The patient specific set of Clear Dental Aligners are indicated for the treatment of tooth malocclusion in patients having permanent dentition. These Aligners position teeth by way of continuous gentle force. These are for self-use by patient with monitoring of improvement by orthodontists.
Device Story
Toothsi & SmileAligners are thermoformed, removable, patient-specific dental appliances; used for orthodontic treatment of malocclusion. Aligners are worn by patients in a prescribed sequence to exert continuous gentle force on teeth, facilitating progressive repositioning. Device is used in a home setting by the patient; orthodontists monitor progress. Aligners are manufactured from BPA-free thermoplastic sheets (0.5 mm to 1.0 mm thickness). Treatment involves wearing aligners for 20-22 hours daily, with changes every 7-14 days per prescription. Output is the physical movement of teeth toward the desired alignment. Benefits include correction of malocclusion through a wearable, removable appliance.
Clinical Evidence
Bench testing only. Testing included surface finishing, thermoformed thickness verification, shape accuracy after 15 days of simulated use, and 3D model shape accuracy for punching effectiveness. All tests were accepted.
Technological Characteristics
Thermoformed, BPA-free thermoplastic dental aligners. Thickness ranges from ~0.5 mm to ~1.0 mm. Non-sterile, single-user device. Biocompatibility compliant with ISO 10993. Mechanical actuation via continuous gentle force.
Indications for Use
Indicated for treatment of tooth malocclusion in patients with permanent dentition.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 9, 2023
AMPA Orthodontics Private Limited % Sanjeev Gupta Managing Consultant Intrust A301/A Green Heritage, Plot 79-80, Sector 20 Kharghar, Maharashtra 410210 India
Re: K223338
Trade/Device Name: Toothsi&SmileAligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC. Dated: October 29, 2022 Received: November 1, 2022
Dear Sanjeev Gupta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael Adjodha, M.ChE.,CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223338
Device Name Toothsi&SmileAligners
Indications for Use (Describe)
The patient specific set of Clear Dental Aligners are indicated for the treatment of tooth malocclusion in patients having permanent dentition. These Aligners position teeth by way of continuous gentle force. These by patient with monitoring of improvement by orthodontists.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "toothsi" in a red, sans-serif font. A curved line is below the "t" in "toothsi", resembling a smile. A four-pointed star is above the "i" in "toothsi".
## 510 k Summary - Substantial Equivalence Discussion K223338
| A. Name and Address of:<br>Manufacturer | AMPA Orthodontics Private Limited<br>1st Floor, Empire Plaza B Wing<br>Lal Bahadur Shastri Marg<br>Vikhroli West<br>Mumbai 400083<br>Maharashtra, India<br>Phone/Fax: +91 9820340245<br>Contact Person: Dr. Manjul Jain (drmanjul@toothsi.in) |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### B. Device Details:
- a) Trade Name: Toothsi & SmileAligners
- b) Common Name: Clear Dental Aligners
- c) Classification Name: Aligner, Sequential
- d) Product Code: NXC
- C. Predicate Device: ClearCorrect System (K113618)
- D. Device Description: The Clear Dental Aligners are thermoformed, orthodontic wearable & removable dental appliances that, when worn in the prescribed sequence and duration, progressively reposition the permanent malocclusion teeth. These aligners could be soft, medium, or hard as per prescription set by an orthodontist, and to exert the desired & continuous gentle force to progressively position the teeth. Each aligner is made of thermoforming BPA free material sheets.
- E. Intended Use The patient specific set of Clear Dental Aligners are indicated for the treatment of tooth malocclusion in patients having permanent dentition. These Aligners position teeth by way of continuous gentle force. These are for self-use by patient with monitoring of improvement by orthodontists.
Ver 01 Date 2022-12-20
013/1
+91 98980000000
1st floor, Empire Plaza B Wing, Lal Bahadur Shashtri Marg, Vikhroli (West), Mumbai - 400083
Info@toothsi.in X 0 www.toothsi.in
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "toothsi" in red font. The "t" in toothsi is lowercase, and the two "o"s are connected. There is a curved line under the first "t" that resembles a smile. There is a four-pointed star to the right of the "i".
## F: Comparison of proposed device with predicate device for
| Equivalence<br>Points | Clear Dental Aligners | ClearCorrect System<br>(K113618) | Comparison |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | AMPA Orthodontics Pvt. Ltd. | CLEARCORRECT LLC | N/A |
| Product Code | NXC | NXC | Equivalent |
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | Equivalent |
| Intended<br>Use/Indications<br>for Use | The patient specific set of Clear<br>Dental Aligners are indicated for<br>the treatment of tooth<br>malocclusion in patients having<br>permanent dentition.<br>These<br>Aligners position teeth by way of<br>continuous gentle force. These are<br>for self-use by patient with<br>monitoring of improvement by<br>orthodontists. | The ClearCorrect System is<br>indicated for the treatment of<br>tooth malocclusion in patients<br>with permanent dentition (i.e.<br>all second molars).<br>The ClearCorrect System<br>positions teeth by way of<br>continuous gentle force | Equivalent |
| Mechanism of<br>Action | Continuous gentle force to<br>move the teeth | Continuous gentle force to<br>move the teeth | Equivalent |
| Material of<br>Construction | Thermoplastic | Thermoplastic | Equivalent |
| Aligner Material<br>Thickness<br>(For different<br>aligner set) | ~0.5 mm to ~1.0 mm as per<br>treatment plan | Soft (0.5 mm) | Equivalent |
| | | Medium (0.62 mm) | |
| | | Hard (0.75 mm)<br>Ultra Hard (1 mm) | |
| Biocompatible | Complies with ISO 10993 (FR# 2-<br>248 & 2-296) | ISO 10993-1 | Equivalent |
| Prescription<br>Steps | Approximately 7-14 days/aligner<br>set as per prescription | Soft (1 Week) | Equivalent |
| | | Medium (1 Week) | |
| | | Hard (1 Week) | |
| Daily Duration of<br>wear | At least 20-22 hours | 17 hours a day | Equivalent |
| Sterile | Non-Sterile | Non-Sterile | Equivalent |
| Single-Use | Single User Use | Single User Use | Equivalent |
#### Ver 01 Date 2022-12-20
(O)
013/2
1* floor, Empire Plaza B Wing, Lal Bahadur Shashtri Marg, Vikhroli (West), Mumbai - 400083
- +91 98980000000
K
X
0
Info@toothsi.in
- www.toothsi.in
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "toothsi" in red font. The "t" in toothsi is stylized with a curved line underneath, resembling a smile. There is a four-pointed star to the right of the "i" in toothsi. The font is rounded and sans-serif.
## G. Performance Testing Bench (Non-clinical)
The following bench testing was performed as part of Design Verification with purpose define in column 2 of Table 2 below. The reports are presented in sequence along with data tables and figures as applicable in 018_Performance Testing Bench. No separate annexure has been created for this section.
| Table 2: Summary of bench testing performed on design samples of clear dental<br>aligners | | | Test Outcome |
|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------|--------------|
| Sr.<br>No | Purpose of Bench Testing | Report Number | |
| 1 | Surface finishing after polishing by two<br>methods | Report No. AO/DV/R1<br>dated 31-12-2021 | Accepted |
| 2 | Thickness of aligners after thermoforming<br>with reference to aligner sheets | Report No.: AO/DV/R2<br>Dated 31-12-2021 | Accepted |
| 3 | Aligner shape accuracy after mimicked use<br>for 15 days | Report No.: AO/DV/R7<br>Dated 10-01-2022 | Accepted |
| 4 | 3D Model shape accuracy for effectiveness<br>of aligner punching as per treatment plan | Report No.: AO/DV/R3<br>Dated 10-01-2022 | Accepted |
The above four tests relate to safety (Sr. No 1 & 4) and performance (Sr. No 2, 3 and 4) of the clear aligners. No comparative testing with predicate device was possible since clear dental aligners are patient specific devices made after treatment plan.
### H. Substantial Equivalence Discussion
Comparison of specifications with predicate device: The Section F-Table 1 above summarizes the comparison of subject device with predicate device.
- a) Similarities: As it emerges from the review of Section F-Table 1 the characteristics related to Intended Use, Mechanism of Action, Material Thickness, Material of Construction, Prescription Steps are equivalent between predicate and subject devices.
- b) Differences: The duration of wearing aligner is around 17 hours in predicate device and around 22 hours for subject device. This difference is subjective and does not alter safety and performance of the devices. Such difference majorly emerges from the user actions.
Ver 01 Date 2022-12-20
+91 98980000000
Info@toothsi.in
1st floor, Empire Plaza B Wing, Lal Bahadur Shashtri Marg, Vikhroli (West), Mumbai - 400083
0 www.toothsi.in
X
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "toothsi" in red font. The "tooth" part of the word is above a curved line, also in red, that resembles a smile. To the right of the "i" is a red star.
- c) Standards Testing: AMPA Orthodontics Private Limited has performed biocompatibility testing of the subject device by application of ISO 10993 standards which has also been followed by the predicate device manufacturer.
Conclusions: We, therefore, conclude that Clear Dental Aligner manufactured by AMPA Orthodontics Private Limited have substantial equivalence with ClearCorrect Aligners (K113618) Manufactured by ClearCorrect LLC.
Ver 01 Date 2022-12-20
(O)
013/4
1st floor, Empire Plaza B Wing, Lal Bahadur Shashtri Marg, Vikhroli (West), Mumbai - 400083
+91 98980000000 Info@toothsi.in
X
www.toothsi.in
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.