K120372 · Always More Marketing, Inc. · LRK · Jun 27, 2012 · Dental
Device Facts
Record ID
K120372
Device Name
DR GREENBURGS HYBRID VACUUM
Applicant
Always More Marketing, Inc.
Product Code
LRK · Dental
Decision Date
Jun 27, 2012
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K120372 · Jun 27, 2012
DR GREENBURGS HYBRID VACUUM
Always More Marketing, Inc.
Jonathan Greenburg's Sleep Study Table (clinical data); Published clinical literature/articles on oral appliance therapy
The sponsor provided clinical data from a sleep study table and referenced published literature to demonstrate that oral appliance therapy, including the use of tongue-retaining components, is effective in managing snoring and mild to moderate obstructive sleep apnea in patients.
Sleep apnea; Oral appliance therapy; Clinical study data; Literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Jonathan Greenburg's Sleep Study Table
Patients with sleep apnea
Not applicable for this study
Improvement in sleep apnea disease
Indications for Use
The Dr Greenburgs Hybrid Vacuum Snoring Device is indicated for persons 18 years or older, who wish to reduce the incidence of snoring and/or mild to moderate obstructive sleep apnea.
Device Story
Removable intraoral device; two-piece upper and lower arch tray system; includes elastic component connecting trays posteriorly. Principle: mandibular advancement and tongue retention; elastic impedes tongue from falling back into airway; trays posture lower jaw forward to increase airway passage opening. Used at home or sleep laboratories; patient-operated. Front opening allows emergency breathing. Non-allergenic, latex-free silicone elastic; copolyester trays. Output: physical repositioning of jaw and tongue. Benefit: reduction of snoring and breathing arrests due to obstructive sleep apnea.
Clinical Evidence
Clinical evidence provided via Jonathan Greenburg's Sleep Study Table illustrating patient improvement in sleep apnea disease. Literature review (SLEEP, Vol. 29, 2006) cited to support historical significance and efficacy of oral appliances for snoring and OSA management, noting they are well-tolerated and effective for mild to severe OSA.
Technological Characteristics
Two-piece intraoral tray system; copolyester trays; non-allergenic, latex-free silicone elastic component. Mandibular advancement and tongue retention principle. Dimensions/form factor: custom-fitted arch trays. No electronic components, software, or energy source.
Indications for Use
Indicated for persons 18 years or older to reduce snoring and/or mild to moderate obstructive sleep apnea.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
K120372
Always More Marketing Inc 4535 W. Sahara Ave, Ste. 200 Las Vegas, Nevada 89102 818 414 4817 igreenburg@earthlink.net
JUN 2 7 2012
# 510(k) SUMMARY
Summary Prepared on 2-3-2012
### Contact Person:
Jonathan Greenburg DDS President Always More Marketing Inc 4535 W. Sahara Ave, Ste. 200 Las Vegas, Nevada 89102 818 4144817 jgreenburg@earthlink.net
Name of Device: Dr Greenburgs Hybrid Vacuum
Common or Usual Name: Anti-Snoring and Apnea Device
Classification Name: Anti-Snoring Device (21CFR872.5570)
Product Code: LRK
## Predicate Devices:
Dr Greenburgs Hybrid (K111680) SomnoGuard (K061688) Respire Blue Series (K111207) Full Breath (K091035) Silent Nite (K972424) EMA (K971794)
{1}------------------------------------------------
Always More Marketing Inc 4535 W. Sahara Ave, Ste. 200 Las Vegas, Nevada 89102 818 414 4817 igreenburg@earthlink.net
### Description of the Device:
The Dr Greenburgs Hybrid Vacuum Anti-Snoring Device is a removable intraoral device for
replainle times in reports a routed is riyond vacum Anti-Snoring Device is a removable intraoral device for
repositioning the upper and lower jaw into a prescribed relationship for a single pat multiple times at home or sleep laboratories a
multiple times at home or sleep laboratories.
The Dr Greenburgs Hybrid Vacuum Device is a two piece, upper and lower arch tray system that
includes an additional elastic component which impedes the tongue from of the st includes an additional elastic component which impedes the tongue from falling back they system to
includes an additional elastic component which impedes the tongue from the airvay stem is designed to posture the lower journal from falling back into have in the ray system that into have in the back into he the intent to increase the airway passage opening, while several elimity back into the
the intent to increase the airway passage opening, while component adds in keeping the arreac the arway passage opening, while the additional elastic componentialistic componential
The combined effect of the advancement of the lower jaw, and the elast intending the tongue and the concert of the advancement of the lower jaw, and the elastic impeding the tongue
from falling back into the airway reduces snoring and breathing arrests due to obstr
Dr Greenburgs Hybrid Vacuum has a front area opening that is large enough for emergency
breathing.
Dr Greenburgs Hybrid Vacuum Snoring appliance consists of two independent trays (Top and
Bottom), and an elastic that is connected between the trave least of any and Bottom), and an elatic that is connected between the independent trays (Top and
Dotton), and an elatic that is connected between the tays located posterioriy. The trays cons of The Jan of State mail scoluted, copolyester that gives strength and studio and structures of the crays consisted on the clays of the class of the elastic consisted suppor The eastic onsists of a non-allergenic, atex free silion of Charles stress are a supe consistent and supply of the elastic of alse sonsellers tral a bays constructures from m molar and toneals of a lon-allergence to imped the silicone. The elage strenches from molar a
the arrosy. The flexibility of the elastic the elague from falling back into the comments at the fishing the elatic allows movement of the tongue, as as are lioning to the tonger, and more the ling lines more the ling lines and and on for. The friedioniny of the helps to prevent any tonguointention. The results is weller in swaller in swaller in swallering in swallering e and the includess slicone helps to prevent any tonges inting in swallowing
elastic is held firmly onto the bottom tray via securing buttons and is locked securely between
The top and bottom trays alignment are determined by the length connecting bands that are
fer and see the attached via buttons on the sides of the top and bottom trays. This system moves the fact a
forward to achieve the most the top and bottom trays. This system moves the jaw forward to achieve the most the sides of the top and be
{2}------------------------------------------------
### Always More Marketing Inc . 4535 W. Sahara Ave, Ste. 200 Las Vegas, Nevada 89102 818 414 4817 igreenburg@earthlink.net
Indications for Use: Dr Greenburgs Hybrid Vacuuniining
years or older, who wish to reduce the include of snoring Device in indicated for persons 18
apnear of the may of e would be for Use: Dr Greenburgs Hybrid Vacuum Anti-Snoring Device is indicated for persons 18
years or older, who wish to reduce the incidence of snoring and/or mild to mo
erate obstructive sleep
Technological Characteristics: Dr Greenburgs Hybrid Vacuum device has the following to
sane intended use. similarities to the previously cleared predicate devices:
Same intended use. Same intended use. Same operating principle.
Same technology.
technology.
Same
Same manufacturing process.
All of the predicates and Dr Greenburgs Hybrid Vacuum act as mandibular repositioners for the
treatment of snoring and mild to moderate Obstructive Sleep Appea and as and predicates and Dr Greenburgs Hybrid Vacuum act as mand.
Dr Greenburgs and mild to moderate Obstructive Sleep Apnea.
Dr Greenburgs Hybrid Vacuum, the Full Breath (K091035), and Dr Greenburgs Hybrid
(K111680) all have tongue retaining components. (K111680) all have tongue retaining components.
Dr Greenburgs Hybrid Vacuum, the Silent Nite (K972424) EMA (K971794) and Dr Greenburgs
g, 1 Hybrid (K111680) use straps/bands held onto button attachments.
Substantial Equivalence is based on non-clinical data.
In addition the 1:
In addition the literature supports the historical significance of oral devices that reposition the
m jaw and reduce and manage supports the historical significance.
Th
The Abstract Oral Appliances for Snoring and
Vol. 20, No. 1
The Abstract Oral Appliances for Snoring and Obstructive Steep Apnea: A Review from SLEEP,
Vol. 2, 2006 Page 259 states. .. "Oveructive Steep Apnea: A Review OS A l Vol. 2018 States Off Shoring and Obstructive Sleep Aprea: A Review from SLE
Vol. 2, 2006 Page 239 states... "Overall, those with mild to severe OSA have a 52%.
effective than - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - e flective than CPAP but may be better accepter asing an mina to severe OSA have a s 2020.
e ffective than CPAP but may be better accepter asmassus and be wholes where subjects used both teatments." Also, "They are well the student of the student of the student were word on word of while where
literature now provides evidence for the treame like and bour treatments." Also, "They are well tolerated by most patients" were world world world wos world world world wouldned world wouldned world model bedished
Downloa
Based on clinical data, it is demonstrated in a...
Based on clinical data, it is demonstrated in a variety of articles that looked at oral appliances
and their use for the treatment of snoring and sleep apnea, oral appliance and the similar data, it is demonstrated in a variety of articles that looked at oral appliances
and the treatment of snoring and sleep and sleep annel opliance therapy is a means as as as the treating and sleep of a ticles mat looked at oral appliances
means by which these conditions can be managed. This effectiveres is embraced from a variety of any by Miron these conditions can be managed.
of aspects including safety, convenience and cost.
Clinical data provided (Jonathan Greenburg's Sleep Study Table) illustrates patients can improve
their sleep apnea disease significantly if they use an intraoral anti-sporin the are and proviced (Johannan Oreenburg's Sleep Study Table) illustrates patients can improve
their sleep apnea disease significantly if they use an intraoral anti-snoring
{3}------------------------------------------------
Always More Marketing Inc 4535 W. Sahara Ave, Ste. 200 Las Vegas, Nevada 89102 818 414 4817 igreenburg@earthlink.net
A risk assessment concluded that there were no new safety concerns raised by the design of Dr Greenburgs Hybrid Vacuum.
In conclusion, a number of studies have shown improvement of the airway during sleep utilizing imaging associated with the use of oral appliances (also referred to as Oral Airway Dilators), aid in the management and reduction of spensand (also reprea. In addition the clinical data that was provided (Jonathan Greenburg's Sleep Study illustrates patients can improve their sleep apnea disease significantly if they use an intractly intestates patients can improve ther sizes and asease
s significantly if they use an intraoral anti-snoring device with a tongue retain
In summary, the device described in this submission is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an image of a stylized eagle with its wings spread. The eagle is facing to the right.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 2 7 2012
Jonathan Greenburg, DDS President Always More Marketing, Incorporated 4535 West Sahara Avenue, Suite 200 Las Vegas, Nevada 89102
Re: K120372
Trade/Device Name: Dr Greenburgs Hybrid Vacuum Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and
Obstructive Sleep Apnea.
Regulatory Class: II Product Code: LRK Dated: May 29, 2012 Received: June 5, 2012
Dear Dr. Greenburg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{5}------------------------------------------------
Page 2- Dr. Greenburg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
ph for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Klao372
## Indications for Use
510(k) Number (if known):
Device Name: Dr Greenburgs Hybrid Vacuum
Indications For Use:
The Dr Greenburgs Hybrid Vacuum Snoring Device is indicated for persons 18 years or older,
who wish to reduce the incidence of snoring and/or mild to moderate
# obstructive sleep apnea
Image /page/6/Picture/8 description: The image shows the words "Prescription Use (Part 21 CFR 801 Subpart D)" on the left side of the image. On the right side of the image, there is a box with a check mark inside of it. The box is outlined in black, and the check mark is also black. The check mark is large and bold.
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Luang
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of
510(k) Number:
) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.