← Product Code [EIH](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH) · K213722

# ICE Ceramics, ICE Stains, Fresco (K213722)

_Zirkonzahn Srl · EIH · Dec 14, 2022 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K213722

## Device Facts

- **Applicant:** Zirkonzahn Srl
- **Product Code:** [EIH](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH.md)
- **Decision Date:** Dec 14, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS DENTINE +, ICE CERAMICS DYNAMIC DENTINE, FRESCO ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and FRESCO DYNAMIC DENTINE are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays or veneers. ICE STAINS, ICE STAINS PRETTAU®, ICE STAINS GLAZE FLUO and ICE STAINS 3D are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays, onlays or veneers.

## Device Story

Dental glass ceramics (powders/pastes) used to coat, stain, and glaze final-sintered zirconia dental restorations; includes ICE Ceramics, ICE Stains, and Fresco product lines. Applied in layers by trained dental personnel in a laboratory setting; requires firing processes followed by manual grinding and polishing. Accessories (liquids/gels) adjust viscosity or refresh materials. Output is a customized, aesthetic dental restoration. Benefits include improved appearance and surface finish of zirconia-based dental prosthetics.

## Clinical Evidence

Bench testing only. Physical properties evaluated per ISO 6872 (flexural strength ≥ 50 MPa; chemical solubility < 100 µg/cm²). Biocompatibility established per ISO 10993-1 and ISO 10993-5.

## Technological Characteristics

Glass ceramic composition with metallic oxides and pigments. Supplied as powders, pastes, and liquids. Classified as Type 1, class 1b per ISO 6872. Biocompatibility per ISO 10993. No electronic components or software.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- Zirkonzahn Ice ([K061851](/device/K061851.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 14, 2022

Zirkonzahn srl Sandra Leitner Regulatory Affairs Responsible Via An der Ahr 7 Gais, BZ 39030 Italy

Re: K213722

Trade/Device Name: ICE Ceramics, ICE Stains, Fresco Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: November 10, 2022 Received: November 14, 2022

Dear Sandra Leitner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Bobak Shirmohammadi -S

For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213722

Device Name ICE Ceramics, ICE Stains and Fresco

#### Indications for Use (Describe)

ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS DENTINE +, ICE CERAMICS DYNAMIC DENTINE, FRESCO ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and FRESCO DYNAMIC DENTINE are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays or veneers.

ICE STAINS, ICE STAINS PRETTAU®, ICE STAINS GLAZE FLUO and ICE STAINS 3D are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays, onlays or veneers.

| Type of Use (Select one or both, as applicable)                                  |                                                                                |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 05 510(k) Summary - K213722

#### 510(k) Summary

## APPLICANT

Zirkonzahn srl Via An der Ahr 7 Gais, ITALY 39030 Phone: +39 0474 066 660 Fax: +39 0474 066 661 E-mail: info@zirkonzahn.com

#### CONTACT PERSON

Julian Steger Chief Executive Officer Zirkonzahn srl Via An der Ahr 7 Gais, ITALY 39030 Phone: +39 0474 066 660 Fax: +39 0474 066 661 E-mail: julian.steger@zirkonzahn.com

#### DATE SUMMARY PREPARED: December 13, 2022

#### DEVICE IDENTIFICATION

| Trade/Proprietary Name: | ICE Ceramics, ICE Stains and Fresco |
|-------------------------|-------------------------------------|
| Regulation Number:      | 21 CFR 872.6660                     |
| Common Name:            | Powder, Porcelain                   |
| Classification Name:    | Porcelain Powder for Clinical use   |
| Class:                  | II                                  |
| Product Code:           | EIH                                 |
| Panel:                  | Dental                              |

{4}------------------------------------------------

#### LEGALLY MARKETED PREDICATE DEVICE

| Company:       | Zirkonzahn srl |
|----------------|----------------|
| Device Name:   | Zirkonzahn Ice |
| Product Code:  | EIH            |
| 510(k) Number: | K061851        |

#### INDICATIONS FOR USE

ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS DENTINE +, ICE CERAMICS DYNAMIC DENTINE, FRESCO ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and FRESCO DYNAMIC DENTINE are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays, onlays or veneers.

ICE STAINS, ICE STAINS PRETTAU®, ICE STAINS GLAZE FLUO and ICE STAINS 3D are type I ceramics for coating, staining and glazing dental restorations made of final-sintered zirconia. They are conceived for patients who require dental restorations in zirconia, such as single crowns, bridges, inlays, onlays or veneers.

#### DEVICE DESCRIPTION

All devices included in this submission are dental glass ceramics. There are different

device types from three device groups:

#### ICE Ceramics Group:

- ICE Ceramics
- ICE Ceramics Tissue
- ICE Ceramics Dentine +
- . ICE Ceramics Dynamic Dentine

#### ICE Stains Group:

- . ICE Stains
- ICE Stains Prettau® ●
- ICE Stains 3D ●
- ICE Stains Glaze Fluo .

#### Fresco Group:

- Fresco ●
- Fresco Gingiva
- Fresco Enamel ●
- Fresco Dentine +
- Fresco Dynamic Dentine ●

The above mentioned devices are supplied in form of powders or pastes and are available in different colors and/ or in different quantity sizes. The users can choose among the different options according to their requirements and preferences. Furthermore specific accessories are supplied to optimize the working procedure:

{5}------------------------------------------------

ICE BUILD UP LIQUID is an accessory used for mixing of ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS DENTINE + and ICE CERAMICS DYNAMIC DENTINE. FRESCO LIQUID and FRESCO GEL are accessories used to change the viscosity of FRESCO ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and FRESCO DYNAMIC DENTINE products.

ICE STAIN LIQUID is an accessory used to freshen up ICE STAINS und ICE STAINS PRETTAU®; REFRESHING LIQUID 3D is an accessory used to freshen up ICE STAINS 3D; REFRESHING LIQUID GLAZE FLUO is an accessory used to freshen up ICE STAINS GLAZE FLUO.

All these devices are used by specifically trained personnel to manufacture customized dental restorations.

### DISCUSSION OF NON CLINCAL TESTS

Non clinical testing was performed to determine the physical properties of the subject device. Performance testing was conducted in accordance with the International Standard ISO 6872 and the results show that the devices are well suited for their intended use. Furthermore, biocompatibility was established in according to ISO 10993-1 and ISO 10993-5.

{6}------------------------------------------------

# COMPARISON TABLE

| Devices                            | New Devices:<br>ICE Ceramics, ICE Stains and Fresco<br>K213722                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate devices:<br>Zirkonzahn Ice<br>K061851                                                                                                                                                            | Comparison                                                                                                                                                                                                                                                        |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company                            | Zirkonzahn srl                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Zirkonzahn srl                                                                                                                                                                                             |                                                                                                                                                                                                                                                                   |
| Product Code                       | EIH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | EIH                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                              |
| Regulation<br>Number               | 872.6660                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 872.6660                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                              |
| Regulation Name                    | Porcelain powder for clinical use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Porcelain powder for clinical use                                                                                                                                                                          | Same                                                                                                                                                                                                                                                              |
| Indications for<br>use             | ICE CERAMICS, ICE CERAMICS TISSUE, ICE CERAMICS<br>DENTINE +, ICE CERAMICS DYNAMIC DENTINE, FRESCO<br>ENAMEL, FRESCO GINGIVA, FRESCO, FRESCO DENTINE + and<br>FRESCO DYNAMIC DENTINE are type I ceramics for coating,<br>staining and glazing dental restorations made of final-sintered<br>zirconia. They are conceived for patients who require dental<br>restorations in zirconia, such as single crowns, bridges, inlays, onlays<br>or veneers.<br>ICE STAINS, ICE STAINS PRETTAU®, ICE STAINS GLAZE FLUO<br>and ICE STAINS 3D are type I ceramics for<br>coating, staining and glazing dental restorations made of final-sintered<br>zirconia. They are conceived for patients who<br>require dental restorations in zirconia, such as single crowns, bridges,<br>inlays, onlays or veneers. | Metal free single posterior crowns,<br>multiple unit anterior<br>crowns/bridges, inlays, onlays<br>bonded dental restorations.                                                                             | Similar<br>Zirkonzahn Ice contains a<br>system of devices<br>(including zirconia blanks,<br>coloring liquids), not only<br>dental ceramics, but the<br>new as well as the<br>predicate devices are<br>intended to veneer /<br>individualize dental<br>frameworks. |
| Sizes of dental<br>frameworks      | All, dependent on zirconia device where applied                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | All, dependent on zirconia device<br>where applied                                                                                                                                                         | Same                                                                                                                                                                                                                                                              |
| Chemical<br>composition            | Glass ceramic, other metallic oxides, with pigments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Glass ceramic, other metallic oxides, with pigments                                                                                                                                                        | Similar                                                                                                                                                                                                                                                           |
| Biocompatibility                   | ISO 10993-1<br>ISO 10993-5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Not specified in K061851                                                                                                                                                                                   | Similar                                                                                                                                                                                                                                                           |
| Tested according<br>to ISO 6872    | Flexural strength ≥ 50 MPa<br>Chemical solubility < 100 µg/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Flexural strength ≥ 50 MPa<br>Chemical solubility < 100 µg/cm²                                                                                                                                             | Similar                                                                                                                                                                                                                                                           |
| Classification<br>acc. to ISO 6872 | Type 1, class 1b                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Type 1, class 1b                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                              |
| Principles of<br>Operation         | The devices can be applied in several layers on final-sintered<br>zirconia structures followed by firing processes. Subsequently,<br>the dental structures are finalized by manual grinding and<br>polishing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | The devices can be applied in several layers on final-sintered<br>zirconia structures followed by firing processes. Subsequently, the<br>dental structures are finalized by manual grinding and polishing. | Same                                                                                                                                                                                                                                                              |
| Shades                             | Dentine shades (Vita dentine shades A1 –D4 etc.)<br>Intensive colors<br>Glazes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Dentine shades (Vita dentine shades A1 -D4 etc.)<br>Intensive colors<br>Glazes                                                                                                                             | Similar                                                                                                                                                                                                                                                           |
| Form                               | Powders, pastes and liquids                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Powders, pastes and liquids                                                                                                                                                                                | Same                                                                                                                                                                                                                                                              |
| Storage<br>Conditions              | In tightly sealed original packaging with a maximal temperature<br>of 35 °C for ICE Stains, Fresco and accessory liquids.<br><br>If stored too cold, the liquid may freeze and must be defrosted<br>before use.<br><br>If stored too warm, the pastes may dry out.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Storage temperature between<br>10 °C and 30 °C                                                                                                                                                             | Differences do not have<br>an effect on final medical<br>device, therefore<br>difference not significant.                                                                                                                                                         |

{7}------------------------------------------------

{8}------------------------------------------------

# SUBSTANTIAL EQUIVALENCE DISCUSSION

The new devices and the predicate devices are similar in function and intended use. The differences noted do not raise any new questions about safety and effectiveness. Therefore, Zirkonzahn concludes that the proposed devices are substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K213722](https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K213722)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
