The FiberDent Corporation Orascope™ is a non diagnostic fiberoptic visualization device used to illuminate and magnify dental surfaces.
Device Story
Orascope is a flexible fiberoptic endoscope for dental visualization. System comprises flexible fiberoptic probe, sterile disposable sheath, high-intensity Xenon-quality light source, CCD color camera, and flat-screen display monitor. Light source transmits illumination through probe fibers to target oral cavity; image transmission fibers carry real-time visual data back to camera. Camera converts optical signal to electronic video for display on monitor mounted on wheeled stand. Used by dentists in clinical settings to enhance visualization of dental surfaces. Device is non-diagnostic; output provides magnified real-time view to assist clinician in oral examination. Sterile sheath provides barrier protection. System benefits patient by facilitating improved visual inspection of dental structures.
Clinical Evidence
Bench testing only. No clinical data provided. Device underwent electrical safety and EMC/EMI testing per EN 61000-4-3, EN 55022, EN 61000-4-6, EN 61000-4-8, EN 61000-4-2, IEC 1000-3-2, IEC 1000-4-11, and EN 61000-4-4. Biocompatibility (USP Class VI) and cytotoxicity testing performed on disposable sheaths. Viral penetration testing confirmed no viral passage through sheath film or heat seals.
Technological Characteristics
Flexible fiberoptic probe (0.9 mm and 1.8 mm diameters); sterile disposable closed-end sheath (USP Class VI biocompatible). Illumination via dual-channel Xenon-quality metal-halide light source. Imaging via 3-chip color CCD camera (768x494 resolution). Display via 15" XGA flat panel monitor. Connectivity via S-Video signal. Power: 12 VDC (camera/monitor), 120 VAC (light source).
Indications for Use
Indicated for use in dental offices as a non-diagnostic visualization tool to illuminate and magnify dental surfaces.
Regulatory Classification
Identification
A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.
K974542 — TRUE VISION, TRUE VISION II · Medical Dynamics, Inc. · Feb 17, 1998
K984332 — SPECTRA VU 1000 SERIES INTRAORAL CAMERA, SPECTRA VU 2000 SERIES INTRAORAL CAMERA, SPECTRA VU HDDI SV I INTRAORAL CAMERA, · International Technology Concepts · Apr 29, 1999
{0}------------------------------------------------
# 510(k) Summary for FIBERDENT CORPORATION ORASCOPE (4) 1011
### 1. SPONSOR
FiberDent Corporation Suite B 3753 Varsity Drive Ann Arbor, MI 48108
| Contact Person: | Neil Haven |
|-----------------|--------------|
| Telephone: | 734-213-7600 |
Date Prepared: March 25, 1999
### 2. DEVICE NAME
| Proprietary Name: | Orascope |
|----------------------|------------------|
| Common/Usual Name: | Dental Endoscope |
| Classification Name: | Endoscope |
### 3. PREDICATE DEVICES
Dental Medical Industries Telicam - K980474
Dental Vision Direct UltraCam - K933671
### DEVICE DESCRIPTION 4.
The Orascope consists of a flexible fiberoptic probe with sterile disposable sheath. The probe feeds magnified images from within the oral cavity. The flexible fiberoptic probe transmits the image information and light to and from the oral cavity. A light source transmits light through the fiber to illuminate the target area. Real time image information from the target area is transmitted through the fiber to a CCD color camera which sends this information to the flat screen display monitor to be viewed by the user. The flat screen display is mounted on a wheeled
{1}------------------------------------------------
flexible stand which can be positioned for optimum viewing comfort. The Orascope enhances the dentist's ability to see within the oral cavity.
The Orascope consists of the fiberoptic endoscope and sterile disposable sheath. A digital camera, monitor and video printer are also offered as accessories to the Orascope.
### Orascope Fiberoptic Dental Endoscopes:
The Orascope consists of a flexible fiberoptic probe which contains both lightemitting fibers and image transmission fibers. The light emitting fibers gather light from a high-intensity light source affixed to the wheeled display stand. The light shines out the probe tip to illuminate the field-of-view for the image transmission fibers. The image transmission fibers transmit the image from the end of the probe tip back to the wheeled display stand where the color video camera creates the electronic image for display on the display screen. The Orascope is offered in both 0.9 mm and 1.8 mm tip outer diameter sizes. Each probe contains light conducting fibers for the illumination of the probe's field-of-view as well as image conducting fibers for optical transmission of the image of the field-of-view to the camera.
### Sterile Disposable Sheath:
The sterile disposable sheath is a single use closed end sheath which covers the end of the probe for about 100 mm. This sheath has been subjected to USP Class VI biocompatibility test results and cytotoxicity testing. Viral Penetration testing has been performed on the sheaths. The test results showed that virus (bacteriophagy) did not penetrate the film or the heat seals of the sheaths tested.
## Camera:
The camera offered with the Orascope is a Panasonic GP-US522 three-chip color CCD camera with remote head. The GP-US522 is a standard three-chip color CCD camera with a resolution of 768 (H) x 494 (V). The camera operates on 12 VDC. The camera is used to capture a video image from the Orascope fiber-optic probe(s) and provide an S-Video signal for display by the video monitor.
## Video Monitor:
The video monitor offered with the Orascope is a National Display Systems 15" XGA Flat Panel Video Display Monitor Model #DSW-X15-N. This flat-screen video display monitor operates from the S-Video signal provided by the camera. The monitor operates on 12 VDC.
{2}------------------------------------------------
### Light Source:
The light source is a Cuda Products Corporation dual channel Xenon-quality fiberoptic light source. The MX2-300 operates on 120 VAC and provides illumination from one of two metal-halide bulbs. Dual channel means that the light source contains two bulbs. If one bulb burns out during a procedure, the light source can be configured so that the second bulb can be used within seconds. The light from the bulb is reflected onto a fiberoptic light-guide which is plugged into the cable receptacle on the front of the MX2-300.
### 5. INTENDED USE
The FiberDent Orascope™ is a non-diagnostic fiberoptic visualization device used to illuminate and magnify dental surfaces.
#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.
The FiberDent Corporation Orascope device is substantially to the Dental/Medical Diagnostic Systems Incorporated Telicam Intraoral Video Camera, subject of K980474 and the Dental Vision Direct, Inc UltraCam, subject of K933671.
The Orascope and the predicate devices are similar in intended use in that they are all intended as non-diagnostic endoscopic cameras to illuminate and magnify dental surfaces. They are all indicated for use in the dental office.
The Orascope and the predicate devices are similar in technological characteristics in that they all consist of a non-diagnostic intraoral video camera and monitor for viewing images. All three devices use a camera and monitor. Both the Orascope and the Telicam provide a sterile disposable sheath with the probe whereas the UltraCam has an autoclavable tip. Both the Orascope and the Telicam require a fiberoptic light source whereas the UltraCam does not use a fiberoptic light source.
The principle of operation of the proposed and predicate devices are the same in that they all transmit and image from the target oral cavity area to camera to a display monitor. Both predicate devices offer a printer for viewing images whereas the Orascope system does not.
{3}------------------------------------------------
### 7. PERFORMANCE TESTING
The Orascope Dental Endoscope and accessories have been subjected to and passed the following electrical safety and EMC/EMI Testing:
| • Radiated Electromagnetic Immunity | EN 61000-4-3 &<br>ENV 0204 |
|--------------------------------------------|----------------------------|
| • Electromagnetic Interference | EN 55022 |
| • Conducted Electromagnetic Immunity | EN 61000-4-6:1996 |
| • Power Frequency Magnetic Field Immunity | EN 61000-4-8:1993 |
| • Electrostatic Discharge Immunity | EN 61000-4-2:1995 |
| • Limits for Harmonic Current Emissions | IEC 1000-3-2:1995 |
| • Voltage Dips, Interrupts and Variations | IEC 1000-4-11:1994 |
| • Surge Immunity | IEC 1000-3-2:1995 |
| • Electrical Fast Transient/Burst Immunity | EN 61000-4-4:1995 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 23 1999
FiberDent Corporation c/o Ms. Mary McNamara-Cullinane, RAC Staff Consultant Medical Device Consultants, Incorporated 49 Plain Street 02760 North Attleboro, Massachuettes
Re : K991011 Orascope™ Trade Name: Requlatory Class: I Product Code: EIA March 25, 1999 Dated: Received: March 26, 1999
Dear Ms. McNamara-Cullinane:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Paqe 2 - Ms. McNamara-Cullinane
this response to your premarket notification Please note: submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): .
Device Name:_FiberDent Corporation Orascope™
Indications For Use:
The FiberDent Corporation Orascope™ is a non diagnostic fiberoptic visualization device used to illuminate and magnify dental surfaces.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Dunn
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital I 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
ાર
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
510(k)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.