← Product Code [NXC](/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC) · K182826

# Ormco Spark Aligner System (K182826)

_Sybron Dental Specialties · NXC · Oct 11, 2018 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC/K182826

## Device Facts

- **Applicant:** Sybron Dental Specialties
- **Product Code:** [NXC](/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC.md)
- **Decision Date:** Oct 11, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Ormco™ Spark™ Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients with permanent dentition (i.e., all second molars). The Ormco™ Spark™ Aligner System positions teeth by way of continuous gentle force.

## Device Story

System consists of series of custom-manufactured, clear plastic removable aligners; designed to move teeth incrementally from original to treated state. Clinician uses web portal to submit patient profiles, digital scans, and prescriptions. Ormco technicians use proprietary 3D design software to create 3D models of teeth and generate transitional treatment stages. Clinician reviews and approves models via Approver software. Fabrication software generates resin molds for thermoforming aligners. Aligners worn in prescribed sequence; held by pressure; removable by patient. Optional attachments, buttons, hooks, bite ramps, and pontics available to facilitate movement or address gaps. Clinician monitors treatment progress. Benefits include orthodontic correction of malocclusion.

## Clinical Evidence

No clinical performance testing completed. Substantial equivalence supported by comprehensive scientific literature review and analysis of predicate device efficacy in treating various malocclusions. Bench testing included material properties (translucency, staining, shelf life, shipping) and feature-specific testing (attachment accuracy, cutout effects, force evaluation). Biocompatibility testing performed per ISO 10993-1.

## Technological Characteristics

Material: Thermoplastic polyurethane-polyester composite resin (TruGEN). Principle: Sequential thermoformed plastic aligners applying continuous force. Connectivity: Web-based portal for prescription/scan submission. Software: Proprietary 3D modeling and design modules. Sterilization: Non-sterile. Form factor: Custom-molded removable trays.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- Invisalign System ([K143630](/device/K143630.md))

## Reference Devices

- ClearCorrect ([K113618](/device/K113618.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 11, 2018

Image /page/0/Picture/1 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services, which features a stylized human figure. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Sybron Dental Specialties % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313

Re: K182826

Trade/Device Name: Ormco Spark Aligner System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: October 3, 2018 Received: October 5, 2018

Dear Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mary S. Runner -S3

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### SECTION 4. INDICATIONS FOR USE STATEMENT

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K 182826

Device Name

### Ormco™ Spark™ Aligner System

Indications for Use (Describe)

The Ormco™ Spark™ Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients with permanent dentition (i.e., all second molars). The Ormco™ Spark™ Aligner System positions teeth by way of continuous gentle force.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801

Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "

FORM FDA 3881 (7/17) Publishing Services (301) 443-6740

PSC

{3}------------------------------------------------

## SECTION 5. 510(k) SUMMARY FOR ORMCO™ SPARK™ ALIGNER SYSTEM

Image /page/3/Picture/1 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters 'sds' in a bold, sans-serif font, with the letters overlapping slightly. The letters are in a dark blue color. Below the letters, the words 'Sybron Dental Specialties' are written in a smaller, sans-serif font, also in dark blue. A curved line extends over the top of the 'sds' letters, adding a dynamic element to the logo.

K182826

## 510(k) SUMMARY for Ormco™ Spark™ Aligner System

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR \$807.92, the following summary of information is provided:

- 1. Submitter Information:
Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687

| Primary Contact:  | Mey Saied    |
|-------------------|--------------|
| Telephone Number: | 909-962-5675 |
| Fax Number:       | 909-962-5694 |

| Alternate Contact: | Valerie Cimmarusti |
|--------------------|--------------------|
| Telephone Number:  | 714-516-7476       |
| Fax Number:        | 909-962-5694       |

Date Prepared: September 12, 2018

- 2. Device Name:

| Proprietary Name:       | Ormco™ Spark™ Aligner System |
|-------------------------|------------------------------|
| Classification Name:    | Aligner, Sequential          |
| Regulation Description: | Aligner, Sequential          |
| CFR Number:             | 872.5470                     |
| Device Class:           | II                           |
| Product Code:           | NXC                          |

- 3. Predicate Device:
The Ormco™ Spark™ Aligner System is substantially equivalent (as defined in 21 U.S.C. \$ 360c(i)) for all regulatory purposes to the legally marketed device Invisalign System K143630 cleared on November 18, 2015, product code NXC, by Align Technology.

{4}------------------------------------------------

#### Description of Device: 4.

The Ormco™ Spark™ Aligner System consists of a series of doctor-prescribed, custom manufactured, clear plastic removable orthodontic appliances (aligners) that incrementally move the patient's teeth from an original state to a treated state. Treatment planning, aligner design and aligner manufacture are supported by a proprietary software system. The Ormco™ SparkTM Aligner system consists of multiple interfacing software modules; Web, Design, Approver and Fabrication. The Spark Aligner Web software module is an online portal, where clinicians can create new patient profiles, plan treatments, manage patients and submit prescriptions, photos, images and digital scans to Ormco. Ormco technicians use a reference Anatomy software and the Ormco™ Spark™ Aligner Design software to create a 3D model of the patient's teeth from dental scans and to generate a 3D image of a final, treated state, as well as 3D aligner transitional treatment stage models, and submits them to the clinician. The clinician uses the Approver Software to review and approve the models, then Ormco uses the Fabrication software to produce resin molds for thermoforming the aligners.

The clinician receives the aligners and provides them, in sequential "stages", to the patient, confirming fit and monitoring treatment from the placement of the first aligner to the removal of the final aligner. The trays are held in place by pressure and can be removed by the patient at any time.

Several treatment options may be integrated into the Ormco Spark/Kappa aligner. Attachments or "buttons" may be prescribed by the dental practitioner to facilitate tooth movement and aligner anchorage. The dental practitioner may choose a standard dental composite and adhesive to bond the attachments to the dentition. The aligners are individually identified and dispensed to patients and are to be worn in a specific, prescribed sequence. Hooks may be designed into the aligner, then connected by an elastic to a toothbonded button on the opposite arch. to apply additional forces. Bite ramps are step features built into the lingual surfaces of the upper aligner arch that used by the clinician to prevent movement of the teeth during overbite correction treatment. Pontics are cavity spaces (that may or may not be "tooth-shaped") built into the aligner per clinician's request to address voids of missing teeth or other gaps during treatment. These treatment options are available in both the Ormco™ Spark™ Aligner as well as the Predicate Invisalign Aligner (please see Table 5.1).

Ormco also intends to trademark the Ormco Spark Aligner System as: Ormco Kappa Aligner System, Ormco Clear Aligner System, and Ormco Spark Clear Aligner System.

#### 5. Indications for Use:

The Ormco™ Spark™ Aligner System is indicated for the alignment of teeth during orthodontic treatment of malocclusion in patients with permanent dentition (i.e., all second molars). The Ormco™ Spark™ Aligner System positions teeth by way of continuous gentle force.

#### 6. Mode of Action:

The mechanism of action is similar to the predicate devices and supports a determination of substantial equivalence (as defined in 21 U.S.C. § 360c(i)). Orthodontic tooth movement 

{5}------------------------------------------------

occurs through forces applied by the appliance to the dentition as each tooth follows the programmed displacement based on a dental clinician's prescription.

#### 7. Description of Substantial Equivalence:

The Ormco™ Spark™ Aligner System is substantially equivalent (as defined in 21 U.S.C. \$ 360c(i)) for all regulatory purposes to the predicate device Invisalign (K143630). For purposes relevant to this regulatory filing, and as discussed in the previous subsections, technological characteristics, performance testing and clinical literature review indicate the substantial equivalence of the Ormco™ Spark™ Aligner System to the Primary Predicate, Invisalign Aligner System. There is sufficient information demonstrating that the proposed subject device provides a reasonable assurance of safety and efficacy. Both products represent new incarnations of an evolving technology with a long history of development.

## a) Technological Characteristics:

The technological principle for both the subject Spark/Kappa Aligner and predicate Invisalign Aligner is the treatment of tooth malocclusions using a series of plastic appliances that incrementally moves teeth to a desired end-state. A comparison between the technological characteristics of the Ormco™ Spark™ Aligner System and that of legally marketed predicate device, Invisalign (K143630) has is summarized in Table 5.1, and demonstrates that the design, technology and composition of the Ormco™ SparkTM Aligner System are substantially equivalent (as defined in 21 U.S.C. § 360c(i)) for all regulatory purposes to the predicate devices.

## b) Performance Testing:

The performance data compares the software, design, treatment characteristics and treatment process of the Spark Aligner to the predicate Invisalign Aligner. The stress relaxation, flexural and tensile strength were compared to the commercially-available Zendura Aligner material, as well as Invisalign's SmartTrack Aligner material. Comparison of mechanical testing of Spark Aligner material (TruGEN) and Invisalign Aligner material (SmarkTrack) indicates the substantial equivalence of their technological characteristics.

Bench testing of aligners was not performed due to the difficulty of evaluating these device types in a laboratory; however, translucency, staining, shelf life and shipping tests were carried on the thermoformed TruGENTM material which showed acceptable properties for all tested samples. Additionally, several aligner features were tested and found to be acceptable including attachment dimensional accuracy, cutout effects on stability, force evaluation of attachments and bite ramp and hook evaluation at specified cycles.

Biocompatibility testing performed on both TruGEN material and Spark aligners per ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process included cytotoxicity, sensitization and irritation, and intracutaneous reactivity evaluation.

The performance data demonstrates substantial equivalence of the subject Spark Aligner device to the Predicate Invisalign Aligner device.

c) Clinical Performance Data:

{6}------------------------------------------------

Clinical performance testing has not been completed for the Ormco™ Spark™ Aligner System. A comprehensive scientific literature review and analysis was conducted by orthodontic specialists on behalf of Ormco (please see Appendix A). Research articles and reports on the predicate device covered various subjects including the efficacy of the aligner on treating different types of orthodontic malocclusions.. There is sufficient information available from scientific literature and from the predicate devices to demonstrate that orthodontic, sequential, clear aligners provide reasonable assurance of safety and effectiveness.

## d) Substantial Equivalence Comparison:

The following table (Table 5.1) compares the Ormco™ Spark™ Aligner System to the predicate device, the Invisalign System (K143630), as well as the reference device ClearCorrect (K113618). There is sufficient evidence to demonstrate substantial equivalence of the proposed Spark™ Aligner device to the predicate Invisalign (K143630) device.

| Element                                 | Proposed<br>Ormco™ Spark™<br>Aligner System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Primary Predicate<br>Invisalign System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Predicate<br>ClearCorrect                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                           | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K143630                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K113618                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| 510(k) Sponsor                          | Sybron Dental<br>Specialties                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Align Technologies                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ClearCorrect, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Regulation Number                       | 21CFR872.5470                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21CFR872.5470                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21CFR872.5470                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Classification<br>Name           | Orthodontic Plastic<br>Bracket                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Orthodontic Plastic<br>Bracket                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Orthodontic Plastic<br>Bracket                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code                            | NXC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | NXC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | NXC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Class                            | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Intended Use                            | Orthodontic tooth<br>movement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Orthodontic tooth<br>movement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Orthodontic tooth<br>movement                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for Use                     | The Ormco™ Spark™<br>Aligner System is<br>indicated for the<br>alignment of teeth<br>during orthodontic<br>treatment of<br>malocclusion in<br>patients with permanent<br>dentition (i.e., all<br>second molars). The<br>Ormco™ Spark™<br>Aligner System<br>positions teeth by way<br>of continuous gentle<br>force.                                                                                                                                                                                                                                       | The Invisalign System<br>is indicated for the<br>alignment of teeth<br>during orthodontic<br>treatment of<br>malocclusion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | ClearCorrect System is<br>indicated for the<br>treatment of tooth<br>malocclusion in<br>patients with permanent<br>dentition (i.e. all<br>second molars), The<br>ClearCorrect System<br>positions teeth by way<br>of continuous gentle<br>force                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Target users                            | Dental Professionals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Dental Professionals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Dental Professionals                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Material                                | Thermoplastic<br>polyurethane -polyester<br>composite resin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Thermoplastic<br>polyurethane -polyester<br>composite resin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Thermoplastic<br>polyurethane resin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Description of<br>Appliance Application | Removable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Removable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Removable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Element                                 | Proposed<br>Ormco™ Spark™<br>Aligner System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Primary Predicate<br>Invisalign System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Reference Predicate<br>ClearCorrect                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Mode of Action                          | Orthodontic tooth<br>movement occurs<br>through forces applied<br>by the appliance to the<br>dentition as each tooth<br>follows the<br>programmed<br>displacement based on<br>a doctor's prescription                                                                                                                                                                                                                                                                                                                                                     | Orthodontic tooth<br>movement occurs<br>through forces applied<br>by the appliance to the<br>dentition as each tooth<br>follows the<br>programmed<br>displacement based on<br>a doctor's prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Orthodontic tooth<br>movement occurs<br>through forces applied<br>by the appliance to the<br>dentition as each tooth<br>follows the<br>programmed<br>displacement based on<br>a doctor's prescription                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Software Used for<br>Ordering Workflow  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Attachments                             | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Hooks/Cutouts                           | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Bite ramps                              | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Pontics                                 | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Single Use                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Non-Sterile Packaging                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Treatment Process                       | Doctor uploads<br>Patient's dental scans<br>into proprietary<br>software; doctor uses<br>software for case<br>viewing and treatment<br>planning; Ormco<br>technician receives case<br>and uses proprietary 3D<br>software to generate<br>models for the aligners<br>based on the<br>prescription (desired<br>outcome); Doctor<br>approves treatment plan<br>and final treated state;<br>manufacturer crafts and<br>produces aligners:<br>aligners and ships<br>them to doctor, who<br>then provides them to<br>the patient, confirming<br>fit and design. | The proprietary<br>Invisalign System 3-D<br>software generates the<br>image of a final, treated<br>state and then<br>interpolates a series of<br>images that represent<br>intermediate teeth<br>states. The dental<br>practitioner then<br>reviews these images to<br>depict, edit, view,<br>monitor, and approve<br>an orthodontic<br>treatment plan. The<br>dental practitioner has<br>the option to reject or<br>request modifications<br>to the set-up prior to<br>approving it for aligner<br>fabrication. Once the<br>doctor approves the set-<br>up, the series of<br>custom-made aligners<br>are then manufactured,<br>packaged, and shipped<br>to the dental<br>practitioner to be<br>dispensed to the patient<br>for treatment. | Doctor sends patient<br>scans, molds and<br>prescription to<br>ClearCorrect; Utilizing<br>standard dental<br>software used for tooth<br>alignment,<br>ClearCorrect designs a<br>series of plastic trays<br>intended to gradually<br>realign the patient's<br>teeth in accordance<br>with the physician's<br>prescription. The<br>prescribing physician<br>reviews and approves<br>the model scheme<br>before the molds are<br>produced. Once<br>approved, ClearCorrect<br>produces the trays,<br>which are formed of<br>clear, thin<br>thermoformed<br>polyurethane plastic.<br>The trays are sent back<br>to the dental health care<br>professional, who then<br>provides them to the<br>patient, confirming fit<br>and design. |

### Table 5.1: Comparison of Proposed and Predicate Devices

{7}------------------------------------------------

{8}------------------------------------------------

### e) Substantial Equivalence Conclusion:

The Ormco™ Spark™ Aligner System is substantially equivalent (as defined in 21 U.S.C. \$ 360c(i)) for all regulatory purposes to the predicate device Invisalign (K143630). For purposes relevant to this regulatory filing, and as discussed in the previous subsections, technological characteristics, performance testing and clinical literature review indicate the substantial equivalence of the Ormco™ Spark™ Aligner System to the Primary Predicate, Invisalign Aligner System. Both products represent new incarnations of an evolving technology with a long history of development. There is sufficient information demonstrating that the proposed subject device provides a reasonable assurance of safety and efficacy.

{9}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC/K182826](https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NXC/K182826)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
