K982793 · Advanced Medical Devices, Inc. · ELC · Oct 28, 1998 · Dental
Device Facts
Record ID
K982793
Device Name
PIEZO TRONIC
Applicant
Advanced Medical Devices, Inc.
Product Code
ELC · Dental
Decision Date
Oct 28, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Piezo Tronic is a device for delivering ultrasonic movement and water to a stainless steel tip that is used by a dentist or dental hygienist. The intended uses are: Periodontal pocket lavage with simultaneous ultrasonic tip movement Scaling and root planing Removal of supra and subginqival calculus and stains from teeth
Device Story
Piezo Tronic is an ultrasonic scaler used by dentists or dental hygienists in clinical settings. Device consists of a main chassis with an ultrasonic generator, water control valve, and power supply; connected to a handpiece with a stainless steel scaler tip and a footswitch. Operator activates device via footswitch, which simultaneously triggers ultrasonic vibration of the tip and water flow. Water flow and ultrasonic energy levels are manually adjustable via front-panel knobs. Device connects to standard office water supply. Output is mechanical ultrasonic movement of the tip combined with water irrigation, facilitating removal of calculus, plaque, and stains, and providing periodontal pocket lavage. Clinical benefit includes efficient dental cleaning and periodontal therapy.
Clinical Evidence
No clinical data provided; bench testing only. Compliance with IEC 601-1 and EMC 89/336/CEE standards for safety and electromagnetic compatibility is cited.
Technological Characteristics
Ultrasonic scaler; stainless steel tip; manual water flow and energy control knobs; single-position footswitch activation. Powered by electric ultrasonic generator. Complies with IEC 601-1 and EMC 89/336/CEE. Standalone device connected to office water supply.
Indications for Use
Indicated for use by dentists or dental hygienists during dental therapy to remove plaque, calculus, and stains from teeth, and for periodontal pocket lavage.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
K132443 — PIEZON 150 · E.M.S Electro Medical Systems S.A · Nov 22, 2013
K991874 — EMS SURFACE MOUNT SCALER · (Ems Sa) Electro Medical Systems · Jul 30, 1999
K033120 — SCALETRON PIEZO ULTRASOUND SCALER · Pac-Dent International (Suzhou), Ltd. · Jan 29, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 2 8 1998
# Summary of Safety and Effectiveness for AMD Piezo Tronic
# 1. Date Summary was Prepared
June 15, 1998
### 2. Submitter's Name and Address
Advanced Medical Devices 14 Canyon Creek Village, Suite 47 Richardson, Texas 75080
# 3. Contact Person
Mr. Greg Holland Holland & Associates 3722 Ave. Sausalito Irvine, CA 92606
Telephone: 949-262-0411 Telefax: 949-552-2821
### 4. Device Name
Trade/Proprietary Name: Piezo Tronic Common Name: Ultrasonic Scaler Classification Name: Ultrasonic Scaler
# 5. Predicate Devices
The legally marketed devices to which equivalence is being claimed are:
- . Piezon Master 400, marketed by Electro Medical Systems SA
- . Dentsply Bobcat® and Cavitron® Model 3000, both marketed by Dentsply International Inc. of York, PA 17405-0872
{1}------------------------------------------------
#### 6. Device Description
The Piezo Tronic consists of a main chassis containing a water control valve, an electric power supply, controls and displays, and ultrasonic generator. A footswitch is connected to the main chassis by a footswitch cord and a handpiece containing an ultrasonic scaler tip is connected to the main chassis by a handpiece cord.
When a dental professional has placed the tip in the proper position, the footswitch is depressed which activates the ultrasonic generator and opens the control valve to initiate water flow. The water flow and the level of ultrasonic energy delivered can be adjusted via operator accessible controls. A circular knob with an inked line on the front panel indicates the power level currently selected. Power level is adjusted by rotating the labeled knob clockwise on the front panel. Water flow is adiusted via a knob also on the front panel. When the foot switch is released, both ultrasonic energy and water flow stop immediately.
Safety of the operator and patient is ensured by compliance to IEC 601-1 and EMC 89/336/CEE.
#### 7. Intended Use
The Piezo Tronic is a device for delivering ultrasonic movement and water to a stainless steel tip that is used by a dentist or dental hygienist. The intended uses are:
- Periodontal pocket lavage with simultaneous ultrasonic tip . movement
- Scaling and root planing ●
- . Removal of supra and subginqival calculus and stains from teeth
{2}------------------------------------------------
#### 8. Device Comparison
The Piezo Tronic is based on the Piezon Master 400 that was found substantially equivalent on August 31. 1990 (Premarket Notification K896749). The ultrasonic generator and the water delivery systems have by modified in this design, but the patient contact parts, the handpiece and the tip are functionally identical to the Piezon Master 400.
There are basically two differences between the Piezo Tronic and the Piezon Master 400. First, the Piezon Master 400 is designed to allow the use of antimicrobial irrigants. This is accomplished by using an electric pump to draw water from a bottle. to which antimicrobials can be added. The Piezo Tronic is not intended for use with antimicrobial irrigants and therefore gets its' water via a connection to the normal office water supply. Secondly, The Piezon Master 400 incorporates a two position foot switch to allow the user to select either water and ultrasound or ultrasound only. The Plezo Tronic has a single position footswitch that turns the water and ultrasound on together. In these respects the Piezo Tronic is similar to another currently marketed device, the Dentsply Cavitron Bobcat® Ultrasonic Scaler which also used tap water and incorporates a single position footswitch.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
OCT 2 8 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Advanced Medical Devices C/O Mr. Greg Holland Consultant to Advanced Medical Devices 3722 Avenue Sausalito Irvine, California 92606
Re : K982793 Piezo Tronic Trade Name: Regulatory Class: II Product Code: ELC Dated: August 7, 1998 August 10, 1998 Received:
Dear Mr. Holland:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Holland
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{5}------------------------------------------------
#### Page 1_ of 1_
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
US. JU FAA 114 Jun 2002
Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications For Use:
1 0 :
The Piezo Tronic is an ultrasonic scaler intended for use during dental The Fiezo Tronio to an axrausi) therapy to remove plaque and calculus cleaning and ponodonial (gent) and ultrasonic vibrating scaler tip to the teeth.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital, Innection (
510(k) Number _
Prescription Use_ (Per 21 CFR 801.109)
"- ", "
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.