This device is intended for the orthodontic movement of teeth. It is used temporarily and is removed after Orthodontic treatment has been completed.
Device Story
Symetri Clear is a single-use, polycrystalline alumina orthodontic bracket; used by orthodontists/dental professionals in clinical settings. Brackets are bonded to teeth using orthodontic adhesive; temporary color-coded plastic axis indicators or labial tie wings aid visual placement; temporary ink dots facilitate tooth designation. Archwires are threaded through slots and secured with ligature ties to exert pressure for tooth movement. Brackets are removed post-treatment using a dedicated debonding instrument. Device benefits include effective tooth movement and aesthetic appearance due to translucent material.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including torque strength, double tie wing strength, shear/tensile bond strength, wire drag, debonding removal fracture, hook strength, and staining. Biocompatibility assessed per ISO 10993 and ISO 14971.
Technological Characteristics
Material: Polycrystalline alumina (translucent). Construction: One-piece base and bracket. Manufacturing: CIM (Ceramic Injection Molding). Features: Laser-etched base, removable plastic axis indicator, temporary ink dot for designation. Dimensions: ISO 27020 compliant. Non-sterile packaging.
Indications for Use
Indicated for orthodontic movement of teeth in patients requiring orthodontic treatment. Used temporarily and removed upon treatment completion.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
January 31, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Ormco Corporation Ms. Tara Bonny Senior Regulatory Affairs Specialist 1717 W. Collins Avenue Orange, California 92867
Re: K172845
Trade/Device Name: Symetri Clear Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: December 20, 2017 Received: December 22, 2017
Dear Tara Bonny:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Mary S. Runner -S
forTina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172845
Device Name
Symetri™ Clear
Indications for Use (Describe)
This device is intended for the orthodontic movement of teeth. It is used temporarily and is removed after Orthodontic treatment has been completed.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Ormco" in a bold, sans-serif font. The word is a dark blue color. There is a trademark symbol to the right of the word.
### 510(k) SUMMARY K172845
This 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92.
| 1. | 510(k) Submitter................................ | Ormco Corporation<br>1717 W. Collins Avenue<br>Orange, California 92867 |
|----|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Contact................................ | Tara Bonny<br>Regulatory Affairs<br>Phone: 909-962-5856<br>Fax: 909-962-5694<br>Email: Tara.Bonny@kavokerr.com |
| 3. | Additional Contact................................ | Valerie A. Cimmarusti<br>Vice President,<br>Regulatory Affairs & Design Assurance (RADA)<br>Phone: 714-516--7476<br>Fax: 909-962-5694<br>Email: Valerie.Cimmarusti@kavokerr.com |
| 4. | Date Prepared................................ | January 2, 2018 |
| 5. | Device | |
| | Proprietary Name................................ | Symetri™ Clear<br>Component of: Symetri™ Straight-Wire™ System |
| | Classification Name................................ | Bracket, Ceramic Dental |
| | Regulation Number................................ | 21 CFR 872.5470 |
| | Product Code................................ | NJM |
| | Medical Specialty Panel................................ | Dental |
| | Classification................................ | Medical Device, Class II |
## Predicate Device 6.
Symetri Clear™ is substantially equivalent to 3M Unitek Corporation's Clarity™ ADVANCED Ceramic Brackets (K102803) cleared on February 17, 2011, product code NJM.
#### 7. Reference Predicate Device
Damon™ 4Clear, K081415
{4}------------------------------------------------
#### 8. Device Description
Symetri Clear, is a single-use orthodontic bracket made from polycrystalline alumina. This device is intended for the orthodontic movement of teeth. It is used temporarily and is removed after Orthodontic treatment has been completed. After following standard office protocols for tooth preparation, the brackets are bonded to the teeth with an orthodontic adhesive or cement. To aid with visual placement of the bracket to the tooth surface, the bracket has a removable, color coded, plastic axis indicator that is lifted away from the bracket following tooth placement or temporary color coded labial tie wings that are removed with a toothbrush and water. To identify tooth designation, the brackets include a temporary ink dot that is removed with a toothbrush and water after the brackets have been placed on the tooth. After the brackets are placed, an archwire is then inserted / threaded through the bracket's archwire slot and is held in place with a ligature tie. These ligatures are tightened, or ligated, around the 'wings' of the bracket and over the archwire. To aid in optimal teeth movement, the individual brackets can also be ligated to other brackets or other orthodontic devices. After the teeth have moved into their proper positions, the orthodontic / dental professional will remove the brackets with the Symetri Clear Debonding Instrument / Pliers. This step is also known as debonding.
The Symetri Clear Debonding Instruments / Plier was included for informational purposes only, not for review for clearance.
#### 9. Indications for Use
This device is intended for the orthodontic movement of teeth. It is used temporarily and is removed after Orthodontic treatment has been completed.
Although Symetri Clear's Indications for Use statement is not identical to the predicate devices, the minor differences in phrasing do not change the intended use.
#### 10. Comparison of Technology
Symetri Clear is a non self-ligating, single-use orthodontic bracket's design has one-piece base and bracket construction, and is made from translucent alumina. To provide optional bond strength and effective bracket debonding, the base has a laseretched pattern. To ensure accurate placement, it uses a removable plastic component that is inserted into the horizontal and vertical slots of the bracket or temporary color which is incorporated into the labial surface instead of the plastic indicator. A temporary ink dot facilitates tooth designation. All ink is removed after the bracket has been placed on the tooth. The bracket is designed to remain intact during debonding.
As with Symetri Clear, the predicate device. Clarity ADVANCED, is also a non selfligating, single-use orthodontic bracket which has a one-piece base and bracket construction, and is made from translucent alumina. It also includes a temporary ink dot to facilitate tooth designation but does not include a removable plastic indicator for bracket placement. Bracket placement is achieved via temporary horizontal and vertical colored ink markers. Like Symetri Clear, all ink is removed after the bracket has been placed on the tooth. It also differs from Symetri Clear in its bracket base; it is a glassgrit, mechanical-locking base with a stress concentrator for predicable squeeze debonding. The stress concentrator on the base, causes the bracket to collapse during debonding by design.
{5}------------------------------------------------
The reference predicate, Damon Clear2 (K081415), is also a single-use orthodontic bracket, with a one-piece base and bracket construction, and is made from translucent alumina. Similar to Symetri Clear, it utilizes a removable plastic indicator for tooth placement and a temporary ink dot for tooth designation. Unlike Symetri Clear and Clarity ADVANCED, this bracket is self-ligating.
Clarity ADVANCED (K102803), Damon Clear / Damon Clear2 (K081415) and Symetri Clear bracket systems (K172845), all offer a hand instrument accessory designed for bracket debonding. While the debonding process is not identical, these variations do not change the indications for use.
Symetri Clear shares the same indications for use as the primary and reference predicate devices. Differences between the devices do not raise any questions of safety and effectiveness.
#### 11. Non-Clinical Performance Data
Non-clinical performance data included testing results for Torque Strength (also known as Wire Torque), Double Tie Wing Strength, Bond Strength Shear (also known as Shear Test), Bond Strength Tensile (also known as Tensile Strength), Wire Drag, Debonding Removal Fracture (also known as Bracket Removal Fracture), Hook Strength and Staining. The testing analysis shows that Symetri Clear brackets perform comparably to the predicate devices.
Metal Ligation and Double Tie bench testing was also conducted, however, not compared to a predicate.
Biocompatibility assessment and testing was performed using standard risk assessment techniques and in consideration of FDA and internationally recognized guidance's.
Symetri Clear brackets were inspected and met the dimensions itemized in ISO 27020. Dentistry - Brackets and tubes for use in orthodontics.
| Element | Clarity™ ADVANCED<br>(K102803) | Damon Clear /<br>Damon Clear2<br>(K081415) | Symetri™ Clear<br>(K172845) |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | K102803 | K081415 | K172845 |
| Trade Name | Clarity ADVANCED<br>Ceramic Brackets | Damon Clear /<br>Damon Clear2 | Symetri Clear |
| Target Users | Licensed Orthodontic /<br>Dental Professionals | Licensed Orthodontic /<br>Dental Professionals | Licensed Orthodontic /<br>Dental Professionals |
| Device<br>Description | Clarity ADVANCED is<br>a single-use<br>polycrystalline alumina | Damon Clear / Damon<br>Clear2 is a single-use<br>polycrystalline alumina | Symetri Clear is a<br>single-use<br>polycrystalline alumina |
| Element | Clarity™ ADVANCED<br>(K102803) | Damon Clear /<br>Damon Clear2<br>(K081415) | Symetri™ Clear<br>(K172845) |
| | orthodontic twin<br>bracket. | orthodontic self-ligating<br>bracket. | orthodontic twin<br>bracket. |
| Indications for<br>Use<br>(Identical to 510k) | Clarity ADVANCED<br>Ceramic Brackets are<br>intended for use in<br>orthodontic treatment.<br>The brackets are affixed<br>to teeth so that pressure<br>can be exerted on the<br>teeth. | Damon 4Clear* is a<br>bracket system intended<br>to aid in the movement<br>of patient teeth during<br>orthodontic treatment.<br>* Name at time of submission. | This device is intended<br>for the orthodontic<br>movement of teeth. It is<br>used temporarily and is<br>removed after<br>Orthodontic treatment<br>has been completed. |
| Common Name | Orthodontic Ceramic<br>Bracket | Orthodontic Ceramic<br>Bracket | Orthodontic Ceramic<br>Bracket |
| Classification<br>Name | Bracket, Ceramic,<br>Dental | Bracket, Ceramic,<br>Dental | Bracket, Ceramic,<br>Dental |
| Class | II | II | II |
| Product Code | NJM | NJM | NJM |
| Material | Alumina (Al2O3) | Alumina (Al2O3) | Alumina (Al2O3) |
| Clarity | Translucent | Translucent | Translucent |
| Bracket Design | 1-Piece Base and<br>Bracket Construction | 1-Piece Base and<br>Bracket Construction<br>and a slide | 1-Piece Base and<br>Bracket Construction |
| Ligation | Non Self-Ligating | Self-Ligating | Non Self-Ligating |
| Manufacturing<br>Method | CIM | CIM | CIM |
| Hooks | CIM Hooks | CIM Hooks | CIM Hooks |
| In/Out | .020"-.045" | .035" - .059" | .020" - .045" |
| Torque | +17° thru -17° | +2° thru -11° | +17° thru -17° |
| Angulation | Up to +8° | Up to +9° | Up to +8° |
| Rotation | Up to 2° | 0° | Up to 2° |
| Bracket<br>Identification<br>and Placement | Temporary color coded<br>placement indicator<br>system to aid in bracket<br>positioning<br>Temporary colored dot | Color coded plastic jig<br>to aid in visual<br>placement | Color coded plastic axis<br>indicator or temporary<br>color coded labial tie<br>wings to aid in visual<br>placement |
| | to indicate<br>Upper/Lower, Left/Right | Temporary colored dot<br>to indicate<br>Upper/Lower, Left/Right | Temporary colored dot<br>to indicate<br>Upper/Lower, Left/Right |
| Element | Clarity™ ADVANCED<br>(K102803) | Damon Clear /<br>Damon Clear2<br>(K081415) | Symetri™ Clear<br>(K172845) |
| Material<br>Composition for<br>Bracket<br>Identification<br>and Placement | Unknown | Bracket Identification –<br>Blue*:<br>-Polypropylene<br>Copolymer<br>- Copper, [29H, 31H-<br>Phthalocyaninato(2-)-<br>N29, N30, N31<br>- Titanium Dioxide<br>Bracket Identification –<br>Red*:<br>-Polypropylene<br>Copolymer<br>-Iron Oxide<br>-Zinc Stearate<br>Bracket Identification –<br>Temporary Acrylic<br>Paint*<br>-Polyethylene Glycol<br>Octylphenyl Ether<br>*Per an independent<br>accredited medical<br>research service company,<br>the finished device is<br>considered to have met<br>biological evaluation and<br>risk assessment<br>requirements (ISO 10993<br>and ISO 14971). | Bracket Identification –<br>Blue*:<br>-Polypropylene<br>Copolymer<br>- Copper, [29H, 31H-<br>Phthalocyaninato(2-)-<br>N29, N30, N31<br>- Titanium Dioxide<br>Bracket Identification –<br>Red*:<br>-Polypropylene<br>Copolymer<br>-Iron Oxide<br>-Zinc Stearate<br>Bracket Identification –<br>Temporary Acrylic<br>Paint*<br>-Polyethylene Glycol<br>Octylphenyl Ether<br>*Per an independent<br>accredited medical<br>research service company,<br>the finished device is<br>considered to have met<br>biological evaluation and<br>risk assessment<br>requirements (ISO 10993<br>and ISO 14971). |
| Bracket Base | Micro-crystalline,<br>mechanical bonding,<br>surface for optimal bond<br>strength | Laser etch pattern<br>strategically cut onto the<br>bottom of the bracket for<br>optimal bond strength<br>between the bracket and<br>tooth | Laser etch pattern<br>strategically cut onto the<br>bottom of the bracket for<br>optimal bond strength<br>between the bracket and<br>tooth |
| Single Use | Yes | Yes | Yes |
| Non-Sterile<br>Packaging | Yes | Yes | Yes |
| Debonding Hand<br>Instrument<br>Accessory | Unitek Self-Ligating<br>Bracket Debonding | Damon Clear / Damon<br>Clear 2 Debonding | Symetri Clear<br>McLaughlin, Bennett &<br>Trevisi – Debonding |
| Element | Clarity TM ADVANCED<br>(K102803) | Damon Clear /<br>Damon Clear2<br>(K081415) | Symetri TM Clear<br>(K172845) |
| *Included for<br>informational<br>purposes only, not<br>for review for<br>clearance | Instrument<br>(REF: 804-170) | Instrument<br>(REF: 866-4019) | Instrument / Pliers<br>(REF: 866-4025) |
#### 12. Predicate, Predicate Reference and Proposed Device Comparison
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### 13. Clinical Performance Data
Clinical performance testing has not been performed for Symetri Clear.
#### Conclusion as to Substantial Equivalence 14.
Based upon similarities in the indications for use and technology, biocompatibility testing and performance testing, Symetri Clear brackets have been shown to be substantially equivalent to Clarity ADVANCED brackets (K102803) and to Damon Clear/Damon Clear2 brackets (K081415).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.