K042638 · Aurident, Inc. · EIH · Dec 17, 2004 · Dental
Device Facts
Record ID
K042638
Device Name
RADIANCE PORCELAIN
Applicant
Aurident, Inc.
Product Code
EIH · Dental
Decision Date
Dec 17, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
Radiance Porcelain is intended for the fabrication of dental restorations, crowns, bridges, veneers, inlays, and onlays, using either dental alloys or its pressable component, Press-I-Dent.
Device Story
Radiance Porcelain is a sodium/potassium alumina silicate leucite glass ceramic system used by dental laboratories to fabricate dental restorations (crowns, bridges, veneers, inlays, onlays). The system functions either by application over dental alloys or via a pressable component, Press-I-Dent. It is a low-fusing porcelain manufactured from fritted glass compositions, designed to exhibit lower opposing occlusal wear compared to traditional feldspar-based porcelains. The device is used in a laboratory setting to create custom dental prosthetics, which are then placed by a clinician to restore patient dentition.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and historical use of components in legally marketed devices.
Technological Characteristics
Sodium/potassium alumina silicate leucite glass ceramic system. Low-fusing porcelain manufactured via fritting of glass compositions. Designed for use with dental alloys or Press-I-Dent pressable component. No specific ASTM standards or software components described.
Indications for Use
Indicated for use by dental laboratories for the fabrication of dental restorations including crowns, bridges, veneers, inlays, and onlays. No specific patient population age or gender restrictions are noted.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
Predicate Devices
Willi Geller Creation Porcelain (k981490)
Finesse Low Fusing Porcelain (k954761)
Vita Omega Porcelain (k921474)
Related Devices
K050838 — THE LF-P CERAMIC SYSTEM · Jensen Industries, Inc. · May 4, 2005
K140848 — LUMINESSE PORCELAIN SYSTEM · Talladium, Inc. · Dec 17, 2014
K982377 — OPC LOWWEAR PORCELAIN · Jeneric/Pentron, Inc. · Aug 18, 1998
K163251 — Life Essence Universal Porcelain System · Blue Sky Bio · Aug 4, 2017
K123613 — SYNSPAR ADVANCED · Ardent, Inc. · Feb 20, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for AURIDENT, INC. The logo consists of a four-pointed star to the left of the company name. The company name is written in bold, sans-serif font.
K642638
P.O. Box 7200 • Fullerton, California 92834
(800) 422-7373 • (714) 523-5544
FAX (714) 870-0608
| DEC 17 2004<br>510(k) Summary: | | |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact: | Bruce D. Spivack | |
| Date Prepared: | September 22, 2004 | |
| Device Classification Name: | Class II, Porcelain Powder for Clinical Use<br>(872.6660), EIH | |
| Trade Name: | Radiance Porcelain | |
| Predicate Devices: | Willi Geller Creation Porcelain, k981490<br>Finesse Low Fusing Porcelain, k954761<br>Vita Omega Porcelain, k921474 | |
| Device Description: | Radiance Porcelain system is a sodium/potassium alumina<br>silicate based leucite glass ceramic, porcelain system for the<br>fabrication of either porcelain fused to metal or all ceramic<br>restorations. Radiance porcelain can only be used with its<br>corresponding leucite glass ceramic pressable, Press-I-Dent for<br>the fabrication of all ceramic systems and therefore constitute<br>one complete porcelain system. | |
| Intended Use: | Radiance Porcelain is intended for the fabrication of dental<br>restorations, crowns, bridges, veneers, inlays, and onlays, using<br>either dental alloys or its pressable component, Press-I-Dent. | |
| Technological Characteristics: | All the components in the Radiance<br>System have been used in legally marketed devices. This<br>system has been on the European market since 1999 and enjoys<br>full CE marking from the manufacturer. Because of the<br>similarities between the Radiance System and the predicated<br>devices, we felt that no further biocompatibility testing was<br>necessary | |
27
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is represented by three parallel lines curving upwards, resembling a stylized wing or feather design. The text is arranged in a circular pattern around the symbol, with "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" clearly visible.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## DEC 17 2004
Mr. Bruce D. Spivack Manager, Sales and New Product Development Aurident, Incorporated 610 South College Boulevard Fullerton, California 92831
Re: K042638
Trade/Device Name: Radiance Porcelain Regulation Number: 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: December 6, 2004 Received: December 9, 2004
Dear Mr. Spivack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Spivack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and instille (2 = = = = forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AURIDENT, INC. The logo consists of a four-pointed star to the left of the company name. The company name is in bold, sans-serif font.
P.O. Box 7200 · Fullerton, California 92834 (800) 422-7373 · (714) 523-5544 FAX (714) 870-0608
## Indications for Use
510(k) Number (if known): K042638
Device Name: Radiance Porcelain
Indications For Use:
Radiance Porcelain is a sodium/potassium alumina silicate leucite glass ceramic porcelain system for the fabrication of dental restorations, crowns, bridges, veneers, inlays, and onlays, using either dental alloys or its pressable component, Press-I-Dent.
This porcelain system is a lower fusing porcelain than feldspar based porcelains and is manufactured from the fritting of manufactured glass compositions. In addition, it has lower opposing occlusal wear characteristics than feldspar based porcelains.
Radiance is intended for use by dental laboratories as an alternative to other porcelains.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Quasar
Asion Sir n-Oif) (Chasion Sign-Oit)
Division of Anesthes Ology, General Hospital.
Division of Anesthes Davices ാന്ത്രിക്കുന്നത്. Dental Dental Dental Dental Dental
Mumber
Page 1 of . -
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.