K201904 · CAO Group, Inc. · QNF · Nov 24, 2020 · Dental
Device Facts
Record ID
K201904
Device Name
Monet Curing Laser
Applicant
CAO Group, Inc.
Product Code
QNF · Dental
Decision Date
Nov 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Monet Curing Laser is a source of illumination for curing photo-active materials and adhesives.
Device Story
Monet Curing Laser delivers light energy to dental restorative materials/adhesives to initiate photopolymerization. Device uses solid-state laser diode (450 ± 5nm) to generate coherent light at max 2.0 watts (2400 mW/cm²). Optical focusing lens collimates energy into parallel beam; consistent intensity regardless of distance. Includes low-intensity aiming beam for site positioning. Operated by dental clinicians in clinical settings. Powered by detachable rechargeable battery; includes charging stand with integrated radiometer for output verification. Simple interface: one on/off switch, fixed 3-second activation time. Metal construction with glass output window. Benefits include consistent curing depth and reliable polymerization of dental composites.
Clinical Evidence
Bench testing only. Depth of cure testing compared Monet laser to VALO curing light using various dental composite brands and shades. Results showed Monet achieved polymerization equal to or better than VALO at lower light intensities.
Technological Characteristics
Solid-state laser diode (450 ± 5nm); max 2.0W output. Metal construction, glass output window. Rechargeable battery power. Optical focusing lens for collimation. Complies with 21 CFR 1040.10/1040.11, IEC 60601-1 (3.1), IEC 60601-1-2 (4), IEC 60601-1-6 (3.1), IEC 60601-2-22 (3), and IEC 60825-1 (2). Non-sterile; uses single-use barrier sleeves.
Indications for Use
Indicated for curing photo-activated dental restorative materials and adhesives in dental patients.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 24, 2020
CAO Group, Inc. Robert Larsen Regulatory Affairs Manager 4628 West Skyhawk Drive West Jordan, Utah 84084
Re: K201904
Trade/Device Name: Monet Curing Laser Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet activator for polymerization Regulatory Class: Class II Product Code: QNF, GEX Dated: August 19, 2020 Received: August 26, 2020
Dear Robert Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201904
Device Name Monet Curing Laser
Indications for Use (Describe)
The Monet Curing Laser is a source of illumination for curing photo-active materials and adhesives.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k) Number: | K201904 |
|------------------------|--------------------------------------------------------|
| Applicant Information: | |
| Company Name: | CAO Group, Inc. |
| Company Address: | 4628 West Skyhawk Drive West Jordan, Utah 84084 U.S.A. |
| Company Phone: | 1-801-256-9282 |
| Company Fax: | 1-801-256-9287 |
| Contact Person: | Robert K. Larsen |
| Preparation Date: | November 20, 2020 |
| Device Name: | |
| Trade Name: | Monet Curing Laser |
| Common Name: | Ultraviolet Activator for Polymerization |
C Product Code: EBZ, GEX 872.6070, 878.4810 Regulation: Product Classification: Class II
## Legally Marketed Predicate Devices for Substantial Equivalence:
PRIMARY PREDICATE DEVICE-Valo Grand, manufactured by Ultradent Products, Inc. (K160551)
#### Description of Submitted Device:
The Monet Curing Laser is a device for delivering light energy to dental restorative materials and adhesives, which utilize the light energy to affect a photopolymerization reaction. This energy is generated by a solid-state laser diode, which provides a consistent and reliable generation of laser energy at 450 ± 5nm for a maximum of 2.0 watts (2400 mW/cm²) of energy output. The laser energy is delivered to the dental material by means of an optical focusing lens that collimates the laser energy to a nearly parallel beam size that remains highly consistent regardless of distance to the target material. The device also includes a low-intensity aiming beam feature that allows the operator to see the delivery site of the energy without risk of premature polymerization of the material.
The device features a metal construction with glass output window that contains all device electronics. The device features a simple on/off switch and set 3-second activation time. The device incorporates a detachable rechargeable battery which is recharged on an accompanying charging stand. The charging stand incorporates a basic radiometer to allow the user to confirm the output intensity of the emitted light.
{4}------------------------------------------------
## Indications for Use of the Submitted Device:
The Monet Curing Laser is a source of illumination for curing photo-activated dental restorative materials and adhesives.
### Rationale for Substantial Equivalence:
The submitted device features an identical Indications for Use as the predicate device. The submitted device shares similarities in materials of construction, intended operator, basic design principle of emitting high intensity blue light to initiate a polymerization of the target dental material or adhesive. The submitted device operates from an internal rechargeable battery, the same as the predicate device. The submitted device includes a charging stand to recharge the batteries, the same as the predicate device. The submitted device is capable of photopolymerizing, or curing, the intended dental materials and adhesives, the same as the predicate device.
The submitted device differs from the predicate device in that the submitted device uses a laser diode to generate coherent, narrow spectrum light photons, whereby the predicate device uses a light-emittingdiode to generate non-coherent, broad spectrum light photons. The submitted device differs from the predicate device where the predicate device features more operational controls (number of user interface buttons and indicators) and choices of different emission times and intensities, where the submitted device provides only one interface button and one indicator, and provides for only one emission time and one emission intensity. The submitted device incorporates a low-intensity aiming beam to aid the clinician in positioning the device prior to high-intensity emissions, a feature that is absent from the predicate device. The submitted device includes laser protective eyewear, where the submitted device provides a filter shield that affixes to the device.
#### Performance Data:
## ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY
The Monet Curing Laser is designed to comply with the performance requirements of 21 CFR 1040.10 and 1040.11, with permissible deviations relative to Laser Notice 50, dated June 24, 2007. The device also complies with the recognized standards of IEC 60601-2-22 Edition 3 and IEC 60825-1 Edition 2. The device is designed in compliance to the entirety of IEC 60601-1: Edition 3.1, IEC 60601-1-2 Edition 4, and IEC 60601-1-6 Edition 3.1.
#### PERFORMANCE BENCH TESTING
Depth of cure testing was performed with the Monet and with the VALO curing light, assessing different popular brands of dental composite and lighter and darker shades of composite from each of these brands. Both light sources were activated for a single exposure of the same amount of time. The Monet is capable of achieving polymerization of the tested composites at a level that is generally equal to or better than the VALO while doing so with a lower light intensity.
#### BIOCOMPATIBILITY
The submitted device features the same nature of incidental patient contact as the predicate device. The submitted device is composed of similar materials to that of the predicate device. The submitted
{5}------------------------------------------------
device is intended as a reusable article, the same as the predicate device, except for the single-use barrier sleeve, which is also similar to that of the predicate device.
# END-USER STERILIZATION
No portions or aspects of the submitted device are intended to be sterilized by the end-user prior to use, nor are any portions provided to the end-user in a sterile state. The submitted device includes single-use barrier sleeves to aid in infection control. The barrier sleeve is not provided in a sterile state, and is not intended to be sterilized by the end-user prior to use.
# Conclusion:
The Monet Curing Laser is substantially equivalent to the listed predicate device without raising any new issues of safety or effectiveness. This device shares an identical intended use, similar principles of basic use and function, and similar performance characteristics. The device is designed to comply with relevant federal and international safety and performance standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.