The Surgical Straight/Angle Handpiece Series is intended for use by licensed professionals for oral surgical and dental use. Typical uses include sectioning and removal of wisdom teeth, oral surgery (such as root tip resection. bone removal, bone shaping) and implant treatment (such as drilling the maxilla and mandible).
Device Story
Surgical Straight/Angle Handpiece series; manual dental/oral surgical instrument. Connects to air or electronic motors (ISO 3964 compliant); accepts friction grip burs (ISO 1797-1). Used by licensed dental professionals in clinical settings for bone cutting, tooth sectioning, and implant site preparation. Device features Bur Lock Ring Chuck Design; external water irrigation nozzle. Non-sterile; requires steam sterilization before use. Provides mechanical rotation to burs for precise bone/tooth manipulation; assists clinicians in surgical access and site preparation; improves procedural efficiency in oral surgery.
Clinical Evidence
Bench testing only. Verification and validation testing performed against internal functional specifications and international standards (ISO 14457, ISO 1797-1, ISO 3964). Sterilization validated per AAMI/ANSI/ISO 17665-1:2006. Biocompatibility evaluated per AAMI/ANSI/ISO 10993-1.
Technological Characteristics
Materials: Stainless steel (including SUS304T water lines) and aluminum. Interface: ISO 3964 motor coupling; ISO 1797-1 Type 2 friction grip burs. Max motor speed: 40,000 min-1. Gear ratios: 1:1 (direct) and 1:2 (increasing). Form factor: Straight and 20-degree angle models; 108.7mm-135mm length. Sterilization: Steam (moist heat). Non-optic.
Indications for Use
Indicated for licensed professionals performing oral surgical and dental procedures, including wisdom tooth sectioning/removal, root tip resection, bone removal/shaping, and implant drilling in the maxilla and mandible.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
Predicate Devices
MontBlanc Surgical Contra-Angle Handpiece and Straight Handpiece (K111532)
{0}------------------------------------------------
November 21, 2017
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a blue square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration".
Nakanishi Inc. c/o Belia Juarez Regulatory Project Manager Ken Block Consulting 1201 Richardson Dr. Richardson, Texas 75080
Re: K171155
Trade/Device Name: Surgical Straight/Angle Handpiece Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: KMW, EGS Dated: October 20, 2017 Received: October 23, 2017
Dear Belia Juarez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mary S. Runner -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K171155
Device Name Surgical Straight/Angle Handpiece
#### Indications for Use (Describe)
The Surgical Straight/Angle Handpiece Series is intended for use by licensed professionals for oral surgical and dental use. Typical uses include sectioning and removal of wisdom teeth, oral surgery (such as root tip resection. bone removal, bone shaping) and implant treatment (such as drilling the maxilla and mandible).
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# NSI
K171155
## ←5. 510(k) SUMMARY
| Submitter: | NAKANISHI INC.<br>700 Shimohinata<br>Kanuma-Shi, Tochigi-Ken Japan 322-8666 | | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person: | NAKANISHI INC.<br>Mr. Kimihiko Satoh<br>General Manager, Regulatory Affairs Dept<br>TEL: +81-0289-64-3380<br>FAX: +81-0289-62-6665<br>k-satoh@nsk-nakanishi.co.jp | | |
| Date Prepared: | November 20, 2017 | | |
| Trade Name: | Surgical Straight/Angle Handpiece | | |
| Common Name: | Handpiece, Rotary Bone Cutting<br>Handpiece, Contra-and Right-Angle Attachment, Dental | | |
| Classification Name: | Bone Cutting Instrument and Accessories | | |
| Regulation: | 21 CFR 872.4120<br>Class II | | |
| Product Code: | KMW and EGS | | |
| Predicate Devices: | Primary: K111532 Anthogyr SAS | MontBlanc Surgical Contra-Angle Handpiece and Straight Handpiece<br>Secondary: K970953 NAKANISHI INC.<br>E-Type Speed Reducer Contra Angle | |
| Device<br>Description: | The Surgical Straight/Angle Handpiece features an allowable maximum motor speed<br>at 40,000min-1. The handpieces are provided non-optic. The handpieces are designed<br>with an angle appearance and a straight appearance. Models are available to be<br>connected directly onto motors compliant to ISO 3964. All the devices are<br>manufactured using stainless steel with some aluminum, and use a Bur Lock Ring<br>Chuck Design. Available models include: Straight (1:1)- SGS-E, SGS-ES; Straight<br>(1:2) - SGS-E2G, SGS-E2S; Angled (1:1) - SGA-E, SGA-ES; Angled (1:2) - SGA-<br>E2G, SGA-E2S. | | |
| Indications<br>for Use: | The Surgical Straight/Angle Handpiece Series is intended for use by licensed<br>professionals for oral surgical and dental use. Typical uses include sectioning and<br>removal of wisdom teeth, cutting teeth, oral surgery (such as root tip resection, bone<br>removal, bone shaping) and implant treatment (such as drilling the maxilla and<br>mandible). | | |
| Summary of<br>Technological<br>Characteristics: | As with the predicate devices referenced above, the Surgical Straight/Angle<br>handpieces are intended for oral surgical and dental use. The handpieces are powered<br>by either an air-motor or electronic micro motors compliant to ISO 14457 with<br>couplings compliant to [ISO 3964 (EN ISO 3964)]. Friction Grip Burs that conform to<br>standard ISO 1797-1 (EN ISO 1797) are to be used with the Surgical Straight/Angle<br>Handpieces. The allowable maximum motor speed rotation for the handpieces should<br>be set at 40,000min-1. The handpieces are provided non-optic. Overall dimensions,<br>weight, angle, and gear ratio vary due to the differences in the device's compatibility. | | |
Image /page/3/Picture/6 description: In the image, a hand wearing a white glove is holding a blue plastic tool. The tool appears to be designed for a specific purpose, possibly related to crafting or assembly. The hand is positioned in a way that suggests the tool is being used or demonstrated. The text "Powerful Partners" is visible at the bottom of the image.
{4}------------------------------------------------
The handpieces overall length varies between 108.7mm-135mm depending on the model. The handpieces feature a Bur Lock Ring Chuck Design. The Surgical Straight/Angle Handpieces are provided non-sterile, so it must be sterilized prior to first use and sterilized after each subsequent use using steam sterilization (moist heat sterilization). All models feature an external water supply irrigation nozzle that is provided with the handpiece. The handpieces are composed of Stainless Steel and Aluminum and the water lines are composed of Stainless Steel (Grade SUS304T.)
The proposed device shares technological characteristics with the predicate devices. The proposed device also has some differences in technological characteristics from those of the predicate devices. Any differences in the technological characteristics are minor and reflect market strategy and/or perceived user preferences and does not impact the substantial equivalence of the device.
| | Proposed: NAKANISHI INC.<br>K171155 | Predicate: ANTHOGYR SAS<br>[K111532] |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Surgical Straight/Angle Handpiece | MontBlanc Surgical Contra-Angle<br>Handpiece and Straight Handpiece |
| Models | SGA-E, SGA-ES, SGA-E2G, SGA-E2S<br>SGS-E, SGS-ES, SGS-E2G, SGS-E2S | 12200X, 12200XL, 12200XLED, 12400X,<br>12400XLED |
| Indication for Use | The Surgical Straight/Angle Handpiece<br>Series is intended for use by licensed<br>professionals for oral surgical and dental<br>use. Typical uses include sectioning and<br>removal of wisdom teeth, cutting teeth,<br>oral surgery (such as root tip resection,<br>bone removal, bone shaping...) and<br>implant treatment (such as drilling the<br>maxilla and mandible). | Indications are very widespread in the<br>field of implantology and surgery. The<br>mentioned handpieces have been<br>developed especially for the following<br>applications:<br>-MontBlac Surgical Contra-angle<br>handpiece: for e.g. hemisection, wisdom<br>tooth extraction.<br>-MontBlac Surgical Straight handpiece:<br>for e.g. application in the area of the<br>front teeth, root tip, resection, bone<br>removal, osteotomy on the upper and<br>lower jaw, preprosthetic surgical<br>osteoplasty, sequestrum removal,<br>fenestration on the alveolar appendix,<br>opical ventilation, bone osteoplasty,<br>bone smoothing |
| Handpiece Type | Straight, 20° Angle | Straight, Contra Angle |
| Operational Mode(s)<br>Motor Connection | ISO 3964 motor connection | ISO 3964 motor connection |
| Weight | 59.5g to 103.5g | 125g straight<br>98g contra-angle |
| Coupling Type | ISO 3964 Coupling | ISO 3964 Coupling |
| Burs | ISO 1797-1<br>Type 2 | EN ISO 1797-1<br>Contra: Type 3 Straight: Type 1 or 2 |
| Max Rotation Speed<br>(motor) | $40,000min^{-1}$ | $40,000min^{-1}$ |
| Max rotation speed<br>(handpiece) | $40,000min^{-1} to 80,000min^{-1}$ | $40,000min^{-1}to 120,000min^{-1}$ |
| Gear Ratio | 1:1 Direct Drive, 1:2 Increasing | 1:1 Direct Drive, 1:3 Increasing |
Summary of Performance Testing:
The Surgical Straight/Angle Handpiece was developed and is produced under considerations of all applicable technical standards, internal specifications, and FDA guidance documents. The product's conformance with applicable international and internal standards was verified in the course of bench testing. Tests were performed on the handpiece including verification/validation testing to internal functional specifications which demonstrated that the device is substantially equivalent. Sterilization has been validated in conformance to the FDA recognized consensus standard AAMI/ANSI/ISO 17665-1:2006; "Sterilization of Health Care Products -
Image /page/4/Picture/6 description: In this image, a hand wearing a white glove is holding a blue medical device. The device appears to be some type of surgical instrument, possibly a clamp or forceps. The hand is positioned as if it is in the process of using the device. The text "Powerful Partners" is visible at the bottom of the image.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for NSK, a Japanese manufacturer of bearings. The logo is white and is set against a blue background. The letters "NSK" are in a bold, sans-serif font, with the "N" and "S" connected by a curved line.
Moist Heat - Part 1: Requirements for the Development. Validation, and Routine Control of a Sterilization Process for Medical Devices." Documentation was provided demonstrating that the Surgical Straight/Angle Handpiece complies with the FDA requirements stated in Guidance for the Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling - Guidance for Industry and Food and Drug Administration Staff. Biocompatibility evaluations were selected in accordance with AAMI/ANSI/ISO 10993-1: "Biological evaluation of medical devices - Part 1: Evaluation and testing" and FDA Guidance "Use on International Standard ISO 10993, "Biological evaluation of medical devices - Part 1: Evaluation and Testing". In addition, testing for electrical safety was performed in accordance with ISO 14457.
Conclusion: NAKANISHI INC. considers the Surgical Straight/Angle Handpiece to be substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.
Section 5: Page 3 of 3
Image /page/5/Picture/3 description: The image shows a hand wearing a white glove holding a blue surgical instrument. The instrument appears to be a type of surgical dissector or similar tool, with a pointed tip and a handle designed for precise manipulation. The text "Powerful Partners" is visible at the bottom of the image, suggesting the instrument is part of a medical product line or partnership.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.