The MESHEET is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Device Story
MESHEET is a series of patient-specific, transparent, thermoplastic aligners used for orthodontic tooth alignment. Input data consists of digital dental scans and a clinician-developed treatment plan. The device is manufactured via thermoforming, utilizing a dual-layer construction: a polyethylene main layer and a thermoplastic polyurethane mesh layer. The aligners are worn by the patient and removed for eating and cleaning; they apply gentle, continuous force to move teeth incrementally toward the planned final position. The system is prescribed by a dental professional who uses commercial scanning, design software, and printing to create the treatment sequence. The device benefits patients by providing a removable, aesthetic alternative to traditional braces for correcting malocclusion.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, irritation, acute systemic toxicity, bacterial reverse mutation, in vitro chromosome aberration, intramuscular tests). Manufacturing process validation evaluated fit, form, and function against software treatment design, meeting all acceptance criteria.
Technological Characteristics
Thermoplastic aligner; dual-layer construction (polyethylene main layer, thermoplastic polyurethane mesh layer). Thermoformed. Patient-specific dimensions. Biocompatible materials per ISO 10993-1. Utilizes digital dental design software for treatment planning and model fabrication.
Indications for Use
Indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients requiring tooth movement.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 22, 2024
Dentis Co., Ltd Gyu Ri Kim Researcher 99, Seongseoseo-ro, Dalseo-gu Daegu. 42718 SOUTH KOREA
Re: K233544
Trade/Device Name: Mesheet Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: November 3, 2023 Received: November 3, 2023
Dear Gyu Ri Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K233544
Device Name
MESHEET
Indications for Use (Describe)
The MESHEET is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "DENTIS" in bold, black font. To the left of the word is an orange graphic that resembles a stylized letter "D" or a curved shape with a circle inside. The graphic is gradient, with the color transitioning from a lighter orange at the top to a darker orange at the bottom.
- Plant 1: 99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea Plant 2: 6, Yulam-ro, Dong-gu, Daegu, Korea Tel : 82.53.582.2804 / Fax : 82.53.583.2806
#### 510(k) Summary
.
#### Submitter
Dentis Co., Ltd. Gyu Ri Kim 99. Seongseoseo-ro. Dalseo-gu Daegu, 42718 Korea Email: kgr1026@dentis.co.kr Tel. +82-53-589-3541 Fax. +82-53-289-7922
#### Device Information
- Trade Name: MESHEET ●
- Common Name: Dental Plastic Aligners ●
- Classification Name: Orthodontic Plastic Brackets - Sequential Aligners
- Product Code: NXC
- Panel: Dental
- Regulation Number: 21 CFR 872.5470
- Device Class: Class II
- . Prepared Date: 02/22/2024
#### Predicate Devices:
K222619, Ensmile by Ensmile Pvt Ltd.
Reference Device: K192847, AUTOLIGN by Diorco Co., Ltd.
#### Indications for Use:
The MESHEET is indicated for the alignment of teeth during orthodontic treatment of malocclusion.
#### Device Description:
MESEEHT is a thermoplastic aligner designed according to the patient to help maintain teeth to the final desired position. This MESHEET is to be removed for eating and for cleaning.
MESHEET is sequence of transparent aligner manufacture from digital scan/plan.
Each aligner delivers a unique combination of minor forces to create the planned tooth movement. MEHSEET has a main layer that material is poly ethylene and mesh layer that material is thermoplastic polyurethane. Among the main layer, mesh layer is added.
Official Correspondent Withus Group Inc. April Lee 106 Superior. Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for DENTIS. The logo consists of an orange swirl shape on the left, followed by the word "DENTIS" in a bold, sans-serif font. The swirl shape is a gradient of orange, with the darker shade at the bottom and the lighter shade at the top. The word "DENTIS" is in black.
Plant 1: 99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea Plant 2: 6, Yulam-ro, Dong-gu, Daegu, Korea Tel : 82.53.582.2804 / Fax : 82.53.583.2806
#### Technical Characteristics
MESEEHT has same technical Characteristics with predicate device(K222619) as manufacture by thermoplastic polymer to create a customized aligner. Subject device and predicate device are thermoformed by and outside laboratory.
A professional prescribes the MESHEET based on and assessment of the patient's teeth, determines a course of treatment with the system. To obtain the dimensions and details of a patient's dentition, a dentist use a commercial scanner, software, printer and model material. The model is fabricated and scanned. Scanned data are directly imported into dental design software for planning. Professional designs the process of treatment by creating a series of the sequential model intended to gradually move the patient's teeth to the desired position in accordance with the dentist's prescription utilizing the dental design software. The doctor reviews and approves the treatment scheme. Professional fabricates a series of models for thermoforming.
When a model is produced, MESHEET produces the orthodontic appliances formed of clear, thin, thermoformed plastic. The aligners provide gentle continuous force to move the patient's teeth in small increments from their original positions to planned positions. MESHEET aligner is held in place by pressure and can be removed by the patient at any time.
#### Non-Clinical Testing
MESHEET passed all the testing in accordance with standards shown below to support substantial equivalence of the subject device:
- Biocompatibility tests such as cytotoxicity, irritation, Acute systemic toxicity, Bacterial reverse mutation, In vitro chromosome aberration, Intramuscular tests were performed in accordance with ISO 10993-1.
- OTS software validation test used to fabricate this subject device was performed according to Guidance for Industry and FDA Staff, Off-The Shelf Software Use in Medical Device. A performance qualification of the available software utilized in the processing steps of the subject device has been included to support substantial equivalence.
- . Manufacturing Process Validation Testing This test was conducted to validate the manufacturing process of the final product. It evaluated the fit, form, and function of the final product by a trained physician as compared to software treatment design. This test has met the pre-established acceptance criteria. The test results showed that the manufacturing process of the subject aligner achieves its intended use, and it is substantially equivalent to the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "DENTIS" in black, bold letters. To the left of the word is an orange graphic that resembles a flame or a stylized letter "D". The logo appears to be for a company named Dentis.
- Plant 1: 99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea . Plant 2: 6, Yulam-ro, Dong-gu, Daegu, Korea . Tel : 82.53.582.2804 / Fax : 82.53.583.2806
#### Substantial Equivalence Comparison
The subject device MESHEET is substantially equivalent to predicate device in intended use, material and technical characteristics. Comparison demonstrating Substantial Equivalence follows:
| | Subject Device | Primary Predicate |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MESHEET | Ensmile |
| Manufacturer | Dentis Co., Ltd | Ensmile Pvt Ltd |
| Regulation | 21 CFR 872.5470 | 21 CFR 872.5470 |
| Classification | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket |
| Product Code | NXC | NXC |
| Device Class | Class II | Class II |
| K number | N/A | K222619 |
| Design | Image: MESHEET aligner | Image: Ensmile aligner |
| Indication for<br>Use | The MESHEET is indicated for the alignment of<br>teeth during orthodontic treatment of malocclusion. | Ensmile is indicated for the alignment of teeth during<br>orthodontic treatment of malocclusion. |
| Device<br>Description | MESEEHT is sequence of transparent aligners<br>created from a digital orthodontic treatment plan<br>and thermoplastic aligners designed according to<br>the patient to help maintain teeth to the final<br>desired position. This MESHEET is to be removed<br>for eating and for cleaning.<br>MESHEET is sequence of transparent aligner<br>manufacture from digital scan/plan. The digital<br>orthodontic treatment plan is created by a dental<br>health professional. Each aligner delivers a unique<br>combination of minor forces to create the planned<br>tooth movement. MESHEET has cross shape holes<br>MEHSEET has main layer that material is poly<br>ethylene and mesh layer that material is<br>thermoplastic polyurethane. Among the main layer.<br>mesh layer is added. | The Ensmile is a sequence of transparent aligners<br>created from a digital orthodontic treatment plan. The<br>Ensmile aligner is fabricated of a transparent<br>thermoformed polyurethane plastic. Each aligner<br>delivers a unique combination of minor forces to<br>create the planned tooth movement. The digital<br>orthodontic treatment plan is created by a dental<br>health professional. The Ensmile Aligner<br>incorporates the use of attachments to create spaces<br>and force points in order to cause minor tooth<br>movement. These force points are located in specific<br>areas and positioned in such a way that they provide<br>a continuous force which slowly dissipates over time<br>on the tooth to be moved for as long as the aligner is<br>worn. |
| Mode of Action | Orthodontic tooth movement occurs through forces<br>applied by the appliance to the dentition as each<br>tooth follows the programmed displacement based<br>on a doctor's prescription. | Orthodontic tooth movement occurs through forces<br>applied by the appliance to the dentition as each tooth<br>follows the programmed displacement based on a<br>doctor's prescription. |
| Anatomy<br>Location | Mouth; mucosal | Mouth; mucosal |
| Size | Patient Specific | Patient Specific |
| Manufacturing<br>Method | Thermoforming | Thermoforming |
| Material | Thermoplastic Polymer | Thermoplastic Polymer |
| Software Used<br>for Ordering | Yes | Yes |
| Substantial<br>Equivalence<br>Comparison | MESHEET has the same indication for use, manufacturing method, fundamental scientific technologies,<br>principle of operation, material and similar design with the identified primary predicate device. There are<br>no difference between our subject device and primary predicate. Therefore, MESHET is substantially<br>equivalent with primary predicate. | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "DENTIS" in bold, black letters. To the left of the word is an orange graphic that looks like a stylized letter "D" or a flame. The graphic is a gradient, with the color fading from dark orange to light orange.
Plant 1: 99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea Plant 2: 6, Yulam-ro, Dong-gu, Daegu, Korea . Tel : 82.53.582.2804 / Fax : 82.53.583.2806
#### Conclusion
The MESHEET moves teeth by way of continuous gentle force through a sequence of clear aligners that follow the treatment plan developed by the clinician as does the primary predicate device, Ensmile Aligner.
●
The conclusions drawn from the data included in this submission, demonstrates that the subject device, MESHEET is substantially equivalent to the primary predicate device in indications for use, design, technological characteristics, mechanism of action, performance, materials and biocompatibility.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.