This device is a ceramic-based material that is used for prefabricated and patient-matched dental restorations such as inlays, artificial teeth, crowns, or bridges (up to maximum three-unit bridges) and processed by dental CAD/CAM.
Device Story
Eclipse Kiss Zirconia is a ceramic-based dental material (porcelain) supplied as blocks or disks; used for fabrication of patient-matched dental restorations including crowns, copings, and bridges (up to three units). Material is processed via dental CAD/CAM systems; milled into precise units based on dentist/physician prescription. Final restorations may undergo additional ceramic layering for aesthetics. Used in dental clinics/laboratories; operated by dental professionals. Provides biocompatible, durable prosthetic support for edentulous patients.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility evaluation per ISO 10993 and comparison of physical/chemical properties (flexural strength, density, thermal expansion) against predicate.
Technological Characteristics
Ceramic-based material (ZrO2+HfO2+Y2O3 ≥ 99 wt%). Form factor: blocks and disks. Sintering temperature: 1450 °C. Flexural strength: ~749 MPa. Density: > 6.0 g/cm³. Coefficient of thermal expansion: 10.3 x 10⁻⁶K⁻¹. Non-sterile. Processed via CAD/CAM.
Indications for Use
Indicated for partial and fully edentulous patients requiring dental restorations including inlays, artificial teeth, crowns, or bridges (up to maximum three-unit bridges).
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 26, 2017
AMS % Ho Dong Yang CEO Onbix Corporation #821 Samil. Plaza, 837 -26 Yeuksam-dong Gangnam-gu Seoul, 135768 REPUBLIC OF KOREA
Re: K163238 Trade/Device Name: Eclipse Kiss Zirconia Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: January 31, 2017 Received: February 6, 2017
Dear Ho Dong Yang: This letter corrects our substantially equivalent letter of May 4, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mary S. Runner -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Change Control Table, Change History
# Change Control Table
| Version | Document Author | Document Approver | Date Approved |
|---------|---------------------|---------------------|---------------|
| 1.00 | Name, Title, Office | Name, Title, Office | MM/DD/YYYY |
Complete Change Control Table (all versions) retained in SWIFT Docs.
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163238
Device Name Eclipse Kiss Zirconia
Indications for Use (Describe)
This device is a ceramic-based material that is used for prefabricated and patient-matched dental restorations such as inlays, artificial teeth, crowns, or bridges (up to maximum three-unit bridges) and processed by dental CAD/CAM.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
| Submitter Information: | AMS<br>#530 Na-Dong Machzone, 117, Hwanggeum-ro<br>Yangchon-eup, Gimpo-si, Gyeonggi-do, Korea (10048) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ho Dong, Yang<br>Onbix Corporation<br>#821 Samil Plaza, 837-26 Yeuksam-dong<br>Gangnam-gu, Seoul, 135,768, Korea<br>Tel: *82-2-566-3360 / Fax: *82-2-6280-3360<br>Email: onbix@naver.com |
| Date Summary Prepared: | Nov 14, 2016 |
#### Device Information:
| Trade Name(s): | Eclipse Kiss Zirconia |
|----------------------------|-----------------------|
| Classification Name: | dental porcelain |
| Panel: | dental |
| Product Code & Regulation: | EIH / 872.6660 |
#### Predicate Device Information:
K150196 / DD Cubex2
#### Device Description:
The Eclipse Kiss Zirconia Dental Material is an integrated system of dental ceramic which support dental restorations and which are milled into prosthetic devices such as Crowns, Bridges, and Copings for partial and fully edentulous patients. The material can be milled into a variety of dental restorations to meet the specific needs of individual dental patients. Each patient is named by a physician or dentists written prescription specifying the type of restoration to be produced.
Eclipse Kiss Zirconia is intended to be used as a biocompatible material to be milled into precise customized dental restoration units for either crowns, copings or bridges (up to maximum three unit bridges); which may be further processed by the addition of ceramic layering to create aesthetic lifelike restorations.
Eclipse Kiss Zirconia is intended to be used as prescribed by a qualified and licensed physician or dentist by a written prescription which names the patient for which each individual dental restoration is to be produced.
#### Intended Use:
This device is a ceramic based material (porcelain) that is used for prefabricated and patient-matched dental restorations such as inlays, artificial teeth, crowns, or bridges (up to maximum three unit bridges) and processed by dental CAD/CAM
#### Comparison to Predicate Device(s):
This device is equivalent to the predicate devices in its intended use and technological characteristics, including:
#### Performance properties
The device was conducted in compliance with the following Standard
{5}------------------------------------------------
ISO 10993 biological evaluation of medical devices
## Summary of the technological characteristics compared to the predicate device Eclipse Kiss Zirconia is substantially equivalent to the predicate device in its technological
characteristics stated in the comparison table provided below
| | Eclipse Kiss Zirconia | DD Cubex2 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | AMS | Dental Direkt Gmbh |
| 510(k) Number | not yet assigned | K150196 |
| Product code | EIH | EIH |
| Regulation<br>number | 872.6660 | 872.6660 |
| Indications for<br>use | Ceramic based material (porcelain) that<br>is used for prefabricated and patient-<br>matched dental restorations such as inlays,<br>artificial teeth, crowns or bridges (up to<br>maximum three unit bridges)and processed<br>by dental CAD/CAM | Dental blanks made from DD cubeX²<br>are indicated for crowns, multi-unit<br>bridges (up to a maximum of three<br>elements) and inlay bridges.<br>Applications include both, anterior<br>and posterior bridges. |
| Chemical<br>composition | ZrO2+HfO2+Y2O3 ≥ 99 wt%<br>Y2O3 < 10 wt%<br>Fe2O3 ≤ 0.0002 wt%1<br>NiO ≤ 0.00008 wt% | ZrO2+HfO2+Y2O3 ≥ 99 wt%<br>Y2O3 < 10 wt%<br>HfO2 < 0.2 wt%<br>Al₂O₃ < 0.1 wt% |
| shape | Blocks, disks | Disks |
| Colour | White | White |
| sterility | Non-sterile | Non-sterile |
| Sintering<br>temperature | 1450 °C | 1450 °C |
| Performance<br>properties | Flexural strength [MPa] ~ 749<br>Density [g/cm³] > 6.0<br>Coefficient of thermal expansion [10⁻⁶K⁻¹]:<br>10.3 | Flexural strength [MPa] > 720<br>Density [g/cm³] > 6.0<br>Coefficient of thermal expansion [10⁻⁶K⁻¹]: 10 |
| Biocompatibility | Suitable | Suitable |
Comparison table is as follows
## Non-Clinical Study performance
Not applicable
### Conclusion
Eclipse Kiss Zirconia is substantially equivalent to the predicate device in its intended use and technological characteristics
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.