DynaFlex® Anti-Snoring & Sleep Apnea Devices are intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults. The devices are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist, and is removable by the patient.
Device Story
Custom-fitted intraoral mandibular repositioning devices; used to treat snoring and mild-to-moderate OSA. Input: patient dental anatomy (custom-molded trays). Operation: trays fit over upper/lower teeth; engage via adjustable lugs or Herbst rod-and-tube mechanism to posture mandible forward; increases pharyngeal space to improve airflow. Output: physical mandibular advancement. Used in home setting; patient-operated (removable). Dentist/physician sets advancement level at fitting. Benefit: reduced snoring and improved airway patency during sleep.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and functional comparison to legally marketed predicate devices.
Indicated for adults with mild to moderate obstructive sleep apnea (OSA) and night time snoring. Prescription use only.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: This image shows a world map. The continents are outlined in black, and the landmasses are filled in with a light gray color. The map shows the general shape and location of the continents, including North and South America, Europe, Africa, Asia, and Australia. The map appears to be a simplified representation of the world, possibly for educational or illustrative purposes.
EMERGO GROUP
ノ07 ラ
FEB 2 3 2011
# SECTION 5 - 510(k) SUMMARY
### Submission Correspondent
Company Name: Company Address:
Company Contact:
Country: Phone: Fax: Website: Submission Date:
#### Submission Sponsor
Company Name: Company Address:
Country: Phone: Fax: Website: Emergo Group, Inc. 611 West 5th Street Third Floor Austin, TX 78701 Stuart R. Goldman Senior Consultant CQM, CQA, CBA, QMS-A 512.600.7616 USA 512.327.9997 512.327.9998 www.emergogroup.com September 29, 2010 -
DynaFlex 10403 International Plaza St. Ann, MO 63074 USA 314.426.4020 314.429.7575 www.dynaflex.com
### Device Classification
#### DynaFlex® Anti-Snoring & Sleep Apnea Devices
| Device Sponsor: | DynaFlex |
|-----------------------------------|---------------------------------------------------------------------------------------------------|
| CDRH Product Classification Name: | Device, Anti-Snoring |
| CDRH Product Code: | LRK |
| CDRH Regulation Name: | Intraoral devices for snoring and intraoral<br>devices for snoring and obstructive sleep<br>apnea |
| CDRH Regulation Number: | 872.5570 |
K103076
{2}------------------------------------------------
Image /page/1/Picture/0 description: This image shows a world map. The map depicts the continents and major landmasses of the world. The continents are shaded in a dark color, while the oceans are white. The map appears to be a simple representation of the world, without detailed geographical features.
K103076 EMERGOGROUP
2007
CDRH Classification Panel:
CDRH Regulatory Class:
Dental Devices
Class II
# Device Description
The proposed DynaFlex® Anti-Snoring & Sleep Apnea Devices are intraoral devices used for treating snoring and sleep apnea, and will consist of two custom fitted trays which fit over the upper and lower teeth and engage by means of adjustable lugs, or by the industry standard Herbst mechanism which is a rod and tube type assembly that orientates the jaws in a predetermined relationship. These devices function as a mandibular re-positioner which acts to increase the patient's pharyngeal space. improving their ability to exchange air during sleep. The devices will be custom made for each patient and have the ability of the adjustment mechanism enabling the amount of mandibular advancement to be set by the dentist or physician at the time of fitting the device. The devices will be sold by prescription only.
The LISA Appliance is a one piece, non-adjustable design. The LISA II Appliance is a dynamic, adjustable design. The Dorsal Appliance is a two-piece design with separate upper and lower acrylic portions that when engaged posture the mandible into protrusive position via acrylic fins built in the lower acrylic portion. The adjustable Herbst Appliance is a one piece construction held together by two adjustment mechanisms on the buccal or outer are of the upper and lower appliance.
Any slight differences in the materials, fit and function of DynaFlex® Anti-Snoring & Sleep Apnea Devices with those of the predicate devices do not raise any new concerns over safety and effectiveness.
# Intended Use
DynaFlex® Anti-Snoring & Sleep Apnea Devices are intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults. The devices are worn while sleeping to support the lower jaw in a forward position prescribed by the dentist, and is removable by the patient.
### Predicate Devices
DynaFlex has selected the following devices as it predicates for its 510(k) submission.
Device Sponsors:
- 1. Somnomed Limited
- 2. Specialty Appliances Works, Inc.
{3}------------------------------------------------
Image /page/2/Picture/0 description: This image shows a world map. The continents of North America, South America, Europe, Africa, Asia, and Australia are visible. The map is a simple black and white representation of the world.
Image /page/2/Picture/1 description: The image shows the text "EMERGO GROUP" with a logo in between the two words. The logo is a stylized letter "E" inside of a circle. To the left of the text is the number "K103076" and to the right is the number "3063".
| Devices: | 3. Airway Management, Inc.<br>1. Somnomed MAS RXA<br>2. Acrylic Splint Herbst Appliance<br>3. TAP T |
|-----------------------------------|-----------------------------------------------------------------------------------------------------|
| 510(k)'s: | 1. K050592<br>2. K083209<br>3. K061732 |
| CDRH Product Classification Name: | Device, Anti-Snoring |
| CDRH Product Code: | LRK |
| CDRH Regulation Name: | Intraoral devices for snoring and intraoral<br>devices for snoring and obstructive sleep<br>apnea |
| CDRH Regulation Number: | 872.5570 |
| CDRH Classification Panel: | Dental Devices |
| CDRH Regulatory Class: | Class II |
### Safety and Effectiveness
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the differences between the DynaFlex® Anti-Snoring & Sleep Apnea Devices and the predicate devices do not raise any questions regarding their safety and effectiveness. DynaFlex® Anti-Snoring & Sleep Apnea Devices, as designed and manufactured, are therefore determined to be substantially equivalent to the referenced predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged vertically along the left side, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" stacked on top of each other. To the right of the text is an abstract symbol consisting of three curved lines that resemble a stylized human figure or a bird in flight.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002
Dynaflex C/O Mr. Stuart R. Goldman Emergo Group Incorporated 611 West 5th Street Third Floor Austin, Texas 78701
# FEB 2 3 2011
Re: K103076
Trade/Device Name: DynaFlex® Anti-Snoring & Sleep Apnea Devices (Lisa Appliance, Lisa II Appliance, Dorsal Appliance, Herbst Adjustable Appliance) Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring And Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: February 15, 2011 Received: February 16, 2011
Dear Mr. Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2- Mr. Goldman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Dunne
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/5/Picture/0 description: This image shows a world map. The map is a simple outline of the continents and islands. The map appears to be a black and white image with a grainy texture.
# EMERGO GROUP
# SECTION 4 - INDICATIONS FOR USE
510(k) Number (if known): م) 7 0 3 0 ) جا
Device Name:
DynaFlex® Anti-Snoring & Sleep Apnea Devices (Lisa Appliance, Lisa II Appliance, Dorsal Appliance, Herbst Adjustable Appliance)
#### Indications for Use:
DynaFlex® Anti-Snoring & Sleep Apnea Devices are intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults. The devices are worn while sleeping to support the lower jaw in a forward ( on ) in advise. The devices dentist, and is removable by the patient.
Prescription Use _ X
AND/OR ·
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital
infection Control, Dental Devices
1 510(k) Number: K103076
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.