K061341 · CAO Group, Inc. · EBZ · Jun 12, 2006 · Dental
Device Facts
Record ID
K061341
Device Name
PALMLIGHT 10, MODEL 001-00030
Applicant
CAO Group, Inc.
Product Code
EBZ · Dental
Decision Date
Jun 12, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6070
Device Class
Class 2
Attributes
Therapeutic
Intended Use
For light activated polymerization of dental materials such as composites, luting cements, adhesives, and sealants using visible and near-UV light.
Device Story
Palmlight 10 is a cordless LED dental curing light used by dental professionals to harden light-curable dental materials. The device emits light in the 390-500nm wavelength range, which is absorbed by photoinitiators in dental compositions to trigger polymerization. It features an internal rechargeable battery for cordless operation, a charging cradle, and an optional power supply connection. Users select from preset time increments and four output modes via an on/off button. The charging cradle includes a power output display to monitor performance. The device benefits patients by enabling efficient, chairside curing of restorative materials.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and compliance with international safety and performance standards, including IEC 60601-1, IEC 60601-1-2, and IEC 60825-1.
Technological Characteristics
LED-based dental curing light; wavelength 390-500nm; internal rechargeable battery; charging cradle with power output display; cordless or corded operation. Complies with IEC 60601-1:1998+A1, IEC 60601-1-2:2004, and IEC 60825-1:1993+A1+A2.
Indications for Use
Indicated for light-activated polymerization of dental materials including composites, luting cements, adhesives, and sealants in the oral cavity.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
Predicate Devices
Bluephase 16i
Palmlight (CAO Group, Inc.)
Palmlight (Ultradent Products, Inc.)
Related Devices
K020057 — PALMLIGHT · CAO Group, Inc. · Aug 27, 2002
K221152 — CuringPen Dental Curing Light · Changzhou Sifary Medical Technology Co., Ltd. · Nov 23, 2022
K092951 — STARLIGHT PRO AND SILIVERLIGHT DENTAL CURING LIGHTS · Mectron S.P.A. · Dec 7, 2009
K052349 — AVANTE LED CURING UNIT, MODEL N44 · Pentron Clinical Technologies · Oct 21, 2005
K031615 — SLED CURING LIGHT SYSTEM · Dentsply International · Jul 21, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Kobi341
### 510(k) Summary of Safety and Effectiveness
CAO Group, Inc. 86863 South 700 West Sandy, UT 84070 Tel: 801-256-9282 Fax: 801-256-9287
JUN 1 2 2006
Robert Larsen, Preparation Date: April 25, 2006
### Device Name:
| Trade Name: | Palmlight 10 |
|-------------------------|-------------------------------------------------------|
| Common Name: | LED Dental Curing Light |
| Product Classification: | Activator, Ultraviolet, For Polymerization (872.6070) |
### Legally Marketed Predicate Devices for Substantial Equivalence:
Bluephase 16i
Palmlight (CAO Group, Inc.)
Palmlight (Ultradent Products, Inc.)
#### Rationale for Substantial Equivalence:
The aforementioned curing light devices share similar indications for use in the oral environment, similar design features including wavelength, output power, and light source. The devices share similar applications with regards to applying light energy to dental materials for the purpose of polymerizing or otherwise energizing said materials.
### Description of Submitted Device:
The Palmlight 10 is a dental curing light source that employs high power light emitting diodes (LED's) to generate light in the wavelengths between 390 and 500nm. This wavelength of light is readily absorbed by photoinitiators in a variety of dental compositions. Such initiators begin a polymerization reaction that causes these dental compositions to harden. The device features an internal rechargeable battery, allowing the device to be used in a cordless fashion. The device can be placed in a charging cradle when not in use, or can be attached to a power supply such that the device can be used or charged while attached to the power supply. The device features the ability to select from preset time increments for activating the device. The user can also select from 4 different
{1}------------------------------------------------
output modes. The device features and on/off button to activate and deactivate the LED's. The device features a power down program such that the unit's display is deactivated if the unit is dormant for an extended period of time. The charging cradle features a power output display to assist the end user in determining the performance of the device.
### Intended Uses of the PalmLight 10 curing light system:
The device is intended to be used on a variety of dental compositions within the oral cavity. See Indications for Use on page 2-1.
### Technological Characteristics and Substantial Equivalence:
The Bluephase 16i uses LED's to generate light energy in the 430-490nm range with a power output of 1600mw/cm². This system features a removable battery back system, allowing the device to be used in a cordless configuration. The system also incorporates a power adaptor, allowing the unit to be attached to power supply and used in this condition. The device is indicated for curing dental compositions such as composites and luting cements. The system features a charging base with a built in power indicator. The system allows the user to select from preset modes and activation times.
The Palmlight, manufactured by CAO Group, Inc. uses LED's to generate light energy in the 400-460nm range with a power output of a minimum 800mW/cm². The system features a light, compact design with a permanent attachment to the power supply. The device is indicated for curing dental compositions such as composites, luting cements, and sealants. The system allows the user to select from preset modes and activation times.
The Palmlight, manufactured by Ultradent Products, Inc. uses LED's to generate light energy in the 400-460nm range with a power output of a minimum 800mW/cm². The system features a light, compact design with a permanent attachment to the power supply. The device is indicated for curing dental compositions such as composites, luting cements, and sealants. The system allows the user to select from preset modes and activation times.
### Performance Standards:
The Palmlight 10 dental curing light complies with the safety and performance requirements of IEC 60601-1:1998+A1, IEC 60601-1-2:2004 and IEC 60825-1:1993+A1+A2.
{2}------------------------------------------------
# Clinical Performance Data
See Part 7: Performance Data
# Conclusion
The Palmlight 10 is substantially equivalent to the listed dental curing light devices without raising any issues of safety or effectiveness. This device shares similar intended uses, and similar functional and performance characteristics. The device is designed to comply with relevant federal and international safety and priformance standards.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half. Inside the circle is a stylized symbol of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 2 2006
Mr. Robert K. Larsen Operations Director CAO Group, Incorporated 8683 South 700 West Sandy, Utah 84070
Re: K061341
Trade/Device Name: Palmlight 10 Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator for Polymerization Regulatory Class: II Product Code: EBZ Dated: April 25, 2006 Received: May 15, 2006
Dear Mr. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 -- Mr. Robert K. Larsen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Suite Y. Michie Ens.
Chiu S. Lin, PhD Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K061341
Device Name: Palmlight 10
Indications For Use:
For light activated polymerization of dental materials such as composites, luting cements, adhesives, and sealants using visible and near-UV light.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Ruane
nesthesiology, General Hospital, ontrol, Dental Devices
Number. K06134
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter Use OR
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.