← Product Code [EIH](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH) · K152103

# IPS Style Ceram, IPS Style Ceram One, IPS Style Press (K152103)

_Ivoclar Vivadent, AG · EIH · May 20, 2016 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K152103

## Device Facts

- **Applicant:** Ivoclar Vivadent, AG
- **Product Code:** [EIH](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH.md)
- **Decision Date:** May 20, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Intended Use: IPS Style Press - Pressing on metal frameworks and superstructures for the fabrication of full-contour crowns and bridges (staining technique) - Pressing on metal frameworks and superstructures for the fabrication of partially anatomic crowns and bridges (cut-back technique) - Pressing of ceramic margins (metal-ceramic) - Pressing of gingiva components (metal-ceramic) - Processing on the most popular dental alloys in the CTE range of 13.8 14.5 10-6/K - (25 500°C) (Cu< 1%, Ag < 10%) - -- Veneering with the layering materials of IPS Style in the conventional layering technique - Characterization with IPS Ivocolor Shade and Essence stains - Glazing with IPS Ivocolor Glaze Intended Use: IPS Style Ceram and IPS Style Ceram Onc - Conventional multi-layer veneering ceramic for the most popular dental alloys (including electroplating) in the CTE range of 13.8 - 15.2 x 10-6/K (25 - 500°C) (IPS Style Ceram) - One-layer veneering ceramic for the most popular dental alloys (including electroplating) in the CTE range of 13.8 - 15.2 x 10-6/K (25 - 500°C) (IPS Style Ceram One) - Veneers on refractory die material (only IPS Style Ceram) - Characterization and veneering of IPS Style Press restorations - Characterization with IPS Ivocolor Shade and Essence stains - Glazing with IPS Ivocolor Glaze

## Device Story

IPS Style is a dental ceramic system comprising powders for conventional veneering and ingots for press-on techniques. Used by dental technicians in laboratories to fabricate metal-ceramic restorations (crowns, bridges, veneers). Inputs include dental alloy frameworks; device transforms these via layering or pressing techniques followed by firing. Outputs are finished dental restorations characterized with stains and glazes. Benefits include aesthetic restoration of tooth structure and function. System integrates conventional veneering and press-on technologies, allowing layering of IPS Style Ceram powders onto IPS Style Press materials. Biocompatible silica-based glass ceramic.

## Clinical Evidence

Bench testing only. Evaluated per ISO 10993-1:2009 for cytotoxicity, genotoxicity, sensitization, systemic toxicity, implantation, and radioactivity. Physical properties tested per ISO 6872 (flexural strength) and ISO 9693 (metal-ceramic bond/debonding strength). Results meet requirements and are comparable to predicates.

## Technological Characteristics

Silica-based dental glass ceramic containing Al, Li, Zr, Y, K, P, Mg, Ce, B, Ca, Na, Ti, Sr, F, Zn, La. Thermal expansion (CTE) ranges 13.8-15.2 x 10^-6/K. Form factors include powders (Ceram/Ceram One) and ingots (Press). Processing involves firing/pressing on metal frameworks. Tested to ISO 6872 and ISO 9693.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- IPS 99 One and IPS 99 Ceram ([K121359](/device/K121359.md))
- IPS InLine PoM System ([K071848](/device/K071848.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular text element and a symbol. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The symbol is a stylized representation of three human profiles facing right, layered on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 20, 2016

Ivoclar Vivadent Ag % Ms. Donna Hartnett Director Regulatory Affairs Ivoclar Vivadent, Inc. 175 Pineview Drive Amherst. New York 14228

Re: K152103

Trade/Device Name: Ips Style Ceram, Ips Style Ceram One, Ips Style Press Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: Class II Product Code: EIH Dated: April 8, 2016 Received: April 11, 2016

Dear Ms. Hartnett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang
-s

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

Device Name

IPS Style Ceram, IPS Style Ceram One, and IPS Style Press

Indications for Use (Describe)

Intended Use: IPS Style Press

- Pressing on metal frameworks and superstructures for the fabrication of full-contour crowns and bridges (staining technique)

- Pressing on metal frameworks and superstructures for the fabrication of partially anatomic crowns and bridges (cut-back technique)

- Pressing of ceramic margins (metal-ceramic)
- Pressing of gingiva components (metal-ceramic)
- Processing on the most popular dental alloys in the CTE range of 13.8 14.5 10-6/K
- (25 500°C) (Cu< 1%, Ag < 10%)
- -- Veneering with the layering materials of IPS Style in the conventional layering technique
- Characterization with IPS Ivocolor Shade and Essence stains
- Glazing with IPS Ivocolor Glaze

Intended Use: IPS Style Ceram and IPS Style Ceram Onc

- Conventional multi-layer veneering ceramic for the most popular dental alloys (including electroplating) in the CTE range of 13.8 - 15.2 x 10-6/K (25 - 500°C)

(IPS Style Ceram)

- One-layer veneering ceramic for the most popular dental alloys (including electroplating)

in the CTE range of 13.8 - 15.2 x 10-6/K (25 - 500°C) (IPS Style Ceram One)

- Veneers on refractory die material (only IPS Style Ceram)
- Characterization and veneering of IPS Style Press restorations
- Characterization with IPS Ivocolor Shade and Essence stains
- Glazing with IPS Ivocolor Glaze

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo for Ivoclar Vivadent. The logo features the company name in a bold, blue sans-serif font. Above the name is an arc of green and blue dots and squares. Below the name is the company's slogan, "passion vision innovation," in a smaller, red sans-serif font.

#### IPS Style Ceram-IPS Style Ceram One-IPS Style Press

| Contact:             | Donna Marie Hartnett                                                                         |
|----------------------|----------------------------------------------------------------------------------------------|
| Company:             | Ivoclar Vivadent, AG<br>Bendererstrasse 2, Schaan, FL-9494, Liechtenstein<br>+423-235-3535   |
| Date Prepared:       | May 16, 2016                                                                                 |
| Proprietary Name:    | IPS Style Ceram<br>IPS Style Ceram One<br>IPS Style Press                                    |
| Classification Name: | Powder, Porcelain for clinical use<br>CFR Regulation (872.6660)<br>(Classification Code EIH) |
| Predicate Device:    | IPS 99 One and IPS 99 Ceram (K121359)<br>IPS InLine PoM System (K071848)                     |

#### Device Description:

IPS Style comprises both materials for the conventional veneering technique and the press technique.

- . IPS Style Ceram: is available as Opaquer 870 in 42 Shades, Opaquer 960 in 42 shades, Margin Powder in 11 Shades, Dentin porcelain in 32 Shades, deep Dentin porcelain in 16 Shades, Incisal Porcelain in 10 Shades, Transparent Porcelain 7 Shades, Mamelon in 3 shades, Occlusal Dentin 2 Shades, Incisal Edge 1 Shade, Opal Effect in 6 Shades, Special Incisal 2 Shades, Inter Incisal 1 Shades, Cerv. Transparent 4 Shades, Add on Ceramic 4 Shades, Bleach Ceramic in 14 Shades, Gingiva Porcelain in 11 Shades and 4 Liquids.
- IPS Style Ceram One: is available in 16 Shades and Shades Guides .
- . IPS Style Press: Press Ingots LT is available in 8 Shades, Press Ingot HT in 8 Shades and Press Ingot Gingiva in 1 Shade.
- . IPS Style Liquids including Build-up and Opaquer which burn out after firing.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is the logo for Ivoclar Vivadent. The logo has the company name in a large, bold, sans-serif font, with "Ivoclar" on top and "Vivadent" below it. Above the company name is a series of green and blue dots and squares. Below the company name is the company's slogan, "passion vision innovation", in a smaller, red, sans-serif font.

IPS Style Ceram-IPS Style Ceram One-IPS Style Press

#### Intended Use: IPS Style Press

– Pressing on metal frameworks and superstructures for the fabrication of full-contour crowns and bridges (staining technique)

– Pressing on metal frameworks and superstructures for the fabrication of partially anatomic crowns and bridges (cut-back technique)

- Pressing of ceramic marqins (metal-ceramic)
- Pressing of gingiva components (metal-ceramic)
- Processing on dental alloys in the CTE range of 13.8 14.5 10-6/K
- (25 500°C) (Cu< 1%, Ag ≤ 10%)
- Veneering with the layering materials of IPS Style in the conventional layering technique
- Characterization with IPS Ivocolor Shade and Essence stains
- Glazing with IPS Ivocolor Glaze

#### IPS Style Ceram and IPS Style Ceram One

- Conventional multi-layer veneering ceramic for dental alloys (including electroplating) in the CTE range of 13.8 - 15.2 x 10-6/K (25 - 500°C) (IPS Style Ceram)

- One-layer veneering ceramic for dental alloys (including electroplating) in the CTE range of 13.8 - 15.2 x 10-6/K (25 - 500°C) (IPS Style Ceram One)

- Veneers on refractory die material (only IPS Style Ceram)
- Characterization and veneering of IPS Style Press restorations
- Characterization with IPS Ivocolor Shade and Essence stains
- Glazing with IPS Ivocolor Glaze

Comparison to Predicate: The predicate devices to which IPS Style Ceram and IPS Style Ceram One has been compared are IPS 99 One and IPS 99 Ceram (K121359) and IPS Style Press has been compared to IPS InLine PoM System (K071848). IPS Style Ceram and IPS Style Ceram One are both ceramic systems for the fabrication of dental restorations on a dental alloy framework. For this application, IPS Style Ceram and IPS Style Ceram One have been compared to its predicate with regard to the chemical compositions, physical properties, operating principals and indications for use. The comparison shows that IPS Style Ceram and IPS Style Ceram One are substantially equivalent to the predicate device.

Biocompatibility: The device was evaluated in connection with ISO 10993-1:2009 Biocompatibility evaluation of medical device – Part 1 and tested for cytotoxicity and genotoxicity and the device possessed no cytotoxic potential and is considered to be nonmutagenic. The device was also evaluated for sensitization, acute and subchronic systemic toxicity, implantation and radioactivity. Based on these results, and a review of cytotoxicity and toxicological data of the predicate device is not expected to cause any adverse mucosal reactions and does not present a toxicological risk for the patient or user.

{5}------------------------------------------------

# passion vision innovation

#### IPS Style Ceram-IPS Style Ceram One-IPS Style Press

|                     | Predicate devices<br>K121359 & K071848                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | IPS Style                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications         | The IPS 99 system contains shades, stains and glaze materials used for the characterization of metal ceramic restorations. <p>In addition:</p> <p>IPS 99 Ceram is a conventional multi-layer veneering ceramic for the fabrication of metal-ceramic restorations using dental alloys in the CTE range of 13.8-15.0 x10<sup>-6</sup>/K (25-500°C).</p> <p>IPS 99 One is a one-layer veneering ceramic for the fabrication of metal-ceramic restorations using dental alloys in the CTE range of 13.8-15.0 x10<sup>-6</sup>/K (25-500°C).</p> <p>IPS 99 is intended to be used for inlays, onlays, veneers and anterior/posterior PFM crowns.</p> | IPS Style Ceram and IPS Style Ceram One is a ceramic system for the fabrication of dental restorations on a dental alloy framework. <p>- Conventional multi-layer veneering ceramic for dental alloys (including electroplating) in the CTE range of 13.8 – 15.2 x 10<sup>-6</sup>/K (25 – 500°C) (IPS Style Ceram)</p> <p>- One-layer veneering ceramic for dental alloys (including electroplating) in the CTE range of 13.8 – 15.2 x 10<sup>-6</sup>/K (25 – 500°C) (IPS Style Ceram One)</p> <p>– Veneers on refractory die material (only IPS Style Ceram)</p> <p>- Characterization and veneering of IPS Style Press restorations</p> <p>- Characterization with IPS Ivocolor Shade and Essence stains</p> <p>- Glazing with IPS Ivocolor Glaze</p> |
|                     | IPS InLine PoM is a leucite ceramic system for pressing fully anatomical restorations on metal crown and bridge frameworks after the application of opaque, mainly in the posterior region. Alloy CTE range: 13.8-14.5 x10 <sup>-6</sup> /K (25-500°C). <p>IPS InLine PoM should be used only with dental alloys having Ag&lt;10%.</p>                                                                                                                                                                                                                                                                                                          | IPS Style Press; <p>Pressing on metal frameworks and superstructures for the fabrication of full-contour crown and bridges (staining technique).</p> <p>Pressing on metal frameworks and superstructures for the fabrication of partially anatomic crowns and bridges (cut-back technique).</p> <p>Pressing of ceramic margins (metal-ceramic).</p> <p>Pressing of gingiva components (metal-ceramic).</p> <p>Processing on dental alloys in the CTE range of 13.8-14.5 10<sup>-6</sup>/K (25-500°C) Cu&lt;1%, Ag&lt;10%).</p> <p>Veneering with the layering materials IPS Style in the conventional layering technique.</p> <p>Characterization with IPS Ivocolor Shade and Essence stains.</p> <p>Glazing with IPS Ivocolor Glaze.</p>                 |
| Summary Indications | The main difference between IPS Style and the IPS 99 range is that the IPS Style Ceram powders can be layered on IPS Style Press materials. In contrast, layering of IPS InLine Dentin, Incisal, Deep Dentin, Margin, Impulse and Gingiva materials on IPS InLine PoM was not included in the indication. The new materials have been specifically developed to be used together and the processing steps are compatible (e.g. firing temperatures).                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Ivoclar Vivadent. The logo features the company name in a stylized font, with "Ivoclar" in green and "Vivadent" in blue. Above the company name is an arc of green and blue dots. Below the company name is the slogan "passion vision innovation" in a smaller font.

#### IPS Style Ceram-IPS Style Ceram One-IPS Style Press

|                                | Predicate devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | IPS Style                                                                                                                                                                                                                                                                                                           |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Working principle              | IPS 99 is used as a conventional veneering<br>material on dental alloy frameworks.<br>IPS InLine PoM is used in the press-on<br>technique on dental alloy frameworks.<br>The products are available in a variety of<br>shades and can be characterized using<br>shades, stains and glaze in the same product<br>family.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | IPS Style includes materials for conventional<br>veneering and pressing-on dental alloy<br>frameworks.<br>The products ae available in a variety of<br>shades.<br>They are characterized using shades, stains<br>and glaze in the IPS Ivocolor universal system                                                     |
| Summary Working<br>Principle   | IPS Style combines conventional veneering and press-on technology in one product family and<br>one 510k application. The products have been developed together and in particular the IPS<br>Style Ceram powders can be layered on IPS Style Press.<br>In contrast to the predicates, IPS Style does not include shade, stain and glaze materials. It is<br>intended that the new universal products IPS Ivocolor are used.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                     |
| Summary Composition            | The key ingredients, is silica-based dental glass ceramic including the chemical elements:<br>aluminum, lithium, zirconium, yttrium, potassium, phosphorous, magnesium, cerium, boron,<br>calcium, sodium, titanium., strontium, fluoride, zinc, lanthanum.<br>The device was evaluated in connection with ISO 10993-1:2009 Biocompatibility evaluation of<br>medical device – Part 1 and tested for cytotoxicity and genotoxicity and the device possessed<br>no cytotoxic potential and is considered to be non-mutagenic. Based on these results, and a<br>review of cytotoxicity and toxicological data of the predicate devices, the device is not expected<br>to cause any adverse mucosal reactions.<br>The biocompatibility of the new formulation was fully assessed and is equivalent to IPS 99.<br>There are some minor differences to the liquids compositions but this only affects handling as<br>the liquids burn out during firing. |                                                                                                                                                                                                                                                                                                                     |
| Physical Properties            | The product has been tested to ISO 6872 and<br>meets the requirements. The product is also<br>tested under ISO 9693 for Debonding/crack<br>initiation strength and meets the requirements.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The product has been tested to ISO 6872 and<br>meets the requirements. IPS Style Ceram,<br>IPS Style Ceram One and IPS Style Press are<br>the same Type and Class material as the<br>predicate. The product is also tested under<br>ISO 9693 for Debonding/crack initiation<br>strength and meets the requirements. |
| Summary Physical<br>properties | The specified flexural strength of IPS Style Press is lower than that of IPS InLine PoM.<br>However the new value fulfils the requirements of the product-specific standard. The test<br>results for flexural strength are significantly above the specification and similar to those of the<br>predicate device IPS InLine PoM.<br>In all other aspects the new products lie in the same ranges as the predicates. The variations<br>are minimal. In addition the new products fulfill the requirements for metal-ceramic bond given<br>in the product-specific standard.                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                     |

#### Conclusion:

The main difference between IPS Style and the IPS 99 range is that the IPS Style Ceram powders can be layered on IPS Style Press materials. In contrast, layering of IPS InLine Dentin, Incisal, Deep Dentin, Margin, Impulse and Gingiva materials on IPS InLine PoM was not included in the indication. It combines conventional veneering and press-on technology in one product family.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Ivoclar Vivadent. The logo is blue and green, with the words "Ivoclar Vivadent" in a large, bold font. Below the company name is the slogan "passion vision innovation" in a smaller font. Above the company name is a series of green and blue dots and squares.

IPS Style Ceram-IPS Style Ceram One-IPS Style Press

Therefore IPS Style Ceram and IPS Style Ceram One and are substantially equivalent to the predicate device, IPS 99 One and IPS 99 Ceram. IPS Style Press is substantially equivalent to the predicate device, IPS InLine PoM System.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K152103](https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIH/K152103)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
