K051946 · Medartis, Inc. · MQN · Aug 22, 2005 · Dental
Device Facts
Record ID
K051946
Device Name
MODUS MODULAR DISTRACTION OSTEOGENESIS SYSTEM
Applicant
Medartis, Inc.
Product Code
MQN · Dental
Decision Date
Aug 22, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Modus Modular Distraction Osteogenesis System is a subcutaneous distractor system intended for use in fractures, osteotomies and arthrodeses including: treatment of hemifacial microsomia, mandibular hypoplasia, microglossia, defects of the alveolar ridge, atrophic bone segments in the mandible and maxilla, mandibular symphysis widening, growth disturbances of the ascending corpus/ascending ramus, LeFort I and LeFort II/III advancements.
Device Story
Subcutaneous distraction osteogenesis system; used for maxillofacial bone reconstruction, fractures, and osteotomies. System components: titanium plates, distraction cylinders (uni/bi-directional), and flexible extensions. Surgeon implants plates and cylinders to facilitate gradual bone distraction; allows vector adjustment via intra-operative bending. Used in clinical settings by surgeons. Provides mechanical force for bone lengthening and correction of skeletal deformities; benefits patients by addressing congenital or acquired maxillofacial growth disturbances and defects.
Clinical Evidence
No clinical data.
Technological Characteristics
Implant plates/ARS distractors: CP titanium Grade 4 (ASTM F 67). Distraction cylinders/extensions: implant steel (ASTM F 138 or F 139). System includes plates, distraction cylinders, and flexible extensions. Mechanical distraction principle. No software or electronic components.
Indications for Use
Indicated for patients requiring bone distraction osteogenesis or corrective osteotomies in the mandible and maxilla, including conditions like hemifacial microsomia, mandibular hypoplasia, microglossia, alveolar ridge defects, atrophic bone segments, symphysis widening, growth disturbances, and LeFort I/II/III advancements.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Submission Summary (Full Text)
{0}------------------------------------------------
510(k): Summary
Modus® Modular Distraction Osteogenesis System
# 510(k): Summary
### ADMINISTRATIVE INFORMATION
| Manufacturer Name: | Medartis, Inc.<br>127 W Street Road, Suite 203<br>Kennett Square, PA 19348<br>Telephone (610) 961-6101<br>FAX (610) 961-6108 |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Kate Gehret |
| Representative/Consultant: | Floyd G. Larson<br>PaxMed International, LLC<br>11234 El Camino Real, Suite 200<br>San Diego, CA 92130<br>Telephone (858) 792-1235<br>FAX (858) 792-1236 |
#### Classification Name: External Mandibular Fixator and/or Distractor Modus Modular Distraction Osteogenesis System Trade/Proprietary Name: Mandibular Fixator and Distractor Common Name:
## ESTABLISHMENT REGISTRATION NUMBER
Medartis, Inc. has submitted an Establishment Registration to FDA. The Establishment Registration number has not yet been assigned. The owner/operator number for Medartis AG, the parent company of Medartis, Inc., is 9033581.
#### DEVICE CLASSIFICATION
FDA has classified bone fixation plates as Class II devices (21 CFR 872.4760). The product code for external mandibular fixator and/or distractor is MQN. This device classification is reviewed by the Dental Devices Branch.
#### INTENDED USE
The Modus Modular Distraction Osteogenesis System is a subcutaneous distractor system intended for use in fractures, osteotomies and arthrodeses including: treatment of hemifacial microsomia, mandibular hypoplasia, microglossia, defects of the alveolar
{1}------------------------------------------------
K 05 1946
#### 510(k): Summary
Modus® Modular Distraction Osteogenesis System
ridge, atrophic bone segments in the mandible and maxilla, mandibular symphysis widening, growth disturbances of the ascending corpus/ascending ramus, LeFort I and LeFort II/III advancements.
#### DEVICE DESCRIPTION
#### Modus Plates
The Modus plates are used with the Modus Distraction Cylinders for various applications of the maxillofacial bone structures. The Modus plates are offered in a variety of sizes and are designed to fit the anatomy of the mandible and/or maxilla and allow adjustment of the vector by intra-operative bending if indicated.
#### Modus Distraction Cylinders
The Modus Distraction Cylinders are used with the Modus plates and screws for gradual bone distraction osteogenesis and for corrective osteotomies of the mandible and maxilla. The Modus distraction cylinders are available in uni-directional or bi-directional with multiple distraction lengths.
#### Modus Flexible Extensions
Flexible extensions are to be used with distraction cylinders if needed. The extensions allow additional access for distraction of the surgical site.
#### Material Composition
The implant plates and ARS distractors are made of CP titanium Grade 4 and comply with ASTM F 67 standards for unalloyed titanium for surgical implant applications. The MDO distraction cylinders and flexible extensions are made of implant steel that complies with ASTM F 138 or ASTM F 139.
#### EQUIVALENCE TO MARKETED PRODUCT
For the purposes of FDA's regulation of medical devices, the Modus Distraction System is substantially equivalent in indications and design principles to the predicate devices, that have been determined by FDA to be substantially equivalent to pre-amendment devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed within a circle that contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement. The text is written in all capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 2 2005
Medartis, Incorporated C/O Mr. Floyd G. Larson PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, California 92130
Re: K051946
Trade/Device Name: Modus Modular Distraction Osteogenesis System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: MQN Dated: July 15, 2005 Received: July 19, 2005
Dear Mr. Larson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Larson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Runner
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ K 051946
Device Name: Modus Modular Distraction Osteogenesis System
Indications for Use:
The Modus® Modular Distraction Osteogenesis System is a subcutaneous distractor system intended for use in fractures, osteotomies and arthrodeses including: treatment of hemifacial microsomia, mandibular hypoplasia, microglossia, defects of the alveolar ridge, atrophic bone segments in the mandible and maxilla, mandibular symphysis widening, growth disturbances of the ascending corpus/ascending ramus, and LeFort I, and LeFort II/III advancements.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
SkenRunger
Division Sian-Off) Division of Anesthesiology, General Hospital. Infection Control, Dental Devic
510(k) Number. K051946
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.