← Product Code [NXC](/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC) · K221097

# SmileSeries (K221097)

_Ordont Orthodontic Laboratories, Inc. · NXC · Jul 14, 2022 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC/K221097

## Device Facts

- **Applicant:** Ordont Orthodontic Laboratories, Inc.
- **Product Code:** [NXC](/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC.md)
- **Decision Date:** Jul 14, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The SmileSeries™ is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in adult and adolescent patients with permanent dentition (i.e. all second molars). The SmileSeries™ positions teeth by way of continuous gentle force.

## Device Story

SmileSeries™ consists of clear, thermoformed, removable plastic aligner trays; designed for orthodontic tooth realignment without wires or brackets. Clinician provides physical or scanned dental impressions; digital setup created for clinician approval. Upon approval, 3D-printed molds are produced; aligners thermoformed on molds and laser-marked. Provided to dental professional for patient fitting. Patient wears aligners ~22 hours/day, removing only for eating; each tray worn 2-3 weeks to exert continuous gentle force. System facilitates progressive tooth movement until final correction. Used in dental clinics; operated by dentists/orthodontists.

## Clinical Evidence

Bench testing only. Validated manufacturing process and accuracy of thermoformed aligners against initial digital scans. No clinical or animal performance testing required for this device category.

## Technological Characteristics

Thermoformed plastic aligner trays; non-sterile. 3D-printed mold fabrication. Software-driven treatment planning and design. No electronic components or energy sources in the final device.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- ClearPath Aligner ([K202857](/device/K202857.md))

## Reference Devices

- Ortho System™, 3Shape A/S ([K180941](/device/K180941.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 14, 2022

Ordont Orthodontic Laboratories, Inc. % Patsy Trisler Regulatory Consultant Trisler Consulting, dba 7949 Beaumont Green East Drive Indianapolis, Indiana 46250

Re: K221097

Trade/Device Name: SmileSeries Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: April 13, 2022 Received: April 14, 2022

Dear Patsy Trisler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221097

Device Name

SmileSeries™

Indications for Use (Describe)

The SmileSeries™ is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in adult and adolescent patients with permanent dentition (i.e. all second molars). The SmileSeries™ positions teeth by way of continuous gentle force.

| Type of Use (Select one or both, as applicable)                                                                   |  |
|-------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> |  |
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div>  |  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# SECTION 5

#### 510(k) SUMMARY

| SUBMITTER                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name:                                                   | Ordont Orthodontic Laboratories, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Submitter Address:                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Contact Person:<br>Email:<br>Telephone:                           | Paul Ruzicka, CDA<br>Paul@ordont.com<br>800-325-3698                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Date Prepared:                                                    | April 13, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| DEVICE                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Trade Name:                                                | SmileSeries™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name:                                                      | Aligner, Sequential (Clear Braces)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification Name<br>Number<br>Product Code<br>Regulatory Class | Orthodontic Plastic Bracket<br>21 CFR 872.5470<br>NXC<br>2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Review Panel                                                      | Dental                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| PREDICATE DEVICE                                                  | Primary Predicate:<br>K202857, ClearPath Aligner, ClearPath Orthodontics<br>Reference Device:<br>K180941 Ortho System™, 3Shape A/S                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| DEVICE<br>DESCRIPTION                                             | The SmileSeries™ is comprised of a series of clear, thin,<br>thermoformed removable aligner trays that are designed to<br>correct tooth malocclusions without the use of conventional<br>wire and bracket orthodontic technology.  SmileSeries™<br>aligners are provided non-sterile and are customized for each<br>patient according to the dental clinician's prescription.<br>The dental health professional (dentist/orthodontist) takes<br>provides physical or scanned impressions of the patient's teeth<br>to SmileSeries™. A digital setup of either the scanned<br>impression or a scan of the physical impression is sent to the<br>clinician for approval.  Upon approval, molds are then created<br>with 3D-printing technology and the clear aligners are<br>thermoformed on the molds and laser marked.<br>The finished, customized aligners are provided to the dental<br>health care professional who provides them to the patient,<br>confirming fit and design. The aligner trays are held in place by<br>pressure and can be removed by the patients at any time. |
| MECHANISM OF<br>ACTION                                            | Each aligner in the set is used for the specified period of time,<br>usually 2-3 weeks, to exert gentle force to achieve progressive<br>realignment of the teeth until the final correction has been<br>achieved. The daily treatment time is usually 22 hours, or<br>except during eating, based on the clinician's prescribed<br>treatment plan.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| SUMMARY OF<br>TECHNOLOGICAL<br>CHARACTERISTICS                    | The thermoplastic material used for the manufacture of the<br>SmileSeries™ aligner is the same material used to make the<br>predicate ClearPath Aligner.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                   | The software system used is Ortho Analyzer, 2019 ver 1.8.1.0<br>by 3Shape A/S (Reference device - K180941). It also is the<br>same as used for the Predicate. It is used for management of<br>3D scanned orthodontic models, orthodontic diagnosis by<br>measuring, analyzing, inspecting and visualizing 3D scanned<br>orthodontic models, virtual planning of orthodontic treatments<br>by simulating tooth movements, and design of orthodontic<br>appliances based on 3D scanned orthodontic models.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| INDICATIONS FOR<br>USE STATEMENT                                  | The SmileSeries™ is a series of clear, lightweight, plastic<br>appliances indicated for the correction of dental malocclusion in<br>adult and adolescent patients with permanent dentition (i.e. all<br>second molars). The SmileSeries™ positions teeth by way of<br>continuous gentle force.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| SAFETY TESTING                                                    | Biocompatibility: Testing of the plastic thermoformable material<br>used to make the SmileSeries™ aligners has been provided in<br>previous 510(k) submissions to FDA.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| PERFORMANCE<br>TESTING                                            | Bench testing was performed to validate the manufacturing<br>process, to ensure the accuracy of the final thermoformed<br>aligner compared to the initial digital scan. A final report was<br>part of the 510(k) package.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                   | In vivo Animal and Human Clinical performance testing are not<br>required for this device category.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| COMPARISON TO<br>THE PREDICATE<br>DEVICE                          | The SmileSeries™ aligner has the same intended use as the<br>predicate device. The thermoplastic material is the same and<br>the design phase makes the use of the same software as the<br>predicate. The manufacturing fabrication of the clear aligner<br>makes use of similar, industry-standard processes with the<br>same or similar machines and materials. Any differences in the<br>specific company processes do not raise new questions of<br>safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| SUBSTANTIAL<br>EQUIVALENCE<br>CONCLUSION                          | The information and data provided in this 510(k) establish that<br>the SmileSeries™ is substantially equivalent to the predicate<br>ClearPath Aligner in the intended use, design, principle of<br>operation, technology, and thermoformable material used to<br>make the aligner. Comparison of all key parameters are<br>presented in the following SE Comparison table                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

{4}------------------------------------------------

{5}------------------------------------------------

| 510(k) Number                                                 | Proposed Device<br>K                                                                                                                                                                                                                                                                                             | Predicate Device<br>K202857                                                                                                                                                                                                                                                                        | Comparison |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Device Name                                                   | SmileSeries™                                                                                                                                                                                                                                                                                                     | ClearPath Aligner                                                                                                                                                                                                                                                                                  | N/A        |
| Manufacturer                                                  | Ordont Orthodontic<br>Laboratories, Inc.                                                                                                                                                                                                                                                                         | ClearPath Orthodontics, Ltd                                                                                                                                                                                                                                                                        | N/A        |
| Classification<br>Regulation<br>Name<br>Product Code<br>Class | 21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>NXC<br>2                                                                                                                                                                                                                                                       | 21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>NXC<br>2                                                                                                                                                                                                                                         | Same       |
| Indications for<br>Use                                        | The SmileSeries™ is a<br>series of clear, lightweight,<br>plastic appliances indicated<br>for the correction of dental<br>malocclusion in adult and<br>adolescent patients with<br>permanent dentition (i.e. all<br>second molars). The<br>SmileSeries™ positions teeth<br>by way of continuous gentle<br>force. | The ClearPath Aligner is a<br>series of clear, lightweight,<br>plastic appliances indicated<br>for the correction of dental<br>malocclusion in patients with<br>permanent dentition (i.e. all<br>second molars). The<br>ClearPath Aligner positions<br>teeth by way of continuous<br>gentle force. | Same       |
| Mode of Action                                                | The removable appliance<br>applies gentle forces on teeth<br>according to the plan<br>prescribed by the doctor.                                                                                                                                                                                                  | The removable appliance<br>applies gentle forces on teeth<br>according to the plan<br>prescribed by the doctor                                                                                                                                                                                     | Same       |
| Description of<br>Use                                         | Each removable preformed<br>plastic tray, prescribed by the<br>Dr, is worn by the patient<br>usually for a few weeks, prior<br>to using the next sequential<br>aligner tray.                                                                                                                                     | Each removable preformed<br>plastic tray, prescribed by the<br>Dr, is worn by the patient<br>usually for a few weeks, prior<br>to using the next sequential<br>aligner tray.                                                                                                                       | Same       |
| Material                                                      | Thermoformed plastic                                                                                                                                                                                                                                                                                             | Thermoformed plastic                                                                                                                                                                                                                                                                               | Same       |
| Manufacturing<br>Process                                      | Forming of plastic sheets on<br>unique dental models using<br>thermoforming machine                                                                                                                                                                                                                              | Forming of plastic sheets on<br>unique dental models using<br>thermoforming machine                                                                                                                                                                                                                | Same       |
| Software Used                                                 | Yes, for treatment planning<br>and 3D printing of models.                                                                                                                                                                                                                                                        | Yes, for treatment planning<br>and 3D printing of models.                                                                                                                                                                                                                                          | Same       |
| Prescription<br>Use                                           | Rx                                                                                                                                                                                                                                                                                                               | Rx                                                                                                                                                                                                                                                                                                 | Same       |
| Biocompatibility                                              | Yes, shown to meet<br>requirements                                                                                                                                                                                                                                                                               | Yes, shown to meet<br>requirements                                                                                                                                                                                                                                                                 | Same       |
| Validation<br>Testing                                         | Yes, performed                                                                                                                                                                                                                                                                                                   | Yes, performed.                                                                                                                                                                                                                                                                                    | Same       |

## Substantial Equivalence Comparison Table

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC/K221097](https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC/K221097)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
