K223518 · Shanghai EA Medical Instruments Co., Ltd. · PNN · Jun 13, 2023 · Dental
Device Facts
Record ID
K223518
Device Name
iOrtho
Applicant
Shanghai EA Medical Instruments Co., Ltd.
Product Code
PNN · Dental
Decision Date
Jun 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
iOrtho is intended for use as a medical front-end device providing tools for management of orthodontic cases, systematic inspection, detailed analysis, treatment simulation and virtual applians (Export of Models, Indirect Bonding Transfer Media, Sequential aligners) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of iOrtho requires the user to have the necessary training and domain knowledge in the practice of orthodonties, as well as to have received a dedicated training in the use of the software.
Device Story
iOrtho is a standalone orthodontic software platform for dental professionals. It ingests 3D digital dental scans (STL files) to facilitate case management, diagnosis, and treatment planning. The software provides tools for auto-segmentation of gums/bone (with manual adjustment), measurement point selection, and 3D treatment simulation. It enables the design of virtual appliances, including sequential aligners and indirect bonding transfer media. The system outputs 3D models for fabrication via 3D printing. Used in clinical settings by orthodontists to monitor treatment progress and verify outcomes against planned objectives. The software aids clinical decision-making by allowing practitioners to visualize tooth movement and refine treatment plans. It benefits patients by enabling customized, precise orthodontic appliance design and progress tracking.
Clinical Evidence
No clinical data. Bench testing only. The device underwent integration, verification, and validation testing in accordance with FDA software guidance and IEC 62304 standards to confirm performance per intended use.
Technological Characteristics
Standalone software; compatible with Windows 7/Mac OS 10.12+; requires WebGL-capable video card, 2GB RAM, and internet access. Features include 3D model import (STL), auto-segmentation with manual adjustment, automated measurement point selection, and virtual appliance design. Operates via web browser (Chrome, Firefox, Safari, Edge).
Indications for Use
Indicated for orthodontic patients requiring management, inspection, analysis, treatment simulation, and virtual appliance design based on 3D dentition models. Applicable before, during, and at the end of orthodontic treatment to evaluate progress and outcomes. Requires trained orthodontic professionals.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 13, 2023
Shanghai EA Medical Instruments Co., Ltd. % Breanne Butler Regulatory Affairs Consultant Prime Path Medtech 1321 Upland Dr. Suite 6792 Houston, Texas 77043
Re: K223518
Trade/Device Name: iOrtho Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN Dated: May 19, 2023 Received: May 19, 2023
Dear Breanne Butler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223518
Device Name iOrtho
#### Indications for Use (Describe)
iOrtho is intended for use as a medical front-end device providing tools for management of orthodontic cases, systematic inspection, detailed analysis, treatment simulation and virtual applians (Export of Models, Indirect Bonding Transfer Media, Sequential aligners) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of iOrtho requires the user to have the necessary training and domain knowledge in the practice of orthodonties, as well as to have received a dedicated training in the use of the software.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY - K223518
A summary of 510(k) substantial equivalence information for this traditional 510(k) in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Shanghai EA Medical Instruments Co., Ltd.<br>No.1619 Huishan Avenue, Huishan Economic Development Zone<br>Wuxi, Jiangsu Province<br>China |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact Person:<br>Phone:<br>Email: | Jessica Luo, Regulatory Affairs, Angel Align<br>+86.0510.83591717(181)<br>luoyuqing@angelalign.com |
| Submission Correspondent:<br>Address:<br>Phone:<br>Email: | Breanne Butler, Regulatory Affairs Consultant<br>1321 Upland Dr. Suite 6792 Houston, TX 77043<br>860-810-5594<br>bbutler@primepathmedtech.com |
| Date Prepared: | June 12, 2023 |
| Trade Name Name: | iOrtho |
| Common Name: | Orthodontic plastic bracket (Software) |
| Classification Name: | Orthodontic plastic bracket (Software) |
| Product Code: | PNN |
| Device Classification: | Class II, 21 CFR 872.5470 |
| Predicate Device: | 3Shape A/S Ortho System (K171634) |
| Reference Device: | Wuxi EA Medical Instruments Technologies Limited Clear Aligner<br>(K203688) |
| Reference Device: | Dentsply Sirona CEREC Ortho Software (K171122) |
## Device Description:
iOrtho (hereafter referred to as "Proposed Device") includes modifications to the currently marketed software included in K203688, cleared October 8, 2021 (hereafter referred to as "Reference Device"). The Proposed Device is an orthodontic appliance design and treatment simulation software is for use by dental professionals to aid in diagnosis and design solutions for patients. Digital scans (3D) of a patient's dentition can be loaded into the software and the dental professional can then create treatment plans for each individual patient and their needs. The system can be used to fabricate 3D dental models
{4}------------------------------------------------
using standard stereolithographic (STL) files for use in 3D printers. These models can then be used as a template for thermoforming aligners or retainers by Angel Align technicians.
## Indications for Use:
iOrtho is intended for use as a medical front-end device providing tools for management of orthodontic cases, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media, Sequential aligners) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of iOrtho requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software. Comparison to Predicate Devices:
The Proposed Device is functionally equivalent to the following device: 3Shape A/S Ortho System (K171634, cleared Jan 17, 2018) (hereafter referred to as "Predicate Device"), and possesses minor differences to the previously cleared Reference Device to allow treating physicians include mandible repositioning and traction accessories during treatment planning and simulation. The following table demonstrates the functional specifications of the Proposed Device are substantially equivalent to the Predicate Device, minorly different to the Reference Device, and raises no new questions regarding safety and effectiveness of the device.
| | Proposed Device: | Predicate Device: | Reference | Reference | |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Specification | iOrtho | 3Shape A/S Ortho<br>System (K171634) | Device: Wuxi<br>EA Medical<br>Instruments<br>Technologies<br>Limited. Clear<br>Aligner<br>(K203688) | Device:<br>Dentsply<br>Sirona<br>CEREC<br>Ortho<br>Software<br>(K171122) | Comparison<br>Result |
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | 21 CFR 872.5470 | 21 CFR<br>872.5470 | Same |
| Classification<br>Name | Orthodontic Plastic<br>Bracket (Software) | Orthodontic Plastic<br>Bracket (Software) | Orthodontic<br>Plastic Bracket | Orthodontic<br>Plastic Bracket<br>(Software) | Same as<br>Predicate |
| Product Code | PNN | PNN | NXC | PNN | Same |
| Classification | Class II | Class II | Class II | Class II | Same |
| Indication for<br>Use | iOrtho | The 3Shape Ortho<br>System™<br>is intended for use<br>as a medical front-end<br>device<br>providing tools for<br>management of<br>orthodontic | Clear Aligners are<br>indicated for the<br>alignment of<br>teeth during<br>orthodontic<br>treatment of<br>malocclusion<br>in patients with | CEREC Ortho<br>Software is<br>intended for<br>use with image<br>data acquired<br>from handheld<br>intra oral 3D<br>cameras and | Similar to<br>predicate |
| | is intended for use<br>as a medical front-<br>end device | is intended for use<br>as a medical front-end<br>device | alignment of | intended for | |
| | providing tools for<br>management of<br>orthodontic cases. | providing tools for<br>management of<br>orthodontic | teeth during | use with image | |
| | orthodontic cases. | orthodontic | in patients with | cameras and | |
| Specification | Proposed Device:<br>iOrtho | Predicate Device:<br>3Shape A/S Ortho<br>System (K171634) | Reference Device: Wuxi<br>EA Medical<br>Instruments<br>Technologies<br>Limited. Clear<br>Aligner<br>(K203688) | Reference Device:<br>Dentsply<br>Sirona<br>CEREC<br>Ortho<br>Software<br>(K171122) | Comparison<br>Result |
| | systematic<br>inspection, detailed<br>analysis, treatment<br>simulation and<br>virtual appliance<br>design options<br>(Export of Models,<br>Indirect Bonding<br>Transfer Media,<br>Sequential aligners)<br>based on 3D models<br>of the patient's<br>dentition before the<br>start of an<br>orthodontic<br>treatment. It can<br>also be applied<br>during the<br>treatment to<br>inspect and analyze<br>the progress of the<br>treatment. It can be<br>used at the end of<br>the treatment to<br>evaluate if the<br>outcome is<br>consistent with the<br>planned/desired<br>treatment<br>objectives. | models, systematic<br>inspection, detailed<br>analysis, treatment<br>simulation and<br>virtual appliance<br>design options<br>(Custom Metal<br>Bands, Export of<br>Models, Indirect<br>Bonding Transfer<br>Media) based on<br>3D models of the<br>patient's dentition<br>before the start of<br>an orthodontic<br>treatment. It can<br>also be applied<br>during the<br>treatment to<br>inspect and analyze<br>the progress of the<br>treatment. It can<br>be used at the end<br>of the treatment to<br>evaluate if the<br>outcome is<br>consistent with the<br>planned/desired<br>treatment<br>objectives. | permanent<br>dentition (i.e. all<br>second molars). | desktop<br>laboratory<br>scanners to<br>create 3D<br>virtual models<br>to be used for<br>data acquisition<br>and modeling<br>analysis for<br>orthodontic<br>patients and<br>conditions. The<br>CEREC Ortho<br>Software 3D<br>model data can<br>be exported to<br>orthodontic<br>design software<br>to aid in the<br>design of<br>orthodontic<br>appliances. | |
| | The use of iOrtho<br><br>requires the user to<br>have the necessary<br>training and domain<br>knowledge in the<br>practice of | The use of the<br>Ortho System™<br><br>requires the user<br>to have the<br>necessary training<br>and domain<br>knowledge in the | | | |
| | | | | | |
| Specification | Proposed Device:<br>iOrtho | Predicate Device:<br>3Shape A/S Ortho<br>System (K171634) | Reference<br>Device: Wuxi<br>EA Medical<br>Instruments<br>Technologies<br>Limited. Clear<br>Aligner<br>(K203688) | Reference<br>Device:<br>Dentsply<br>Sirona<br>CEREC<br>Ortho<br>Software<br>(K171122) | Comparison<br>Result |
| | orthodontics, as<br>well as to have<br>received a<br>dedicated training<br>in the use of the<br>software. | practice of<br>orthodontics, as<br>well as to have<br>received a<br>dedicated training<br>in the use of the<br>software. | | | |
| Technological<br>Features | Stand Alone Software Imports Digital Dental Data Useful for as an aid in Diagnosis, and treatment planning Virtual Planning of tooth movement Patient follow-up monitoring and management Cephalometry | Stand Alone Software Imports Digital Patient Scans Can be used to design Dental Casts Useful for Diagnosis, treatment planning, and CAD design Virtual Planning of tooth movement | Produce 3D-model file of the PVS impression or digital scan. Identifies the individual teeth that will require treatment (i.e. repositioning) Creates a treatment plan (i.e. 3-D models that represent the treatment plan). The treating dental practitioner reviews these images using software and has the option to reject or request modifications | Stand Alone Software Imports Digital Patient Scans Can be used to design Dental Models | Similar to<br>Predicate |
| Specification | Proposed Device:<br>iOrtho | Predicate Device:<br>3Shape A/S Ortho<br>System (K171634) | Reference<br>Device: Wuxi<br>EA Medical<br>Instruments<br>Technologies<br>Limited. Clear<br>Aligner<br>(K203688) | Reference<br>Device:<br>Dentsply<br>Sirona<br>CEREC<br>Ortho<br>Software<br>(K171122) | Comparison<br>Result |
| Minimum<br>Hardware/Soft<br>ware<br>Requirements | • OS: Windows<br>7 )/Mac OS 10.12<br>• RAM: 2 GB<br>• Video Card: Any<br>that supports<br>webGL<br>• Hard Drive Space:<br>50MB<br>• Web browser<br>(iOrtho): Chrome<br>63+, Firefox 58+,<br>Safari 12+, Edge<br>79+<br>• Internet Access | • OS: Windows 7,<br>8, 10 64-bit<br>• RAM: 8 GB<br>• Video Card<br>Memory: 1 GB<br>• Hard Drive<br>Space: 250 GB<br>• CPU: Intel Core<br>i5 or equivalent<br>• Mouse: with<br>wheel button | • OS: Windows<br>7, 10, Mac OS<br>X<br>• RAM: 2 GB<br>(minimum)<br>• Hard Drive<br>Space: 50 MB<br>• Video Card:<br>Graphics<br>Display Card<br>supporting<br>webGL<br>• Web Browser:<br>Chrome 9+,<br>Firefox 4+,<br>Safari 5.1+,<br>Edge 5.1+,<br>IE11+<br>• Internet<br>Access | • OS:<br>Windows 7,<br>64-bit<br>• RAM: 8 GB<br>• Video Card<br>Memory: 1<br>GB<br>• Hard Drive<br>Space: 250<br>GB<br>• CPU: Intel<br>QuadCore<br>1.6 GHz<br>processor | Similar |
| Login Method | Username and<br>password | Username and<br>password | Username and<br>password | Unknown | Same |
| Supported<br>Anatomic Areas | Maxilla/Mandible | Maxilla/Mandible | Maxilla/Mandibl<br>e | Maxilla/Mandib<br>le | Same |
| Intended Use | | | | | |
| Specification | Proposed Device:<br>iOrtho | Predicate Device:<br>3Shape A/S Ortho<br>System (K171634) | Reference<br>Device: Wuxi<br>EA Medical<br>Instruments<br>Technologies<br>Limited. Clear<br>Aligner<br>(K203688) | Reference<br>Device:<br>Dentsply<br>Sirona<br>CEREC<br>Ortho<br>Software<br>(K171122) | Comparison<br>Result |
| Managing<br>Patient and<br>case base data | Yes | Yes | Yes | Yes | Same |
| Collection of<br>study material | Yes | Yes | Yes | Yes | Same |
| Segmenting of<br>study material<br>(Segmenting<br>gums/bone) | Yes — auto-<br>segmented with<br>manual adjustment | Yes — manual<br>selection | Yes — manual<br>selection | Yes — auto-<br>segmented<br>with manual<br>adjustment | Similar |
| Alignment of<br>study material | Yes | Yes | Yes | Yes | Same |
| Measuring<br>study material | Yes | Yes | Yes | Yes | Same |
| Measurement<br>point selection | Automatic with<br>manual adjustment | Manual selection | Manual selection | Automatic with<br>manual<br>adjustment | Similar |
| Analyzing study<br>material | Yes | Yes | Yes | Yes | Same |
| Treatment<br>simulation | Yes | Yes | Yes | N/A | Same |
| Treatment<br>Options | Up to unlimited<br>number of stages | No limit on number<br>of stages | Up to unlimited<br>number of stages | N/A | Similar |
| Virtual<br>appliance<br>design | Yes | Yes | Yes | Yes | Same |
| Surface scan<br>for intraoral<br>scanner | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.