K3

K152830 · Osstem Implant Co., Ltd. · EIA · Apr 20, 2017 · Dental

Device Facts

Record IDK152830
Device NameK3
ApplicantOsstem Implant Co., Ltd.
Product CodeEIA · Dental
Decision DateApr 20, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6640
Device ClassClass 1
AttributesTherapeutic, Pediatric

Indications for Use

K3 is intended to supply power to and serve as a base for dental devices and accessories. This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants.

Device Story

K3 is a dental operative unit and chair designed for dental treatment of children and adults. It serves as a base for dental devices and accessories, supplying power, compressed air, and water. Components include a patient chair, dentist's element, assistant's element, dental light, monitor arm, and console. The device features an ergonomic design with adjustable touch panels, 3-way syringe, and assistant controls. It is operated by trained dentists or dental assistants in a dental clinic environment. The system electronically controls chair movements and water unit functions via integrated software. The device provides a stable platform for dental procedures, facilitating patient positioning and instrument access, thereby supporting dental care delivery.

Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-5), sterilization validation (ISO 17665), waterline cleaning/disinfection validation (ISO 16954, ISO 19458, ISO/TS 11080), electrical safety (IEC 60601-1), EMC (EN/IEC 60601-1-2), and usability (IEC 62366).

Technological Characteristics

AC-powered (100-120/220-240V) dental operative unit. Includes hydraulic electromotor for chair movement. Components: chair, unit, table, seat, stool, 3-way syringe, monitor arm, console. Materials: upholstery, waterlines, airlines (biocompatibility tested per ISO 10993). Connectivity: electronic control of chair and water functions. Sterilization: steam sterilization for reusable components (ISO 17665). Standards: ISO 7494-1/2, ISO 6875, IEC 60601-1, IEC 60601-1-2, IEC 62366.

Indications for Use

Indicated for use in dental clinics by trained dentists and/or dental assistants for the treatment of children and adults.

Regulatory Classification

Identification

A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 20, 2017 Osstem Implant Co. Ltd. c/o Mr. David Kim Hiossen Inc. QA/RA Manager 85 Ben Fairless Dr. Fairless Hills, PA 19030 Re: K152830 Trade/Device Name: K3 Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: Class I Product Code: EIA Dated: March 7, 2017 Received: March 21, 2017 Dear Mr. David Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Michael J. Ryan -S for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-012<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------| | 510(k) Number (if known) | K152830 | | Device Name | K3 | | Indications for Use (Describe) | | | K3 is intended to supply power to and serve as a base for dental devices and accessories. This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants. | | | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ st floor, B-dong, 135 Gasan digital 2-ro,Geumcheon-gu, Seoul, Republic of Korea. 153-759 ## 510(k) Summary Date: April 20, 2017 ## 1. Company and Correspondent making the submission: - Submitter's Name: - Address - Contact : - Phone: - Correspondent's Name: - Address: - Contact: - Phone: ## 2. Device: Trade or (Proprietary) Name : Common or usual name : Classification Name : Regulation Number : Device Classification: Subsequent Product Code: Osstem Implant Co., Ltd. 1st floor, B-dong, 135 Gasan digital 2-ro, Geumcheon-gu, Seoul, Republic of Korea. 153-759 Mr. Kim, Byunghoon +82-70-4394-5157 HIOSSEN Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 DAVID KIM 267 759 7031 ## K3 Dental Unit and Chair Dental operative unit and accessories 21CFR872.6640 Class I EIA ## 3. Predicate Device: Sirona C8 (K983242) ## 4. Description K3, Dental Unit and Chair, is designed for dental treatment of children and adults. It is a dental treatment unit in accordance with ISO 7494-1, ISO 7494-2 and ISO 6875. The 3way syringe is a dental instrument in accordance with EN 1639. It aids the dental application in the mouth of the patient by supplying air, water or spray. This product is designed for use in dentistry only and may only be used by trained medical personnel K3 is similar to other commercially available products based on the intended use, the technology used, the claims, the electrical power and performance characteristics. It is substantially equivalent in design, function and intended use to the predicate devices. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, with a small orange circle to the right of the word. Below "OSSTEM" is the word "IMPLANT" in gray. The logo is simple and modern. 1st floor, B-dong, 135 Gasan digital 2-ro,Geumcheon-gu, Seoul, Republic of Korea. 153-759 ## (1) Design Feature K3 dental operative unit has features of ergonomic, optimal position for treatment. Adjusted touch panel, 3way syringe and assistant touch panel, easy opening and closing arm rest. - (2) Main component of K3 K3 Mount and K3 Cart consist of the chair, unit, table, seat, stool, 3-way syringe, monitor arm, foot control and console. - (3) Technological Characteristics K3 is AC powered dental operative units with accessories that are intended to supply power to, and serve as a base for dental device and accessories. Including a treatment chair, the dentist's element, the assistant's element and a dental light, offer several additional options and electronically control chair movements with software and water unit functions. ## 5. Substantial Equivalence Discussion The K3 is substantial equivalent to the predicate device (K983242). The following comparison table is presented to demonstrate substantial equivalence. | | Predicate Device | Subject Device | |-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Name | Sirona C8 | K3 | | 510(k) | K983242 | N/A | | Manufacturer | Sirona Dental Systems GmbH | OSSTEM IMPLANT Co., Ltd. | | Intended use | The Sirona C8 is an electronically<br>controlled dental operative unit with<br>accessories that are intended to<br>supply power to, and serve as a base<br>for dental devices and accessories. | K3 is intended to supply power to and<br>serve as a base for dental devices and<br>accessories. This product includes a dental<br>chair. The dental treatment unit is intended<br>for use in the dental clinic environment<br>and is used by trained dentists and/or<br>dental assistants. | | Code | EIA | EIA | | Power and<br>Utility supply | 110V/220V AC electrical supply,<br>compressed air and water | AC 100-120/220-240V, 50/60Hz, compressed<br>air and water | | Protection Class | Class 1 equipment | Class 1 equipment | | Applied parts | Type B | Type B | | Components | Doctors element, assistants element, flex<br>arm, utility center with optional<br>cuspidor, patient chair, treatment light, | Chair, Unit, Table, Seat, Stool, Chair, Unit,<br>Table, Seat, Stool, Monitor Arm*, Hanaro<br>Console* | Comparison Table {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Osstem Implant. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. The logo is simple and modern, and it is likely used to promote the company's dental implant products. 1st floor, B-dong, 135 Gasan digital 2-ro,Geumcheon-gu, Seoul, Republic of Korea. 153-759 | | floor box. | (*K3 Cart model applied) | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------| | Installation | Available with chair mounted and cart<br>mounted dentist element Installed by<br>trained technicians. | Same as | | Syringe | 3-way syringe /<br>Sprayvit | 3-way syringe | | Curing light | Mini LED | N/A | | Control of Air<br>and water | Uses pneumatically controlled vales to<br>Water control the flow of air and water.<br>On/off and intensity controlled by foot<br>pedal. | Same as | | Air Pressure | 550 kPa/750 kPa (min/max) | 500kPa(min)/750kPa(max) | | Water<br>Pressure | 250 kPa/600 kPa(min/max) | 250kPa(min)/600 kPa(max) | | Water System | User may select self-Contained water<br>system or city water supply | City water supply | | Suction<br>devices | HVE (High volumeevacuator)<br>Saliva Ejectors | HVE (High volume evacuator)<br>Saliva Ejectors | | Patient load | Max. 135kg | Same as | | Chair Height | Max.755mm<br>Min. 395mm | Max. 790±10mm<br>Min. 450±10mm | | Back Rest | 0° to 60° | 0° to 66° | | Head Rest | -10° to 45° | -10° to 45° | | LiftMotor | Motor SL (electricmotor) | Hydraulic electromotor | | Electrical<br>Safety | Comply with IEC 60601-1 | Same as | | EMC | Comply with IEC 60601-1-2 | Same as | | Place of Use | Dental office, hospital | Same as | | Intended<br>Users | Dentists, Dental Hygienists, Dental<br>Equivalent Assistants | Same as | As shown in the comparison chart for K3 of OSSTEM IMPLANT Co., Ltd., predicate device, they are substantially equivalent in terms of indication for use, technology and performance specifications. The external design of the subject device might be slightly different from the subject devices, however, the performance testing results provided in this submission supports that the subject device performs as well as the predicate devices for its intended use. {6}------------------------------------------------ 1st floor, B-dong, 135 Gasan digital 2-ro,Geumcheon-gu, Seoul, Republic of Korea. 153-759 ## 6. Indication for use K3 is intended to supply power to and serve as a base for dental devices and accessories. This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants. ## 7. Summary of Non-Clinical Performance Testing The K3 has been subjected to extensive safety, performance, and product validations prior to release. Tests including EMC have been performed to ensure the devices comply with the US regulations. ## Biocompatibility Testing The biocompatibility evaluation for K3 components was conducted in accordance with the FDA Guidance Document and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The upholstery, waterlines, and airlines were tested for cytotoxicity (ISO 10993-5). The components of the K3 are considered external communicating device for a duration of less than 24 hours. ## Cleaning and Sterilization Validation A representative sample of the3-Way syringe components were tested to validate that the components can withstand the steam sterilization process and that acceptable sterility is achieved using the recommended sterilization protocols. The sterilization testing was conducted according to ISO 17665-1:2006 and ISO 17665-2 and it validated that the reusable K3 components can be sterilized to reach an acceptable sterility assurance level. In addition, cleaning/disinfection was conducted on the waterlines of the subject device. Validation was conducted using the following standards: - ISO 16954:2015 Dentistry Test methods for dental unit waterline biofilm treatment - · ISO 19458:2006 Water quality-Sampling for microbiological analysis - · ISO/TS 11080:2009 Dentistry Essential characteristics of test methods for the evaluation of treatment methods intended to improve or maintain the microbiological quality of dental unit procedural water The microbiological simulation test and physical & chemical test are performed to ensure that the Dental chair waterline is effectively cleaned. ## Electrical Safety and Electromagnetic Compatibility (EMC) Comprehensive performance testing has been conducted on the K3 in accordance FDA recognized standards. EMC testing was conducted in accordance with Standard EN/IEC 60601-1-2. Electrical, mechanical, and environmental safety testing according to Standard EN/EC 60601-1 was performed. Usability testing was conducted in accordance with Standard EN/EC 60601-1-6 and EN/IEC 62366. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo and address of OSSTEM Implant Co., Ltd. The logo features the word "OSSTEM" in orange, with the word "IMPLANT" in gray below it. The address is listed as 1st floor, B-dong, 135 Gasan digital 2-ro, Geumcheon-gu, Seoul, Republic of Korea. The postal code is 153-759. #### Software and System Verification and Validation Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. | No. | Test Item | Standard | |-----|-------------------------------------------------------------|---------------------------------------------------------| | 1 | Cytotoxicity | ISO 10993-5 | | 2 | EMC Test | EN 60601-1-2 , EN55011 , EN 61000-3-2<br>EN 61000-3-3 | | 3 | Software Validation | ISO 14971 , IEC62304 , IEC60601-1 | | 4 | Basic safety & essential performance | IEC 60601-1 | | 5 | General requirement & test method | ISO 7494-1 | | 6 | Air, Water, suction & wastewater<br>system | ISO 7494-2 | | 7 | Dentistry-Patient chair | ISO 6875 | | 8 | Basic safety & essential performance<br>of dental equipment | IEC 80 601-2-60 | | 9 | Collateral Standard : Usability | IEC 60601-1 | | 10 | Usability engineering | IEC 62366 | | 11 | Hydraulic motor durability test | ISO 13485 | | 12 | Dental Hand piece | ISO 15606 | | 13 | Cleaning Validation | ISO 16954 , ISO 19458 , ISO/TS 11080 ,<br>ASTM E2871*12 | | 14 | Sterilization Validation | ISO 17665-1, ISO/TS 17665-2 | ## 8. Summary of clinical testing No clinical studies are submitted {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for OSSTEM IMPLANT. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. The logo is simple and modern. 1st floor, B-dong, 135 Gasan digital 2-ro,Geumcheon-gu, Seoul, Republic of Korea. 153-759 ## 9. Review The K3 has the similar device characteristics as the predicate device ; intended use, Power and Utility supply, Syringe, design and use concept are similar. The K3 has a slight difference in components of unit except for curing light. The K3 has been subjected to extensive safety, performance, and product validations prior to release. Safety tests including EMC have been performed to ensure the devices comply with the US regulations. Based on the comparison of intended use and technical features, the K3 is substantial equivalent to the predicate devices. ## 9. Conclusions In accordance with the above data and based on the information provided in this premarket notification OSSTEM IMPLANT Co., Ltd. concludes that the K3 is substantially equivalent to predicate devices as described herein. END
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...