← Product Code [JEQ](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/JEQ) · K040416

# SONICARE ADVANCE TOOTHBRUSH, MODEL 4900 (K040416)

_Philips Oral Healthcare, Inc. · JEQ · Jul 2, 2004 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/JEQ/K040416

## Device Facts

- **Applicant:** Philips Oral Healthcare, Inc.
- **Product Code:** [JEQ](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/JEQ.md)
- **Decision Date:** Jul 2, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6865
- **Device Class:** Class 1
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

For promotion of good oral hygiene, including the reduction of dental plaque and gingivitis.

## Device Story

Sonicare Advance (Model 4900) is a powered toothbrush for home use. Device consists of a brush head, power handle, and induction charger base. Operation involves a stator coil in the handle generating a 261-Hz oscillating magnetic flux, which drives magnets mounted on a resonator arm within the brush head. This oscillation causes the bristle plate to vibrate at a resonant frequency, facilitating plaque removal. The system is powered by internal nickel-cadmium rechargeable batteries. The device is operated by the user to improve oral hygiene and reduce gingivitis. Clinical benefits include reduction of gingival inflammation and plaque.

## Clinical Evidence

Seven clinical trials (six controlled) were conducted on the Sonicare Advance line. Endpoints included soft tissue safety, gingival inflammation indices, pocket probing depth, and clinical attachment level. Results demonstrated that the device effectively reduces gingivitis.

## Technological Characteristics

Powered toothbrush; 261-Hz oscillating magnetic flux drive; resonator arm and torsion bar mechanism; nickel-cadmium rechargeable batteries; induction charging; solid-state electronics; 3 W maximum power dissipation; 120 V, 60 Hz input; 65 kHz charging signal.

## Regulatory Identification

A powered toothbrush is an AC-powered or battery-powered device that consists of a handle containing a motor that provides mechanical movement to a brush intended to be applied to the teeth. The device is intended to remove adherent plaque and food debris from the teeth to reduce tooth decay.

## Predicate Devices

- TX Electric Toothbrush ([K921773](/device/K921773.md))
- Dental Air Force Home Dental Cleaning System ([K001493](/device/K001493.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K040416

JUL 0 2 2004

### 510(k) SUMMARY

Name of 510(k) sponsor:

Philips Oral Healthcare, Inc.

Address:

35301 SE Center Street Snoqualmie, WA 98065 Telephone: 425 396-2000 Fax: 425 396-4825

Contact information:

Ms. Rose DeGiacomo 425 396-2308 Telephone: 425 396-4825 Fax: February 18, 2004 Date summary prepared: Sonicare® Advance Toothbrush (Model 4900) Proprietary name of device: Toothbrush, powered Generic/classification name:

JEQ (Class 1, 21 C.F.R. § 872.6865) Product code (classification):

#### Legally Marketed Predicate Devices:

TX Electric Toothbrush (K921773)

Dental Air Force Home Dental Cleaning System (K001493)

## Device Description and Technological Characteristics:

Sonicare® Advance Model 4900 consists of three primary parts: brush head, power handle, and charger base. Central to the brush head are the resonator arm and the torsion bar. The brush head's bristles are mounted in a bristle plate on one end of the arm, two magnets are mounted on the other, and the torsion bar is perpendicular in the middle. A 261-Hz oscillating magnetic flux generated inside the power handle forces the magnets into motion, which then causes the resonator arm to oscillate about the torsion bar. The torsion bar, magnets and bristle plate have a resonant frequency close to the 261-Hz driving frequency, and this enables the system to produce a desired amplitude effect. The bristles vibrate in the direction perpendicular to the bristles' length and nominally perpendicular to the brush's axis.

The oscillating magnetic flux is generated by a stator coil in the power handle, in close proximity to the attachment of the brush head. The flux is controlled by solid-state electronics and powered from two nickel-cadmium rechargeable batteries elsewhere within the handle. Energy for recharging the batteries is transferred from the charger base through a system of induction

{1}------------------------------------------------

coils within the power handle and the charger base. Maximum power dissipation by the handle is 3 W.

The charger base runs on standard North American 120 V, 60 Hz power. Internal electronics convert the input power to a 65 kHz output signal for more efficient transfer of power to the handle.

#### Intended Use

For promotion of good oral hygiene, including the reduction of dental plaque and gingivitis,

#### Testing

Various models of the Sonicare® Advance toothbrushes (identical or closely similar in specifications to the Model 4900) have been studied in a total of seven clinical trials (six controlled). Endpoints evaluated in these trials include examination of soft tissues for evidence of injury and indices of gingival inflammation (six trials), as well as pocket probing depth and clinical attachment level (four trials). Collectively, the studies demonstrate that the Sonicares Advance toothbrushes reduce gingivitis.

#### Conclusions

The results from these tests support the safety and effectiveness of the 4900 model of the Sonicare® Advance toothbrush and its substantial equivalence to the predicate devices.

: :

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a bird-like figure, composed of three curved lines that suggest wings or feathers.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 0 2 2004

Ms. Rose DeGiacomo Corporate Counsel Philips Oral Healthcare 35301 S.E. Center Street Snoqualmie, Washington 98065

Re: K040416

Trade/Device Name: Sonicare® Advanced Toothbrush (Model 4900) Regulation Number: 872.6865 Regulation Name: Powered Toothbrush Regulatory Class: I Product Code: JEQ Dated: June 14, 2004 Received: June 18, 2004

Dear Ms. DeGiacomo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. DeGiacomo

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a handwritten string of characters, "K040416". The characters are written in black ink on a white background. The characters are all connected, and they are all the same size. There is a line drawn underneath the string of characters.

# INDICATIONS FOR USE STATEMENT

| Applicant:           | Philips Oral Healthcare, Inc.                          |
|----------------------|--------------------------------------------------------|
| 510(k) Number:       | K040416                                                |
| Device Name:         | Sonicare <sup>®</sup> Advanced Toothbrush (Model 4900) |
| Indications for Use: |                                                        |

For promotion of good oral hygiene, including the reduction of dental plaque and gingivitis.

## (PLEASE DO NOT WRITE BELOW THIS LINE)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use _____________________________________________________________________________________________________________________________________________________________

or

Over-the Counter Use _L

Susan Rums

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

510(k) Number:

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/JEQ/K040416](https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/JEQ/K040416)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
