DENTEK EASY BRUSH WITH FLUORIDE

K130186 · Den Tek Oral Care, Inc. · EFW · Jul 15, 2013 · Dental

Device Facts

Record IDK130186
Device NameDENTEK EASY BRUSH WITH FLUORIDE
ApplicantDen Tek Oral Care, Inc.
Product CodeEFW · Dental
Decision DateJul 15, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6855
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The Easy Brush with fluoride is intended to mechanically remove plaque and food particles from between teeth to reduce tooth decay. It is to be used with an in and out motion to clean interdental spaces, as it is done with dental floss.

Device Story

Manual interdental brush; consists of handle, wire, nylon bristles, and cap. Wire twisted around filaments, molded to handle, coated with flavor and sodium fluoride (approx. 0.15mg per brush). Used by patients for interdental cleaning via in-and-out motion; functions similarly to dental floss or toothpicks. Mechanical action removes plaque/debris; fluoride coating provides anti-caries benefit.

Clinical Evidence

No clinical data. Biocompatibility supported by reference to previously cleared devices containing sodium fluoride and established safety of fluoride in oral care applications.

Technological Characteristics

Manual interdental brush. Components: plastic handle, wire, nylon filaments, cap. Coated with flavor and sodium fluoride solution. Non-powered. No software.

Indications for Use

Indicated for individuals requiring mechanical removal of plaque and food particles from interdental spaces to reduce tooth decay. Suitable for over-the-counter use.

Regulatory Classification

Identification

A manual toothbrush is a device composed of a shaft with either natural or synthetic bristles at one end intended to remove adherent plaque and food debris from the teeth to reduce tooth decay.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) SUMMARY-Revised ### DenTek Oral Care Inc.'s Easy Brush with fluoride (K130186) ### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Linda Giles Date Prepared: January 24, 2013 Revision Date: June 12, 2013 #### Name of Device and Name/Address of Sponsor Easy Brush with fluoride DenTek Oral Care, Inc. 307 Excellence Way Maryville, TN 37801 Phone: (865) 983-1300 Facsimile: (865) 983-2444 #### Common or Usual Name Interdental brush #### Classification Name Manual toothbrush #### Classification Product Code EFW, Class I #### Predicate Devices Oral-B Woodsticks with Fluoride (K942633) and "Johnson & Johnson Fluoride Dental Floss, Mint Waxed" (K935440). #### Purpose of the Traditional 510(k) notice. The Easy Brush with fluoride is a modification to DenTek Oral Care Inc.'s Easy Brush. FDA device listing number B094614, on which a flavor and fluoride solution has been applied. Each brush contains approximately 0.15mg sodium fluoride in the flavor coating. {1}------------------------------------------------ ## Intended Use The Easy Brush with fluoride is intended to mechanically remove plaque and food particles from between teeth to reduce tooth decay. It is to be used with an in and out motion to clean between interdental space as it is done with floss, and other interdental cleaners such as tooth picks. # Technological Characteristics The Easy Brush with fluoride is an interdental brush designed to clean between teeth. It is composed of 4 parts (a handle, a wire, bristles, and a cap). The wire is twisted around the nylon filaments to create the brush, and then is molded onto a plastic handle. It is then coated with a solution containing flavor and sodium fluoride and then capped off. ## Substantial Equivalence The Easy Brush with fluoride is intended to mechanically remove plaque and food particles from between teeth to reduce tooth decay. As described in more detail in the Substantial Equivalence Chart-Revised, (page 4 of this submission), which lists the devices' similarities and differences, the Easy Brush with fluoride has the same intended use as "Oral-B Interdental Woodsticks with Fluoride" (K942633) and "Johnson & Johnson Fluoride Dental Floss, Mint Waxed" (K935440). ## Biocompatibility The biocompatibility of the device relies upon testing done on approved devices such as DENTSPLY 5% Sodium Fluoride Vamish K122331, which was reported to have no biocompatibility issues, and upon the predicate devices mentioned above which contain an amount of sodium fluoride that is substantially equivalent to the subject Easy Brush with fluoride. In addition, bottled drinking water and tap water are also approved to contain sodium fluoride in amounts proven to be safe and beneficial to the public as an anti-caries agent. Therefore, the Easy Brush with fluoride at a level of approximately 0.15 mg per brush is also biocompatible. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 #### July 15, 2013 DenTek Oral Care, Incorporated Ms. Linda Giles Regulatory Specialist 307 Excellence Way MARYVILLE TN 37801 Re: K130186 Trade/Device Name: Easy Brush with Fluoride Regulation Number: 21 CFR 872.6855 Regulation Name: Manual Toothbrush Regulatory Class: 1 Product Code: EFW Dated: June 12, 2013 Received: June 13, 2013 Dear Ms. Giles: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Giles Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mary S. Runner -S Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use Statement 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: Easy Brush with fluoride Indications for Use: The Easy Brush with fluoride is intended to mechanically remove plaque and food particles from between teeth to reduce tooth decay. It is to be used with an in and out motion to clean interdental spaces, as it is done with dental floss. Prescription Use . (Per 21 C.F.R. 801.109) AND/OR Over-The-Counter Use_X_ (Per 21 C.F.R. 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mary S. Runner -S | | <img alt="Signature" src="signature.png"/> | |--|--------------------------------------------| | | Susan Runner DDS MA | | | 2013.07.12 | |--|------------------| | | 14:13:46 -04'00' | (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices | 510(k) Number: | K130186 | |----------------|---------| |----------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...