FASTBRACES CERAMIC BRACKETS

K130446 · Orthodontic Design and Production, Inc. · NJM · Jun 7, 2013 · Dental

Device Facts

Record IDK130446
Device NameFASTBRACES CERAMIC BRACKETS
ApplicantOrthodontic Design and Production, Inc.
Product CodeNJM · Dental
Decision DateJun 7, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

FastBraces® Ceramic Brackets are intended for orthodontic movement of natural teeth.

Device Story

FastBraces® Ceramic Brackets are orthodontic appliances used to move teeth into desired positions. The device consists of molded polycrystalline alumina (ceramic) brackets bonded to the surface of natural teeth. Brackets feature archwire slots, tie wing undercuts for ligatures, and distal-gingival hooks. Orthodontists operate the device by bonding brackets to teeth and threading orthodontic wires through the slots. Tensioned ligating wires and elastic ligatures exert force on the brackets to guide tooth movement along a pre-determined path. The device is used in a clinical setting for the duration of orthodontic treatment (up to 30+ months). The aesthetic ceramic material provides a tooth-colored alternative to metal brackets. Benefits include effective tooth alignment and improved patient aesthetics.

Clinical Evidence

No clinical performance testing was conducted. Substantial equivalence is supported by non-clinical bench testing, including shear strength, wire torque, wire drag, and bracket removal tests, alongside a literature review for biocompatibility per ISO 10993-1:2009.

Technological Characteristics

Material: Polycrystalline alumina (ceramic). Design: True-twin tie wings, archwire slots (.018" or .022"), base flanges for adhesion, and distal-gingival hooks. Biocompatibility: ISO 10993-1:2009 compliant. Form factor: Molded ceramic body with rounded edges. Connectivity: None. Energy source: None (mechanical). Sterilization: Delivered non-sterile.

Indications for Use

Indicated for orthodontic movement of natural teeth in patients of all ages when prescribed by an orthodontist.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo with the letters 'odp' inside of a circle. The letters 'odp' are in a sans-serif font, with the 'o' on the left, the 'd' in the middle, and the 'p' on the right. The circle is surrounded by a larger circle, and the entire logo has a slightly distressed or textured appearance. The background is white. **JUN 07 2013** # 510(k) Summary 510(k) Submitter: Orthodontic Design & Production, Inc. 1370 Decision Street, Suite D Vista, California 92081 Official Contact Person: Richard Merrell Requlatory Affairs Manager Telephone: (760) 734-3995 ext. 25 Facsimile: (760) 734-1735 E-mail: rm@odpinc.com Date Summary was Prepared: February 8, 2013 Device Trade Name: FastBraces® Ceramic Brackets Common Name: Orthodontic Ceramic Bracket Classification Name: Orthodontic Ceramic Bracket, Class II (21 CFR 872.5470, Product Code: NJM) Predicate Device: K922499, Reflections Ceramic Dental Bracket #### Description of Device: Ceramic orthodontic brackets are small devices that are intended to teeth, upon which an orthodontic wire is placed to move the teeth to desired positions. They are indicated for orthodontic treatment in patients of all ages when prescribed by an orthodontist. Ceramic orthodontic brackets are primarily offered as an aesthetic alternative to metal orthodontic brackets. They are close to natural tooth coloring, and blend in well enough that they are not as visible as metal brackets. This aesthetic look is popular with many patients, and especially older patients. Ceramic orthodontic brackets have been in use throughout the orthodontic industry for approximately thirty years. Like its predicate, FastBraces® ceramic brackets are manufactured from polycrystalline alumina (ceramic) material, and have bases that are designed to provide maximum adhesion to the tooth while still allowing for easy and complete removal when necessary. The brackets incorporate a water soluble color placement dot as an indicator for correct selection of brackets for each tooth. Like its predicate. FastBraces® ceramic brackets do not incorporate medicinal substances, tissues, or blood products. They do not include software or accessories, and are delivered non- 1370 Decision Street, Suite D Vista, CA 92081 Phone (760) 734-3995 Fax (760) 734-1735 USA Toll Free 1-800-383-5301 www.odpinc.com {2}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo with the letters 'odp' inside a circular design. The 'o' is on the left, the 'd' is in the middle, and the 'p' is on the right. The letters are in a bold, sans-serif font. The logo has a circular border around the letters, and there is a faint outer circle around the border. The logo appears to be slightly distressed or aged, with some imperfections in the image quality. sterile to the end user. Ceramic brackets are intended to be used only once by a single patient. Product labels contain appropriate "do not reuse" symbols. Orthodontic ceramic brackets are used for the duration of orthodontic treatment, which can last more than 30 months. During this time, the devices remain in direct contact with the patient's oral cavity. Because the intended purpose of the device is clearly understood by licensed orthodontists, instructions for use are not provided. With the exception of the materials from which they are constructed, the form, fit, and function of orthodontic ceramic brackets are identical to those of traditional metal orthodontic brackets. Orthodontic ceramic brackets, like metal brackets achieve their intended purpose (to move teeth into a desired position) through industry standard "prescriptions" that are programmed into the design of the brackets. Specific torques, angulations, and distal offset dimensions are designed into each bracket, along with archwire slots that are designed to accommodate the correct size archwire (typically .018" or .022" thick). Ceramic brackets are designed with tie wing undercut spaces for orthodontic ligatures. They have a molded ceramic bracket body with rounded corners and edges, and rounded hooks on the distal-gingival tie wings to accommodate ligation during orthodontic treatment. These design characteristics allow a tensioned ligating wire to move the brackets, which are securely bonded to the teeth, along a pre-determined path until the desired tooth position is reached. Elastic ligatures may be used on the tie wings and hooks to further facilitate tooth movement, and to secure the orthodontic wire into the bracket's archwire slot. #### Intended Use: FastBraces® Ceramic Brackets are intended for orthodontic movement of natural teeth. #### Technological Characteristics: The design, material, and intended use of FastBraces® Ceramic Brackets are similar to the predicate device. Regarding design, FastBraces® ceramic brackets and their predicate both incorporate specific torques, anqulations, and distal offset dimensions, along with archwire slots that are designed to accommodate the correct size archwire (typically .018" or .022" thick). They both feature tie wing undercut spaces for orthodontic ligatures, have a molded ceramic bracket bodies with rounded corners and edges for patient comfort, and rounded hooks on the distal-gingival tie wings to accommodate ligation during orthodontic treatment. Regarding materials, FastBraces® ceramic brackets and its predicate are manufactured from polycrystalline alumina (ceramic) material, which is of known biocompatibility in the oral environment. Regarding intended use, the design characteristics of FastBraces® ceramic brackets and its predicate allow a tensioned ligating wire to move the brackets, which are securely bonded to the teeth, along a pre-determined path until the desired tooth position is reached. Elastic ligatures may be used on the tie wings and hooks to further facilitate tooth movement, and to secure the orthodontic wire into the bracket's archwire slot. 1370 Decision Street, Suite D Vista, CA 92081 Phone (760) 734-3995 Fax (760) 734-1735 USA Toll Free 1-800-383-5301 www.odpinc.com Page 2 of 2 {3}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white logo with the letters "odp" inside of a circle. The letters are white, and the circle is black. There is a semi-circle around the top and left side of the circle. The logo is simple and modern. ### Substantial Equivalence: FastBraces® Ceramic Brackets are safe and effective for their intended use in orthodontic treatment, and perform at least as well as the predicate device listed above. FastBraces® ceramic brackets are engineered to be substantially equivalent to the predicate with respect to intended use, technological characteristics, device design, materials, performance, safety, effectiveness, and biocompatibility. Minor design changes are the only variations as compared to the predicate. There are no changes in the intended use or the fundamental scientific technology. #### Device Material: The FastBraces® Ceramic Bracket and its predicate are both made of ceramic tri polycrystalline alumina, which has a well-documented history of biocompatibility within the oral environment. ODP has prepared a biocompatibility summary report in accordance with ISO 10993-1:2009, which is located in Section 15 of this submission. Because the materials from which the devices are constructed are well established in the orthodontic industry, product testing in regards to biocompatibility was not performed. Instead, a literature review has been conducted in accordance with ISO 10993-1:2009, Annex C. #### Device Design: FastBraces® Ceramic Brackets and its predicate both have tie wing undercut spaces for orthodontic ligatures. They each have true-twin tie wings, i.e. four tiewings, for versatile use with auxiliaries. FastBraces® Ceramic Brackets and its predicate also contain base flanges for bracket placement and adhesive flash cleanup. FastBraces® Ceramic Brackets and its predicate contain a molded ceramic bracket body with rounded corners and edges, and a round hook on the distalgingival tiewings. Like its predicate, certain FastBraces® Ceramic brackets contain vertical slot and stress concentrators to facilitate debonding of the bracket from the tooth. #### Nonclinical Performance Testing: The nonclinical performance testing analysis shows that FastBraces® Ceramic Brackets perform comparably to the predicate device as follows: - Shear Test measures shear strength required to remove a bonded bracket from a 1. substrate when a force is applied in the occlusal direction. The test results showed that the bond strength of FastBraces® Ceramic Brackets are comparable to the predicate, and exceed the minimum bond strength required to affix the bracket to the tooth. - Wire Torque Test measures the torsional force required to break a bonded bracket when 2. a rectanqular wire is twisted in the wire slot. The test results showed comparable bracket strengths, with the Fast Braces® wire torque test averaging 3,473 gm Force, and the predicate device averaging 3,333 gm Force before breakage occurred. - ని. Wire Drag Test measures the force required to drag a ligated stainless steel wire through the slot of a bonded bracket. The test results showed lower forces were required to drag the Fast Braces® ceramic bracket along a ligated stainless steel wire than the predicate, indicating better sliding mechanics during treatment. - Bracket Removal Test evaluates the visual condition of a bonded bracket after twisting it ধ .. বাংলাদেশে বাংলাদেশ করে আমাকে বিশ্বকাপে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে পারে করে পারে পারে করে পারে ক off the substrate with pliers. The Fast Braces® bracket removal test resulted in less bracket fracturing as compared to the predicate device when removing the bracket from a substrate using pliers. 1370 Decision Street, Suite D Vista, CA 92081 Phone (760) 734-3995 Fax (760) 734-1735 USA Toll Free 1-800-383-5301 www.odpinc.com Page 3 of 3 {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo with the letters 'odp' inside of a circle. The letters 'dp' are in white font on a black background. The letter 'o' is in black font on a white background. The circle is surrounded by a gray ring. ### Clinical Performance Testing: No clinical performance testing was conducted on FastBraces® Ceramic Brackets. #### Conclusion: ODP's FastBraces® Ceramic Brackets are standard orthodontic appliances that are similar to those that have been legally marketed and used safely and effectively for many years in the clinical environment. FastBraces® ceramic brackets are engineered to be substantially equivalent to the predicate with respect to intended use, technological characteristics, device design, materials, performance, safety, effectiveness, and biocompatibility. There are no major differences between the FastBraces® Ceramic Brackets and the predicate device cited. Therefore, FastBraces® Ceramic Brackets raise no new issues of safety or effectiveness. FastBraces® Ceramic Brackets are designed and manufactured to industry-standard specifications using materials that have a well established and documented history of biocompatibility within the oral environment. As designed, FastBraces® Ceramic Brackets are as safe and effective as the predicate device, and are determined to be substantially equivalent to the referenced predicate device currently on the market. Taking these factors into account, it can be safely concluded that ODP's FastBraces® Ceramic Brackets are of low risk to the end user, are clinically safe, and perform at least as well as the predicate device referred to herein. 1370 Decision Street, Suite D Vista, CA 92081 Phone (760) 734-3995 Fax (760) 734-1735 USA Toll Free 1-800-383-5301 www.odpinc.com Page 4 of 4 {5}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 ### June 7, 2013 Mr. Richard Merrell Regulatory Affairs Manager Orthodontic Design and Production, Inc. . 1370 Decision Street, Suite D Vista, CA 92081 Re: K130446 Trade/Device Name: FastBraces® Ceramic Brackets Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NJM Dated: April 10, 2013 Received: April 15, 2013 Dear Mr. Merrell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ ### Page 2 - Mr. Merrell Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events} {21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Susan Runner DDS, PA Mary S. Runner-S Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows a logo with the letters 'odp' inside a circle. The letters are in a sans-serif font and are white against a black background. The circle is also black, with a white outline. The logo has a slightly distressed or textured appearance, as if it has been printed or stamped. Indications for Use (130446 510(k) Number: Device Name: FastBraces® Ceramic Brackets Indications for Use: FastBraces® Ceramic Brackets are indicated for orthodontic movement of natural teeth. Prescription Use_ X (Part 21 CFR 801 Subpart D) Over the Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR ### Concurrence of CDRH, Office of Device Evaluation (ODE) Andrew 1. Steen 2013.06.07 10:48:08 -04'00" (Division Sign-Off) (Division Sign-Off) Division of Anesthesiology, General Hospital Division of Anesthesiology, General Devices Division of Anesthesions of Anesthology Infection Control, Dental Devices 510(k) Number: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1370 Decision Street, Suite D Vista, CA 92081 Phone (760) 734-3995 Fax (760) 734-1735 USA Toll Free 1-800-383-5301 www.odpinc.com ... .. 5 6 2 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 5 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 3 6 5 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 ・・・ ﻭ ﺔ ﺍﻟ . ్రాల : --------------------------------------------------------------------------------------------------------------------------------------------------------------------------- ----- ﻤﻴﺔ ﺳﺎ Carles Carders 14 15 Page 1 of 1 .. .
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...