Firstar Dental Unit

K231297 · Firstar Dental Company · EIA · Jan 27, 2025 · Dental

Device Facts

Record IDK231297
Device NameFirstar Dental Unit
ApplicantFirstar Dental Company
Product CodeEIA · Dental
Decision DateJan 27, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6640
Device ClassClass 1

Indications for Use

The Firstar Dental Unit is intended to serve as a base for ancillary dental devices and accessories by providing air, water, vacuum, and low voltage electrical power to hand-held dental instruments. The Firstar Dental Unit is intended for use by dental practitioners in a clinical environment.

Device Story

Firstar Dental Unit serves as delivery system for dental instruments; provides air, water, vacuum, and low voltage electrical power to hand-held devices. Components include doctor control head, foot switch, cuspidor, side box, high volume evacuator, saliva ejector, and air/water syringe. Used in clinical environments by dental practitioners. Utility box houses shut-off valves, pressure regulators, and transformers; internal tubing routes utilities to control head and instruments. Control head manages handpiece pressure and electrical power; integrates chair controls if configured. Supports ancillary devices via ISO 9168-20 connections (e.g., pneumatic/electric handpieces, scalers, curing lights, intraoral cameras). Benefits patient by enabling organized, powered delivery of dental treatment tools.

Clinical Evidence

Bench testing only. Verification and validation testing performed to confirm conformance with IEC 80601-2-60:2019, ISO 7494-1:2018, ISO 7494-2:2015, and ISO 10993-1. Results met all acceptance criteria.

Technological Characteristics

Stationary or mobile dental delivery unit. Provides air, water, vacuum, and 24VAC power. Features include control head, foot switch, cuspidor, and side box. Connectivity via ISO 9168-20 standard connections. Complies with IEC 80601-2-60, ISO 7494-1, ISO 7494-2, and ISO 10993-1. No software-based algorithms described.

Indications for Use

Indicated for use by dental practitioners in a clinical environment as a base for ancillary dental devices and accessories, providing air, water, vacuum, and low voltage electrical power to hand-held dental instruments.

Regulatory Classification

Identification

A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 27, 2025 Firstar Dental Company David Yee Vice President 18915 East Valley Highway Kent, Washington 98032 Re: K231297 Trade/Device Name: Firstar Dental Unit Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit And Accessories Regulatory Class: Class I, reserved Product Code: EIA Dated: December 31, 2024 Received: January 2, 2025 Dear David Yee: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use Submission Number (if known) K231297 Device Name Firstar Dental Unit Indications for Use (Describe) The Firstar Dental Unit is intended to serve as a base for ancillary dental devices and accessories by providing air, water, vacuum, and low voltage electrical power to hand-held dental instruments. The Firstar Dental Unit is intended for use by dental practitioners in a clinical environment. Type of Use (Select one or both, as applicable) > | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ | 510(k) #: | K231297 | 510(k) Summary | Prepared on: 2024-12-31 | |-----------|---------|----------------|-------------------------| |-----------|---------|----------------|-------------------------| Contact Details | | 21 CFR 807.92(a)(1) | |-----------------------------|----------------------------------------------------| | Applicant Name | Firstar Dental Company | | Applicant Address | 18915 E Valley Highway Kent WA 98032 United States | | Applicant Contact Telephone | (425) 473-6928 | | Applicant Contact | Mr. DAVID YEE | | Applicant Contact Email | david@firstardental.com | Device Name | | 21 CFR 807.92(a)(2) | |---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name | Firstar Dental Unit: 37H-DU, 38H-DU, 39H-DU, 50H-DU, 37H-DU1, 38H-DU1, 39H-DU1, 50H-DU1, 37H-DU2, 38H-DU2, 39H-DU2, 50H-DU2, 37H-DU3, 38H-DU3, 39H-DU3, 50H-DU3, SMU-A1, SMU-A2, SMU-A3, WMU-B1, WMU-B2, WMU-B3, DCU-C1, DCU-C2, DCU-C3, ASCU-D1, ASCU-D2, ASCU-D3 | | Common Name | Dental operative unit and accessories | | Classification Name | Unit, Operative Dental | | Regulation Number | 872.6640 | | Product Code(s) | EIA | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | |--|---------------------| |--|---------------------| | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |-------------|----------------------------------------------------------|--------------| | K173608 | Forest Dental Operative Unit, 4485SI | EIA | Device Description Summary | | 21 CFR 807.92(a)(4) | |--|---------------------| |--|---------------------| The Firstar Dental Unit (also known as the delivery unit or delivery system) is a combination of dental equipment components used to carry, position, and control the devices used in the practice of dentistry. Delivery systems provide for hand pieces and other variousaccessories by delivering air, water, vacuum and low voltage electricity to the dentist during procedures. Depending on the configuration, it includes combinations of a Doctor Control Head (delivery head), Delivery System Foot Switch (sometimes referred to as a rheostat), Cuspidor, Side Box (Control Box), High Volume Evacuator (HVE), Saliva Ejector (SE), and an Air and Water Multifunction Syringe. Air, water, vacuum, drainage, and electrical power utilities are delivered through the shut-off valves, pressure regulators, and electrical transformers located in the post mounted utility box. Subsequently, internal tubing delivers the air, water and vacuum to the control head and to the dental hand pieces and multifunction syringe. Low voltage electricity is delivered to the control head for operating items such as scalers, electric hand pieces and fiber optic light sources. The Firstar Dental Unit is available in various configurations and mounting styles including dental chair mounted, mobile cart mounted, and wall mounted on a dental chair, the Firstar Dental Unit can be configured in two distinct styles - a pivot mounted unit which allows for ambidextrous access to the patient, and a post mounted unit which includes a side box located on one side of the dental chair that can mount a cuspitor (spittoon) on it. Despite the various applications, the key elements of the Dental Unit such as the internal plumbing, tubing, materials, power sources, utility connections, delivery head internal components, and overall operating principles remain the same. {5}------------------------------------------------ The control head allows operators to place surgical equipment on top of the tray and it provides controls for hand piece water pressure, air pressure and electrical power. Additionally, controls for a Firstar Dental Chair are often integrated on the control head if the dental unit is configured to mount on a dental chair. If equipped, the cuspidor bowl allows 90 decrees ande rotation that provides access to collect the patient's saliva during dental treatment. The side box provides controls for the cuspidor cup fill, bowl rinse, air, and pressure regulators for the air and water quick disconnect outlets. Various ancillary devices can be connected to the Firstar Dental Unit by means of standard ISO 9168-20 connections including but not limited to: pneumatic handpieces, electric micromotors and handpieces, air/water syringes, piezo or magnetostrictive dental scalers, curing lights, intraoral cameras, saliva ejectors, and high volume evacuators. 24VAC power supplied from the Junction Box Power Supply can be used to power electrically operated ancillary equipment such as electrical handpieces, dental does not manufacture the aforementioned devices except for the air/water syringes, saliva ejectors, and high volume evacuators. # Intended Use/Indications for Use ## 21 CFR 807.92(a)(5) The Firstar Dental Unit is intended to serve as a base for ancillary dental devices and accessories by providing air, watum, and low voltage electrical power to hand-held dental instar Dental Unit is intended for use by dental practitioners in a clinical environment. ## Indications for Use Comparison The indications for use are equivalent to the predicate device. # Technological Comparison The subject device and predicate device share the same technical characteristics with the following differences: - Relative humidity range - · Operating temperature range - · Air supply pressure range - · Air/oil separator - · Water supply pressure range - Standards applied - Syringe water flow control - Syringe air flow control - Maximum flex arm load The subject device complies with relevant demonstrating equivalent performance. The above differences do not raise any different questions regarding substantial equivalence compared to the predicate device. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Non-Clinical verification and validation tests have been performed with regards to the intended use, requirement specifications and the risk management results.The non-clinical testing was performed to demonstrate verlication testing in conformance with the acceptance criteria of test methods and recognized consensus standards shown below. Testing was performed based on the following standards: • IEC 80601-2-60:2019 Medical electrical equipment — Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment. - ISO 7494-1:2018 Dentistry Stationary dental units and dental patient chairs Part 1: General requirements. - · ISO 7494-2:2015 Dentistry Dental units Part 2: Air, water, suction and wastewater systems. • ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. The test results demonstrate that the Firstar Dental Unit conforms with the relevant FDA-recognized consensus standards and quidance documents and meets the acceptance criteria. Therefore, the subject device, Firstar Dental Unit, is substantially equivalent to the predicate device. ## 21 CFR 807.92(a)(6)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...