← Product Code [EIA](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA) · K242404

# BDC Dental Unit (K242404)

_Bdc Dental Corporation , Ltd. · EIA · Feb 7, 2025 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA/K242404

## Device Facts

- **Applicant:** Bdc Dental Corporation , Ltd.
- **Product Code:** [EIA](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA.md)
- **Decision Date:** Feb 7, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6640
- **Device Class:** Class 1
- **Review Panel:** Dental

## Indications for Use

The BDC Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.

## Device Story

BDC Dental Unit serves as a centralized control station for dental procedures; provides air, water, vacuum, and electrical power to connected dental handpieces and accessories. System components include patient chair, dentist element (tray table), assistant element (3-way syringe, high-volume evacuator, saliva ejector), side cabinet, foot control, cuspidor, and operating light. Used in dental clinics by dentists and assistants. Device transforms pneumatic and electrical inputs into controlled outputs for patient positioning, light operation, and instrument function via foot pedal and panel controls. Benefits include ergonomic patient support and integrated utility management for clinical workflows. Components are designed for use with disposable barriers and reprocessing (cleaning/disinfection/sterilization).

## Clinical Evidence

No clinical data. Substantial equivalence established via bench testing, including biocompatibility (ISO 10993-5/10), reprocessing validation, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing per ISO 7494-1/2 and ISO 9168.

## Technological Characteristics

Dental operative unit; pneumatic/electric power; includes patient chair, dentist/assistant elements, cuspidor, light. Materials: biocompatible components (ISO 10993). Connectivity: standalone utility supply. Software: basic level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 80601-2-60, ISO 7494-1/2, ISO 9168, ISO 16954 (waterline biofilm).

## Regulatory Identification

A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.

## Predicate Devices

- A-dec 500 ([K213932](/device/K213932.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health and Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym in a blue square and the words "U.S. Food & Drug Administration" in blue font.

February 7, 2025

BDC Dental Corporation Ltd. % Dave Kim FDA Regulatory Consultant Mtech Group LLC 7505 Fannin St. Suite 610 Houston, Texas 77054

Re: K242404

Trade/Device Name: BDC Dental Unit Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: Class I, reserved Product Code: EIA, KLC Dated: January 10, 2025 Received: January 10, 2025

Dear Dave Kim:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# MICHAEL E. ADJODHA -S

Michael Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

| Submission Number (if known) |  |
|------------------------------|--|
|------------------------------|--|

#### K242404

Device Name

BDC Dental Unit

Indications for Use (Describe)

The BDC Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(K) SUMMARY K242404

# I. SUBMITTER

| Date:                   | February 4, 2025                                                                                                 |
|-------------------------|------------------------------------------------------------------------------------------------------------------|
| Submitter's Name:       | BDC Dental Corporation Ltd.                                                                                      |
| Submitter's HQ Address: | Part 3, No. 1 Guanchong Section, Shilian Rd., Shiqi Town, Panyu District, Guangzhou, 511450 Guangdong, P.R.China |
| Submitter's Telephone:  | 020-32052929                                                                                                     |
| Contact Person:         | Yanfei Liu, mr@o-bdc.com, RA Manager                                                                             |

## II. PROPOSED DEVICE

| Trade/proprietary Name: | BDC Dental Unit                                                              |
|-------------------------|------------------------------------------------------------------------------|
| Model Name: :           | N1, N2, N3, N1 Plus, N2 Plus, N3 Plus, S1, S2, S3, S1 Plus, S2 Plus, S3 Plus |
| Regulation Name:        | Dental Operative Unit and Accessories                                        |
| Regulation Number:      | 21 CFR 872.6640                                                              |
| Product Code:           | EIA, KLC                                                                     |
| Regulatory Class:       | Class I, reserved                                                            |

## III. PREDICATE DEVICE

| 510(k) Number:        | K213932                               |
|-----------------------|---------------------------------------|
| Primary Manufacturer: | A-dec Inc.                            |
| Device Name:          | A-dec 500                             |
| Regulation Name:      | Dental Operative Unit and Accessories |
| Regulation Number:    | 21 CFR 872.6640                       |
| Product Code:         | EIA                                   |
| Regulatory Class:     | Class I, reserved                     |

#### IV. DEVICE DESCRIPTION

The BDC dental unit is intended to supply air, water, vacuum and electrical power to allow the dental practitioner an intuitive control center for all common and normal patient treatment procedures performed in the dental operatory.

The BDC dental unit consists of a patient chair, dentist element (tray table), assistant element (3-way syringe, high volume evacuator and saliva ejector), side cabinet support center, foot control, cuspidor and a dental operating light, which for patient to sit during the dental diagnosis, treatment and/or operation.

Components: The BDC dental unit consists of a patient chair, dentist element (tray table), assistant element (3-way syringe, high volume evacuator and saliva ejector), side cabinet support center, foot control, cuspidor and a dental operating light, which for patient to sit during the dental diagnosis, treatment and/or operation, these components are devices such as dental handpieces can be connected to the dental unit. Dental handpieces are not included in the BDC Dental Unit. The components of the BDC Dental unit are intended to be covered with an FDA cleared disposable barrier and reprocessed (e.g., cleaning, disinfection, and/or sterilization) per the instructions for use.

{5}------------------------------------------------

# Brief of differences between all models

| Model   | Dentist<br>Element<br>(Traditional) | Dentist<br>Element<br>(Traditional<br>Cart Type) | Dentist<br>Element<br>(Continental) | Dentist<br>Element<br>(Continental<br>Cart Type) | Left/right<br>Arm<br>Swivel | Disc<br>foot<br>control | Lever<br>foot<br>control |
|---------|-------------------------------------|--------------------------------------------------|-------------------------------------|--------------------------------------------------|-----------------------------|-------------------------|--------------------------|
| N1      | √                                   |                                                  |                                     |                                                  |                             | √                       |                          |
| N2      |                                     | √                                                |                                     |                                                  |                             | √                       |                          |
| N3      | √                                   |                                                  |                                     |                                                  |                             |                         | √                        |
| N1 Plus |                                     |                                                  | √                                   |                                                  |                             | √                       |                          |
| N2 Plus |                                     |                                                  |                                     | √                                                |                             | √                       |                          |
| N3 Plus |                                     |                                                  | √                                   |                                                  |                             |                         | √                        |
| S1      | √                                   |                                                  |                                     |                                                  |                             | √                       |                          |
| S2      |                                     | √                                                |                                     |                                                  | √                           | √                       |                          |
| S3      | √                                   |                                                  |                                     |                                                  | √                           |                         | √                        |
| S1 Plus |                                     |                                                  | √                                   |                                                  | √                           | √                       |                          |
| S2 Plus |                                     |                                                  |                                     | √                                                | √                           | √                       |                          |
| S3 Plus |                                     |                                                  | √                                   |                                                  | √                           |                         | √                        |

## V. INDICATIONS FOR USE

The BDC Dental Unit is intended to supply power to and serve as a base for dental devices and accessories. This device includes a dental chair and is intended for use in the dental clinic environment and is designed for use by trained dental professionals, dentists and/or dental assistants.

#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Table 2

| Items                                  | Subject device                                                                                                                                                                                                                                                                                                              | Predicate device                                                                                                                                                                                                                                                                                                                                                               | Comparison |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer                           | BDC Dental Corporation Ltd.                                                                                                                                                                                                                                                                                                 | A-dec Inc.                                                                                                                                                                                                                                                                                                                                                                     |            |
| 510(k) Number                          | K242404                                                                                                                                                                                                                                                                                                                     | K213932                                                                                                                                                                                                                                                                                                                                                                        |            |
| Product name                           | Dental Unit                                                                                                                                                                                                                                                                                                                 | A-dec 500 Delivery System                                                                                                                                                                                                                                                                                                                                                      |            |
| Model                                  | N1, N2, N3, N1 Plus, N2 Plus, N3<br>Plus, S1, S2, S3, S1 Plus, S2 Plus, S3<br>Plus                                                                                                                                                                                                                                          | A-dec 500                                                                                                                                                                                                                                                                                                                                                                      |            |
| Design                                 | The BDC dental unit is mainly<br>composed of a patient chair, dentist<br>element(tray table, 3-way syringe),<br>assistant element (3-way syringe,<br>high volume evacuator and saliva<br>ejector), side cabinet support center,<br>foot control, cuspidor and a dental<br>operating light.                                  | A-dec 500 dental system is<br>composed of dental chair, delivery<br>system, assistant arm, dental light,<br>cuspidor.                                                                                                                                                                                                                                                          | Similar    |
| Indications for Use                    | The BDC Dental Unit is intended to<br>supply power to and serve as a base<br>for dental devices and accessories.<br>This device includes a dental chair<br>and is intended for use in the dental<br>clinic environment and is designed<br>for use by trained dental<br>professionals, dentists and/or dental<br>assistants. | The A-dec Delivery System is<br>intended to provide a mounting<br>location in addition to providing<br>air, water, vacuum, and electrical<br>power to dental devices for use<br>during diagnostic and therapeutic<br>treatment by licensed health  care<br>professionals. Delivery systems<br>may be mounted to dental chairs,<br>dental carts,<br>dental cabinets, and walls. | Similar    |
| Dentist Element                        | - Control of water and air supply,<br>Table Height, Patient Chair<br>Positioning, Light On/Off, Handpiece<br>Function, Timer, Mode Selection                                                                                                                                                                                | -Turn on/Turn off the air, water,<br>and power, Position the Delivery<br>System, Position the Handpiece<br>Holders, Operate Touchpad<br>Lockout, User Setting and<br>Auxiliary Device Control,<br>Position the Chair, Operate the<br>Dental Light, Operate the<br>Cuspidor, Operate Handpiece                                                                                  | Same       |
| Assist Table<br>(Assistant<br>Element) | - Components: 3-way Syringe, Saliva<br>Ejector, High Volume Evacuator<br>(Large Saliva Ejector)<br>- Control of Light On/Off, Cup/<br>Spittoon (Cuspidor) Water Dispenser,<br>Changing the Settings, Position<br>memories                                                                                                   | -Components: Syringe, Saliva<br>Ejector, High Volume Evacuator<br>-Position the Handpiece<br>Holders, Adjust the<br>Worksurface and<br>Instrumentation Height, Operate<br>the Dental Light,<br>Operate the Cuspidor                                                                                                                                                            | Same       |
| Main Components                        | The BDC dental unit is mainly<br>composed of a patient chair, dentist<br>element(tray table, 3-way syringe),<br>assistant element (3-way syringe,<br>high volume evacuator and saliva<br>ejector), side cabinet support center,<br>foot control, cuspidor and a dental<br>operating light.                                  | Dental system is composed of<br>dental Chair , delivery system,<br>assistant arm, dental light ,<br>cuspidor.                                                                                                                                                                                                                                                                  | Same       |
| Control of water<br>and air            | Uses pneumatically controlled vales<br>to water control the flow of air and<br>water on /off and intensity controlled<br>by foot control.                                                                                                                                                                                   | Use air coolant and water<br>coolant to adjust Air and water,<br>speed controlled by foot control.                                                                                                                                                                                                                                                                             | Same       |
| Software-<br>controlled<br>components  | Software functions are reflected in<br>the dentist element and assistant<br>element panel .Software can control<br>dental patient chair up, down, tilt<br>forward, tilt backward and set<br>treatment position, dental operating<br>light on/off, cuspidor flush and cup-<br>filler.                                        | Software functions are reflected<br>in the delivery system and<br>assistant arm panel .<br>Software can control dental chair<br>base up, down, and set treatment<br>position, dental light on/off/<br>intensity, bowl flush and<br>cupfill, handpieces, electric<br>handpiece.                                                                                                 | Similar    |
| Table<br>(Dentist Element)             | - Control of water supply, scaler<br>vibration power, table height,<br>patient<br>position, System Power, Film viewer,<br>Patient Chair Positioning, Light<br>On/Off, Handpiece function, Timer,<br>Mode Selection, LED Display                                                                                             | - Control of water and air supply,<br>Table Height, Patient Chair<br>Positioning, Light On/Off,<br>Handpiece Function, Timer,<br>Mode Selection                                                                                                                                                                                                                                | Similar    |
| Assist Table<br>(Assistant<br>Element) | - Accessories: Saliva Ejector<br>(Small,<br>Large), 3-way Syringe - Control of<br>Light On/Off, Cup/Spittoon Water<br>Dispenser, Changing the Settings,<br>Position memories                                                                                                                                                | - Components: 3-way Syringe,<br>Saliva Ejector, High Volume<br>Evacuator (Large Saliva<br>Ejector)<br>- Control of Light On/Off, Cup/<br>Spittoon (Cuspidor) Water<br>Dispenser, Changing the<br>Settings, Position memories                                                                                                                                                   | Same       |
| Main Components                        | Chair, Unit, Table, Seat, Stool, Monitor<br>Arm, Operation table                                                                                                                                                                                                                                                            | Chair, Unit (side cabinet support<br>center), Table (Assistant Element),<br>Seat (Upholstery Seat), Stool,<br>Operation table (Dentist Element)                                                                                                                                                                                                                                | Same       |
| Power& Utility<br>Supply               | AC110- 120 / 220-240V,<br>50/60Hz, compressed air and water                                                                                                                                                                                                                                                                 | AC100- 120/220-240V, 50/60Hz,<br>compressed air and water                                                                                                                                                                                                                                                                                                                      | Same       |
| Syringe                                | 3-way syringe                                                                                                                                                                                                                                                                                                               | 3- way syringe                                                                                                                                                                                                                                                                                                                                                                 | Same       |
| Control of water<br>and air            | Uses pneumatically controlled vales to<br>water control the flow of air and water<br>on/off and intensity controlled by foot<br>pedal                                                                                                                                                                                       | Uses pneumatically controlled vales<br>to water control the flow of air and<br>water on /off and intensity<br>controlled by foot pedal.                                                                                                                                                                                                                                        | Same       |
| Air Pressure                           | 0.55-0.8MPa / 5.5-8.0bar                                                                                                                                                                                                                                                                                                    | 550kPa(min)/800kPa(max)                                                                                                                                                                                                                                                                                                                                                        | Similar    |
| Water Pressure                         | 0.20-0.40MPa / 2.0-4.0bar                                                                                                                                                                                                                                                                                                   | 410±140 kPa                                                                                                                                                                                                                                                                                                                                                                    | Similar    |
| Dental unit<br>handpiece hose          | Air flow rate:300Kpa,max.68NL/min<br>Spray air<br>flowrate:250Kpa,max.18NL/min<br>Water flow rate:250Kpa,max.75mL/min                                                                                                                                                                                                       | N/A                                                                                                                                                                                                                                                                                                                                                                            |            |
| Water System                           | City water supply                                                                                                                                                                                                                                                                                                           | City water supply                                                                                                                                                                                                                                                                                                                                                              | Same       |
| Water Heater                           | Max. (< 42 ° C)                                                                                                                                                                                                                                                                                                             | Max. (< 42 ° C)                                                                                                                                                                                                                                                                                                                                                                | Same       |
| Water Sanitation<br>System             | Distilled water container added                                                                                                                                                                                                                                                                                             | Distilled water container added                                                                                                                                                                                                                                                                                                                                                | Same       |
| Suction                                | HVE (High Volume Evacuator) Saliva<br>Ejector                                                                                                                                                                                                                                                                               | HVE (High Volume Evacuator)<br>Saliva Ejector                                                                                                                                                                                                                                                                                                                                  | Same       |
| Patient Load                           | Max. 150kg                                                                                                                                                                                                                                                                                                                  | Max. 227kg                                                                                                                                                                                                                                                                                                                                                                     | Similar    |
| Chair Height                           | Max.780±10mm, Min.380±20mm                                                                                                                                                                                                                                                                                                  | Max.780±10mm, Min.460±20mm                                                                                                                                                                                                                                                                                                                                                     | Similar    |
| Backrest range                         | 110°-180°                                                                                                                                                                                                                                                                                                                   | N/A                                                                                                                                                                                                                                                                                                                                                                            |            |
| Head rest range                        | 0-100mm                                                                                                                                                                                                                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                            |            |
| Lift Motor                             | Hydraulic electromotor                                                                                                                                                                                                                                                                                                      | Electromotor                                                                                                                                                                                                                                                                                                                                                                   | Similar    |
| Dental light                           | Available                                                                                                                                                                                                                                                                                                                   | Available                                                                                                                                                                                                                                                                                                                                                                      | Same       |
| Foot control                           | Standard                                                                                                                                                                                                                                                                                                                    | Disc Type                                                                                                                                                                                                                                                                                                                                                                      | Similar    |
| Electrical Safety                      | Complied with IEC 60601- 1                                                                                                                                                                                                                                                                                                  | Complied with IEC 60601- 1                                                                                                                                                                                                                                                                                                                                                     | Same       |
| Electromagnetic<br>compatibility       | Complied with IEC 60601- 1-2                                                                                                                                                                                                                                                                                                | Complied with IEC 60601- 1-2                                                                                                                                                                                                                                                                                                                                                   | Same       |

{6}------------------------------------------------

{7}------------------------------------------------

# Substantial Equivalence Discussion

The proposed BDC and predicate device are similar in all the items in the comparison chart except water pressure, patient load, chair height, back/head rest, lift motor, foot control and models. These differences do not affect substantial equivalence, also the subject device safety was verified with testing according to IEC 80601-2-60 and IEC 60601-1.

As the models/types do not differ significantly in purpose, design, materials, energy source, function or any other feature related to substantial equivalence.

On the basis of the discussion above, it is concluded that the subject device is substantially equivalent to the predicate device.

{8}------------------------------------------------

# VII. Non-Clinical Test Data

The following performance data were provided in support of the substantial equivalence determination.

#### Biocompatibility testing

The biocompatibility evaluation for the 3-way syringe , high-volume suction, saliva ejector, and tubing were conducted with the following performance standards:

ISO 10993-5: 2009 Biological evaluation of medical devices — Part 5: Tests for in vitro cvtotoxicity ISO 10993- 10: 2010 Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization

The 3-way syringe, high-volume suction, saliva ejector, and tubing of testing included the following tests:

- Cytotoxicity
- Sensitization
- Irritation

The tubing and syringes body are considered as indirect tissue contacting.

Such like armrest pad, dental operating light handles, upholstery seat, 3-way syringe, high volume evacuator and saliva ejector, we recommended using a barrier sleeves and barrier film which is non-sterile and intended for single patient use only.

## Reprocessing Validation

Reprocessing validation was conducted per FDA Guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling - Guidance for Industry and Food and Drug Administration Staff (fda.gov)" including cleaning, disinfection, and sterilization validation for the components of the BDC Dental Unit.

As to the body/head of the 3-way syringe, high-volume evacuator, and saliva ejector, even though we conducted the Cleaning, intermediate disinfection testing on them, due to their connection which is fixed, the user can not disassembly them from the tubing, so we recommended in our IFU not only the cleaning and disinfection method, but also we recommend to wear compatible FDA-cleared disposable barriers when using them. Cleaning and Sterilization validation was conducted for the 3-way syringe tip.

In addition, cleaning and disinfection validation was provided for the surfaces of the subject device. Cleaning and disinfection vas conducted on the waterlines and water bottles of the subject device. Validation was conducted using the standard 16954:2015 Dentistry - Test methods for dental unit waterline biofilm treatment.

## Electrical safety and electromagnetic compatibility (EMC)

The Electrical Safety and Electromagnetic compatibility tests were performed in accordance with the following standards. Comprehensive performance testing has been conducted on the BDC dental unit in accordance FDA recognized standards. EMC testing was conducted in accordance with Standard EN/IEC 60601- 1-2. Electrical, and environmental safety testing according to Standard EN/IEC 60601- 1 was performed and IEC 80601-2-60 Particular requirements for the basic safety and essential performance of dental equipment.

{9}------------------------------------------------

#### Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions." The software for this device was considered as a "BASIC" level of concern, since a failure or latent flaw in the software would not directly result in serious injury or death to the patient or operator.

#### Performance Test

The performance tests were conducted as bench test and the test results met the pre-set criteria.

- ISO 7494-1 Third edition 2018-06 - Dentistry - Dental units - Part 1: General requirements and test methods

- ISO 7494-2 Second edition 2015-01 - Dental units - Part 2: Water and air supply - ISO 9168 - 2009-07- 15 - Dentistry - Hose connectors for air driven dental handpieces devices

## Clinical Testing

Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product.

#### VIII. CONCLUSIONS

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, the BDC dental unit is substantially equivalent to K3, the predicate device, in intended Use/indications for use, material composition, fundamental engineering technology, the principle of operation and basic design.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA/K242404](https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA/K242404)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
