Vatech Clismile is indicated for orthodontic movement of natural teeth, excluding mandibular bicuspid teeth.
Device Story
Vatech Clismile is a self-ligating orthodontic bracket system used to treat malocclusion. The device consists of a polycrystalline alumina ceramic body and a nickel-cobalt alloy metal cap. It is applied to teeth by an orthodontist or dental professional to transmit physical force via orthodontic wires to correct tooth alignment. The device functions by securing the archwire within the bracket slot using the integrated metal cap mechanism. It is available in MBT and ROTH designs. By providing a stable attachment point for orthodontic forces, the device facilitates controlled tooth movement, aiding in the correction of dental malocclusions.
Clinical Evidence
Bench testing only. Performance evaluation included dimensional analysis per ISO 27020, biological stability testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), and ISO 10993-11 (systemic toxicity), and corrosion/hazardous element testing per ISO 22674. Results confirmed the device meets design specifications and is equivalent to the predicate.
Technological Characteristics
Materials: Polycrystalline Alumina (100%) body, Nickel-Cobalt Alloy (MP35N) cap. Design: Self-ligating bracket, MBT/ROTH configurations. Standards: ISO 27020 (dimensions), ISO 10993 (biocompatibility), ISO 22674 (corrosion). Form factor: Half-transparent ceramic body with metal cap. Non-sterile packaging.
Indications for Use
Indicated for orthodontic movement of natural teeth, excluding mandibular bicuspid teeth.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 6, 2022
Rayence Co.,Ltd Dave Kim Medical Device Regulatory Affairs Mtech Group 7505 Fannin St. Ste 610 Houston, Texas 77054
Re: K220689
Trade/Device Name: Vatech Clismile Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: February 28, 2022 Received: March 9, 2022
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K220689
Device Name Vatech Clismile
Indications for Use (Describe)
Vatech Clismile is indicated for orthodontic movement of natural teeth, excluding mandibular bicuspid teeth.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K220689
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date 510k summary prepared: January 28th, 2022
#### Submitter's Name, address, telephone number, a contact person:
| Submitter's Name : | Rayence Co., Ltd. |
|-------------------------|---------------------------------------------------------|
| Submitter's Address: | 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea |
| Submitter's Telephone: | +82-31-8015-6459 |
| Contact person: | Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459 |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net) |
| Address: | 7505 Fannin St. Ste 610-V111, Houston, TX 77054 |
| Telephone: | +713-467-2607 |
Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
| Trade/proprietary name | : Vatech Clismile |
|------------------------|------------------------------------|
| Common Name | : Orthodontic ceramic bracket |
| Regulation Number | : 21 CFR 872.5470 |
| Regulation Name | : Bracket, Ceramic, Orthodontic |
| Regulatory Class | : Class II |
| Product Code | : NJM(Orthodontic plastic bracket) |
#### Predicate Device :
| Trade/Device Name | : In-Ovation C |
|-------------------|---------------------------------|
| Common Name | : Orthodontic ceramic bracket |
| 510(k) Number | : K060837 |
| Regulation Number | : 21 CFR 872.5470 |
| Regulation Name | : Bracket, Ceramic, Orthodontic |
| Regulatory Class | : Class II |
| Product Code | : NJM |
{4}------------------------------------------------
### 2. Device Description
This product is an orthodontic bracket and is used with orthodontic wire materials and other orthodontic products to treat malocclusion of the upper and lower teeth.
It is used to correct teeth by transmitting physical force to each tooth, and is a selfligation bracket using a ceramic body and a metal cap made of a metal material. Shelf-life is not applicable because of the low likelihood of time-dependent product degradation.
### 3. Indication for use
Vatech Clismile is indicated for orthodontic movement of natural teeth, excluding mandibular bicuspid teeth.
## 4. Summary of Design Control Risk management
Vatech Clismile orthodontic brackets were developed with the same structure, material / operation method, and purpose as In-Ovation C.
The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.
{5}------------------------------------------------
# 5. Summary of the technological characteristics of the device compared to the predicate device:
Vatech Clismile described in this 510(k) has the same indications for use and similar technical characteristics as its predicate devices, IN-OVATION C (K060837).
| Characteristic | Proposed Device | Predicate Device | Similarity |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Rayence CO., LTD | DENTSPLY International | - |
| Product Name | Vatech Clismile | IN-OVATION C | - |
| Picture | Image: Vatech Clismile | Image: IN-OVATION C | Similar |
| 510(k) number | K220689 | K060837 | - |
| Indications for<br>use | Vatech Clismile is indicated for<br>orthodontic movement of natural<br>teeth, excluding mandibular<br>bicuspid teeth. | In-Ovation C is indicated for<br>orthodontic movement of natural<br>teeth, excluding mandibular<br>bicuspid teeth. | Same |
| Material<br>composition<br>of Bracket | Polycrystalline Alumina(100%) | Polycrystalline Alumina(100%) | Same |
| Material<br>composition<br>of Door | Nikel-Cobalt Alloy(100%) | Nikel-Cobalt Alloy(100%) | Same |
| Transparency | Half-transparency | Half-transparency | Same |
| Bracket design | MBT, ROTH designs with<br>and without hook,<br>conforming to ISO<br>27020:2010 Dentistry –<br>Brackets and tube for Use in<br>Orthodontics | MBT, ROTH designs with<br>and without hook,<br>conforming to ISO<br>27020:2010 Dentistry –<br>Brackets and tube for Use in<br>Orthodontics | Same |
| Self-ligating<br>mechanism | Yes | Yes | Same |
| Orientation<br>marking | Yes | Yes | Same |
| Non-sterile<br>packaging | Yes | Yes | Same |
| In-out(mm) | 1.05 to 1.54 | 1.05 to 1.55 | Same |
| Torque(°) | -11 to +17 | -11 to +17 | Same |
| Angulation(°) | 0 to 13 | 0 to 13 | Same |
| slot sizes | 0.022 inch | 0.018" / 0.022 inch | Similar |
#### Comparison table
{6}------------------------------------------------
### 6. Summary of Performance Testing
IN-OVATION C and Vatech Clismile have the same composition and raw material of the components constituting the device, and the purpose and principle of operation of the calibration bracket are the same. Bracket generally uses a one-kit product consisting of 20 brackets. The K100-552-81 model of In-Ovation-C is composed of 16 ceramic and 4 metal brackets, and the Clismile is also composed of the same number of sets.
The ceramic bracket is composed of alumina (Body) and Ni-Co alloy (Cap, MP35N)
The design characteristics of the product were evaluated for seven items specified in ISO 27020, which is a bracket standard. As representative products, dimensional analysis of "Clismile 12-1223" and "IN-OVATION C H100-132-81" corresponding to the maximum canine was conducted, and it was confirmed that the dimensions were the same within the error range.
### 7. Summary for any testing and reference guidance:
> Dimension evaluation of prescription of orthodontic bracket according to standard ISO27020 (Brackets and tubes for use in orthodontics)
> Evaluation report on biological stability according to standard ISO10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin sensitization, ), ISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
> Evaluation of hazardous element content and corrosion test for metal cap bracket parts according to standard ISO 22674(Metallic materials for fixed and removable restorations and appliances)
#### 8. Conclusions:
In accordance with the performance outcomes. Vatech Clismile demonstrated equivalent performance compared to In-Ovation-C. Therefore, Rayence claims the substantial equivalency between the proposed device and predicate device in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.