FS ALIGNER

K093821 · Allesee Orthodontic Appliances · NXC · Aug 27, 2010 · Dental

Device Facts

Record IDK093821
Device NameFS ALIGNER
ApplicantAllesee Orthodontic Appliances
Product CodeNXC · Dental
Decision DateAug 27, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The FS Aligner System is intended for minor anterior tooth movement in patients with permanent dentition (second molars). The FS Aligner System positions teeth by way of continuous gentle force.

Device Story

FS Aligner is a removable orthodontic aligner system for correcting minor anterior mal-alignments. Input consists of dental impressions or physical models provided by a dental practitioner. A laboratory technician uses a computerized setup as a 'blueprint' to guide the manual, sequential repositioning of teeth on physical molds. For each stage of progressive correction, a clear, lightweight, flexible plastic aligner is fabricated. The system typically involves 5-7 aligners per iteration; complex cases may require 1-8 iterations. The practitioner delivers the aligners to the patient, who wears them to apply continuous gentle force to reposition teeth. The process is iterative, requiring new impressions or bite registrations for subsequent sequences. The device benefits patients by providing an aesthetic, removable alternative to conventional wire and bracket orthodontics.

Clinical Evidence

No clinical data. Substantial equivalence is supported by biocompatibility testing (cytotoxicity, irritation, and sensitization) and comparison of technological characteristics.

Technological Characteristics

Removable clear aligners made of thin thermoformed plastic. Operates via mechanical force application for tooth movement. Manufacturing involves manual repositioning of teeth on physical molds guided by a computerized setup 'blueprint'. No electronic or energy-based components. Standalone device.

Indications for Use

Indicated for patients with permanent dentition (second molars) requiring minor anterior tooth movement.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # Section III - 510(k) Summary of Safety and Effectiveness #### Submitter: Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, California 92867 (714) 516-7602 - Phone (714) 516-7488 - Facsimile Wendy Garman - Contact Person Date Summary Prepared: August 2010 #### Device Name: - Trade Name FS Aligner . - . Common Name - Aligner, Sequential - Classification Name Orthodontic Plastic Bracket, per 21 CFR § 872.5470 . #### Devices for Which Substantial Equivalence is Claimed: - . Align Technology, Invisalign - Red, White & Blue, Allessee Orthodontic Appliances ● #### Device Description: FS Aligner is an aligner system which offers a solution to those patients who want a simple, aesthetic system of removable aligners to correct minor anterior mal-alignments in patients with permanent dentition (second molars) without the use of conventional wire and bracket orthodontic technology. The system consists of a series of clear, lightweight, flexible, plastic aligners. Each aligner applies incremental progressive force to reposition teeth to a more ideal alignment as prescribed by the treating dental practitioner. The practitioner will make a dental impression of the patient's teeth and select the teeth that are to be repositioned (corrected). The impressions, or models of the teeth made from the impression, are sent to AOA with a prescription form. The mold will be scanned and the resulting electronic mold with be altered within the computer program to both determine the {2}------------------------------------------------ degree of correction and to offer the practitioner the opportunity to review, alter, and approve of the digital corrections: Once approved, the teeth on the physical mold(s) will be sequentially repositioned by hand followed by the fabrication of an aligner for that stage of progressive correction. After 5-7 progressive corrections the resulting aligners are returned to the practitioner for delivery to the patient. Once that sequence of treatment is completed the practitioner will submit a new impression or bite registration to AOA laboratories for the manufacture of the next group or sequence of progressive corrections and aligners. Depending on the complexity of the mal-alignment it is possible that a patient may require 1-8 such iterations of 5-7 aligners. The FS Aligner system will be mailed (shipped) to the practitioner, who in turn, will provide them to their patient with instructions for use. # Intended Use of the Device: The FS Aligner System is intended for minor anterior tooth movement in patients with permanent dentition (second molars). The FS Aligner System positions teeth by way of continuous gentle force. # Summary of Technological Characteristics: FS Aligner is substantially equivalent to other legally marketed devices in the United States. FS Aligner functions in a manner similar to and is intended for the same use as Invisalign that is currently marketed by Align Technology and Red, White & Blue that is currently marketed by Allessee Orthodontic Appliances. | Features | FS Aligner | Invisalign | Red, White & Blue | |--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | A series of clear plastic<br>aligners intended for<br>treatment of tooth<br>malocclusion by way of a<br>continuous gentle force. Each<br>aligner is produced using<br>manually created individual<br>molds each incorporating a<br>progressive correction. | A series of clear, plastic<br>aligners (from 1 – to<br>over 100) intended for<br>treatment of tooth<br>malocclusion by way of<br>a continuous gentle<br>force. Each aligner<br>produced through<br>computerized<br>generation of individual<br>molds each<br>incorporating a<br>progressive correction. | A series of 3 clear,<br>plastic retainers<br>intended to be used to<br>correct minor to<br>intermediate tooth mal-<br>alignments by moving<br>teeth progressively to a<br>final, treated state. | | Aligner<br>Material | Thin thermo formed plastic<br>material. | Thin thermo formed<br>plastic material. | Thin thermo formed<br>plastic material. | | Mode of Use | | | | | | Each appliance is worn by the patient as determined by the treating dental practitioner, generally 2-4 weeks prior to being replaced by the next aligner in sequence. | Each appliance is worn by the patient as determined by the treating dental practitioner, generally 2weeks prior to being replaced by the next aligner in sequence. | Each appliance is worn by the patient for 2-4 weeks, or until the patient feels the appliance is passive, in the appropriate sequence of Red, White & Blue retainers. | | Description of Appliance Application | Removable | Removable | Removable | | Manufacturing Method | Each set up and aligner is manually formed. The technician will cut the desired teeth from the mold and will progressively reposition the teeth required to meet the prescript goal of alignment. The set up portion of the process will be guided by a computerized version of the set up providing information similar to a "blueprint" for the technician. After each repositioning the technician will create a corresponding aligner for a total of 5-7 aligners. These will be returned to the practitioner and delivered to the patient. Following the completion of wear by the patient a new bite registration or impression is returned to the laboratory (AOA) for the next sequence of aligners to be fabricated: This system of "reboot" and acquiring benchmarks to further correct the mal- alignment is continued as needed. | The aligners are manufactured through the use of three- dimensional laser scanning technology - scanning the patient's mold into computer software. From the scanned image, and following a practitioner's prescription, the software generates the image of the final, treated state and then interpolates a series of images that represent intermediate states of alignment. The resulting computer "set ups" relay this information to rapid prototyping machines that produce the physical molds and the aligners are produced from each mold to produce the sequence of aligners. | The retainers are manually formed. The technician will cut the desired teeth from the mold and will progressively repositio the teeth. After each repositioning the technician will create a corresponding appliance for a total of active appliances. | {3}------------------------------------------------ and the state of the state of the states and 1 : {4}------------------------------------------------ # Non-Clinical Test Data: Biocompatibility studies (cytotoxicity, irritation and sensitization) have been completed which demonstrates that the material used to produce the FS Aligner is safe for its intended use. ### Clinical Test Data: Clinical testing has not been conducted on this product. #### Conclusion: Based upon the biocompatibility and similar technological characteristics to the predicate devices, the performance of the FS Aligner is deemed to be substantially equivalent to Invisalign and Red, White & Blue. {5}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Allesee Orthodontic Appliances C/O Ms. Wendy Garman Sybron Dental Specialties, Incorporated 1717 West Collins Avenue Orange, California 92867 Re: K093821 AUG 2 7 2010 Trade/Device Name: FS Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NXC Dated: July 30, 2010 Received: August 2, 2010 Dear Ms. Garman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ #### Page 2- Ms. Garman Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Susan Russo Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {7}------------------------------------------------ K093821 # Indications for Use 510(k) Number (if known): Device Name: FS Aligner Indications For Use: The FS Aligner System is intended for minor anterior tooth movement in patients with permanent dentition (second molars). The FS Aligner System positions teeth by way of continuous gentle force. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) R.S. Butz DDS for Dr. Susan Kummer . Ision Sign-Off Page 1 of 1 Islon of Anesthesiology, General Hospital ifection Control, Dental Devices 10(k) Number:
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...