← Product Code [NJM](/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NJM) · K080906

# CDB CLIP (K080906)

_Cdb Corporation · NJM · Oct 17, 2008 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NJM/K080906

## Device Facts

- **Applicant:** Cdb Corporation
- **Product Code:** [NJM](/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NJM.md)
- **Decision Date:** Oct 17, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The CDB Clip, commonly known as an orthodontic ligation clip, is an orthodontic bracket accessory. It attaches to the orthodontic bracket to hold the archwire firmly in the slot. Together they facilitate the orthodontic movement of teeth.

## Device Story

CDB Clip is a passive ligation clip; attaches to orthodontic brackets to retain archwire in slot; facilitates orthodontic tooth movement. Device used in clinical orthodontic settings by dental professionals. Material: Celcon MT 24U01 Acetal (POM).

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Passive ligation clip; material: Celcon MT 24U01 Acetal (POM) without fillers; mechanical retention of orthodontic archwire.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- NEO-CLIP ORTHODONTIC BRACKET ACCESSORY ([K042905](/device/K042905.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(K) SUMMARY

OCT 1 7 2008

| Submitter:                    | CDB Corporation<br>9201 Industrial Blvd.<br>Leland, NC 28451 USA<br>Phone: 910 383-6464<br>Fax: 910 383-6465<br>Contact: Jens Rumsfeld<br>Title: Mgr. Managing Director<br>e-mail: jrumsfeld@cdbcorp.net                                                                                        |                        |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Date Prepared:                | January 22, 2008                                                                                                                                                                                                                                                                                |                        |
| Name of device:               | The CDB Clip - Orthodontic Bracket Accessory                                                                                                                                                                                                                                                    |                        |
| Classification Name           | Bracket, Ceramic, Orthodontic                                                                                                                                                                                                                                                                   |                        |
| Device Classification         | Regulatory Class II<br>Product Code: NJM<br>Classification Panel: Dental<br>Regulation Number: 21 CFR 872.5470                                                                                                                                                                                  |                        |
| Predicate Device(s)           |                                                                                                                                                                                                                                                                                                 |                        |
| 510(k) Number                 | Device                                                                                                                                                                                                                                                                                          | Manufacturer           |
| K042905                       | NEO-CLIP ORTHODONTIC<br>BRACKET ACCESSORY                                                                                                                                                                                                                                                       | Dentsply International |
| Device Description            | The CDB Clip is a passive ligation clip which functions to retain the<br>orthodontic archwire in the slot of an orthodontic bracket. It is made of<br>Celcon MT 24U01 Acetal (POM) with no fillers.                                                                                             |                        |
| Indications                   | The CDB Clip, commonly known as an orthodontic ligation clip, is an<br>orthodontic bracket accessory. It attaches to the orthodontic bracket to hold the<br>archwire firmly in the slot. Together they facilitate the orthodontic movement of<br>teeth.                                         |                        |
| Technological Characteristics | The function and performance of the CDB Clip is identical to the predicate<br>device listed above. There is no difference in the fundamental technology.<br>The CDB Clip is made of the same material as the predicate device currently<br>on the market and has the same overall intended use. |                        |

{1}------------------------------------------------

#### Conclusion By definition, a device is substantially equivalent when the device has the same intended use and the same technological characteristics as the predicate device.

There are no major differences between the CDB Clip and the predicate device cited, therefore, the CDB Clip does not raise any questions regarding the safety and effectiveness.

The CDB Clip, as designed, is as safe and effective as the predicate device and the device is determined to be substantially equivalent to the referenced predicate device currently on the market.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with a more abstract and modern design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## OCT 1 7 2008

Mr. Jens Rumsfeld Managing Director CDB Corporation 9201 Industrial Boulevard Leland, North Carolina 28451

Re: K080906

Trade/Device Name: CDB Clip Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NJM Dated: September 12, 2008 Received: September 16, 2008

Dear Mr. Rumsfeld:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2 – Mr. Rumsfeld

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Th. Spraguelsted, m.D. for "

Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use

510(k) Number (if known): K080906

Device Name: CDB Clip

Indications For Use:

The CDB Clip, commonly known as an orthodontic ligating clip, is an orthodontic bracket accessory. It attaches to the orthodontic bracket to hold the archwire firmly in the slot. Together they facilitate the orthodontic movement of teeth.

Susan Punaes

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K080996

Prescription Use X (Part 21 CFR 80'1 Subpart D) ANDIOR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (OD E)

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NJM/K080906](https://fda.innolitics.com/submissions/DE/subpart-f%E2%80%94therapeutic-devices/NJM/K080906)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
