K200214 · 3D Smile USA, Inc. · NXC · Feb 24, 2020 · Dental
Device Facts
Record ID
K200214
Device Name
3D Predict
Applicant
3D Smile USA, Inc.
Product Code
NXC · Dental
Decision Date
Feb 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
3D Smile Predict aligners are indicated for the alignment of teeth in patients with permanent dentition (i.e. all second molars) during orthodontic treatment of malocclusion.
Device Story
3D Smile Predict is a prescription-only, clear, removable thermoplastic aligner system. Input: digital scan or traditional PVS impression of patient dentition. Process: specialized orthodontic CAD/CAM software generates 3D models of initial, intermediate, and final tooth states based on clinician prescription; clinician reviews and approves treatment plan; software converts approved plan into 3D files for mold production; aligners are fabricated by thermoforming plastic over 3D-printed molds. Output: series of sequential aligners. Usage: clinic setting; operated by orthodontist or dentist. Clinician monitors treatment throughout; patient wears aligners sequentially to apply continuous gentle force for tooth repositioning. Benefit: incremental correction of malocclusion.
Clinical Evidence
No clinical data included. Performance supported by established clinical history of sequential aligners (product code NXC) and manufacturing validation of dimensional accuracy (3D printed molds, thermoformed adaptation). Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Non-sterile, biocompatible thermoplastic material. Sensing/actuation: continuous gentle force via sequential aligner displacement. Connectivity: CAD/CAM software-based treatment planning. Manufacturing: 3D printing of molds followed by thermoforming. Software: orthodontic CAD/CAM for 3D modeling and treatment planning.
Indications for Use
Indicated for alignment of teeth in patients with permanent dentition (all second molars) during orthodontic treatment of malocclusion.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
K191823 — Clear-Aligners · Cdb Corporation · Dec 20, 2019
K182329 — SureCure Orthodontic Aligner System · Digital Orthodontic Care · Nov 8, 2018
K211537 — QuickAligners · Oral Image, Inc. · Oct 26, 2021
K192244 — Arcad SmileStudio and Aligner System · Arcadlab · May 11, 2020
K212803 — DailyMate Orthodontic Aligner System · 3D Global Biotech, Inc. · May 23, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
February 24, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in smaller blue letters below.
3D Smile USA, Inc % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K200214
Trade/Device Name: 3D Predict Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: January 24, 2020 Received: January 28, 2020
Dear Prithul Bom:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Srivinas 'Nandu' Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# 510(k) Summary 3D Smile USA, Inc 3D Smile Predict 12/24/2019
### 5.1 ADMINISTRATIVE INFORMATION
| Manufacturer Name | 3D Smile USA, Inc |
|-------------------|------------------------|
| | 3675 Market Street |
| | Suite #200 |
| | Philadelphia, PA 19104 |
| | USA |
| Telephone: | +1 (920) 332 4344 |
| Fax: | n/a |
| Official Contact | Marina Domracheva, CEO |
|------------------|--------------------------------|
| Email: | marina.domracheva@3d-smile.com |
### 5.2 DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | 3D Smile Predict |
|-------------------------|----------------------|
| Common Name: | Aligners, sequential |
| Classification Name: | Orthodontic Plastic Bracket |
|-----------------------------|----------------------------------------------------------------------------------------------|
| Classification Regulations: | 21 CFR 872.5470 |
| Device Class: | Class II |
| Product Code: | NXC |
| Review Panel: | Dental Products Panel |
| Reviewing Branch: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)<br>Dental Devices (DHT1B) |
#### PREDICATE DEVICE INFORMATION 5.3
The devices within this submission are substantially equivalent in indications, intended use and design principles to the following predicate and reference devices:
| 510(k) | Predicate Device Name | Company Name |
|---------|-----------------------|------------------------|
| K191990 | OSW Aligner System | OSW Manufacturing, LLC |
## 5.4 DEVICE DESCRIPTION
The 3D Smile Predict aligner system is a series of prescription-only clear plastic removable aligners intended to incrementally move a patient's teeth from an initial state to a different end state using a software-generated sequence of intermediate states. 3D Smile Predict Aligners sequentially reposition teeth by way of continuous gentle force.
A digital or traditional mold impression of the patient's teeth is provided by a dental health professional (e.g. orthodontist or dentist). From the digital data created of the patient's teeth, specialized orthodontic CAD/CAM software is used to develop treatment plans. Using the software, dental technicians design a series of intermediate models corresponding to each stage of treatment, gradually realigning the patient's teeth according to the dental health professional's prescription.
{3}------------------------------------------------
The prescribing doctor reviews and approves the model scheme and treatment plan before the molds/models are produced.
Once approved, the specialized orthodontic software is used to generate standard format 3D files which are used to physically produce each model/mold in the treatment plan for aligner fabrication. 3D Smile USA, Inc produces the aligner trays by thermoforming a plastic sheet over each model in the treatment plan. The aligner trays are sent to the dental health professional who then delivers them to the patient, confirming fit and design of the first stage of treatment. Additional trays are used sequentially by the patient to gradually move the teeth to the desired position. The dental health professional monitors treatment from the moment the first aligner is delivered to when the final aligner is finished and treatment complete. The aligners are held in place by pressure and can be removed by the patient at any time.
The technology is identical to that used by the Predicate device, OSW Aligner System (K191990) and a number of other sequential aligner systems currently being legally marketed.
### 5.5 INDICATIONS FOR USE
3D Smile Predict aligners are indicated for the alignment of teeth in patients with permanent dentition (i.e. all second molars) during orthodontic treatment of malocclusion.
### EQUIVALENCE TO MARKETED DEVICE 5.6
Overall, the Subject device is substantially equivalent to the Predicate device with respect to Indications for Use and technological principles. The Comparison table below compares parameters and characteristics of the Subject device and Predicate/reference devices.
| Parameter | Subject Device | Predicate Device |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | 3D Smile Predict<br>3D Smile USA, Inc | OSW Aligner System<br>OSW Manufacturing, LLC<br>K180241 |
| Regulation # | 21 CFR 872.5470 | 21 CFR 872.5470 |
| Classification Name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket |
| Product Code | NXC | NXC |
| Classification | Class II | Class II |
| Indications for Use | 3D Smile Predict aligners are indicated for the alignment of teeth in patients with permanent dentition (i.e. all second molars) during orthodontic treatment of malocclusion. | OSW Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion. |
| Mode of action | Orthodontic movement occurs through continuous gentle forces applied to the dentition as each tooth follows the programmed displacement based on a doctor's prescription. | Orthodontic movement occurs through continuous gentle forces applied to the dentition as each tooth follows the programmed displacement based on a doctor's prescription. |
| Method of use | Each appliance is worn by the patient as determined by the dental practitioner, generally 2 weeks prior to being replaced by the next aligner in sequence. | Each appliance is worn by the patient as determined by the dental practitioner, generally 2 weeks prior to being replaced by the next aligner in sequence. |
| 3D Software<br>Description | The 3D Smile Predict 3-D Software uses a scan of a PVS impression, physical or a digital scan (which represents an untreated state) to generate the image of a final, provisional treated state and then interprets a series of images that represent intermediate teeth states. The dental practitioner then reviews these images and has the option to reject or request modifications to the set-up prior to approving it for aligner fabrication. Once the dental practitioner approves the treatment plan, the software converts the files to produce the series of custom-made aligners | The OSW Manufacturing 3-D Software uses a scan of a PVS impression or a digital scan (which represents an untreated state) to generate the image of a final, provisional treated state and then interprets a series of images that represent intermediate teeth states. The dental practitioner then reviews these images and has the option to reject or request modifications to the set-up prior to approving it for aligner fabrication. Once the dental practitioner approves the treatment plan, the software converts the files to produce the series of custom-made aligners |
## Predicate Device Comparison Table
{4}------------------------------------------------
| Parameter | Subject Device | Predicate Device |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 3D Smile Predict | OSW Aligner System |
| | 3D Smile USA, Inc | OSW Manufacturing, LLC |
| | | K180241 |
| Function of the 3D<br>software | The 3D Smile 3-D Software performs the following<br>operations:<br>• Produce 3D-model file of the patient's dentition<br>from a digital scan.<br>• Generates a treatment plan (i.e. 3-D models that<br>represent the treatment plan) called a 3D plan based<br>on the prescribing doctor's instructions. The treating<br>dental practitioner reviews these images using 3D<br>Smile software and has the option to reject or<br>request modifications to the set-up prior to<br>approval. | OSW Aligner System 3-D software performs the<br>following operations:<br>• Produce 3D-model file of the PVS impression or<br>digital scan.<br>• Identifies the individual teeth that will require<br>treatment (i.e. repositioning).<br>• Creates a treatment plan (i.e. 3-D models that<br>represent the treatment plan). The treating dental<br>practitioner reviews these images using OSWDP<br>software and has the option to reject or request<br>modifications to the set-up prior to approval. |
| Material | Thermoplastic | 0.03" thick, thermoformed polyurethane |
| Appliance Application | Removable | Removable |
| Design | Image: 3D Smile Predict Aligner | Image: OSW Aligner System |
| Biocompatible | Yes | Yes |
| OTC or Rx | Rx | Rx |
| Sterile | Non-sterile | Non-sterile |
The wording of the Indications for Use of the Subject device is slightly different than that of the Predicate device. At this time, the Subject device is indicated for patients with permanent dentition. This does not change the intended use of both Subject and Predicate devices to be used in the alignment of teeth during orthodontic treatment of malocclusion.
## 5.7 TECHNOLOGICAL CHARACTERISTICS
Orthodontic tooth movement occurs through forces applied to the teeth by the appliance as each tooth follows the predetermined displacement based on a dental health professional's prescription. The Subject device mode of action and method of use are identical to the Predicate device and supports a determination of substantial equivalence.
Even if the specific features or process of the Software used in treatment planning differs, the use of Software in treatment planning supports a determination of substantial equivalence. Most importantly, both verified and validated software programs rely on the prescribing doctor's approval of the treatment plan. The Subject and Predicate devices are both fabricated of a non-sterile, biocompatible thermoplastic material which supports a determination of substantial equivalence.
## 5.8 PERFORMANCE DATA
Due to the difficulty in evaluating this type of dental device in a laboratory environment, no direct performance bench testing of the aligners was performed. The use of thermoplastic materials for sequential aligners intended to treat malocclusions have been well documented in scientific literature regarding incremental tooth moving forces.
A manufacturing validation was performed to demonstrate the dimensional accuracy of the manufacturing process for 3D Smile Predict Aligners. Three critical aspects of the manufacturing process were assessed for accuracy: 3D printed molds, thermoformed aligner adaptation to the mold, thermoformed aligner to the software output.
Inspection software was used to perform point-to-point and critical displacement measurements.
{5}------------------------------------------------
All measurements were within the acceptance criteria. These test have met the pre-established acceptance criteria to demonstrate dimensional accuracy.
An aligner fit validation was performed to confirm fit of aligners on the patient.
## 5.9 CLINICAL TESTING
The performance of sequential aligners in the clinical environment has been well established since the first such devices were cleared by the FDA in 1998 under product code NXC. There no clinical testing is required to support 3D Smile Predict Aligners, as the Indications for Use is equivalent to the Predicate device, which also was not subject to clinical testing. No clinical data is included in this submission.
## 5.10 BIOLOGICAL TESTING
Biocompatibility testing for the aligner material, the only patient contacting material, was conducted in accordance with International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". The results of the tests satisfy the requirements of the study protocols and comply with ISO 10993-1 for the intended use.
The following biological tests were performed:
| Biological Endpoint | Relevant Standard |
|------------------------------|-------------------|
| Cytotoxicity | ISO 10993-5:2009 |
| Sensitization | ISO 10993-10:2010 |
| Irritation | ISO 10993-10:2010 |
| Subacute/SubChronic Toxicity | ISO 10993-11:2006 |
| Genotoxicity | ISO 10993-3:2014 |
## 5.11 CONCLUSION
Overall, the Indications for Use statement for the Subject and Predicate devices are substantially equivalent.
Overall, the Technological Characteristics, Materials, Prescription Use and Non-sterile status of the Subject device are substantially equivalent to the Predicate device. The use of Software to produce the Subject and Predicate devices is substantially equivalent.
Overall, the 3D Smile Predict aligners are substantially equivalent to the Predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.