The DCI Edge Dental Chair with operative unit is a device intended to position the patient in such a manner to provide to the dental practitioner visual and manual access to the oral cavity. This device should provide for the ergonomic necessities of most patients, both adults and children, throughout its range of movement. The device shall also be the means of supporting delivery systems for the purpose of standard dental procedures.
Device Story
Dental chair with integrated operative unit; includes adjustable headrest, backrest, seat height, and armrests. Operative unit features floor box, dental light, cuspidor, cup filler, control panel, foot control, 3-way syringe, and saliva ejector. Used in dental clinics by dentists, technicians, and assistants. Device provides patient positioning for oral cavity access and supports delivery systems. Software controls chair movement (up/down, recline/upright) via pre-programmed settings activated by push-button. Movement is slow, limited by sensors and mechanical hard stops to ensure patient safety. Healthcare providers use the chair to facilitate ergonomic access during standard dental procedures.
Clinical Evidence
Bench testing only. Compliance with ISO 14971 (risk management), ISO 10993-1 (biocompatibility), ANSI/AAMI ES60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 62366 (usability), and IEC 80601-2-60 (dental equipment safety).
Technological Characteristics
Materials: metal structural support, sheet metal, plastic covers. Energy: electrical (chair movement). Connectivity: standalone. Software: pre-programmed control for chair positioning. Safety: sensors and mechanical hard stops. Standards: ISO 14971, ISO 10993-1, ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 62366, IEC 80601-2-60.
Indications for Use
Indicated for positioning adult and pediatric patients to provide dental practitioners visual and manual access to the oral cavity and to support delivery systems for standard dental procedures.
Regulatory Classification
Identification
A dental chair and accessories is a device, usually AC-powered, in which a patient sits. The device is intended to properly position a patient to perform dental procedures. A dental operative unit may be attached.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 1, 2016
DCI % Mr. Robert Bellingham Quality & Regulatory Manager 305 N. Springbrook Rd. Newberg, Oregon 97132
Re: K151987
Trade/Device Name: DCI Edge Dental Chair with Operative Unit Regulation Number: 21 CFR 872.6250 Regulation Name: Dental Chair and Accessories Regulatory Class: I Product Code: KLC, EIA Dated: May 2, 2016 Received: May 2, 2016
Dear Mr. Bellingham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
in filmləri Alinan Alinan Ali Qaran Qırmanın Qaran Qaran Qırmanın Qaran Qırmızı Qaran Qırmızı Qaran Qırmızı Qaran Qırmızı Qaran Qırmızı Qaran Qırmızı Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151987
Device Name
DCI Edge Dental Chair with operative unit
### Indications for Use (Describe)
The DCI Edge Dental Chair with operative unit is a device intended to position the patient in such a manner to provide to the dental practitioner visual and manual access to the oral cavity. This device should provide for the ergonomic necessities of most patients, both adults and children, throughout its range of movement. The device shall also be the means of supporting delivery systems for the purpose of standard dental procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
305 N. Springbrook Road, Newberg, Oregon, USA 97132
## K151987 510(k) Summary DCI Edge Dental Chair with Operative Unit
#### 1 Sponsor
Dental Components Inc. 305 N. Springbrook Road Newberg, Oregon 97128 Telephone: 503-537-2431 Date Prepared: May 31, 2016
#### 2 Device Name
Proprietary Names: DCI Edge Dental Chair with Operative Unit Common/Usual Name: Chair, Dental with Operative Unit Classification: Chair, dental, with operative unit Regulation Number: 21 CFR 872.6250 Class: 1 Product Code: KLC Secondary Product Code: EIA Establishment Registration Number: 3029780
#### 3 Predicate Device
Foshion Chair Mounted Dental Unit (K071353)
#### 4 Intended Use
The DCI Edge Dental Chair with operative unit is a device intended to position the patient in such a manner to provide to the dental practitioner visual and manual access to the oral cavity. This device should provide for the ergonomic necessities of most patients, both adults and children, throughout its range of movement. The device shall also be the means of supporting delivery systems for the purpose of standard dental procedures.
#### 5 Device Description and Function
The DCI Edge Chair with Operative Unit is a dental chair with an operative unit attached. There are three main parts to the proposed device: a dental chair, dental light and dental chair has adjustable headrest, backrest, seat height and armrests. The dental chair unit consists of a floor box to which are attached a number of operative units: lights, cuspidor, cup filling device, control panel, and foot control. The instrument panel also includes a 3-way syringe and salvia ejector.
#### Level of Concern 6
The software was determined to be of a moderate level of concern. The software controls the movement of the dental chair, and it is preprogrammed and activated by the push of a button. The chair then moves slowly up and down, and the backrest moved slowly from an upright or a reclined position. The movement is very slow and the movement is limited by a stop sensor and mechanical hard stop. There are sensors and hard stops to prevent the chair from going beyond its pre-programmed range.
{4}------------------------------------------------
#### 7 Technological Characteristics Summary
The technological characteristics between the predicate and proposed device are similar physically and technically. While the wording of the Indications for Use differs, both the subject and predicate devices share the same intended use, as well as equivalent material composition, fundamental scientific technology, principles of operation, and basic design. The predicate device does offer several accessories as standard, such as x-ray viewer and dental handpiece. The DCI Edge chair does not offer these as standard. However, the intended use and the fundamental scientific technology are still the same. The x-ray viewer and dental handpieces are available from other companies as aftermarket devices.
| PARAMETER | DCI EDGE CHAIR<br>W/OPERATIVE UNIT<br>(K151987) | DCI DENTAL UNIT<br>(K944271) | FOSHION DENTAL<br>CHAIR W/OPERATIVE UNIT (K071353) |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The DCI Edge Dental Chair<br>with operative unit is a device<br>intended to position the patient<br>in such a manner to provide<br>to the dental practitioner<br>visual and manual access<br>to the oral cavity. This device<br>should provide for the ergonomic<br>necessities of most patients, both<br>adults and children, throughout<br>its range of movement. The<br>device shall also be the means<br>of supporting delivery systems<br>for the purpose of standard<br>dental procedures. | The DCI Dental Unit is<br>intended to supply power<br>and serve as a base for<br>dental devices and accessories.<br>The product includes a dental<br>unit, 3-way syringe and air<br>foot control. The unit is<br>intended for use in the<br>dental office or clinic and is<br>used by trained dentists<br>and/or technicians and<br>assistants. It can be attached<br>to a chair or stand alone. | The Foshion Chair Mounted<br>Dental Unit is intended to<br>supply power and serve as a<br>base for dental devices and<br>accessories. The product<br>includes a dental chair,<br>operating light x-ray viewer,<br>control panel, low and high<br>speed turbine handpieces,<br>3- way syringe and air foot<br>control. The unit is intended<br>for use in the dental office<br>or clinic and is used by<br>trained dentists and/or<br>dental technicians and<br>assistants. |
| Features | | | |
| Seat and backrest<br>positioning | Pre-programmed settings. | N/A | Pre-programmed settings. |
| Multiple Handpiece<br>Automatic Control<br>System | Three handpiece controls are<br>standard, up to five handpieces<br>optional. Handpieces are not<br>sold with the unit. | Three handpiece controls are<br>standard, up to five<br>handpieces optional.<br>Handpieces are not sold with<br>the unit. | Three handpiece controls<br>are standard, up to five<br>handpiece optional. Unit is<br>sold with handpieces. |
| Water Control System<br>for Handpieces and<br>Syringes | Non-retracting water system for<br>handpieces and syringes.<br>Retracting available with check-<br>valve. | Non-retracting water system<br>for handpieces and syringes.<br>Retracting available with<br>check-valve. | Non-retracting water<br>system for handpieces and<br>syringes. Retracting<br>available with check-valve. |
| 3-Way Syringe | 3-Way syringe with quick<br>disconnect tip and<br>autoclavable tips standard. | 3-Way syringe with quick<br>disconnect tip and<br>autoclavable tips standard. | 3-Way syringe with quick<br>disconnect tip and<br>autoclavable tips<br>standard. |
| Instrument Holders | Multiple position holder. | Multiple position | Multiple position holder. |
| | | holder. | |
| Assistant Holders | Multiple position holder. | Multiple position | Multiple position holder. |
| | | holder. | |
| Gravity Drain Cuspidor | Cuspidor available. | Cuspidor available. | Cuspidor available. |
| Mounting | Chair and cart. | Chair and cart. | Chair and cart. |
| Vacuum System | HVE and saliva | HVE and saliva ejectors | HVE and saliva ejectors |
| | ejectors available. | available. | available. |
| Foot Control | Standard. | Standard. | Standard. |
| Utility Box with Air and | Standard. | Optional w/Chair | Standard. |
| Water Shut-off, | | | |
| Regulators and Filters | | | |
| Control Pad for | Standard | Optional w/Chair | Standard |
| Chair Movement | | | |
| and Light | | | |
| Heated Water Supply | Available. | Optional w/Chair | Available. |
| Dental Light | Available. | Available | Available. |
| Turbines | Can be added. | Can be added | Standard |
| Upholstery | Standard | Standard if ordered | Standard |
| | | w/Chair | |
| Arm Rest | Standard | Standard if ordered | Standard |
| | | w/Chair | |
| Materials | | | |
| Construction | Metal structural support, | Metal structural support, | Metal structural support, |
| | sheet metal, plastic | sheet metal plastic | sheet metal, plastic covers. |
| | covers. Details contain in | covers. | |
| | product manual. | | |
{5}------------------------------------------------
## 305 N. Springbrook Road, Newberg, Oregon, USA 97132
#### 8 Non Clinical Testing
The DCI Edge Dental Chair with Operative Unit complies with the following performance standards:
- ISO 14971 Second Edition 2007-03-01 Medical devices Application of risk management to medical devices.
- · AAMI ANSI ISO 10993-1:209/(R) 2013 Biological evaluation of medical devices-Part1: Evaluation and testing within a risk management process.
- ANSI/AAMI ES60601-1:2005/(R)2012 Issued: 2012/01/17 Medical Electrical Equipment—Part 1: General requirements for basic safety and essential performance with C1:2009 (R)2012 and A2:2010/(R)2012
- · ISO 60601-1-2:2007 Medical electrical equipment –Part1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility-Requirements and test.
- · IEC 62366 Edition 1.1 Medical devices-Application of usability engineering to medical devices.
- · IEC 80601-2-60:2012 Medical electrical equipment Part 2-60: Particular requirements for basic safety and essential performance of dental equipment.
#### 9 Conclusion
DCI, ITL has demonstrated that, for the purposes of FDA's regulation of medical devices, the DCI Edge Dental Chair with Operative Unit is substantially equivalent to the predicate devices in intended use, material composition, fundamental scientific technology, principles of operation, and basic design.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.