KREATIV MACH 5, MACH 5 PLUS, MACH M, MACH 6

K980216 · Kreativ, Inc. · KOJ · Jun 16, 1998 · Dental

Device Facts

Record IDK980216
Device NameKREATIV MACH 5, MACH 5 PLUS, MACH M, MACH 6
ApplicantKreativ, Inc.
Product CodeKOJ · Dental
Decision DateJun 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Kreativ Mach Series Air Abrasion systems are used for cavity preparation in Classes I, II, III, IV, and V. The uses include removal of tooth structure and restorative dental materials, and site preparation for pit and fissure sealant therapy and bonding of porcelain and ceramic. They are also used for restoration prophylaxis.

Device Story

Pneumatic air abrasion system; utilizes pressurized air and aluminum oxide powder to generate high-velocity particle stream; delivered via handpiece similar in size to high-speed dental drill; used by dentists in clinical settings for cavity preparation, caries removal, and surface preparation for restorative adhesion; refillable particulate canisters; facilitates dental restorative procedures; improves ease of service and adaptability compared to manual methods.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological and intended use comparison to predicate devices.

Technological Characteristics

Pneumatic air abrasion system; utilizes pressurized air and aluminum oxide abrasive powder; handpiece-based delivery; refillable particulate canisters; mechanical operation.

Indications for Use

Indicated for dental cavity preparation (Classes I-V), removal of tooth structure and restorative materials, site preparation for pit and fissure sealants, bonding of porcelain/ceramic, and restoration prophylaxis.

Regulatory Classification

Identification

An airbrush is an AC-powered device intended for use in conjunction with articulation paper. The device uses air-driven particles to roughen the surfaces of dental restorations. Uneven areas of the restorations are then identified by use of articulation paper.

Special Controls

The special control for this device is International Electrotechnical Commission's IEC 60601-1-AM2 (1995-03), Amendment 2, “Medical Electrical Equipment—Part 1: General Requirements for Safety.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## JUN 1 6 1998 2980214 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Regulatory Authority: Safe Medical Devices Act of 1990, CFR 807.87 Company Name: Kreativ, Inc. 1517 Industrial Way SW Albany, OR 97321 Company Contact, Regulatory Affairs Jeff Lane Kreativ, Inc. 1517 Industrial Way. SW Albany, OR 97321 Phone: 541/924 2480 ### Device Name: Mach 5, Mach 5 plus, Mach M, and Mach 6 ### Predicate Devices: KV 1, K#940776 (Kreativ, Inc., Albany, OR), KCP 2000, K#921748 (American Dental Technologies, Troy, MI). and the MicroPrep Cavity Preparation System, K#932997, (Sunrise Technologies, Inc.). ### Device and Indications for use: The Mach 5, Mach 5 plus, Mach 6 are pneumatic devices which combine pressurized air and aluminum oxide powder to produce a high velocity stream of particles to perform dental restorative procedures, including preparation for pit and fissure sealant and composite restorations. The units are capable of removing dental caries, old restoration materials, as well as healthy enamel and dentin to prepare the tooth surface for subsequent adhesion of restorative materials. The abrasive particulate is delivered via a small handpiece which is approximately the size of a high speed dental drill. The systems are designed for ease of service and maintenance with readily accessed and refillable canisters for the particulate supply. ## Discussion: Since the intended use and technical specifications of the Mach 5, Mach 5 plus, Mach M. and the Mach 6 are virtually identical to the predicate devices, and the differences in the devices only make them easier to use and more adaptable to a variety of dental practice situations, we believe that the Mach Series devices are substantially equivalent to the predicate devices and can be marketed under Section 510 (k) of the FD&C Act. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three human profiles facing to the right. ుండి 15:15 గా నిర్మాలు గారా Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 1 6 1998 Mr. Joe Forehand · Vice President of Operations Kreativ, Incorporated 1517 Industrial Way S.W. Albany, Oregon 97321 Re : K980216 Trade Name: Kreativ Mach 5, Mach 5 Plus, Mach M, and Mach 6 Regulatory Class: III Product Code: KOJ January 19, 1998 Dated: Received: January 21, 1998 #### Dear Mr. Forehand: We have reviewed your Section 510(k)· notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may resulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of {2}------------------------------------------------ Page 2 - Mr. Forehand the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdaygov/cdrh/dsmamain.html". Sincerely yours Timothy A. Ulinski TimothyA. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation-Center for Devices and Radiological Health Enclosure ్ స్టేష్ ੀ {3}------------------------------------------------ ## Exhibit 3a # Indications Statement 510(k) submission: Kreativ Mach Series Air Abrasion Systems: Mach 5, Mach 5+, Mach M, and Mach 6 ## Indications for use: The Kreativ Mach Series Air Abrasion systems are used for cavity preparation in Classes I, II, III, IV, and V. The uses include removal of tooth structure and restorative dental materials, and site preparation for pit and fissure sealant therapy and bonding of porcelain and ceramic. They are also used for restoration prophylaxis. യുടെ പുറത്തേ പണ്ട Susan Runyon (Division Sign-Off) vision of Dental, Infection Control, General Hospital Deyic 510(k) Number .
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...