device have been evaluated according to the FDA recognized standards. | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for SOGA. The logo is in dark gray, with the letter A in the logo having a green triangle in the middle. Below the logo is the company name in Chinese characters. There is a horizontal line above and below the Chinese characters.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for SOGA. The logo is in a dark gray color, with the letter A having a green triangle in the middle. Below the logo is the company name in Chinese characters.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for SOGA. The logo is in a dark gray color, with the letter A having a green triangle in the middle. Underneath the logo is the company name in Chinese characters, with a line on either side of the name.
## 5.6.1 Comparison summary
First, the subject device (Model: ILaser I) enjoys identical classification and intended use with the predicate device, which forms the foundation of their substantial equivalence.
Secondly, the most technological characteristics have substantial equivalence difference is below:
| Difference item |
|--------------------------|
| Shipping list |
| Size |
| Wavelength |
| Aiming Beam |
| Frequency |
| Output power |
| N.O.H.D. |
| Enclosure material |
| Handpiece cover material |
## 5.7 Discussion of Tests Performed
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for SOGA. The logo is in a dark gray color, with the letter A in the logo having a green triangle in the middle. Below the logo is the company name in Chinese.
# 5.7.1 Clinical Test
Clinical testing was not performed for SOGALaser (Model: ILaser I) as part of the submission.
5.7.2 Non-Clinical Tests
The subject device was tested/analyzed according to the following standards in order to ensure its effectiveness and safety:
| | Standard<br>Designation<br>Number | Title of Standard |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety | IEC 60601-1 | Medical electrical<br>equipment Part 1:<br>General requirements<br>for basic safety and<br>essential performance |
| Electromagnetic<br>Compatibility | IEC 60601-1-2 | Medical Electrical<br>Equipment - Part 1-2:<br>General Requirements<br>For Basic Safety And<br>Essential Performance<br>– Collateral Standard:<br>Electromagnetic<br>Compatibility -<br>Requirements And<br>Tests |
| Performance<br>Safety and<br>Effectiveness | IEC 60825-1 | Safety of laser<br>products - Part 1:<br>Equipment<br>classification, and<br>requirements |
| | IEC 60601-2-22 | Medical electrical<br>equipment - Part 2-22: |
| | Particular<br>requirements for basic<br>safety and essential<br>performance of<br>surgical, cosmetic,<br>therapeutic and<br>diagnostic laser<br>equipment | |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for SOGA. The logo is in a dark gray color, except for a small green triangle in the upper right corner of the A. Below the logo is the company name in Chinese characters.
# 5.8 Conclusion
From the above analysis, it is proper to conclude that the subject device (Model: ILaser I) will be as safe and effective for usage as the listed predicate devices that have already been on the U.S. market.
5.9 Release date
Apr. 11, 2022
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.