ApnoDent Appliance

K181123 · Apnomed, Inc. · LRK · Nov 21, 2018 · Dental

Device Facts

Record IDK181123
Device NameApnoDent Appliance
ApplicantApnomed, Inc.
Product CodeLRK · Dental
Decision DateNov 21, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ApnoDent® appliance is intended to reduce nighttime snoring and/or mild to moderate obstructive sleep apnea (OSA) in adult patients 18 years of age or older.

Device Story

ApnoDent is a customized, patient-specific oral appliance used to treat snoring and mild to moderate OSA. The device consists of upper and lower interlocking trays that guide the mandible forward during sleep, preventing tongue and soft tissue collapse into the airway. It is prescribed by a dentist and manufactured based on patient dental models. The device features adjustable screws on both sides that allow for incremental mandibular advancement to optimize patient comfort and therapeutic efficacy. The design maximizes tongue space and allows for jaw movement (opening/closing) during wear. It is intended for use by adults, including those with CPAP intolerance.

Clinical Evidence

Bench testing only. Evidence includes a biocompatibility and physical properties assessment based on material composition of the primary predicate, and a risk assessment comparing the subject device to the predicates.

Technological Characteristics

Customized oral appliance consisting of upper and lower interlocking trays. Materials include industry-standard dental acrylic and stainless steel orthodontic wires/adjustment screws. Mandibular advancement is achieved via screw adjustment (6-8 mm range). Non-sterile, prescription-only device.

Indications for Use

Indicated for adult patients 18+ with nighttime snoring and/or mild to moderate obstructive sleep apnea (OSA).

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. November 24, 2018 ApnoMed, Inc. % Colette Cozean Regulatory Consultant Dr. Colette Cozean, PhD 21581 Midcrest Drive Lake Forest, California 92630 Re: K181123 Trade/Device Name: ApnoDent Appliance Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea Regulatory Class: Class II Product Code: LRK, LOZ Dated: October 18, 2018 Received: October 19, 2018 Dear Colette Cozean: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Digitally signed by Mar Mary S. S Runner - 53 Runner - S3 - Date: 2018.11.21 11:29:46 -05'00 For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181123 Device Name ApnoDent Indications for Use (Describe) The ApnoDent® appliance is intended to reduce nighttime snoring and/or mild to moderate obstructive sleep apnea (OSA) in adult patients 18 years of age or older. Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 5. 510(k) Summary | Applicant: | ApnoMed, Inc.<br>1515 116th Ave NE. Suite 105C<br>Bellevue WA, 98004<br>(425) 443-4020 | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Colette Cozean, PhD<br>21581 Midcrest Drive<br>Lake Forest, CA 92630<br>(949) 855-2885<br>colettecozean@gmail.com | | Date Prepared | August 31, 2018 | | Proprietary Name | ApnoDent® Appliance | | Common Name | Device, Snoring and Mild to Moderate Sleep Apnea | | Classification Name | Intraoral devices for snoring; intraoral devices for snoring and obstructive sleep apnea<br>(Class II, 21 CFR872.5570, Product Code LRK, LQZ) | | Primary Predicate Device | mRNA (K130067) | | Reference Predicate Devices | DynaFlex (K103076), OASYS-Oral Airway (K030440), Apinator-Oral Airway (K160123) | Description: The ApnoDent® appliance is a customized oral device featuring both lower and upper trays and interlocking system. The product is non-sterile, biocompatible, and provided in a sealed box with instructions for use. The ApnoDent® appliance reduces snoring and mild to moderate sleep apnea by guiding the mandible forward during sleep, preventing the tongue and soft tissues of the throat from collapsing into the airway. Designed as a patient-specific device, the ApnoDent® series are manufactured to the dentist prescriber's requested advancement positions to provide a selection of gentle adjustments according to patient comfort and need. As such, prescribed advancements can be achieved by simply turning the advancing screws on both sides which are in contact with the inside of the lower teeth and guide the lower jaw forward. The ApnoDent® design vields a small and a comfortable patient-specific mandibular advancement device. The design of the device maximizes tongue space and mandibular movement resulting in the ability to open and close during wear for patient comfort. Indications for Use: The ApnoDent® appliance is intended to reduce nighttime snoring and/or mild to moderate obstructive sleep apnea (OSA) in adult patients 18 years of age or older. Technological Characteristics: The ApnoDent® appliance consists of upper and lower interlocking. customized trays. The ApnoDent® appliance is customized on models of the patient's teeth, using standard orthodontic acrylics and standard orthodontic wires for clasps and retention. The ApnoDent® appliance allows for inter locking of the upper and lower trays to adjust the mandibular position of the user. The technical characteristics are identical to the predicate devices. All predicate devices (mRNA, DynaFlex, OASYS, Apinator-Oral Airway) also function as mandibular advancement devices to increase the patient's pharyngeal space and improve the ability to exchange air during {4}------------------------------------------------ sleep. They each have customized upper and lower trays that interlock to advance the mandible. In all technological characteristics the subject device is substantially equivalent to the predicate devices. | | Subject Device | Primary<br>Predicate | Reference<br>Predicate #1 | Reference<br>Predicate #2 | Reference<br>Predicate #3 | |---------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------| | | ApnoDent® | mRNA Appliance | DynaFlex Anti-<br>Snoring and<br>Sleep Apnea<br>Devices | Apinator-Oral<br>Airway System | OASYS-Oral<br>Airway System | | 510(k) Number | K181123 | K130067 | KI03076 | K160123 | K030440 | | Indication for<br>Use | Intended to<br>reduce night<br>time snoring and<br>mild to<br>moderate<br>obstructive sleep<br>apnea (OSA) in<br>adults. | Intended to<br>reduce night<br>time snoring and<br>mild to<br>moderate<br>obstructive sleep<br>apnea (OSA) in<br>adults. | Intended to<br>reduce night<br>time snoring and<br>mild to<br>moderate<br>obstructive sleep<br>apnea (OSA) in<br>adults. | Intended for use<br>on adult patients<br>18 years of age<br>and older as an<br>aid for the<br>reduction and/or<br>alleviation of<br>snoring and mild<br>to moderate<br>obstructive sleep<br>apnea | Intended to<br>reduce night<br>time snoring and<br>mild to<br>moderate<br>obstructive sleep<br>apnea (OSA) in<br>adults. | | Classification<br>Name | Device, Anti-<br>Snoring and<br>Obstructive<br>Sleep Apnea | Device, Anti-<br>Snoring and<br>Obstructive<br>Sleep Apnea | Device, Anti-<br>Snoring and<br>Obstructive<br>Sleep Apnea | Device, Anti-<br>Snoring and<br>Obstructive<br>Sleep Apnea | Device, Anti-<br>Snoring and<br>Obstructive<br>Sleep Apnea | | Class | II | II | II | II | II | | Product Code | LRK, LQZ | LRK, LQZ | LRK, LQZ | LRK, LQZ | LRK, LQZ | | May be used in<br>СРАР<br>Intolerance | yes | yes | yes | yes | yes | | Area of Use | mouth | mouth | mouth | mouth | mouth | ## Technological Characteristics {5}------------------------------------------------ | Construction | Industry-<br>standard dental<br>acrylic, stainless<br>steel orthodontic<br>wires,<br>orthodontic<br>adjustment<br>screws | Industry-<br>standard dental<br>acrylic, stainless<br>steel orthodontic<br>wires,<br>orthodontic<br>adjustment<br>screws | Industry-<br>standard dental<br>acrylic, stainless<br>steel orthodontic<br>wires,<br>orthodontic<br>adjustment<br>screws | Industry-<br>standard dental<br>acrylic, stainless<br>steel orthodontic<br>wires,<br>orthodontic<br>adjustment<br>screws | Industry-<br>standard dental<br>acrylic, stainless<br>steel orthodontic<br>wires,<br>orthodontic<br>adjustment<br>screws | |---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------| | Fabrication | Customized | Customized | Customized | Customized | Customized | | Mechanism of<br>Action | Advancing the<br>mandible | Advancing the<br>mandible | Advancing the<br>mandible | Advancing the<br>mandible | Advancing the<br>mandible | | Method of<br>positioning of<br>the mandible | Screw<br>adjustment | Screw<br>adjustment | Screw<br>adjustment | Screw<br>adjustment | Screw<br>adjustment | | Amount of<br>mandible<br>protrusion | 6-8 mm | 6-8 mm | 6-8 mm | 6-8 mm | 6-8 mm | | Prescription only | yes | yes | yes | yes | yes | Clinical and Non-Clinical Data: A biocompatibility and physical properties assessment was completed based on the material composition of the primary predicate, which concluded that the subject device was substantially equivalent to the primary predicate device. A risk assessment has also been conducted with the subject device, which concluded there are no additional risks as compared to the predicate device(s). Summary: Based on the intended use, technical characteristics, and other data provided in this submission, the ApnoDent® appliance demonstrates substantial equivalence to the predicate devices in both safety and
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...