BRIUS Software Suite is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, treatment simulation and virtual appliance design options (Patient-Specific Nitinol Wires, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of the BRIUS Software Suite requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
Device Story
BRIUS Software Suite is a browser-based orthodontic CAD/CAM software for treatment planning and appliance design. Input consists of 3D digital dental scans (STL files). Users (orthodontists/technicians) perform systematic inspection, treatment simulation, and virtual appliance design. Output includes 2D DXF files for laser-cutting nitinol sheets and 3D STL files for fabricating high-temperature molds to shape-set nitinol wires. The software facilitates the creation of patient-specific nitinol wires and indirect bonding transfer media. Clinicians use the software to plan tooth movement, approve designs, and evaluate treatment progress or outcomes against objectives. The system supports clinical decision-making by providing a digital environment for virtual orthodontic planning and custom device fabrication.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including integration, verification, and validation testing performed in accordance with the FDA guidance document 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices'.
Technological Characteristics
Standalone software module; browser-based (Windows/Mac). Supports STL file import/export. CAD/CAM technology for virtual tooth movement planning and appliance design. No hardware components; requires Windows 10 64-bit, 8GB RAM, 10GB disk space. Login via username/password.
Indications for Use
Indicated for orthodontic patients requiring management of dental models, treatment simulation, and design of patient-specific nitinol wires, indirect bonding transfer media, and model exports. Requires trained orthodontic professionals.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue.
April 17, 2020
BRIUS % Breanne Butler Consultant Medavice. Inc 1321 Upland Suite 6792 Dr. Houston, Texas 77043
Re: K191720
Trade/Device Name: BRIUS Software Suite Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN Dated: March 16, 2020 Received: March 17, 2020
Dear Breanne Butler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191720
Device Name BRIUS Software Suite
#### Indications for Use (Describe)
BRIUS Software Suite is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, treatment simulation and virtual applians (Patient-Specific Nitinol Wires, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the BRIUS Software Suite requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
| Submitter: | BRIUS<br>4553 Westgrove Dr.<br>Addison TX 75001 |
|------------------------------|-----------------------------------------------------------------|
| Company Contact<br>Person: | Mehdi Roein-Peikar |
| Phone: | (888) 755.8320 |
| Email: | mehdi@brius.com |
| Submission<br>Correspondent: | Medavice, Inc.<br>Breanne Butler, Regulatory Affairs Consultant |
| Address: | 1321 Upland Suite 6792 Dr. Houston TX 77043 |
| Phone: | (860) 810-5594 |
| Email: | bbutler@medavice.com |
| Date Prepared: | 04/16/2020 |
| Proprietary Name: | BRIUS Software Suite |
| Common Name: | Orthodontic Software |
| Product Code: | PNN – Orthodontic Software |
| Device Classification: | Class II, 21 CFR 872.5470 |
| Predicate Devices: | 3Shape OrthoSystem (K171634) |
## Device Description:
The BRIUS Software Suite is an orthodontic appliance design and treatment simulation software. Until recently, this software was used only as a manufacturing software for nitinol wires, indirect bonding transfer media, and export of models. This software is now for use by professional technicians or physicians to design solutions for patients. Digital scans (3D) of a patient denture can be loaded into the software and a technician or physician can then create treatment plans for each individual patient and their needs. After approval by the patient's physician, the system can be used to fabricate patientspecific nitinol wires, indirect bonding transfer media, and export of models using standard stereolithographic (STL) and drawing exchange format (DXF) files for the design of custom shape-set nitinol wires. A 2D rendering of the nitinol wires needed for the patient can be created in the form of a DXF file and a 3D representation of the ideal tooth
{4}------------------------------------------------
alignment for the patient can be created in the form of an STL file. The 2D rendering can then be used to laser cut a nitinol sheet to form a 2D version of the patient-specific nitinol wire shape. Finally, a high-temperature mold can be fabricated from the STL file to shape set the 2D cut nitinol wire for a specific fit for patients.
## Indications for Use:
BRIUS Software Suite is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, treatment simulation and virtual appliance design options (Patient-Specific Nitinol Wires, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the BRIUS Software Suite requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
## Comparison to Predicate Devices:
BRIUS Software Suite is functionally equivalent to the following predicate device: 3Shape OrthoSystem (K171634 cleared January 31st, 2017).
The following table demonstrates the functional specifications of BRIUS Software Suite are substantially equivalent to the predicate devices.
| Specificati<br>on | BRIUS Software Suite | 3Shape OrthoSystem | Comparison<br>Result |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Indication<br>for Use | BRIUS Software Suite is<br>intended for use as a medical<br>front-end device providing tools<br>for management of orthodontic<br>models, systematic inspection,<br>treatment simulation and virtual<br>appliance design options<br>(Patient-Specific Nitinol Wires,<br>Export of Models, Indirect<br>Bonding Transfer Media) based<br>on 3D models of the patient's<br>dentition before the start of an<br>orthodontic treatment. It can<br>also be applied during the<br>treatment to inspect and | The 3Shape Ortho System™<br>is intended for use as a<br>medical front-end device<br>providing tools for<br>management of orthodontic<br>models, systematic inspection,<br>detailed analysis, treatment<br>simulation and virtual<br>appliance design options<br>(Custom Metal Bands, Export<br>of Models, Indirect Bonding<br>Transfer Media) based on 3D<br>models of the patient's<br>dentition before the start of an<br>orthodontic treatment. It can | Similar |
| | analyze the progress of the<br>treatment. It can be used at the<br>end of the treatment to<br>evaluate if the outcome is<br>consistent with the<br>planned/desired treatment<br>objectives.<br>The use of BRIUS Software<br>Suite requires the user to have<br>the necessary training and<br>domain knowledge in the<br>practice of orthodontics, as well<br>as to have received a<br>dedicated training in the use of<br>the software. | also be applied during the<br>treatment to inspect and<br>analyze the progress of the<br>treatment. It can be used at<br>the end of the treatment to<br>evaluate if the outcome is<br>consistent with the<br>planned/desired treatment<br>objectives.<br>The use of the Ortho<br>SystemTM requires the user to<br>have the necessary training<br>and domain knowledge in the<br>practice of orthodontics, as<br>well as to have received a<br>dedicated training in the use of<br>the software. | |
| Technology<br>Features | Stand Alone Software<br>Module Imports Digital Patient<br>Scans Can be used to design<br>Dental Casts Useful for Diagnosis,<br>treatment planning, and<br>CAD design Virtual Planning of tooth<br>movement Supports STL Files | Stand Alone Software<br>Module Imports Digital Patient<br>Scans Can be used to design<br>Dental Casts Useful for Diagnosis,<br>treatment planning, and<br>CAD design Virtual Planning of tooth<br>movement Supports STL Files | Same |
| Minimum<br>Hardware/Soft<br>ware<br>Requirements | OS: Windows 10 64-bit RAM: 8 GB Monitor Resolution:1280 X 800 Video Card Memory: 1 GB Hard Drive Space: 10 GB CPU: Intel compatible 2.6<br>GHz/Dual or Quad core 2.6 GHz Mouse: Any Mouse with scrolling<br>wheel or button | OS: Windows 7, 8, 10 64-bit RAM: 8 GB Monitor Resolution:1280 X<br>800 Video Card Memory: 1 GB Hard Drive Space: 10 GB CPU: Intel Core i5 or equivalent Mouse: with wheel button | Similar; No effect to<br>the indication for use. |
| Login Method | Username and Password | Username and Password | Same |
## Table 1: Functional Specification Comparison
{5}------------------------------------------------
| | Table 2. Feature Comparison Table for BRIUS and 3Shape |
|--|--------------------------------------------------------|
|--|--------------------------------------------------------|
| Feature Comparison | BRIUS | 3Shape (K171634) |
|-----------------------------|------------------|------------------|
| Supported anatomic<br>areas | Maxilla/Mandible | Maxilla/Mandible |
{6}------------------------------------------------
| | Intended Use | |
|------------------------------------------------------------|----------------------------------------|---------|
| Managing Patient and<br>case base data | Yes | Yes |
| Collection of study<br>material | Yes | Yes |
| Alignment of study<br>material | Yes | Yes |
| Measuring study<br>material | Yes | Yes |
| Analyzing Study<br>Material | Yes | Yes |
| Treatment Simulation | Yes | Yes |
| Virtual Appliance<br>Design | Yes | Yes |
| Supported PC formats | Windows/Mac: Internet<br>Browser-based | Windows |
| Managing patient and case base data | | |
| Creating, editing,<br>deleting and copying<br>patient data | Yes | Yes |
| Creating, editing,<br>deleting and copying<br>case data | Yes | Yes |
| Collection of study material | | |
| Surface scan for<br>intraoral scanner | Yes | Yes |
| Surface scan from STL<br>file | Yes | Yes |
| CT image data<br>(DICOM) | No | Yes |
| 2D overlay (PNG, JPG,<br>BMP) | No | Yes |
| Alignment of study material | | |
| Aligning surface scan<br>and CT image | No | Yes |
| Aligning Cephalometric<br>Images | No | Yes |
| Alignment of 2D<br>overlays | No | Yes |
| Ability to check/adjust<br>DICOM visibility | No | Yes |
| DICOM scan<br>Segmentation | No | No |
| 2D Measurement tool<br>box | No | Yes |
| 3D Measurement tool<br>box | No | Yes |
| Analyzing study material | | |
| Arch shape | Yes | Yes |
| Wire length | No | Yes |
| Tooth width | No | Yes |
| Bolton | No | Yes |
| Space Analysis | No | Yes |
| Overjet/Overbite | Yes | Yes |
| Occlusion Map | Yes | Yes |
| Treatment Simulation | | |
| 2D | Yes | Yes |
| 3D | Yes | Yes |
| Virtual Appliance Design | | |
| Orthodontic Appliance<br>Search | No | Yes |
| Orthodontic Appliance<br>Virtual Preparation | Yes | Yes |
| Orthodontic Appliance<br>Design | Yes | Yes |
| Orthodontic appliance<br>Export | Yes | Yes |
{7}------------------------------------------------
## Comparison of Indications for Use to Predicate Devices:
Based on the above comparison, the indications for use of the BRIUS Software Suite is similar to that of the 3Shape OrthoSystem. The 3Shape system is different in that it produces custom metal bands while the BRIUS system produces patient-specific nitinol wires. Otherwise, the indications for use for both devices are similar. Therefore, the BRIUS Software Suite can be considered substantially equivalent to its predicate device.
## Comparison of Technological Features to Predicate Devices:
the above comparison, the design. construction, Based on and performance characteristics of the BRIUS Software Suite is similar to that of 3Shape OrthoSystem. The BRIUS system does not include some measurement and search tools which are included in the 3Shape system. These tools do not affect the ability of users to perform necessary functions within the software. Additionally, both systems use CAD/CAM technology to produce digital models patient-specific orthodontic devices. Therefore, the BRIUS Software Suite can be considered substantially equivalent to its predicate devices.
## Summary of Performance Data and Substantial Equivalence:
Utilizing FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2015), the BRIUS Software Suite underwent appropriate integration, verification, and validation testing.
## Conclusion:
Based on comparison of indications for use, technological features, performance testing, and software validation testing, the BRIUS Software Suite have been shown to be appropriate for its indications for use and is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.