← Product Code [EIH](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH) · K193055

# Purzir Dental Zirconia (K193055)

_3DBiocad · EIH · May 13, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K193055

## Device Facts

- **Applicant:** 3DBiocad
- **Product Code:** [EIH](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH.md)
- **Decision Date:** May 13, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Pürzir Dental Zirconia blanks are used for dental restorations using different CAD / CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.

## Device Story

Pürzir Dental Zirconia consists of zirconia oxide discs in various shades and dimensions. Used by dental laboratories and professionals to fabricate dental restorations (crowns, bridges, veneers, inlays, onlays). Input is anatomical rendering of patient teeth; processed via CAD/CAM or manual milling machines to create custom restorations. Final restorations require sintering at temperatures > 1500 °C. Device provides a durable, biocompatible material for dental prosthetics, benefiting patients by restoring tooth function and aesthetics.

## Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-1, -3, -5, -10, and -11. Non-clinical performance testing conducted per ISO 6872, including chemical solubility, flexural strength, and coefficient of linear thermal expansion.

## Technological Characteristics

Zirconia oxide (ZrO2, Y2O3, HfO2, Al2O3) with optional inorganic pigments. Disc form factor. Sintering required (> 1500 °C). Tested to ISO 6872 (dentistry ceramic materials) and ISO 10993 (biocompatibility). Non-sterile, single-use.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- Upcera Dental Zirconia Blank and Dental Zirconia Pre-Shaded Blank ([K141724](/device/K141724.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 13, 2020

3DBioCAD % Stuart Goldman Sr. Consultant RA/QA EMERGO Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, Texas 78746

Re: K193055

Trade/Device Name: Purzir Dental Zirconia Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: February 7, 2020 Received: February 13, 2020

Dear Stuart Goldman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193055

Device Name Pürzir Dental Zirconia

Indications for Use (Describe)

Pürzir Dental Zirconia blanks are used for dental restorations using different CAD / CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# Pürzir Dental Zirconia

# 1. Submission Sponsor

3DBioCAD 1525 N 4th St. Renton, WA 98057 USA Contact: Gyu D. Cho Title: President

#### 2. Submission Correspondent

Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Email: LST.AUS.ProjectManagement@ul.com Contact: Stuart R. Goldman Title: Sr. Consultant RA/QA

#### 3. Date Prepared

May 11, 2020

#### 4. Device Identification

| Trade/Proprietary Name: | Pürzir Dental Zirconia            |
|-------------------------|-----------------------------------|
| Common/Usual Name:      | Powder, Porcelain                 |
| Classification Name:    | Porcelain powder for clinical use |
| Regulation Number:      | CFR 872.6660                      |
| Product Code:           | EIH                               |
| Class:                  | II                                |
| Classification Panel:   | Dental                            |

## 5. Legally Marketed Predicate Device

| Device name:   | Upcera Dental Zirconia Blank and Dental Zirconia Pre-Shaded Blank |
|----------------|-------------------------------------------------------------------|
| 510(k) number: | K141724                                                           |
| Manufacturer:  | Liaoning Upcera Co., Ltd.                                         |

{4}------------------------------------------------

#### 6. Indication for Use

Pürzir Dental Zirconia blanks are used for dental restorations using different CAD / CAM or manual milling machines. All blanks are processed through dental laboratories or by dental professionals.

#### 7. Device Description

Pürzir Dental Zirconia are disc shaped zirconia oxide blanks made available in different versions and chemical compositions of various colors, shades and dimensions and sold to dental labs that will further process the discs by milling them to make final dental restorations such as crowns, bridges, veneers, inlays and on lays based on the anatomical rendering of the patient's teeth using Computer Aided Drafting / Computer Aided Machining (CAD / CAM) equipment. The subject device is made available in six versions: 1. HS (High Strength), 2. HT (High Translucent), 3. HT Plus (High Translucent Plus), 4. ST (Super Translucent), 5. MMT x2 (Max Multi Translucent), and 6. MMT Plus x2 (Max Multi Translucent Plus).

#### 8. Substantial Equivalence Comparison

Table 5-1 compares Pürzir Dental Zirconia discs to the predicate device (K141724) with respect to regulatory information, intended use, indications for use, technological characteristics, and non-clinical performance testing and provides detailed information regarding the basis for the determination of substantial equivalence between the subject and predicate device.

| Attributes          | Subject Device                                                                                                                                                                                                          | Predicate Device                                                                                                                                                                                                                                               | Similarities / Differences    |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Device Name         | Pürzir Dental Zirconia                                                                                                                                                                                                  | Upcera Dental Zirconia Blank and<br>Dental Zirconia Pre-Shaded Blank                                                                                                                                                                                           | -                             |
| Manufacturer        | 3DBioCAD                                                                                                                                                                                                                | Liaoning Upcera                                                                                                                                                                                                                                                | -                             |
| 510(k) #            | K193055                                                                                                                                                                                                                 | K141724                                                                                                                                                                                                                                                        | -                             |
| Product Code        | EIH                                                                                                                                                                                                                     | EIH                                                                                                                                                                                                                                                            | Same                          |
| Regulation          | CFR 872.6660                                                                                                                                                                                                            | CFR 872.6660                                                                                                                                                                                                                                                   | Same                          |
| Regulation Name     | powder, porcelain                                                                                                                                                                                                       | powder, porcelain                                                                                                                                                                                                                                              | Same                          |
| Class               | II                                                                                                                                                                                                                      | II                                                                                                                                                                                                                                                             | Same                          |
| Review Panel        | Dental                                                                                                                                                                                                                  | Dental                                                                                                                                                                                                                                                         | Same                          |
| Indications for Use | Pürzir Dental Zirconia blanks are<br>used for dental restorations using<br>different CAD / CAM or manual<br>milling machines. All blanks are<br>processed through dental<br>laboratories or by dental<br>professionals. | Upcera Dental Zirconia Blank and<br>Dental Zirconia Pre-Shaded Blank are<br>used for dental restorations using<br>different CAD / CAM or manual<br>milling machines. All blanks are<br>processed through dental<br>laboratories or by dental<br>professionals. | Same                          |
| Form                | Disc form.                                                                                                                                                                                                              | Block, disc and rod form.                                                                                                                                                                                                                                      | The subject<br>device is only |

#### Table 5-1 – Substantial Equivalence Comparison of Pürzir Dental Zirconia vs. Predicate Device (K141724)

{5}------------------------------------------------

| Attributes               | Subject Device                                                                                                                                                                   | Predicate Device                                                                                                                                                                 | Similarities / Differences   |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
|                          |                                                                                                                                                                                  |                                                                                                                                                                                  | made available in disc form. |
| Dimensions               | Various                                                                                                                                                                          | Various                                                                                                                                                                          | Similar                      |
| Material                 | Regular:<br>Zirconia (ZrO2 + Y2O3 + HfO2 + Al2O3 ≥ 99.0%)<br>Pre-Shaded:<br>Zirconia (ZrO2 + Y2O3 + HfO2 + Al2O3 ≥ 98.0%)<br>Inorganic pigments (Fe2O3, Pr2O3, and Er2O3 < 2.0%) | Regular:<br>Zirconia (ZrO2 + Y2O3 + HfO2 + Al2O3 ≥ 99.0%)<br>Pre-Shaded:<br>Zirconia (ZrO2 + Y2O3 + HfO2 + Al2O3 ≥ 98.0%)<br>Inorganic pigments (Fe2O3, Pr2O3, and Er2O3 < 2.0%) | Same                         |
| Color                    | None, and Pre-shaded (for pre-shaded series).                                                                                                                                    | None, and Pre-shaded (for pre-shaded series).                                                                                                                                    | Same                         |
| Processing               | Sintering at temperature > 1500 °C                                                                                                                                               | Sintering at temperature > 1500 °C                                                                                                                                               | Same                         |
| Conditions of Use        | Professional use for the fabrication of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers.                                                 | Professional use for the fabrication of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers.                                                 | Same                         |
| Single Use               | Yes                                                                                                                                                                              | Yes                                                                                                                                                                              | Same                         |
| Supplied Sterile         | No                                                                                                                                                                               | No                                                                                                                                                                               | Same                         |
| Packaging                | Single blank (disc) per box.                                                                                                                                                     | Single blank (disc) per box.                                                                                                                                                     | Same                         |
| Shelf Life               | 5 years from date of manufacture                                                                                                                                                 | 5 years from date of manufacture                                                                                                                                                 | Same                         |
| Biocompatibility Testing | Tested to ISO 10993-1                                                                                                                                                            | Tested to ISO 10993-1                                                                                                                                                            | Conforms                     |
| Performance Testing      | Tested to ISO 6872                                                                                                                                                               | Tested to ISO 6872                                                                                                                                                               | Conforms                     |

## 9. Summary of Non-Clinical Performance Testing

As part of demonstrating substantial equivalence of Pürzir Dental Zirconia discs to the predicate device, 3DBioCAD tested final finished samples of their device in accordance with the applicable parts of the following FDA guidance documents and voluntary recognized consensus standards. Results confirm that the design inputs and performance specifications for the subject device have been met.

- . Biocompatibility Testing:
	- o FDA Guidance Document Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process

{6}------------------------------------------------

- ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and testing within o a risk management process:
	- " ISO 10993-3 (genotoxicity)
	- 트 ISO 10993-5 (cytotoxicity)
	- " ISO 10993-10 (skin sensitization and irritation)
	- I ISO 10993-11 (toxicity)
- Non-Clinical Performance Testing:
	- o ISO 6872, Dentistry Ceramic materials
		- I chemical solubility
		- 트 flexural strength
		- coefficient of linear thermal expansion

#### 10. Clinical Testing

Clinical testing on the subject device have not been performed.

## 11. Statement of Substantial Equivalence

Pürzir Dental Zirconia has the same intended use and indications for use as Upcera Dental Zirconia Blank and Dental Zirconia Pre-Shaded Blank (K141724). Any minor differences in the technological features of the subject device when compared to the predicate device have been successfully evaluated through nonclinical performance testing and other verification activities such that the information submitted to the FDA demonstrates that Pürzir Dental Zirconia is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K193055](https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K193055)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
