← Product Code [PNN](/submissions/DE/subpart-f%E2%80%94therapeutic-devices/PNN) · K181112

# Orchestrate 3D (K181112)

_Orchestrate 3D · PNN · Nov 26, 2018 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/PNN/K181112

## Device Facts

- **Applicant:** Orchestrate 3D
- **Product Code:** [PNN](/submissions/DE/subpart-f%E2%80%94therapeutic-devices/PNN.md)
- **Decision Date:** Nov 26, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Software as a Medical Device

## Indications for Use

The Orchestrate 3D Orthodontic Software is indicated for use as a front-end software tool for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options, including dental casts, which may be used for sequential aligner trays or retainers. These applications are based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of the Orchestrate 3D requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

## Device Story

Orchestrate 3D is orthodontic software for dental professionals. It imports 3D digital scans of patient dentition to facilitate diagnosis, treatment planning, and virtual appliance design. The software enables systematic inspection, detailed analysis, and treatment simulation. Users create treatment plans and design dental casts, which are exported as STL files for 3D printing to manufacture sequential aligner trays or retainers. Used in clinical settings by orthodontists or trained dental staff to monitor treatment progress and evaluate final outcomes against planned objectives. The software provides tools for managing patient case data, aligning study materials, and performing virtual tooth movement. It serves as a front-end tool to assist clinicians in designing orthodontic solutions, ultimately benefiting patients by facilitating customized treatment plans and appliance fabrication.

## Clinical Evidence

Bench testing only. The device underwent integration, verification, and software validation testing in accordance with FDA guidance for software contained in medical devices. No clinical data was required or provided.

## Technological Characteristics

Standalone software module; Windows 10 64-bit environment. Supports STL file imports for 3D dental models. Features include virtual tooth movement, treatment simulation, and CAD design for dental casts. Connectivity via standard PC hardware (Intel 2.6 GHz, 8GB RAM).

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- 3Shape Ortho System ([K171634](/device/K171634.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

Orchestrate 3D % Bill Jacqmein Consultant JCO Consulting 11218 Zest Court NE Blaine, Minnesota 55449

Re: K181112

Trade/Device Name: Orchestrate 3D Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN Second Product Code: LLZ Dated: October 17, 2018 Received: October 23, 2018

Dear Bill Jacqmein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

11.26.18

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Mary S. Mary S. Runner -S3 Runner -S3 07:57:41 -05'00'

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K181112

Device Name Orchestrate 3D

#### Indications for Use (Describe)

The Orchestrate 3D Orthodontic Software is indicated for use as a front-end software tool for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options, including dental casts, which may be used for sequential aligner trays or retainers. These applications are based on 3D models of the patient's dentition before the start of an orthodontic treatment.

It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.

The use of the Orchestrate 3D requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

Type of Use (Select one or both, as applicable)

| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)       </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)       </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5. 510(k) Summary

# K181112

## 510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

| Submitter:                    |                              | Orchestrate 3D<br>1747 North Riverside Ave<br>Rialto, CA 92376                      |
|-------------------------------|------------------------------|-------------------------------------------------------------------------------------|
| Company Contact<br>Person:    |                              | Todd Ehrler, DDS, MS                                                                |
|                               | Phone:<br>Email:             | (888) 755.8320<br>Todd.Erhler@orchestrate3d.com                                     |
| Submission<br>Correspondent:  |                              | Bill Jacqmein, Regulatory Affairs Consultant                                        |
|                               | Address:<br>Phone:<br>Email: | 11218 Zest Ct. NE, Blaine, MN 55449<br>(404) 216-6190<br>bjacqmein@rcqsolutions.com |
| Date Prepared:                |                              | October 17, 2018                                                                    |
| Proprietary Name:             |                              | Orchestrate 3D Orthodontic Software                                                 |
| Common Name:                  |                              | Orthodontic Plastic Brackets (Software)                                             |
| Product Code:                 |                              | PNN – Orthodontic Plastic Brackets, LLZ – System, Image<br>Processing, Radiological |
| Device Classification:        |                              | Class II, 21 CFR 872.5470                                                           |
| Primary Predicate<br>Devices: |                              | 3Shape Ortho System (K171634)                                                       |

### Device Description:

The Orchestrate 3D Orthodontic Software is an orthodontic appliance design and treatment simulation software. This software is for use by Dental professionals to diagnose and design solutions for patients. Digital scans (3D) of a patient denture can be loaded into the software and the dental professional can then create treatment plans for each individual patient and their needs. The system can be used to fabricate dental casts using standard stereolithographic (STL) files for use in 3D printers. Dental casts printed can then be used to manufacture sequential aligner trays and retainers.

{4}------------------------------------------------

### Indications for Use:

The Orchestrate 3D Orthodontic Software is indicated for use as a front-end software tool for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options, including dental casts, which may be used for sequential aligner trays or retainers. These applications are based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.

The use of the Orchestrate 3D requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.

### Comparison to Predicate Devices:

Orchestrate 3D Orthodontic Software is functionally equivalent to the following predicate device: 3Shape Ortho System (K171634) cleared January 17th, 2018).

The following table demonstrates the functional specifications of Orchestrate 3D Orthodontic Software are substantially equivalent to the predicate devices.

| Specification         | Orchestrate 3D Orthodontic Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 3Shape Ortho System (K171634)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comparison<br>Result |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Indication for<br>Use | The Orchestrate 3D Orthodontic Software is indicated for use as a front-end software tool for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options, including dental casts, which may be used for sequential aligner trays or retainers. These applications are based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of the Orchestrate 3D requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have | The 3Shape Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Custom Metal Bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of the Ortho System™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have | Similar              |

### Table 1: Functional Specification Comparison

{5}------------------------------------------------

|                                              | received a dedicated training in the<br>use of the software.                                                                                                                                                           | received a dedicated training in the<br>use of the software.                                                                                                                                                |                                                 |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Technology<br>Features                       | Stand Alone Software Module Imports Digital Patient Scans Can be used to design Dental Casts Useful for Diagnosis, treatment planning, and CAD design Virtual Planning of tooth movement Supports STL Files            | Stand Alone Software Module Imports Digital Patient Scans Can be used to design Dental Casts Useful for Diagnosis, treatment planning, and CAD design Virtual Planning of tooth movement Supports STL Files | Same                                            |
| Minimum<br>Hardware/Software<br>Requirements | OS: Windows 10 64-bit RAM: 8 GB Monitor Resolution: 1280 X 800 Video Card Memory: 1 GB Hard Drive Space: 10 GB CPU: Intel compatible 2.6 GHz/Dual or Quad core 2.6 GHz Mouse: Any Mouse with scrolling wheel or button | OS: Windows 7, 8, 10 64-bit RAM: 8 GB Monitor Resolution: 1280 X 800 Video Card Memory: 1 GB Hard Drive Space: 10 GB CPU: Intel Core i5 or equivalent Mouse: with wheel button                              | Same                                            |
| Login Method                                 | None Required; Uses Windows 10<br>Log in Security for User ID                                                                                                                                                          | Username and Password                                                                                                                                                                                       | Similar; No effect to the<br>indication for use |

| Feature Comparison                                         | Orchestrate 3D    | 3Shape (K171634) |     |
|------------------------------------------------------------|-------------------|------------------|-----|
| Supported anatomic<br>areas                                | Maxilla/Mandible  | Maxilla/Mandible |     |
| Intended Use                                               |                   |                  |     |
| Managing Patient and<br>case base data                     | Yes               | Yes              |     |
| Collection of study<br>material                            | Yes               | Yes              |     |
| Alignment of study<br>material                             | Yes               | Yes              |     |
| Measuring study<br>material                                | Yes               | Yes              |     |
| Analyzing Study<br>Material                                | Yes               | Yes              |     |
| Treatment Simulation                                       | Yes               | Yes              |     |
| Virtual Appliance<br>Design                                | Yes               | Yes              |     |
| Supported PC formats                                       | Windows           | Windows          |     |
| Managing patient and case base data                        |                   |                  |     |
| Creating, editing,<br>deleting and copying<br>patient data | Yes               | Yes              |     |
| Creating, editing,<br>deleting and copying<br>case data    | Yes               | Yes              |     |
| Collection of study material                               |                   |                  |     |
| Surface scan for                                           | intraoral scanner | Yes              | Yes |
| Surface scan from STL<br>file                              |                   | Yes              | Yes |
| CT image data<br>(DICOM)                                   |                   | No               | Yes |
| 2D overlay (PNG, JPG,<br>BMP)                              |                   | No               | Yes |
| Alignment of study material                                |                   |                  |     |
| Aligning surface scan<br>and CT image                      |                   | No               | Yes |
| Aligning Cephalometric<br>Images                           |                   | No               | Yes |
| Alignment of 2D<br>overlays                                |                   | No               | Yes |
| Ability to check/adjust<br>DICOM visibility                |                   | No               | Yes |
| DICOM scan<br>Segmentation                                 |                   | No               | No  |
| 2D Measurement tool<br>box                                 |                   | No               | Yes |
| 3D Measurement tool<br>box                                 |                   | No               | Yes |
| Analyzing study material                                   |                   |                  |     |
| Arch shape                                                 |                   | Yes              | Yes |
| Wire length                                                |                   | No               | Yes |
| Tooth width                                                |                   | No               | Yes |
| Bolton                                                     |                   | No               | Yes |
| Space Analysis                                             |                   | No               | Yes |
| Overjet/Overbite                                           |                   | Yes              | Yes |
| Occlusion Map                                              |                   | Yes              | Yes |
| Treatment Simulation                                       |                   |                  |     |
| 2D                                                         |                   | No               | Yes |
| 3D                                                         |                   | Yes              | Yes |
| Virtual Appliance Design                                   |                   |                  |     |
| Orthodontic Appliance<br>Search                            |                   | No               | Yes |
| Orthodontic Appliance<br>Virtual Preparation               |                   | Yes              | Yes |
| Orthodontic Appliance<br>Design                            |                   | Yes              | Yes |
| Orthodontic appliance<br>Export                            |                   | Yes              | Yes |

{6}------------------------------------------------

### Comparison of Indications for Use to Predicate Devices:

Based on the above comparison, the indications for use of the Orchestrate 3D Orthodontic Software is similar to that of the 3Shape Software. The Orchestrate 3D Orthodontic Software can be considered substantially equivalent to its predicate device.

## Comparison of Technological Features to Predicate Devices:

Based on the above comparison, the design, construction, and performance characteristics of the

{7}------------------------------------------------

Orchestrate 3D Orthodontic Software is similar to that of 3Shape Software. Therefore, the Orchestrate 3D Orthodontic Software can be considered substantially equivalent to its predicate devices.

#### Summary of Performance Data and Substantial Equivalence:

Utilizing FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2015), the Orchestrate 3D Orthodontic Software underwent appropriate integration, verification testing. The software passed the testing and performed per its intended use.

#### Conclusion:

Based on comparison of indications for use, technological features, performance testing, and software validation testing, the Orchestrate 3D Orthodontic Software have been shown to be appropriate for its indications for use and is substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/PNN/K181112](https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/PNN/K181112)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
