K213159 · Siesta Medical, Inc. · ORY · Dec 22, 2021 · Dental
Device Facts
Record ID
K213159
Device Name
Encore System
Applicant
Siesta Medical, Inc.
Product Code
ORY · Dental
Decision Date
Dec 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Encore System is intended for anterior advancement of the tongue base and hyoid suspension. It is indicated for the treatment of obstructive sleep apnea (OSA) and /or snoring.
Device Story
Encore System is a surgical implant kit for anterior tongue base and hyoid suspension. Components include integrated suture passer with #2-0 braided polyester suture, bone screws, bone screw inserters, suspension line lock tool, threading tool, drill bit, and suspension lines (#2 braided polyester or #2 braided polyethylene). Device is used by surgeons to fix soft tissue of tongue base and hyoid bone to the mandible. Procedure aims to reduce airway obstruction in OSA patients. System is non-electronic; no software or algorithms involved. MRI safety labeling updated to allow scanning under specific 1.5T or 3T conditions; drill bit length increased from 5mm to 10mm.
Clinical Evidence
Bench testing only. Performance testing was conducted to verify MRI safety and compatibility of the implantable metallic components under specified magnetic field conditions.
Technological Characteristics
Surgical kit containing bone screws, suspension lines (#2 braided polyester or polyethylene), suture passer, and associated manual surgical tools. Drill bit length is 10 mm. Device is non-electronic, mechanical, and implantable. MRI safety parameters defined for 1.5T and 3T environments.
Indications for Use
Indicated for patients requiring anterior advancement of the tongue base and hyoid suspension for the treatment of obstructive sleep apnea (OSA) and/or snoring.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 22, 2021
Siesta Medical, Inc. Michael Kolber Vice President 101 Church Street Los Gatos, California 95030
Re: K213159
Trade/Device Name: Encore System Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: ORY Dated: September 25, 2021 Received: September 28, 2021
Dear Michael Kolber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213159
Device Name Encore System
Indications for Use (Describe)
The Encore System is intended for anterior advancement of the tongue base and hyoid suspension. It is indicated for the treatment of obstructive sleep apnea (OSA) and/or snoring.
**Type of Use (Select one or both, as applicable)**
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Siesta Medical. The word "Siesta" is written in a bold, sans-serif font, with a curved line above it. Below "Siesta" is the word "Medical" in a smaller, lighter font. The logo is simple and modern.
#### 510(k) Summary ection 5
| 510(k) Number | K213159 |
|----------------------------|-----------------------------------------------------------------|
| Submitter Name and Address | |
| Name | Siesta Medical, Inc. |
| Contact | Michael Kolber |
| | Vice President, Regulatory Affairs |
| Address | 101 Church Street, Suite 3 |
| | Los Gatos, CA 95030 |
| Telephone | 408-505-6626 |
| Fax | 408-399-7600 |
| Date Prepared | September 25, 2021 |
| General Device Information | |
| Product Name | Encore™ System |
| Common Name | Intraoral device for snoring and obstructive sleep apnea |
| Classification | 21CFR872.5570 |
| | Intraoral devices for snoring and intraoral devices for snoring |
| | and obstructive sleep apnea. |
| Device Class | Class II |
| Product Code | ORY |
| Predicate Device Manufacturer | Device Name | 510(k) Number |
|-------------------------------|---------------|---------------|
| Siesta Medical, Inc. | Encore System | K201238 |
Device Description
The Encore System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone and hyoid bone suspension to the mandible bone using a bone screw and suspension lines. The Encore System consists of 1) an integrated suture passer pre-loaded with size #2-0 braided polyester suture, 2) three (3) bone screws and two (2) bone screw inserters, 3) a suspension line lock tool, 4) a threading tool, and 5) a drill bit. In addition, the following suspension lines are provided depending on the model number: 1) #2 braided polyester suspension line and 2) #2 braided polyethylene suspension line.
## Intended Use (Indications)
The Encore System is intended for anterior advancement of the tongue base and hyoid suspension. It is indicated for the treatment of obstructive sleep apnea (OSA) and /or snoring.
## Comparison to the Predicate Device
Labeling for the Encore System has been modified with respect to the MRI statement for the Encore Systems packaged with the #2 braided polyester suspension line and the #2 braided polyethylene suspension line. The labeling has changed as follows:
## From:
MRI Safety Information: The ENCORE System has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration or image artifact in the MR environment. The safety of the ENCORE System in the MR environment is unknown. Scanning a patient who has this device may result in patient injury.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Siesta Medical. The word "Siesta" is in bold, black font, with a gray wave-like line above it. Below "Siesta" is the word "Medical" in a smaller, gray font. The logo is simple and modern.
| To: | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| MRI Safety Information:<br>A patient implanted with the Encore Bone Screw may be safely scanned under the following<br>conditions. Failure to follow these conditions may result in injury to the patient. | |
| Nominal value(s) of Static<br>Magnetic Field [T] | 1.5 T ro 3 T |
| Maximum Spatial Field<br>Gradient<br>[T/m and gauss/cm] | 30 T/m (3000 gauss/cm) |
| RF Excitation | Circularly Polarized (CP) |
| RF Transmit Coil Type | Whole body transmit coil, head RF transmit-receive coil |
| Maximum Whole Body SAR<br>[W/kg] | 4.0 W/kg (First Level Operating Mode) |
| Limits on Scan Duration | 4.0 W/kg whole body average SAR for 15 minutes of continuous<br>RF (a sequence or back to back series/scan without breaks) |
| MR Image Artifact | The presence of this implant may produce an image artifact of<br>12mm |
| If information about a specific parameter is not included, there are no conditions associated with | |
that parameter.
In addition, the drill bit has been lengthened from 5 mm to 10 mm.
Summary of Non-Clinical Testing
Performance testing was conducted to confirm that the implantable metallic components of the Encore System do not pose a significant risk to a patient in a clinical MRI environment under the conditions presented in the labeling.
Statement of Equivalence
The Encore System has the same design, indications for use and the same technological characteristics as the predicate device. Based on this and the data provided in this pre-market notification, the subject device and the predicate device have been shown to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.