6MS Invisible Aligner is indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients with permanent dentition (i.e. all second molars). The 6MS Invisible Aligner positions teeth by way of continuous gentle force.
Device Story
6MS Invisible Aligner consists of customized, clear, removable thermoplastic polyurethane-polyester composite orthodontic trays. Input: intraoral scans or physical impressions of patient teeth. Process: dental clinician uses commercial dental software to plan tooth movement; digital plan approved by clinician; SMS OPCO manufactures trays via thermoforming on molds. Output: series of sequential aligner trays. Usage: clinician-prescribed; patient wears trays incrementally to apply continuous gentle force to teeth, repositioning them from misalignment to aligned state. Benefit: gradual, incremental orthodontic correction of malocclusion.
Clinical Evidence
No clinical data. Bench testing only. Non-clinical testing included mechanical property verification (ASTM D638: tensile strength, elongation, tensile modulus) for 5-year shelf life, packaging moisture barrier validation, and process validation. Biocompatibility testing performed per ISO 10993-1, -5, and -10 (cytotoxicity, sensitization, irritation, intracutaneous reactivity).
Technological Characteristics
Material: thermoplastic polyurethane-polyester composite. Form factor: customized thermoformed clear aligner trays. Manufacturing: thermoforming on molds. Connectivity: uses commercial dental software (3Shape K180981, NemoCast K193003) for design. Biocompatibility: ISO 10993 compliant. Non-sterile packaging.
Indications for Use
Indicated for alignment of teeth during orthodontic treatment of malocclusion in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 2, 2021
Sms Opco, LLC % Patsy Trisler Regulatory Consultant Qserve Group US, Inc. 7949 Beaumont Green East Drive Indianapolis, Indiana 46250
## Re: K210652
Trade/Device Name: 6MS Invisible Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: April 19, 2021 Received: April 21, 2021
## Dear Patsy Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210652
Device Name
6MS Invisible Aligner
Indications for Use (Describe)
6MS Invisible Aligner is indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients with permanent dentition (i.e. all second molars). The 6MS Invisible Aligner positions teeth by way of continuous gentle force.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary – K210652
| Submitter Name: | SMS OPCO, LLC |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 672 Morning Star Drive, Suite 120<br>The Colony, Texas 75056 |
| Phone Number: | 866-957-7645 |
| Contact Person: | Mr. Bruce Page, QA/RA Manager |
| Date Prepared: | April 19, 2021 |
| Device Trade Name: | 6MS Invisible Aligner |
| Common Name | Aligner, Sequential |
| Classification Name<br>Number<br>Product Code<br>Regulatory Class | Orthodontic Plastic Bracket<br>21 CFR 872.5470<br>NXC<br>2 |
| Primary Predicate<br>Device: | K182826, Ormco™ Spark Aligner System, Sybron Dental Specialties |
| Reference Device: | None |
| Indications for Use | 6MS Invisible Aligner is indicated for the alignment of teeth during<br>orthodontic treatment of malocclusion in patients with permanent<br>dentition (i.e. all second molars). The 6MS Invisible Aligner<br>positions teeth by way of continuous gentle force. |
| Device Description,<br>Mechanism of Action<br>and Summary of<br>Technological<br>Characteristic | 6MS Invisible Aligner consists of a series of dental-clinician<br>prescribed customized clear plastic removable orthodontic aligner<br>trays that are made from a clear, thin thermoformed polyurethane-<br>polyester composite. The aligners are designed to gradually move<br>the patient's teeth incrementally, repositioning them from their<br>original misalignment to a more aligned state. This occurs through<br>forces applied by the appliance to the teeth over time until final<br>correction, according to the dental clinician's prescription has been<br>achieved.<br>A dental clinician prescribes the 6MS Invisible Aligner based on an<br>evaluation of the patient's teeth. Either intraoral scans or physical<br>impressions of the patient's teeth are taken, after which the clinician<br>determine course of treatment with the system. The clinician<br>completes a prescription form using standard dental software used<br>for tooth alignment. The series of plastic aligner trays are designed<br>according to commercially available dental software for planning the<br>tooth alignments.<br>After the digital plan is developed and approved by the prescribing<br>dental clinician, SMS OPCO produces the aligner trays. The<br>customized aligner trays are then provided to the dental clinician<br>who distributes them to the patient assuring fit and function. |
{4}------------------------------------------------
The clear, thin, thermoplastic polyurethane-polyester composite resin material used for fabrication of the 6MS Invisible Aligners is commonly used in many dental and orthodontic appliances, including the predicate clear aligners.
Technological The subject device is compared to the Predicate, K182826 in the comparison table on the final page of this Summary. It shows the Comparison: following:
- Intended use: Same
- Thermoplastic material composition: Same ●
- . Mechanism of Action: Same
- Biocompatible according to ISO 10993 requirements: Same ●
- Design of aligners: Similar ●
- Method used to manufacture the customized aligners by . thermoforming: Similar
- . Software use during design and manufacturing: Similar
- Process validation testing: Similar .
The differences in the specific manufacturing methods used, including the use of different design and manufacturing software products, and the variability in the manufacturing process validation testing, do not raise new questions of safety and effectiveness.
{5}------------------------------------------------
| | Subject Device:<br>6MS Invisible Aligner | Predicate Device:<br>Ormco™ Spark™ Aligner System | Comparison |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| 510(k) Number | K210652 | K182826 | N/A |
| Manufacturer | SMS OPCO, LLC | Sybron Dental Specialties | N/A |
| Classification #<br>and Name<br>Product Code<br>Class | 21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>NXC<br>2 | 21 CFR 872.5470<br>Orthodontic Plastic Bracket<br>NXC<br>2 | Same |
| Indications for<br>Use | 6MS Invisible Aligner is<br>indicated for the alignment of<br>teeth during orthodontic<br>treatment of malocclusion in<br>patients with permanent<br>dentition (i.e. all second<br>molars). The 6MS Invisible<br>Aligner positions teeth by<br>way of continuous gentle<br>force. | The Ormco Spark System is<br>indicated for the alignment of<br>teeth during orthodontic<br>treatment of malocclusion in<br>patients with permanent<br>dentition (i.e. all second<br>molars). The Ormco Spark<br>Aligner System positions<br>teeth by way of continuous<br>gentle force. | Same Intended<br>Use |
| Mode of Action | Alignment of teeth by<br>application of continuous<br>gentle force, by sequential<br>use of preformed plastic<br>trays. | Same, alignment of teeth by<br>application of continuous<br>gentle force, by sequential<br>use of preformed plastic<br>trays. | Same Mode of<br>action |
| Description of<br>Use | Each preformed plastic tray<br>is worn by the patient as<br>prescribed by the dental<br>practitioner, usually a few<br>weeks prior to using the next<br>sequential aligner tray. | Same, each preformed<br>plastic tray is worn by the<br>patient as prescribed by the<br>dental practitioner, usually a<br>few weeks prior to using the next<br>sequential aligner tray. | Same Use |
| Material | Thin thermoplastic sheet of<br>polyurethane polyester<br>composite resin | Thin thermoplastic sheet of<br>polyurethane polyester<br>composite resin | Same material |
| Biocompatible | Meets ISO 10993-1, -5, -10<br>requirements: Cytotoxicity,<br>Sensitization-Irritation,<br>Intracutaneous Reactivity | Meets ISO 10993-1, -5, -10<br>requirements: Cytotoxicity,<br>Sensitization-Irritation,<br>Intracutaneous Reactivity | Same |
| Manufacturing<br>Process | Thermoforming on molds | Thermoforming on molds | Similar method |
| Software Used<br>for Treatment<br>Planning and<br>Manufacturing | 3Shape K180981 and<br>NemoCast K193003 | Spark Aligner Software | Similar - Software<br>products are<br>different, but<br>designed to<br>perform similar<br>functions |
| Non-Sterile<br>Packaging | Yes | Yes | Similar - Exact<br>packaging of<br>Predicate is<br>unknown |
| OTC or Rx | Rx | Rx | Same |
# Substantial Equivalence Comparison Table
{6}------------------------------------------------
#### Device Testing Non-clinical Testing
Test data were submitted to:
- . assure the mechanical properties of the aligner material meet specifications for up to 5 years shelf life [according to ASTM D638 standard's testing for: tensile strength (PSI): elongation (%), tensile@ yield (PSI), elongation @ yield (PSI) and tensile modulus (PSI)];
- . assure the aligner material packaging retains the required moisture barrier properties;
- . validate the processes used for the design and manufacture of the customized aligners, to ensure consistency between the aligner's design and the manufactured aligners.
All testing met the pre-determined acceptance criteria.
### Biocompatibilitv
The thermoplastic polyurethane used for making the aligner series has been tested according to Good Laboratory Practices for its biocompatibility according to ISO 10993, as follows:
- Part 5 (Cytotoxicity Elution MEM),
- Part 10 (Intracutaneous/Intradermal Reactivity).
- Part 10 (Oral Mucosa Irritation),
- Part 10 (Maximization for Delayed-Type Hypersensitivity)
In addition, the finished customized aligner was tested according to ISO 10993, Part 5, Cytotoxicity. Results show it is noncytotoxic. All testing showed that the thermoplastic material and the manufactured finished aligner met the requirements of the test methods and are safe and biocompatible for the stated intended use.
Clinical Testing
Neither animal nor human testing are required for this product because it is composed of the same materials, is designed similarly, and is manufactured by a similar method as the predicate device.
- Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe and effective as the legally marketed device, K182826.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.