K142233 · Wieland Dental + Technik GmbH & Co. KG · EIH · Jan 5, 2015 · Dental
Device Facts
Record ID
K142233
Device Name
zenostar MO, Zenostar T, Zenostar VisualiZr
Applicant
Wieland Dental + Technik GmbH & Co. KG
Product Code
EIH · Dental
Decision Date
Jan 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Intended Use
Zenostar MO (Medium Opacity) are milling discs intended for the fabrication of crown frameworks and multi-unit bridge frameworks for use as dental restorations. Zenostar T (Translucent) are milling blanks intended for making single-tooth and multiunit restorations for use as dental restorations. Depending on the indication, frameworks or monolithic (full contour) restorations can also be made. Zenostar VisualiZr is for the temporary dyeing of Zenostar Color Zr solutions according to the Instructions for Use.
Device Story
Zenostar MO and T are pre-sintered zirconia discs used as raw material for dental prosthetics; processed via CAD/CAM milling systems; sintered at high temperatures to achieve final density and physical properties. Zenostar VisualiZr is an accessory liquid used to temporarily dye coloring solutions. Used in dental laboratories by technicians to fabricate crowns, onlays, and bridges. MO grade is typically used for frameworks requiring layering ceramics; T grade is used for monolithic restorations due to higher translucency. Final restorations are placed by dentists to restore tooth function and aesthetics.
Clinical Evidence
Bench testing only. Testing performed in accordance with ISO 6872:2008 for flexural strength, chemical solubility, coefficient of thermal expansion, and radioactivity. Results demonstrate equivalence to predicate devices. No clinical data submitted.
Technological Characteristics
Pre-sintered zirconia milling discs (98.5mm diameter, various thicknesses). Material properties tested per ISO 6872:2008. Designed for use with wet CAM milling systems. VisualiZr is a liquid additive that burns out during sintering.
Indications for Use
Indicated for patients requiring full ceramic dental restorations, including crowns, onlays, and bridges (anterior and molar). Zenostar VisualiZr is indicated for temporary dyeing of Zenostar Color Zr solutions.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a way that they appear to be connected or overlapping.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 5,2015
Wieland Dental + Technik Gmbh & Co. Kg c/o Ms. Donna Marie Hartnett Director OA/Regulatory Affairs Ivoclar Vivadent, Inc. 175 Pineview Drive Amherst, New York 14228
Re: K142233
> Trade/Device Name: Zenostar Mo, Zenostar T, Zenostar VisualiZr Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: II Product Code: EIH Dated: November 26, 2014 Received: December 5, 2014
Dear Ms. Hartnett,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runner DDS, MA
Erin Keith Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): K142233
Device Name: Zenostar MO, Zenostar T, Zenostar VisualiZr
Indications For Use:
Zenostar MO and Zenostar T consist of machinable zirconia discs for the preparation of full ceramic crowns, onlays and bridges (anterior and molar).
Zenostar VisualiZr is for the temporary dyeing of Zenostar Color Zr solutions according to the Instructions for Use.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
{3}------------------------------------------------
## 510(K) SUMMARY
Image /page/3/Picture/1 description: The image shows the logo for Wieland Dental. The text "WIELAND" is in large, bold, blue letters. Below that, in smaller blue letters, is the word "DENTAL" with a trademark symbol next to it.
A MEMBER OF THE IVOCLAR VIVADENT GROUP
Contact: Donna Marie Hartnett
Company: lvoclar Vivadent. AG Bendererstrasse 2, Schaan, FL-9494, Liechtenstein +423-235-3535
January 2, 2015 Date Prepared:
Proprietary Name:
## Zenostar MO. Zenostar T. Zenostar VisualiZr
Classification Name: Powder, Porcelain (872.6660) (Classification Code EIH)
IPS e.max ZirCAD (K051705) by Ivoclar Vivadent, AG Predicate Device: (Liechtenstein) and Zenotec Zr Bridge, Zenostar Zr Translucent, Zenotec Color Zr, Zenostar Color Zr, by Wieland Dental (K112710).
Device Description: Zenostar MO and Zenostar T are pre-sintered Zirconia discs with 98.5mm width and various thickness for use in the fabrication of dental prosthesis through the CAD/CAM milling technology. After milling, they are to be sintered at a high temperature to its full density. in order to achieve its expected physical properties. In addition, visualizer coloring liguid concentrates for dyeing Zenostar Color Zr (K11270) Solutions.
Indications for Use:
Zenostar MO and Zenostar T consist of machinable zirconia discs for the preparation of full ceramic crowns, onlays and bridges (anterior and molar). Zenostar VisualiZr is for the temporary dyeing of Zenostar Color Zr solutions according to the Instructions for Use.
Intended Use:
Zenostar MO (Medium Opacity) are milling discs intended for the fabrication of crown frameworks and multi-unit bridge frameworks for use as dental restorations.
Zenostar T (Translucent) are milling blanks intended for making single-tooth and multiunit restorations for use as dental restorations. Depending on the indication, frameworks or monolithic (full contour) restorations can also be made
Zenostar VisualiZr is for the temporary dyeing of Zenostar Color Zr solutions according to the Instructions for Use.
Comparison to Predicate: The predicate device to which Zenostar MO has been compared is IPS e.max ZirCAD (K051705) and Zenostar T and Zenostar VisualiZr have been compared to the predicate Zenotec Zr Bridge, Zenostar Zr Translucent, Zenotec Color Zr, Zenostar Color Zr, by Wieland Dental (K112710). For this application, Zenostar MO, Zenostar T have been compared to its predicate with regard to chemical composition, performance data and
510K Summary 5-1
{4}------------------------------------------------
## 510(K) SUMMARY
Image /page/4/Picture/1 description: The image shows the logo for Wieland Dental. The logo is in blue and features the word "WIELAND" in large, bold letters. Below the word "WIELAND" is the word "DENTAL" in smaller letters. There is a trademark symbol to the right of the word "DENTAL".
A MEMBER OF THE IVOCLAR VIVADENT GROUP indications for use. The comparison shows that Zenostar MO, Zenostar T are substantially equivalent to the predicate device. Zenostar VisualiZr is an accessory only and bears no effects on the device or its comparison to the predicate. The predicate device e.max ZirCAD is available in 1 shade and 1 block size. The subject device is available in 5-7 shares and 7 different disc thicknesses.
The intended use of both Zenostar MO and T are the same. They are used as restorative materials for crowns and bridges. However aesthetic criteria which have to be fulfilled vary depending on the esthetic requirements in the mouth. The availability of Zenostar in two grades of translucency allows the product to be selected which is best suited to the esthetic situation. The physical properties of both products are the same and both MO and T could be used for the identical indications, however due to the esthetic consideration we describe the indications which are most appropriate for each version. The medium opacity material "MO" is most appropriate used as a framework material, which will be covered with a more esthetic layering ceramic. The translucent material "T" is appropriate for monolithic solutions due to the good translucency.
Technological Characteristics: The device design, i.e. delivery form, and intended use of Zenostar MO, Zenostar T and the predicate device are the same. Zenostar VisualiZr bears no effects on the technological characteristics of the device. The predicate device e.max ZirCAD was delivered in block shape for use in the Sirona In-Lab dry CAD/CAM System. The subject device is sold in 98.5mm disc shape for use in the Wieland wet CAM milling system.
Testing Summary: The device has been designed and tested in accordance with ISO 6872:2008 Dentistry: Ceramic Materials for Flexural Strength, Chemical Solubility, Co-efficient of thermal expansion and Radioactivity. The test results are equivalent to those of the predicate device. Biocompatibility data are not submitted because the ingredients of the subject device and the predicate are equivalent. For the VisualiZr Colors, the liquid burns out almost completely during sintering. Any remaining molecules are bound in the molecular structure of the sintered restoration.
CONCLUSION: The above data and analysis demonstrates that Zenostar MO, Zenostar T and Zenostar VisualiZr are substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.