RADx Intraoral Appliance for Snoring and Sleep Apnea
Applicant
Achaemenid, LLC
Product Code
LQZ · Dental
Decision Date
Jul 12, 2023
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended to reduce or alleviate snoring, mild to moderate Obstructive Sleep Apnea (OSA) while sleeping in adults.
Device Story
Custom-fabricated intraoral mandibular repositioning device; fits over upper and lower teeth; maintains lower jaw in forward position during sleep. Mechanical protrusion increases pharyngeal space; improves air exchange. Prescription device; fitted by dentist/physician. Features adjustable rods/arms for mandibular advancement; increments of 0.5 mm. Two variants: RADx White Night (removable progressive arms) and RADx White Night-p (fixed upper tray-arm combination). Used in home setting by patient during sleep. Benefits include reduction of snoring and OSA symptoms.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-3, -5, -10, -11 and USP <151>/USP 88 standards, confirming non-cytotoxic, non-sensitizing, non-irritant, non-toxic, non-mutagenic, and pyrogen-free status.
Technological Characteristics
Custom-fabricated intraoral appliance; materials: BioMed Clear and dental LT Clear V2 resins. Mechanical mandibular advancement via adjustable rods/arms (0.5 mm increments). Two configurations: removable arms or fixed upper tray-arm. Biocompatibility per ISO 10993 and USP standards.
Indications for Use
Indicated for adults with snoring and mild to moderate Obstructive Sleep Apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 12, 2023
Achaemenid LLC % Joseph Azary Regulatory Consultant Aztech Regulatory & Quality LLC 543 Long Hill Avenue Shelton, Connecticut 06484
Re: K230532
Trade/Device Name: RADx Intraoral Appliance for Snoring and Sleep Apnea Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LQZ, LRK Dated: June 15, 2023 Received: June 22, 2023
Dear Joseph Azary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha
# -5
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230532
### Device Name
RADx Intraoral Appliance for Snoring and Sleep Apnea
Indications for Use (Describe)
Intended to reduce or alleviate snoring, mild to moderate Obstructive Sleep Apnea (OSA) while sleeping in adults.
Type of Use (Select one or both, as applicable) ري Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
Page 4-1
{3}------------------------------------------------
K230532
Image /page/3/Picture/1 description: The image shows a black and white emblem. The emblem features a bird with outstretched wings, a sun above its head, and two swords crossed below. The bird has a geometric pattern on its chest, and the swords have snakes wrapped around them.
# 510(k) Summary ACHAEMENID RADx INTRAORAL APPLIANCE FOR SNORING & SLEEP APNEA
# 1. SUBMITTER/510(K) HOLDER
Achaemenid LLC 2318 Main Street Stratford, CT 06615
Contact Name: Joseph Azary (Aztech Regulatory & Quality LLC) Email: jazary@erols.com Telephone: (203) 242-6670
Date Prepared: February 23, 2023
#### 2. DEVICE NAME
| Proprietary Name: | RADx Intraoral Appliance for Snoring and Sleep Apnea |
|------------------------------|--------------------------------------------------------------|
| Common/Usual Name: | Intraoral Devices for Snoring and Obstructive Sleep<br>Apnea |
| Classification Name: | Intraoral Devices for Snoring and Obstructive Sleep<br>Apnea |
| Classification Regulation: | 21 CFR 872.5570 |
| Classification Product code: | LQZ |
| Subsequent Product Code: | LRK |
| Classification: | Class 2 |
| Medical Specialty (Panel): | Dental |
#### 3. PREDICATE DEVICES
- Primary Predicate Device - Slow Wave DS8, 510(k) K191320
- Reference Predicate Device Panthera Anti-Snoring Device, 510(k) K143244 ●
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a black and white emblem featuring an eagle-like bird with outstretched wings. Above the bird's head is a radiant sun-like symbol. Below the bird, there are two caduceus symbols, each with a staff entwined by two snakes and topped with wings. The bird has a patterned chest, and the overall design appears to be a stylized or heraldic representation.
#### 4. DEVICE DESCRIPTION
The RADx White Night and RADx White Night-p Intraoral Appliance for Snoring and Mild to Moderate Sleep Apnea are removable intraoral devices used for treating snoring and mild to moderate obstructive sleep apnea. It consists of two custom fabricated adjustable rods that fit separately over the upper and lower teeth. The device functions as a mandibular repositioner, maintaining the lower jaw in a forward position during sleep. This mechanical protrusion acts to increase the patient's pharyngeal space, improving their ability to exchange air during sleep. The device is a prescription device customized for each patient and has an adjustment mechanism enabling the amount of mandibular advancement to be set by the dentist or physician at the time of fitting the device. The device can be adjusted only by the user and / or the dentist in increments of 0.5 mm.
The custom oral appliance is checked for proper and comfortable fit as well as the appropriate initial progressive arm attachment. The dentist will provide the patient with subsequent progressive arms from 1.0 mm up to 5.0mm in 0.5mm increments.
There are 9 different arms each identified with a number for each side of the jaw. The arms are also identified as either "L" for left or "R" for right.
# Variants / Configurations
There are two variants of the device.
The RADx White Night has removable progressive arms with the female housing active as the receptacle for the arms.
The RADx White Night-p was designed with the upper housing and arm as one piece. In this version, the fixed upper tray-arm combination are the progressive parts of the entire apparatus.
#### 5. INTENDED USE
Intended to reduce or alleviate snoring, mild to moderate Obstructive Sleep Apnea (OSA) while sleeping in adults.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a black and white emblem with an eagle-like bird in the center. Above the bird's head is a radiant sun symbol. The bird is flanked by stylized wings, and below it are two crossed staffs entwined with snakes, resembling caducei. The overall design has a symmetrical and heraldic appearance.
#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.
The subject device has the same intended use, materials and operating principles as the predicate devices. Additionally, the differences in technological characteristics do not raise new questions of safety or effectiveness. Therefore, based on the substantial equivalence evaluation, the company concludes that the subject device is equivalent to the predicate devices.
#### 7. PERFORMANCE TESTING
The following standards were either used as a reference and/or for testing of the subject device:
- . Cytotoxicity Testing per ISO 10993-5 Biological Evaluation of Medical Devices - Part 5: Tests for in vitro Cytotoxicity.
- Intracutaneous Irritation Testing per ISO 10993-10 Biological Evaluation of . Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
- Sensitization Testing per ISO 10993-10 Biological Evaluation of Medical ● Devices - Part 10: Tests for Irritation and Skin Sensitization.
- Systemic Toxicity Testing per ISO 10993-11 Biological Evaluation of Medical . Devices - Part 11: Tests for Systemic Toxicity.
- . Pyrogen Testing per United States Pharmacopeia 43, Formulary 38, General Chapter <151>, Pyrogen Test.
- Genotoxicity Testing per ISO 10993-3 Biological Evaluation of Medical Devices . - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity.
- Acute Systemic Toxicity per United States Pharmacopeia 35, Biological . Reactivity Tests, In Vivo
- . Implantation per United States Pharmacopeia 35, Biological Reactivity Tests, In Vivo
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a black and white emblem. The emblem features a bird with outstretched wings, a sun above its head, and two staffs with snakes intertwined at the bottom. The bird has a triangular pattern on its chest. The sun has rays emanating from it.
# Summary of Testing
| Standard | Description of Test | Results |
|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| ISO 10993-5 (2009)<br>Cytotoxicity<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | Biological Evaluation of Medical<br>Devices - Part 5: Tests for in vitro<br>cytotoxicity<br>ISO Elution Method | Non-Cytotoxic |
| ISO 10993-5 (2009)<br>Cytotoxicity<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | Biological Evaluation of Medical<br>Devices - Part 5: Tests for in vitro<br>cytotoxicity<br>MTT Cytotoxicity Test | No cytotoxic potential |
| ISO 10993-10 (2010)<br>Sensitization<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | Biological Evaluation of Medical<br>Devices - Part 10: Tests for irritation<br>and skin sensitization<br>ISO Guinea Pig Maximization<br>Sensitization Study | Non-sensitizer |
| ISO 10993-10 (2010)<br>Irritation<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | Biological Evaluation of Medical<br>Devices - Part 10: Tests for irritation<br>and skin sensitization<br>USP Intracutaneous Study in Rabbits | Non-Irritant |
| ISO 10993-10 (2010)<br>Irritation<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | Biological Evaluation of Medical<br>Devices - Part 10: Tests for irritation<br>and skin sensitization<br>ISO Oral Mucosal Irritation Study in<br>Hamsters 14 day | Non-Irritant |
| USP 88<br>Acute Systemic Toxicity<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | United States Pharmacopeia 35,<br>Biological Reactivity Tests, In Vivo<br>Acute Systemic Toxicity | No evidence of systemic toxicity |
| USP 88<br>Implantation<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | United States Pharmacopeia 35,<br>Biological Reactivity Tests, In Vivo<br>Modified USP Muscle Implantation<br>Study in Rabbits - 7 day | Macroscopic reaction was not significant as<br>compared to the negative control article. |
| ISO 10993-11 (2017)<br>Systemic Toxicity<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | Biological Evaluation of Medical<br>Devices - Part 11: Test for systemic<br>toxicity | Non-Toxic |
| USP 43, National<br>Formulary 38, General<br>Chapter <151> (2020)<br>Pyrogen Test | Pyrogen Test | Absence of Pyrogens |
| BioMed Clear and<br>dental LT Clear V2<br>Resins | | |
| ISO 10993-3 (2003)<br>Genotoxicity<br>BioMed Clear and<br>dental LT Clear V2<br>Resins | Biological Evaluation of Medical<br>Devices - Part 3: Tests for<br>genotoxicity, carcinogenicity and<br>reproductive toxicity | Non-Mutagenic |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a black and white emblem or logo. At the center is a stylized eagle with outstretched wings. Above the eagle is a radiant sun or halo-like symbol. The eagle's body has a geometric pattern. Below the eagle are two crossed staffs, each entwined with snakes, resembling caduceus symbols. The overall design appears to be symbolic and possibly related to medicine or healing.
# 8. SAFETY AND EFFICACY
Biocompatibility: A biocompatibility assessment of the device was performed. The purpose of the biocompatibility assessment was to ensure that biocompatibility had been established for the device. The device is biocompatible, based on the identical use of materials of construction of the predicate devices.
Clinical Testing: Human clinical studies were not deemed necessary to evaluate the safety or effectiveness of the subject device. The subject device uses similar designs as the predicate devices, uses identical materials, had same indications for use, and equivalent technological characteristics.
Risks: As with the predicate devices, the subject device can have risks for dental or soft tissue soreness, TMJ dysfunction, obstruction to oral breathing, dental movement or changes in dental occlusion. The Instructions for Use include warnings, precautions and contraindications to address such risks.
The device is manufactured using the same equipment and same materials as the predicate devices.
## 9. CONCLUSION
Information presented supports substantial equivalence of the RADx Intraoral Appliance for Snoring and Sleep Apnea to the predicate devices based on similarities in intended use, design, principles of operation, material composition and performance specifications.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a black and white emblem featuring an eagle-like bird with outstretched wings. Above the bird's head is a radiant sun-like symbol. Below the bird, there are two caduceus symbols, each with a staff entwined by two snakes and topped with wings. The bird has a patterned chest, and the overall design appears to be a stylized or heraldic representation.
Achaemenid LLC believes that based on the indications for use, technological characteristics, and comparison to predicate devices the RADx Intraoral Appliance for Snoring and Sleep has been shown to be substantially equivalent to the predicate and is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.