K110888 · Dr. Fresh, Inc. · EEG · Jan 12, 2012 · Dental
Device Facts
Record ID
K110888
Device Name
ABSOLUTE WHITE LIGHT
Applicant
Dr. Fresh, Inc.
Product Code
EEG · Dental
Decision Date
Jan 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6475
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Absolute White Light emits visible blue light in the 400 nanometer spectrum and is intended to provide a light source for bleaching teeth.
Device Story
Absolute White Light is a battery-operated, hand-held device for consumer use. It emits visible blue light at 400 nm to activate photoactive substances in teeth, generating minimal heat. Safety features include a 2-minute automatic shut-off and a 'Light Guard' to prevent ocular exposure. The device is used in the oral cavity to assist in tooth whitening procedures. It functions as a light source in conjunction with a bleaching agent.
Clinical Evidence
Bench testing only. The device was evaluated for compliance with EN 61000-6-1 and EN 61000-6-3 electromagnetic compatibility standards.
Technological Characteristics
Hand-held, battery-operated LED light source. Emits 400 nm blue light. Includes automatic 2-minute shut-off timer and physical light guard. Manufactured per 21 CFR 820, DIN EN ISO 13485:2003, and DIN EN ISO 9001:2000. Complies with EN 61000-6-1 and EN 61000-6-3 EMC standards.
Indications for Use
Indicated for use as a light source for bleaching teeth in the oral cavity.
Regulatory Classification
Identification
A heat source for bleaching teeth is an AC-powered device that consists of a light or an electric heater intended to apply heat to a tooth after it is treated with a bleaching agent.
Predicate Devices
Dentovations Luster Whitening System (formerly South Beach Smile Light and Klear Action) (K042153)
Related Devices
K042153 — SOUTH BEACH SMILE LIGHT WHITENING SYSTEM · Dentovations, Inc. · Nov 8, 2004
K063704 — BRITEWHITE TEETH WHITENING SYSTEM, MODEL BW1101 · Beks Incorporated · Feb 20, 2007
K052040 — QUICKSMILE · Cosmetic Dental Materials, Inc. · Oct 18, 2005
K992620 — BLULAZE · Premier Laser Systems, Inc. · Aug 20, 1999
K122725 — TANDA PEARL, ULTRAWYTE, PEARL · Syneron Beauty · Nov 27, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
K1108888
## ATTACHMENT II
# 510(K) SUMMARY (AS PER 21 CFR 807.92)
# JAN 1 2 2012
### I. GENERAL INFORMATION
Device Generic Name: Whitening Light
Trade Name: Absolute White Light
Device Classification: Heat Source for Bleaching Teeth (21 CFR 872.6475)
Product Code: EEG
Applicant Name and Address:
Dr. Fresh, Inc. 6645 Caballero Blvd. Buena Park, CA 90620
Contact Name: Gary Pendyala 714-690-1573
Date of Submission: December 15, 2011
510(k) Number: K110888
### II. Device Description
The Absolute White Light is intended for use by the end consumer. The Absolute White Light is a battery operated, hand held device that contains a hun vosible light in the 400 nanometer spectrum, which can penetrate the tooth and activate the photoactive substances within the tooth yielding a minimal amount of heat. Built in safety features include automatic light shut off after two minutes and "Light Guard" that prevents light exposure to the user's eyes.
### III. Indications for Use
The Absolute White Light emits visible blue light in the 400 nanometer spectrum and is intended to provide a light source for bleaching teeth.
{1}------------------------------------------------
### IV. Predicate Device
The Absolute White Light is substantially equivalent to other tooth whitening lights currently in commercial distribution such as the Dentify while in the Luster Whitening System (formerly South Beach Smile Light and Klear Action) as cleared by the FDA under Accession Number K042153).
### Summary of the Technical Characteristics of the Absolute White Light as > Related to the Referenced Predicate Devices.
The Absolute White Light and the aforementioned predicate device are light sources for bleaching teeth as defined in 21 CFR 872.6475.
The Absolute White Light is a conventional tooth whitening light which works in conjunction with a bleaching agent to provide a light which which works in The Absolute White Light has the same intended use and similar technological characteristics to the predicate device.
| Product | Absolute White<br>Light | Dentovations Luster<br>Light |
|------------------------|---------------------------------|----------------------------------|
| Intended Use | Light source Teeth<br>Bleaching | Light source, Teeth<br>Bleaching |
| Method of Use | Hand Held | Hand Held |
| Treatment<br>Time | 2 minutes | 2 minutes |
| Treatment(s) | As Needed | As Needed |
| Area of Use | Oral Cavity | Oral Cavity |
| Light Source | LED | LED |
| Wavelength of<br>Light | 400 nm | 350 – 700 nm |
| Color of Light | Blue | Blue |
| Presentation | Non-Sterile | Non-Sterile |
## VI. Performance Date
The device is manufactured in compliance with Quality System meeting the requirements of the Food and Drug Ministration's Quality System Regulations (21 CFR 820), DIN EN ISO13485:2003 and / or DIN EN ISO 9001:2000. The device is manufactured to comply with recognized consensus and safety standards listed below:
- EN 61000-6-1 Electromagnetic compatibility (EMC) Part 6-1: . Generic standards - Immunity standard for residential, commercial and light-industrial environments
{2}------------------------------------------------
- EN61000-6-3 Electromagnetic compatibility (EMC) Part 6-3: . Generic standards ~ Emission standard for residential, commercial and light-industrial environments
# Conclusion
Pursuant to the testing and comparison to the predicate device, the Absolute White Light has the same intended use, with similar function and performance characteristics. Therefore, the Company believes that the Absolute White Light is substantially equivalent to currently marketed devices, such as the Dentovations Luster Whitening System.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Dr. Fresh, Incorporated C/O Ms. Joyce Heinrich Regulatory Consultant Texas Applied Biomedical Services 12101 Cullen Boulevard Houston, Texas 77047-1131
JAN 1 2 2012
Re: K110888
Trade/Device Name: Absolute White Light Regulation Number: 21 CFR 872.6475 Regulation Name: Heat Source for Bleaching Teeth Regulatory Class: I Product Code: EEG Dated: December 20, 2011 Received: January 5, 2012
Dear Ms. Heinrich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Ms. Heinrich
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
## ATTACHMENT I
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K110888
Device Name:
Absolute White Light
### Indications for Use:
The Absolute White Light emits visible blue light in the 400 nanometer spectrum and is intended to provide a light source for bleaching teeth.
AND/OR Over the Counter Use: Prescription Use: J X (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODPE)
Susan Ranne
(Division Sign-Off) Division of Anesthesiology, General Hospltal Infection Control, Dental Devices
510(k) Number: JC110555
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.