K121574 · Synthes, Inc. · JEY · Jun 29, 2012 · Dental
Device Facts
Record ID
K121574
Device Name
SYNTHES MATRIXMADIBLE PLATE AND SCREW SYSTEM
Applicant
Synthes, Inc.
Product Code
JEY · Dental
Decision Date
Jun 29, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Synthes MatrixMANDIBLE plate and screw system is intended for oral, maxillofacial surgery: - Trauma - Reconstructive surgery - Orthognathic surgery (surgical correction of dentofacial deformities)
Device Story
MatrixMANDIBLE Plate and Screw System consists of titanium or titanium alloy (Ti-6Al-7Nb) plates and screws for internal bone fixation. Screws are 2.0 mm diameter, 4-8 mm length, 0.5 mm thread pitch. Used by surgeons in oral and maxillofacial procedures to stabilize bone segments in trauma, reconstruction, or orthognathic cases. Devices are non-sterile, single-use, and require sterilization before use. Mechanical performance verified via bench testing (strip-out resistance, pull-out strength, yield/insertion torque).
Clinical Evidence
No clinical testing was performed. Substantial equivalence supported by non-clinical mechanical bench testing comparing strip-out resistance, pull-out strength, yield torque, insertion torque, and insertion factor of safety against predicate devices.
Technological Characteristics
Materials: Titanium or titanium alloy (Ti-6Al-7Nb). Form factor: Plates of various shapes/sizes; 2.0 mm diameter screws. Principle: Internal bone fixation via cortex screws. Sterilization: Non-sterile, requires user sterilization. Single-use.
Indications for Use
Indicated for patients undergoing oral, maxillofacial surgery including trauma, reconstructive surgery, and orthognathic surgery for correction of dentofacial deformities.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
K032442 — KLS-MARTIN MANDIBULAR/RECONSTRUCTION SYSTEM II · KLS-Martin L.P. · Oct 29, 2003
K082335 — SYNTHES MATRIXMANDIBLE PLATE AND SCREW SYSTEM · Synthes (Usa) · Nov 10, 2008
K113567 — SYNTHES MATRIXMADIBLE PLATE AND SCREW SYSTEM · Synthes, Inc. · Nov 19, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
JUN 2 g 2012
| 5 | 510(k) Summary | Page 1 of 2 |
|---|----------------|-------------|
|---|----------------|-------------|
| Date Prepared | April 23, 2012 |
|---------------------|------------------------------------------------------------------------------------------------|
| Submitter | Synthes (USA) |
| | 1301 Goshen Parkway |
| | West Chester, PA 19380 |
| | United States of America |
| Contact | Alan T. Haley |
| | haley.alan@synthes.com |
| | (484) 356-9763 |
| Trade Name | MatrixMANDIBLE Plate and Screw System |
| Common Name | Bone Plate |
| Classification Name | Bone Plate, 21 CFR 872.4760, Product Code JEY |
| Predicate Devices | MatrixMANDIBLE Plate and Screw System (K063790) |
| | Synthes Mandibular Modular Fixation System (K954385) |
| Intended Use | The Synthes MatrixMANDIBLE plate and screw system is intended for oral, maxillofacial surgery: |
- Trauma
- Reconstructive surgery
- Orthognathic surgery (surgical correction of dentofacial deformities)
### Device Description
The Synthes MatrixMANDIBLE Plate and Screw System consists of a variety of plates offered in multiple shapes and sizes and a variety of screws offered in multiple diameters and lengths to meet the anatomical needs of the patient. System implants are manufactured in either titanium or titanium alloy and are intended for single use only.
The MatrixMANDIBLE screws that are the subject of this premarket notification are made from titanium alloy (Ti-6Al-7Nb) and are available in a diameter of 2.0 mm and lengths ranging from 4 mm to 8 mm, and have a thread pitch of 0.5 mm. These screws work with all plates within the MatrixMANDIBLE Plate and Screw System.
These devices are offered non-sterile and must be sterilized prior to use. MatrixMANDIBLE screws are intended for single use.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a logo that appears to be a circle with a design inside. To the right of the word is a small circle with an R inside, indicating a registered trademark. A horizontal line is present under the logo and the word.
## 510(k) Summary
Page 2 of 2
### Technological Characteristics
The proposed MatrixMANDIBLE devices are similar to the predicate devices in terms of indications, dimensions, principles of operation, and design (i.e. cortex screws for internal fixation of bone). The non-clinical testing data discussed below show that the subject devices have equivalent or better mechanical performance when compared to the predicate devices and that the minor differences in device geometry do not raise new issues of safety and effectiveness.
### Clinical Testing
No clinical testing was performed to support this submission.
#### Non-Clinical Testing
Mechanical testing was performed to compare the proposed devices to the predicates to measure:
- . Strip-out resistance (N m)
- . Pull-out strength (N·m)
- . Yield Torque (N·m)
- . Insertion Torque (N·m)
- Insertion Factor of Safety .
The non-clinical test results demonstrate that the mechanical performance of the proposed Synthes MatrixMANDIBLE screws is equivalent to or better than the predicate devices and support the substantial equivalence to the predicate devices.
#### Substantial Equivalence to Predicate Devices
In conclusion, the proposed Synthes MatrixMANDIBLE Plate and Screw System devices have the same intended use as, and technological characteristics similar to, the legally marketed predicate devices. Non-clinical testing data demonstrate that differences in the technological characteristics do not affect safety or effectiveness. The information presented supports substantial equivalence of the proposed devices to the predicate devices.
#### (end of summary)
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Alan T. Haley Regulatory Affairs Specialist Synthes. Incorporated 1301 Goshen Parkway West Chester, Pennsylvania 19380
JUN 2 9 2012
Re: K121574
Trade/Device Name: MatrixMANDIBLE Plate and Screw System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: May 25, 2012 Received: May 30, 2012
Dear Mr. Haley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2- Mr. Haley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
hn fac
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. A horizontal line is present below the word "SYNTHES". The "®" symbol is located to the right of the word.
#### Indications for Use Statement 4
510(k) Number (if known):
Device Name:
Indications for Use:
MatrixMANDIBLE Plate and Screw System
The Synthes MatrixMANDIBLE plate and screw system is intended for oral, maxillofacial surgery:
- . Trauma
- Reconstructive surgery .
- Orthognathic surgery (surgical correction of dentofacial ● deformities)
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices
510(k) Number: K121574
14
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.