The intended use of composite brackets and buttons is to be bonded to the teeth to provide orthodontic movement of natural teeth. The main indication for use of orthodontic brackets and buttons is a malocclusion, such as misaligned teeth, overcrowding, spacing issues, and bite irregularities.
Device Story
Composite orthodontic brackets and buttons are aesthetic, plastic-based devices used by orthodontists/dentists to correct dental malocclusions. The device is bonded to the tooth surface using dental adhesive; orthodontic arch wires are then inserted into the bracket slot. The slot guides the wires, which—in conjunction with elastic bands and other accessories—exert gentle, controlled forces on the teeth to move them into the desired alignment. Treatment duration typically ranges from 6 months to 2.5 years. The device serves as an integral component of a broader orthodontic system; it does not include the adhesives, wires, or bands themselves. By providing a mechanical interface for force application, the device enables the gradual correction of tooth position, benefiting patients by addressing aesthetic and functional dental irregularities.
Clinical Evidence
No clinical data or animal studies were required. Substantial equivalence is supported by bench testing, including dimensional, hardness, shear, friction force, gate pull, torque, wear abrasion, water absorption, and discoloration tests. Testing demonstrated the device functions as intended and meets performance specifications.
Technological Characteristics
Materials: Polyurethane (biocompatible plastic). Design: Low-profile bracket body, mechanical retention base, slot for arch wires; some models include hooks. Dimensions/Specs: Torque, angulation, and rotation values per defined prescriptions. Energy Source: None (passive mechanical device). Connectivity: None. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for patients requiring orthodontic treatment for malocclusions, including misaligned teeth, overcrowding, spacing issues, and bite irregularities.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 21, 2025
EC Certification Service GmbH
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Drive,
Suite 510k
Saint Paul, Minnesota 55114
Re: K251196
Trade/Device Name: Composite Orthodontic Brackets and Buttons
Regulation Number: 21 CFR 872.5470
Regulation Name: Orthodontic plastic bracket
Regulatory Class: Class II
Product Code: DYW
Dated: April 17, 2025
Received: April 17, 2025
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K251196 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}
K251196 - Prithul Bom
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K251196
Device Name
ECC Premium Line
Indications for Use (Describe)
The intended use of composite brackets and buttons is to be bonded to the teeth to provide orthodontic movement of natural teeth.
The main indication for use of orthodontic brackets and buttons is a malocclusion, such as misaligned teeth, overcrowding, spacing issues, and bite irregularities.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”
{5}
| 510(k) #: K251196 | 510(k) Summary | Prepared on: 2025-04-11 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | EC Certification Service GmbH | |
| Applicant Address | Sandgasse 39a Sankt Veit an der Glan AT 9300 Austria | |
| Applicant Contact Telephone | +4342126094 | |
| Applicant Contact | Mrs. Elisabeth Moschik | |
| Applicant Contact Email | office@ec-c.at | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | ECC Premium Line | |
| Common Name | Composite Orthodontic Brackets and Buttons | |
| Classification Name | Bracket, Plastic, Orthodontic | |
| Regulation Number | 872.5470 | |
| Product Code(s) | DYW | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K973776 | Reflections | DYW |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| Composite orthodontic brackets and buttons present a clear and aesthetically pleasing choice for orthodontic patients undergoing orthodontic treatment. The Composite Bracket and Button is engineered to offer simple and aesthetic devices that address both minor and complex misalignments in patients.
Composite orthodontic brackets and buttons are classified as Class II medical devices. In accordance with ISO 10993-1:2018, based on the nature of body contact, orthodontic brackets and buttons are categorized as having mucosal contact, as they are affixed to the teeth and maintain prolonged contact with the oral mucosa. Furthermore, the contact duration is categorized as "permanent exposure" (=more than thirty days), as brackets and buttons are expected to remain in place for extended periods during orthodontic treatment.
Utilized by orthodontic professionals, these brackets and buttons serve as integral components - together with adhesives, wires, bands, elastics and other attachments in the alignment and correction of dental malocclusions and misalignments. The duration of the treatment depends on the degree of misalignment and is between 6 months and a maximum of 2,5 years.
Brackets and buttons are strategically placed and affixed to the surface of the teeth using dental adhesive. This bonding process ensures that the brackets and buttons remain securely attached to the teeth throughout the orthodontic treatment. Then Orthodontic arch wires are pushed into the slot of the bracket. The slot acts as an aid to hold and guide the orthodontic wires that - in conjunction with elastic bands and other orthodontic accessories - exert gentle forces to the teeth, gradually, moving them into their desired position. These direct and indirect orthodontic bonding techniques with orthodontic adhesives is thoroughly documented in orthodontic literature. These additional products required for orthodontic treatment are chosen by the orthodontist as required and are hence not part of this 510(k) application. The removal of the brackets/buttons follow standard procedures that do not require additional training.
Composite Orthodontic Products are only to be used by Orthodontists/Dentists. As educated specialists they know exactly how to use | | |
{6}
brackets and buttons. They have received training in order to analyse and decide with which other products they are best compatible (wires, adhesives, elastics and instruments). This 510(k) application is only for composite orthodontic brackets and buttons.
## Intended Use/Indications for Use
**21 CFR 807.92(a)(5)**
The intended use of composite brackets and buttons is to be bonded to the teeth to provide orthodontic movement of natural teeth.
The main indication for use of orthodontic brackets and buttons is a malocclusion, such as misaligned teeth, overcrowding, spacing issues, and bite irregularities.
## Indications for Use Comparison
**21 CFR 807.92(a)(5)**
The main predicate device "K973776" states following as intended use/Indications for Use:
Orthodontic brackets are usually cemented to the front surface of the tooth where they direct the mechanical forces that urge teeth into correct alignment. A curved arch wire that is bent or twisted to a particular prescription before installation is forced into a slot in each bracket. The resulting torque or restoring force causes the teeth to shift into the desired alignment. Depending on the shape and twist of the arch wire and the orientation of the bracket slot, it is possible to apply forces, which can shift, rotate, or tip the teeth in any desired direction.
Stainless steel, ceramic and plastic brackets are in common use. Plastic brackets are more aesthetically pleasing than the stainless steel which give a metallic smile. They are less expensive than ceramic.
The Intended Use is substantially Equivalent. The explanation provided for the predicate in its 510k application indicates through its explanation that the intention of the medical device is to create tooth movement ("...force causes the teeth to shift into the desired alignment"). Hence ECC and CDB share this intended use.
## Technological Comparison
**21 CFR 807.92(a)(6)**
The FDA Guidance "510(k) Programm: Evaluating Substantial Equivalence in Premarket Notifications (510(k))" was consulted on the topic of Substantial Equivalence.
Brackets and Buttons from EC Certification Service fulfill the same technological characteristics as the predicate device from CDB Corp (K973776) regarding:
Design: All brackets do have a bracket body with low profile, possibility for hooks on cuspids and bicuspids, a mechanical retention base eliminating the need for plastic condition pretreatment, a slot to place wires, Torque / Angulation and Rotation values according defined prescriptions. Products do not show significant differences that would suggest one bracket type outperforms another in terms of technical specifications.
Orthodontic buttons are single wing brackets without torque, angulation, rotation and hook. Some of them have a slot to place wires.
Material and Chemical composition: The subject devices as well as the predicated ones are made of Polyurethane – which is a Composite / Plastic material and ensures an esthetic look.
The Categorization of Medical Devices according to ISO 10993-1:2018 is the same.
The patient and tissue contact materials is in both cases biocompatible plastic (polyurethane).
Clinical characteristics: All Orthodontic Brackets and buttons show same clinical conditions, intended use, principal of operations, are used in the same site in the body, are in contact with the mucosa for the same duration and effect the same population.
The subject device, compared to the predicate device do not raise any issues of safety and effectiveness.
The comparison concludes that the ECC Premium Line is same or substantially equivalent in all areas of question.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
**21 CFR 807.92(b)**
A series of bench tests were completed to show that the medical devices function as intended and meet the specified performance characteristics, as well as to ensure that the medical device does not pose a risk to the user or patient. The following tests were completed: Dimensional test, Hardness test, Shear Test, Friction Force Test, Gate Pull Test, Torque Test, Wear Abrasion Test, Water Absorption Test and Discoloration Test.
{7}
The dimensional test was done according to ISO 27020:2019 (FR 4-313).
Also the FDA Guidance "Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions" was taken into consideration.
No animal studies or clinical testing were required for these devices.
The results of the performed tests demonstrate that the medical device functions as intended and meets the specified performance characteristics. Furthermore, they ensure that the medical device withstands the stresses encountered during use as well as that they do not pose a risk to the user or the patient. Thus, the device is equally safe, effective, and performs as well as its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.