K232025 · Good Sleep CO Pte , Ltd. · LRK · May 9, 2024 · Dental
Device Facts
Record ID
K232025
Device Name
Hushd Pro Avera
Applicant
Good Sleep CO Pte , Ltd.
Product Code
LRK · Dental
Decision Date
May 9, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hushd Pro Avera device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Device Story
Removable, patient-specific intraoral device; treats snoring and mild to moderate OSA. Consists of custom-made maxillary and mandibular splints; fits over teeth. Functions as mandibular repositioner; holds mandible forward during sleep; prevents tongue and soft tissue collapse into airway. CAD/CAM designed and manufactured; sequential devices provide gentle mandibular advancement based on dentist/physician prescription. Maximum protrusion 10mm. No mechanical adjustment mechanisms (pistons, straps, elastics); advancement achieved by replacing current splints with next in series. Used in home setting; patient-operated. Healthcare provider uses output to manage airway patency; benefits patient by reducing OSA symptoms and snoring.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including mechanical property evaluation (flexural modulus, flexural strength, impact strength) and biocompatibility testing (cytotoxicity, irritation, sensitization, systemic toxicity, pyrogenicity) per ISO 10993-1.
Technological Characteristics
Custom-made intraoral splints; material: polyamide type 12 polymer. CAD/CAM manufacturing; selective laser sintering. Non-sterile. Mandibular repositioning via twin-mated posts. Max advancement 10mm. No active energy source. Standalone device.
Indications for Use
Indicated for adults with snoring and mild to moderate obstructive sleep apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 9, 2024
Good Sleep Co Pte Ltd. % Na Zhang Project Manager evo820, LLC 1 Bay Street Rancho Mission Viejo, California 92694
Re: K232025
Trade/Device Name: Hushd Pro Avera Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea Regulatory Class: Class II Product Code: LRK Dated: June 27, 2023 Received: July 7, 2023
Dear Na Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232025
Device Name Hushd Pro Avera
Indications for Use (Describe)
The Hushd Pro Avera device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K232025 510(k) Summary
# SUBMITTER
| Date Prepared: | May 7, 2023 |
|-------------------|------------------------------------------------------------------------------|
| Submitter: | Good Sleep Co Pte Ltd.<br>226 Ocean Drive<br>Singapore, 098617<br>Singapore |
| Official Contact: | Joel Simpson<br>Director<br>Tel:+658752941<br>Email: joel@goodsleepco.health |
# DEVICE
| Trade/Proprietary Name: | Hushd Pro Avera |
|-----------------------------|---------------------------------------------------------------------------------|
| Classification Name : | Device Anti-Snoring |
| Common Name : | Intraoral device for snoring and<br>mild to moderate obstructive<br>sleep apnea |
| Classification Regulations: | 21CFR 872.5570 |
| Product Code: | LRK |
| Device Classification: | Class II |
| Classification Panel: | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
# PRIDICATE DEVICE
| Predicate Device: | K143244 | The Panthera Anti-Snoring Device | Panthera Dental, Inc. |
|-------------------|---------|---------------------------------------|--------------------------|
| Reference Device: | K133683 | MicrO2 Obstructive Sleep Apnea Device | MicroDental laboratories |
| | K220330 | Soundly Mandibular Advancement Device | Greystone IP Ltd |
| | K171576 | Panthera Anti-Snoring X3 Device | Panthera Dental, Inc. |
{5}------------------------------------------------
# DEVICE DESCRIPTION
The Hushd Pro Avera device series are a set of removable intraoral patient specific device used for treating snoring and mild to moderate obstructive sleep apnea. The device consists of number of custom made maxillary and mandibular splints fit over the upper and lower teeth. when interfaced together, functions as a mandibular repositioner holding the mandible forward during sleep thus preventing the togue and soft tissues of the throat from collapsing into the airway.
The Hushd Pro Avera device series are patient-specific devices, consisting of one or more maxillary device(s) and one or more mandibular device(s) that mates together. The sequential devices are CAD/CAM designed and manufactured according to the requested mandibular advancements position prescribed by the dentist or physician to provide a selection of gentle advancements according to patient comfort and need. As such, prescribed advancements can be achieved by simply removing current maxillary or mandibular device and inserting the next maxillary or mandibular device in the advancements series. The maximum protrusion of the device is 10mm.The Hushd Pro Avera device does not have any adjustment mechanisms to modify or maintain the mandibular position such as pistons, straps or repositioning elastics.
# INDICATIONS FOR USE
The Hushd Pro Avera device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
### COMPARISON OF TECHNOLOGICAL WITH THE PREDICATE DEVICE
The subject device is substantial equivalent in intended use and technological characteristics to the predicate devices shown above. Below is a summary table comparing the subject device with the primary predicate and reference devices.
| Features | Submission<br>Device | Predicate<br>Device | Reference<br>Device | Reference<br>Device | Reference<br>Device |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacture | Good Sleep Co<br>Pte Ltd. | Panthera Dental,<br>Inc. | MicroDental<br>laboratories | Greystone IP Ltd | Panthera Dental Inc. |
| Trade Name | Hushd Pro Avera | The Panthera<br>Anti-Snoring<br>Device | MicrO2<br>Obstructive Sleep<br>Apnea Device | Soundly<br>Mandibular<br>Advancement<br>Device | Panthera Anti-<br>Snoring X3 Device |
| 510(k) Number | K232025 | K143244 | K133683 | K220330 | K171576 |
| Regulation<br>Number | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 |
| Classifications | Class II | Class II | Class II | Class II | Class II |
| Product Code | LRK | LRK | LRK | LRK, LQZ | LRK |
| Indications for<br>Use | The Hushd Pro<br>Avera device is | The Panthera<br>Anti-Snoring | The<br>MICRODENTAL | The Soundly<br>Mandibular | The Panthera Anti-<br>Snoring X3 Device |
| | intended to reduce<br>or alleviate<br>snoring and mild<br>to moderate<br>obstructive sleep<br>apnea (OSA) in<br>adults. | device is intended<br>to reduce or<br>alleviate snoring<br>and mild to<br>moderate<br>obstructive sleep<br>apnea (OSA) in<br>adults. | , Inc, micr02<br>device is intended<br>to reduce night<br>time snoring and<br>mild to moderate<br>obstructive sleep<br>apnea (OSA) in<br>adults. | Advancement<br>Device is intended<br>to reduce or<br>alleviate snoring<br>and mild to<br>moderate<br>Obstructive Sleep<br>Apnea(OSA)<br>whilst sleeping in<br>adults. | is intended to<br>reduce or alleviate<br>snoring and mild to<br>moderate<br>obstructive sleep<br>Apnea(OSA) in<br>adults. |
| Principle of<br>Operation | Mandibular<br>Repositioners,<br>Repositions the<br>lower jaw forward<br>to increase the<br>patency of the | Mandibular<br>Repositioners,<br>Repositions the<br>lower jaw forward<br>to increase the<br>patency of the | Mandibular<br>Repositioners,<br>Repositions the<br>lower jaw forward<br>to increase the<br>patency of the | Mandibular<br>Repositioners,<br>Repositions the<br>lower jaw forward<br>to increase the<br>patency of the | Mandibular<br>Repositioners,<br>Repositions the<br>lower jaw forward<br>to increase the<br>patency of the |
| Adjustment<br>mechanism | airway<br>Advancements<br>can be achieved<br>by simply<br>removing the<br>current upper or<br>lower device and<br>inserting the next<br>upper or lower<br>device in the<br>mandibular<br>advancing series. | airway<br>Adjusted via the<br>use of interlocking<br>rods placed on<br>each side of the<br>mandible splint.<br>The shorter the<br>rod, the further<br>mandible is<br>advanced. The<br>dentist can select a<br>shorter connecting<br>rod until optimal<br>advancement is<br>achieved | airway<br>Advancements<br>can be achieved<br>by simply<br>removing the<br>current upper or<br>lower device and<br>inserting the next<br>upper or lower<br>device in the<br>mandibular<br>advancing series. | airway<br>User choose<br>which of the 3<br>lower splints to<br>wear-self titration | airway<br>Adjusted via the<br>adjustable clip<br>assembly placed on<br>each side of the<br>maxillary splint.<br>The longer the stop-<br>clip is, the further<br>the mandible is<br>advanced. The<br>dentist can select a<br>longer stop-clip<br>until optimal<br>advancement is<br>achieved. |
| Mandibular<br>advancement<br>range | Up to 10mm<br>50-90% of the<br>patient's<br>maximum<br>comfortable<br>mandibular<br>advancement at<br>1mm increment | Up to 15mm at<br>1mm increment | Up to 6.0mm | 3 positions at<br>40%,60% & 70%<br>of the patient's<br>maximum<br>comfortable<br>mandibular<br>advancement<br>Maximum<br>advancement is<br>12mm | Up to 5mm at 1mm<br>increments |
| Material | Made from<br>polymers<br>(polyamide type<br>12), Supplied by<br>EOS | Made from<br>polymers<br>(polyamide type<br>12), Supplied by<br>EOS | Hard PMMA<br>material | Form labs Dental<br>LT Clear V2 | Made from<br>polymers(polyamide<br>type12), supplied by<br>EOS. |
| Biocompatible | Yes | Yes | Yes | Yes | Yes |
| Rx or OTC | Rx | Rx | Rx | Rx | Rx |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Non-sterile |
| Target<br>Population | Adults | Adults | Adults | Adults | Adults |
| Design | Two customized<br>intra oral splints | Two customized<br>intra oral splints | Two customized<br>intra oral splints | Two customized<br>intra oral splints | Two customized<br>splints that fit |
| | that fit separately<br>over the upper and<br>lower teeth. | that fit separately<br>over the upper and<br>lower teeth. | that fit separately<br>over the upper and<br>lower teeth. | that fit separately<br>over the upper and<br>lower teeth. | separately over the<br>upper and lower<br>teeth inside the<br>mouth. |
| | These separate<br>maxillary and<br>mandible splints<br>are designed with<br>twin-mated posts,<br>when interface<br>together,<br>functioning as<br>mandible<br>repositioner<br>holding the<br>mandible forward<br>during sleep | The mandible<br>splint contains a<br>triangular<br>protrusion<br>allowing the<br>splints to engage<br>by means of<br>interlocking rods<br>on the sides. | These separate<br>maxillary and<br>mandible splints<br>are designed with<br>twin-mated posts,<br>when interface<br>together,<br>functioning as<br>mandible<br>repositioner<br>holding the<br>mandible forward<br>during sleep | The mandibular<br>advancement is<br>achieved by the<br>interlocking of<br>opposing wedge<br>faces on the upper<br>and lower jaw<br>position, There are<br>no locking side<br>arms, so the<br>splints can be<br>disengaged by a<br>small forward and<br>downward<br>movement of<br>lower jaw. The<br>upper splint is<br>supplied with 3<br>sizes of<br>mandibular<br>advancement, tied<br>to the patient's<br>maximum<br>mandibular<br>advancement. | The mandible splint<br>contains wing<br>protrusions that<br>interface with the<br>incline blocks built<br>into the buccal of<br>the maxillary splint. |
| Manufacture<br>Technology | Use the computer<br>-aided design<br>(CAD) and<br>Computer-aided<br>manufacturing(CA<br>M)<br>Use CAD that<br>enables a high<br>degree of<br>customization<br>according to the<br>physician or<br>dentists<br>prescription to<br>accommodate the<br>complex dental<br>anatomy of<br>individual<br>patients. The<br>CAM and<br>selective laser<br>sintering<br>guarantees<br>precision,<br>accuracy and | Use the computer<br>-aided design<br>(CAD) and<br>Computer-aided<br>manufacturing(CA<br>M)<br>Use CAD that<br>enables a high<br>degree of<br>customization<br>according to the<br>physician or<br>dentists<br>prescription to<br>accommodate the<br>complex dental<br>anatomy of<br>individual<br>patients. The<br>CAM and<br>selective laser<br>sintering<br>guarantees<br>precision,<br>accuracy and | Use the computer<br>-aided design<br>(CAD) and<br>Computer-aided<br>manufacturing(CA<br>M) | Use the computer<br>-aided design<br>(CAD) and<br>Computer-aided<br>manufacturing(CA<br>M) | Uses computer-<br>aided(CAD) and<br>computer-aided<br>manufacturing(CA<br>M).<br>Uses CAD that<br>enables a high<br>degree of<br>customization<br>according to the<br>physician or dentist<br>prescription. The<br>CAM and selective<br>laser sintering<br>guarantee precision,<br>accuracy and<br>consistency of each<br>patient. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| | consistency for | consistency for | | |
|--|-----------------|-----------------|--|--|
| | each patient | each patient | | |
Compared to the predicate device , the Hushd Pro Avera has the same intended use/Indications for use , principle of operation , fabricated by using the same material and a similar manufacturing technology except its adjustment mechanism , however the same adjustment mechanism was adopted by the reference device. The slight difference in twin-mate posts design between the subject device and reference device does not affect the equivalence in function and intended use of the devices. Therefore, the subject device is considered to be substantially equivalent to the predicate device and/or the reference device based on a comparison of intended use and technological characteristics.
### Non-Clinical PERFORMANCE DATA
Non-clinical testing have been conducted to verify that the Hushd Pro Avera meets all design specification which supports the conclusion that it's substantially equivalent (SE) to the predicate device.
A risk analysis on Hushd Pro Avera in accordance with ISO 14971:2007 and FDA guidance document "Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea" was conducted. All identified risks have been addressed through device design, biocompatibility and bench testing or through labeling provided to the consumer.
A risk assessment included assessment of the mechanical properties of the material of construction to evaluate its ability to achieve its intended use. The mechanical properties testing of the material of construction were performed by the manufacturer for :
- 트 Flexural Modulus
- Flexural Strength
- 트 Impact Strength
Biocompatibility testing for the final finished device was conducted in accordance with International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The final finished device was tested for the following biocompatibility parameters:
- 트 Cytotoxicity
- Irritation
- 트 Sensitization
- 트 System Toxicity
- 트 Pyrogen
The subject device is able to achieve its intended use based on the same materials of construction, similar manufactory process and the identical application to the predicate device. There are no new questions of safety or efficacy raised by the Hushd Pro Avera devices compared to the predicate device.
{9}------------------------------------------------
## CLINICAL PERFORMANCE DATA
The clinical performance of the subject device is deemed not necessary . The Hushd Pro Avera device has equivalent indication and method of use to its predicated and reference devices, therefore there was no clinical testing to support this device.
No clinical study was performed in association with this submission.
## CONCLUSION
The subject device has very similar technological characteristics (i.e. design, function, principle of operation, materials, biocompatibility and sterilization) to the predicate device . The intended use and indications for use of the subject device are the same as the predicate device.
In conclusion, the device is substantially equivalent based on a comparison of intended use, and technological characteristics, the device is safe and effective for its intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.