Anti Snoring Mouthpiece (L08-BP, L08-PT, L08-BT, L08-TP, L08-TB, L08-PB)

K253845 · Ram.Shaw Pte. , Ltd. · LRK · Mar 3, 2026 · Dental

Device Facts

Record IDK253845
Device NameAnti Snoring Mouthpiece (L08-BP, L08-PT, L08-BT, L08-TP, L08-TB, L08-PB)
ApplicantRam.Shaw Pte. , Ltd.
Product CodeLRK · Dental
Decision DateMar 3, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Anti Snoring Mouthpiece is designed to advance the user's lower jaw and maintain this position during use, opening the airway to reduce snoring during sleep. It is intended for adults with at least 18 years old.

Device Story

Intraoral mandibular repositioning device; used at home by adults to reduce snoring. Device consists of upper and lower tray assemblies; user selects and combines trays to achieve 0-6 mm mandibular advancement. Mechanism increases pharyngeal space to improve air exchange and decrease turbulence. Device is heat-sensitive and moldable for custom fit. Output is physical jaw positioning; affects clinical decision-making by providing non-invasive snoring management. Benefits patient by reducing snoring-related airway obstruction.

Clinical Evidence

No clinical data. Bench testing only. Biocompatibility testing (cytotoxicity, irritation, sensitization) performed per ISO 10993-1. Physical property testing performed per ASTM D790-17, ASTM D638-22, ASTM D570-22, ASTM D3418-21, and ASTM D792-20.

Technological Characteristics

Mandibular repositioning device; materials: Polypropylene and Ethylene vinyl acetate copolymer. Custom-fitted, heat-sensitive, moldable. Dimensions: Large and Standard sizes. Mechanical advancement: 0-6 mm. Non-sterile. Biocompatibility per ISO 10993. Physical standards: ASTM D790-17 (flexural), ASTM D638-22 (tensile), ASTM D570-22 (water absorption), ASTM D3418-21 (melting), ASTM D792-20 (density).

Indications for Use

Indicated for adults 18+ years old to reduce snoring during sleep by advancing the lower jaw to open the airway.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 3, 2026 Ram.Shaw Pte., Ltd. Emma Chen Applicant/Sponsor 112 Robinson Rd. # 03-01 Robinson 112, Singapore Singapore, 068902 SINGAPORE Re: K253845 Trade/Device Name: Anti Snoring Mouthpiece (L08-BP, L08-PT, L08-BT, L08-TP, L08-TB, L08-PB) Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: December 2, 2025 Received: December 2, 2025 Dear Emma Chen: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253845 - Emma Chen Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253845 - Emma Chen Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253845 | ☑ | | Please provide the device trade name(s). | | ☑ | | Anti Snoring Mouthpiece (L08-BP, L08-PT, L08-BT, L08-TP, L08-TB, L08-PB) | | | | Please provide your Indications for Use below. | | ☑ | | The Anti Snoring Mouthpiece is designed to advance the user's lower jaw and maintain this position during use, opening the airway to reduce snoring during sleep. It is intended for adults with at least 18 years old. | | | | Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ☑ | {4} 510(k) Summary # 510(k) Summary - K253845 This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92. ## Type of submission Traditional ## Date prepared 2025/11/27 ## Submitter's Information Company: RAM.SHAW PTE.LTD Address: 112 ROBINSON ROAD #03-01 ROBINSON 112 SINGAPORE (068902) Submitter: Emma Chen Telephone: +65-62262377 Email: compliance@complianceglobe.com ## Device Information Device name: Anti Snoring Mouthpiece Model: L08-BP, L08-PT, L08-BT, L08-TP, L08-TB, L08-PB Size: Large (L) and Standard (S) Production code: LRK Regulation number: 21 CFR 872.5570 Classification: Class II ## Predicate Device Information 510(k) Number: K170825 Manufacturer: Apnea Sciences Corporation Device Name: SnoreRx Device Class: Class II Product Code: LRK Regulation Number: 21 CFR 872.5570 {5} 510(k) Summary ## Indications for use The Anti Snoring Mouthpiece is designed to advance the user's lower jaw and maintain this position during use, opening the airway to reduce snoring during sleep. It is intended for adults with at least 18 years old. ## Device Description The L08 series Anti Snoring Mouthpiece is an intraoral device that is used to maintain the lower jaw in a forward position to increase pharyngeal space so as to improve the ability to exchange air and decrease air turbulence, a causative factor in snoring. L08 series Anti Snoring Mouthpiece includes three upper trays (A0, A2, A4) and three lower trays (B0, B1, B2). One upper tray and one lower tray are combined together for use. By randomly combining one upper tray with one lower tray, the device can be used for advancing lower jaw from 0mm to 6mm. Considering that the size of the oral cavity varies among different groups of people, L08 series Anti Snoring Mouthpiece is designed to be available for two size (L and S). Even though the dimensions of these sizes of Anti Snoring Mouthpiece are different, by positioning the upper protrusion and the lower protrusion at different locations within the upper tray assembly and the lower tray assembly respectively, advancement distance of 0-6 mm can still be realized. 2 / 5 {6} 510(k) Summary # Substantial Equivalence Discussion The table below is the comparison of subject device and predicate device: Table 1 Substantial equivalence comparison | Item | Subject Device | Predicate Device | Remark | | --- | --- | --- | --- | | Device Name | Anti Snoring Mouthpiece | SnoreRx | - | | Model | L08-BP, L08-PT, L08-BT, L08-TP, L08-TB, L08-PB | / | - | | Product Code | LRK | LRK | - | | Regulation No. | 21 CFR 872.5570 | 21 CFR 872.5570 | - | | Classification Name | Intraoral Device for Snoring and Obstructive Sleep Apnea | Intraoral Device for Snoring and Obstructive Sleep Apnea | - | | Technological Features | Mandibular repositioning device that advances the lower jaw to increase pharyngeal space. | Mandibular repositioning device that advances the lower jaw to increase pharyngeal space. | Same | | Intended Use | Minimize air turbulence that causes snoring. Device is intended for Over The-Counter use. | Minimize air turbulence that causes snoring. Device is intended for Over The-Counter use. | Same | | Material | -Polypropylene -Ethylene vinyl acetate copolymer | -Polycarbonate resin -Ethylene vinyl acetate copolymer | Similar The hard support of predicate device is made of Polycarbonate, while the support of subject device is made of Polypropylene. Both material are common plastic. | {7} 510(k) Summary | | | | This difference won’t raise any safety and effectiveness issue. | | --- | --- | --- | --- | | Desirable Characteristics | Home use, heat sensitive / moldable, adjustable jaw advancement position | Home use, heat sensitive / moldable, adjustable jaw advancement position | Same | | Specifications: - Physical: - Mechanical: - Single use | -Custom-fitted intraoral device -Repositions mandible anteriorly up to 6 mm -Reusable | -Custom-fitted intraoral device -Repositions mandible anteriorly up to 6 mm -Reusable | Same | | Sterility | Non-sterile | Non-sterile | Same | | Biocompatibility | ISO 10993: Cytotoxicity Irritation Sensitization | ISO 10993: Cytotoxicity Irritation Sensitization | Same | 4 / 5 {8} 510(k) Summary # Performance Testing - Clinical Not Applicable # Performance Testing - Animal Not Applicable # Non-clinical testing Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: Biocompatibility test The material were test for cytotoxicity, irritation, and sensitization. Bio-compatibility assessment was conducted in accordance with ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Physical properties ASTM D790-17 for flexural strength and flexural modulus. ASTM D638-22 for Tensile Strength and Elongation at break. ASTM D570-22 for water absorption. ASTM D3418-21 for melting temperature. ASTMD 792-20 for Density. # Conclusion Based on the comparison of the indications for use, technical characteristics and performance, we find that the L08 series Anti Snoring Mouthpiece is substantially equivalent to the predicate SnoreRx with 510(K) number K170285.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...