← Product Code [EIA](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA) · K143696

# Spirit (K143696)

_Pelton & Crane · EIA · Sep 18, 2015 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA/K143696

## Device Facts

- **Applicant:** Pelton & Crane
- **Product Code:** [EIA](/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA.md)
- **Decision Date:** Sep 18, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.6640
- **Device Class:** Class 1
- **Review Panel:** Dental

## Indications for Use

The Spirit Dental Operative Units are intended to supply power to and serve as a base for other dental devices and accessories by providing air, water, vacuum and low voltage electrical power to hand held dental instruments. The Spirit Dental Operative Units are intended for use by professional dental practitioners in providing treatment in a dental operatory.

## Device Story

Spirit Dental Operative Units serve as a base for dental instruments; they deliver air, water, vacuum, and low-voltage electrical power to handpieces, scalers, intra-oral cameras, and curing lights. The system includes a doctor's unit, assistant's unit, cuspidor, mount arms, foot control, and a junction box. Operation is pneumatic; the foot control activates handpiece drive air and coolant. Suction and syringes are manually controlled via valves on the instruments. The unit can house optional electric motors (using previously cleared Kaltenbach & Voigt components) for power-driven handpieces. Used in dental operatories by professional practitioners, the device supports clinical treatment by providing necessary utilities to hand-held tools. The provider uses the delivery head controls to adjust flow and select instruments. The system benefits patients by enabling efficient, powered dental procedures.

## Clinical Evidence

No clinical data was conducted. Substantial equivalence is established through non-clinical bench testing, including electrical safety, EMC, and software validation per AAMI ANSI IEC 62304. Biocompatibility was confirmed per ISO 10993-1.

## Technological Characteristics

Dental delivery unit providing air, water, vacuum, and low-voltage power. Features include pneumatic control block, LCD screen, and optional integrated electric motors. Complies with ISO 7494-1/2, AAMI ES60601-1, and IEC 60601-1-2. Connectivity includes internal 10-pin port for software updates. Materials are biocompatible per ISO 10993-1.

## Regulatory Identification

A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.

## Predicate Devices

- ELEVANCE ([K120239](/device/K120239.md))

## Reference Devices

- ELECTROtorque TLC ([K103027](/device/K103027.md))
- COMFORTronic 4894 ([K080677](/device/K080677.md))
- COMFORTdrive 200XDA Handpiece ([K080677](/device/K080677.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem depicts a stylized human figure in profile, with three overlapping heads suggesting a sense of community or interconnectedness. The figure is rendered in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 18, 2015

Pelton & Crane Mr. Frank Ray Regulatory Affairs Manager 11727 Fruehauf Dr. Charlotte, North Carolina 28273

Re: K143696

Trade/Device Name: Spirit Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: I Product Code: EIA Dated: August 19, 2015 Received: August 21, 2015

Dear Mr. Rav:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Frank Ray

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K143696

Device Name

Spirit

Indications for Use (Describe)

The Spirit Dental Operative Units are intended to supply power to and serve as a base for other dental devices and accessories by providing air. water, vacuum and low voltage electrical power to hand held dental instruments. The Spirit Dental Operative Units are intended for use by professional dental practitioners in providing treatment in a dental operatory.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "Pelton & Crane" in a stylized, cursive font. The text is a dark blue color, and the ampersand symbol is used between the two names. The font is elegant and flowing, giving the text a sophisticated appearance. The overall impression is that of a professional logo or brand name.

# Section V - 510(k) Summary for Spirit

- 1. Submitter Information: Pelton & Crane 11727 Fruehauf Drive Charlotte, NC 28273

| Contact Person:   | Frank Ray      |
|-------------------|----------------|
| Telephone Number: | (704) 587-7227 |
| Fax Number:       | (704) 587-7250 |

Date Prepared: December 19, 2014

- 2. Device Name:
	- Proprietary Name: ●
	- SPIRIT Common Name: ● Dental Delivery Unit
		- Classification Name: Unit, Operative Dental
	- CFR Number: ●
		- 872.6640 Device Class:
		- I ● Product Code: EIA ●
- 3. Predicate Device:
	- ELEVANCE (K120239) Proprietary Name: ●

I

- . Common Name: Dental Delivery Unit
- Classification Name: Unit, Operative Dental ●
- CFR Number: 872.6640 ●
- Device Class: ●
- EIA . Product Code:
- 4. Description of Device:

The Spirit Dental Operative Units serves as a base that includes components to deliver air, water, electrical power, and vacuum to dental handpieces, instruments, and accessories. The controls are contained in a Doctor's Unit, an Assistant's Unit, and a Cuspidor. Additional parts include mount arms, foot control, and a junction box that houses a power supply and air/water regulators. Various Handpieces and accessories can be added to the Spirit Dental Operative Unit which Pelton & Crane does not manufacture but does provide a means to connect them into the Sprit Dental Operative Units. These include, but not limited to, pneumatic handpieces,

{4}------------------------------------------------

electric motors with handpieces, scalers, intra-oral cameras, curing lights, air/water syringes, SE and HVE vacuum instruments.

Per the Guidance for Industry and FDA Staff; Bundling Multiple Devices or Multiple Indications in a Single Submission, dated June 22, 2007, Pelton & Crane is bundling the Spirit dental operative unit models listed in Table 5.1 below as the models do not differ significantly in purpose, design, materials, energy source, function or any other feature related to substantial equivalence. The device description and intended use are identical for all configuration models listed in Table 5.1 below. The differences between the three series and mounting configurations are cosmetic in nature such as size/shape of the delivery head, mount arms, and instrument hangers. All critical components within the Spirit Dental Operative units are common.

The Spirit Dental Operative Units are available in three different series (3000, 2000, and 1000) which are offered in various mounting configurations such as chair mount traditional delivery, chair ellipse delivery, cabinet mount delivery, wall mount delivery, cart delivery, flexible work station delivery, and mobile work station delivery. The various mounting configurations for the Spirit dental operative units are illustrated within the Use and Care manuals. The differences between the Spirit 3000, 2000, and 1000 series are the Spirit 3000 series are the premier level dental operative units, the Spirit 2000 series are the mid-level dental operative units, and the Spirit 1000 series are the economy level dental operative units.

| Series | Model  | Description                                     | Integrated Dual Electric Motors (previously cleared under K103027 and K080677) |
|--------|--------|-------------------------------------------------|--------------------------------------------------------------------------------|
| 3000   | SDW 30 | Spirit 3000 Wall Delivery                       | Optional                                                                       |
| 3000   | SDC 30 | Spirit 3000 Cabinet Delivery                    | Optional                                                                       |
| 3000   | SCT30  | Spirit 3000 OTP w/Traditional Delivery          | Optional                                                                       |
| 3000   | SET30  | Spirit 3000 OTP Ellipse w/Traditional Delivery  | Optional                                                                       |
| 3000   | FWS30  | Spirit 3000 Flexible Work Station (Renaissance) | Optional                                                                       |
| 2000   | SCT20  | Spirit 2000 OTP w/Traditional Delivery          | N/A                                                                            |
| 2000   | SET20  | Spirit 2000 OTP Ellipse w/Traditional Delivery  | N/A                                                                            |
| 2000   | SDW-D  | Spirit 2000 Wall Delivery                       | N/A                                                                            |
| 2000   | SDC-D  | Spirit 2000 Cabinet Delivery                    | N/A                                                                            |
| 2000   | MWS-C  | Spirit 2500 Mobile Work Station                 | N/A                                                                            |
| 2000   | FWS-C  | Spirit 2000 Flexible Work Station (Centennial)  | N/A                                                                            |
| 2000   | FCT-C  | Spirit 2000 Flexible Cart                       | N/A                                                                            |
| 1000   | SCT15  | Spirit 1500 OTP w/Traditional Delivery          | N/A                                                                            |
| 1000   | SET15  | Spirit 1500 OTP Ellipse w/Traditional Delivery  | N/A                                                                            |
| 1000   | SCE15  | Spirit 1500 OTP w/Euro Delivery                 | N/A                                                                            |
| 1000   | SEE15  | Spirit 1500 OTP Ellipse w/Euro Delivery         | N/A                                                                            |
| 1000   | CRT15  | Spirit 1500 Cart                                | N/A                                                                            |
| 1000   | SDWD15 | Spirit 1500 Wall Delivery                       | N/A                                                                            |
| 1000   | SDCD15 | Spirit 1500 Cabinet Delivery                    | N/A                                                                            |
| 1000   | CD15   | Spirit 1500 Cabinet Delivery                    | N/A                                                                            |
| 1000   | SCT17  | Spirit 1700 OTP w/Traditional Delivery          | N/A                                                                            |
| 1000   | SET17  | Spirit 1700 OTP Ellipse w/Traditional Delivery  | N/A                                                                            |
| 1000   | RE15   | Spirit Cuspidor/PMU/Assistant's Unit            | N/A                                                                            |

### Table 5.1

{5}------------------------------------------------

| 2000 |       |                                          | N/A |
|------|-------|------------------------------------------|-----|
| 1000 | AV    | Spirit Cabinet Mount Assistant's Unit    |     |
| 3000 |       |                                          | N/A |
| 2000 |       |                                          |     |
| 1000 | RM    | Spirit Rear Chair Mount Assistant's Unit |     |
| 1000 | EX 15 | Spirit Executive Dental Unit             | N/A |

The Pelton & Crane Spirit dental operative units can also be equipped with available and already marketed Pelton & Crane products. These products include:

- Tip-A-Dilly Tip ●
- Tip-A-Dilly ●
- Tip - A
- Tip - D
- Tip E ●
- Tip F ● .
- Tip C

Optional accessories (devices) from other manufacturers that are integrated/attached to the Spirit dental operative units have already been cleared by the FDA.

#### Principle of Operation:

The delivery head is mounted to an arm mechanism for support and positioning of the delivery head around the patient. The types of mounting configurations include chair, cabinet, wall, cart, and floor mounts as illustrated within the Use and Care manuals. Depending on mounting configuration, a junction box or utility center provides housing for connections to the facility air, water, vacuum, and power sources, regulators for air and water, and transformers for optional integrated accessories. Regulated air and water, source vacuum, and power tubing and cables are routed through the mounting arms to the delivery head where the utilities are distributed to the individual dental instruments with a handpiece control system contained in the delivery head.

The handpiece control system is a pneumatic control system that distributes utilities to desired instruments that acts as "selected" instrument once removed from the respective instrument holder. The foot control activates the handpiece drive air, coolant air and water. Individual handpiece drive air and water flow adjustments are individually controlled by the operator via the control block and control valves. The delivery head is provided with master on-off switch to control air/water flow.

An optional foot control may be integrated into the system that acts as an intermediate manual user control for air and water activation to a selected instrument.

Suction instruments are not activated with the handpiece control system as the vacuum supply is manually controlled by the user with a flow control valve integrated in the instrument body. Also, air and water syringes are not activated with the handpiece control system as air and water flow is manually opened and closed with button actuated valves integrated into the syringe body.

{6}------------------------------------------------

Some handpieces may be power driven. The delivery head is provided with low voltage power to drive these handpieces. For some single and dual electrically driven motor configurations, the delivery system utilizes components and software from the device manufacturer, Kaltenbach & Voigt GMBH which was approved under 510(k) K103027 (ELECTROtorque TLC) and K080677 (COMFORTronic 4894 and COMFORTdrive 200XDA Handpiece) to control features and functionality of the motor(s). The Spirit dental operative unit provides the means to internally mount and house the electrically driven motor control and handpiece components and the utilities needed for operation (air, water, and low-voltage electricity). For the dual electrically driven motor configuration, an air activated switching board is used to drive the motors from one control board.

- 5. Indications for Use:
The Spirit dental operative units are intended to supply power to and serve as a base for other dental devices and accessories by providing air, water, vacuum and low voltage electrical power to hand held dental instruments. The Spirit dental operative units are intended for use by professional dental practitioners in providing treatment to dental patients in a dental operatory.

#### 6. Description of Substantial Equivalence: Technological Characteristics:

The Spirit Dental Operative Units, manufactured by Pelton & Crane, is being modified to have the ability to control two (2) electric motors. The Spirit Dental Operative Units has the same intended use as previously cleared Elevance Dental Operative Unit (K120239), manufactured by Midmark Corporation, but does have different technological characteristics. However, these different technological characteristics do not raise new concerns of substantial equivalence. The performance data and testing of the Pelton & Crane Spirit Dental Operative Units demonstrates substantial equivalence to the Midmark Elevance Dental Delivery Unit. Hence, the device is substantially equivalent.

# Table 5.2

| Feature             | Midmark Elevance Dental<br>Delivery Unit (K120239)                                                                                                                                                                                                                          | Pelton & Crane Spirit Dental<br>Operative Unit                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Midmark instrument delivery<br>systems are intended to<br>provide dental professionals<br>with air, water, and suction<br>along with low-voltage<br>electricity to operate dental<br>handpieces, syringes, and<br>accessories during dental<br>examinations and treatments. | The Spirit Dental Operative Units<br>are intended to supply power to<br>and serve as a base for other<br>dental devices and accessories by<br>providing air, water, vacuum and<br>low voltage electrical power to<br>hand held dental instruments. The<br>Spirit Dental Operative Units are<br>intended for use by professional<br>dental practitioners in providing<br>treatment to dental patients in a<br>dental operatory. |
| Regulation Number   | 21 CFR 872.6640                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Requlation Title    | Dental operative unit and<br>accessories                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                           |

{7}------------------------------------------------

| Feature                                 | Midmark Elevance Dental<br>Delivery Unit (K120239)                                                     | Pelton & Crane Spirit Dental<br>Operative Unit |  |  |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------|--|--|
| Regulation Class                        | I                                                                                                      | Same                                           |  |  |
| Product Code                            | EIA                                                                                                    | Same                                           |  |  |
| Device Classifications<br>(Electrical)  | Class I, Type B applied part,<br>IPX0, continuous operation                                            | Same                                           |  |  |
|                                         | Utilities and Standards                                                                                |                                                |  |  |
| Transportation /<br>Storage temperature | 23°F to 100°F                                                                                          | -68°F to 122°F                                 |  |  |
| Relative humidity<br>range              | 10% to 90%                                                                                             | Same                                           |  |  |
| Operating temperature<br>range          | 59°F to 95°F                                                                                           | 68°F to 76°F                                   |  |  |
| Air supply pressure<br>range            | 80-100 psi                                                                                             | 80-105 psi                                     |  |  |
| Air/oil separator                       | Gauze                                                                                                  | Same                                           |  |  |
| Water supply pressure<br>range          | 30-50 psi                                                                                              | 40-80 psi                                      |  |  |
| Isolated water bottle<br>system         | Optional                                                                                               | Same                                           |  |  |
| Standards                               | EN 60601-1-2:2007 Part 1-2                                                                             | EN 60601-1-2:2007 Part 1-2                     |  |  |
|                                         | EN 61000-3-:2006+A1:2009<br>+A2:2009 Part 3-2                                                          | EN 61000-3-2:2006+A1:2009<br>+A2:2009 Part 3-2 |  |  |
|                                         | IEC 60601-1 Part 1                                                                                     | ES 60601 -1 Part 1                             |  |  |
|                                         | ISO 7494-1:2004                                                                                        | ISO 7494-1:20041                               |  |  |
|                                         | ISO 7494-2:2003                                                                                        | ISO 7494-2:20031                               |  |  |
|                                         | User / Service Interface                                                                               |                                                |  |  |
| Number of user<br>accounts              | Three                                                                                                  | One2                                           |  |  |
| Setting display                         | LED screen                                                                                             | LCD screen2                                    |  |  |
| Screen navigation                       | Navigation arrows                                                                                      | Same                                           |  |  |
| Software updates                        | Via external USB port                                                                                  | Via internal 10-pin port2                      |  |  |
| Error tracking                          | Offered                                                                                                | Same                                           |  |  |
| Built-in diagnostics                    | Offered                                                                                                | Same                                           |  |  |
| Hand Held Devices                       |                                                                                                        |                                                |  |  |
| Optional accessories                    | Air/water syringe<br>Saliva ejector<br>HVE<br>Up to 2 micro motors<br>Scaler<br>Camera<br>Curing light | Same                                           |  |  |

{8}------------------------------------------------

| Feature                                              | Midmark Elevance Dental<br>Delivery Unit (K120239) | Pelton & Crane Spirit Dental<br>Operative Unit |
|------------------------------------------------------|----------------------------------------------------|------------------------------------------------|
| Accessory connection                                 | Pneumatic motor                                    |                                                |
| Accessory connection                                 | Integrated                                         | Same                                           |
| Number of hand piece<br>locations                    | 4-6                                                | Same                                           |
| Hand piece control<br>system                         | Kink Valve                                         | Valve block                                    |
| Syringe water flow<br>control                        | Adjustable                                         | Same                                           |
| Syringe air flow<br>control                          | Adjustable                                         | Same                                           |
| Coolant air flow<br>control                          | Adjustable                                         | Same                                           |
| Hand piece air and<br>water bypass                   | Offered                                            | Not applicable3                                |
| Number of hand piece<br>presets                      | Five                                               | Six2                                           |
| Remote hand piece<br>activation with water<br>toggle | Via foot control                                   | Same                                           |
| Positioning                                          |                                                    |                                                |
| Delivery unit head<br>positioning                    | Flex arm                                           | Same                                           |
| Flex arm brake release                               | Integrated                                         | Same                                           |
| Maximum load on flex<br>arm mounted units            | 10 lbs.                                            | Same4                                          |
| Additional Features                                  |                                                    |                                                |
| Endodontic capability                                | Offered                                            | Same                                           |
| Hand piece flush                                     | Standard feature                                   | Same                                           |
| Air/Water quick<br>connect ports                     | Offered                                            | Same                                           |
| Light control                                        | Offered                                            | Same                                           |
| Tray options                                         | Three                                              | One2                                           |
| Unit configurations for<br>dominant hand             | Left/Right                                         | Same                                           |

1 The Spirit Dental Operative Unit meets all requirements found in ISO 7494-1 and ISO 7494-2

² These differences do not affect substantial equivalence as they are only differences in marketing features between the proposed device and the predicate device

3 Hand piece air and water by pass is not necessary for the Spirit Dental Operative Unit as both air and water are pneumatically controlled rather than electronically controlled.

4 In the Spirit Dental Operative Unit manuals, it states: "The maximum weight capacity for the control head is 3 lbs." However, this only notes the limit of additional weight that can be placed on the control head. When considering the load of the control head itself and included accessories, the maximum load on the flex arm is 10 lbs.

{9}------------------------------------------------

#### Performance Data:

Electrical, mechanical, and performance testing according to standard AAMI ES60601-1, was performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. The Pelton & Crane Spirit dental operative units passed all tests. In addition to testing governed by requlatory standards, additional testing was conducted to ensure the technological characteristic differences between the Pelton & Crane Spirit Dental Operative Units and the predicate device, the Midmark Elevance Dental Delivery Unit (K120239), did not present any new concerns about substantial equivalence. The Pelton & Crane Spirit Dental Operative Unit passed all tests. Hence, the device demonstrates substantial equivalence.

Substantial equivalence of the Spirit dental operative unit with dual electric motors has been successfully evaluated with passing results via validation and verification testing.

Electrical Safety and EMC testing on the Spirit dental operative unit with dual electric motors was performed to confirm conformance.

Biocompatibility evaluation was conducted on patient contacting parts and found to be in conformance with ISO 10993-1.

Additionally, the Spirit Dental operative unit software was successfully validated to confirm the performance of the device per AMMI ANSI IEC 62304:2006 Medical Device Software. Software for the dual electric motors from the device manufacturer, Kaltenbach & Voiqt GMBH has already been cleared under 510(k) K103027 (ELECTROtorque TLC) and K080677 (COMFORTronic 4894 and COMFORTdrive 200XDA Handpiece). These two 510(k) devices are being integrated into the Spirit Dental Operative units and are listed in the optional accessories (devices) table. The testing also considered FDA Software Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

#### Clinical Performance Data:

Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product.

#### Conclusion as to Substantial Equivalence:

Based on the comparison of intended use, technological characteristics, and performance data, the minor differences between the Spirit dental operative units with dual electric motors, and the predicate device, ELEVANCE Delivery Unit (K120239), do not raise new concerns regarding substantial equivalence for the proposed indications of use. Pelton & Crane concludes that the Spirit dental operative units are substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA/K143696](https://fda.innolitics.com/submissions/DE/subpart-g%E2%80%94miscellaneous-devices/EIA/K143696)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
