The PINNACLE Driver is intended for driving screws and for drilling into bone, in conjunction with dental, orthognathic, and mandibular surgical procedures.
Device Story
Handheld, cordless, battery-operated driver/drill for high-speed screw insertion and pilot hole drilling. Operates in forward and reverse directions. Used in healthcare facilities by clinicians. Input: user activation via forward/reverse buttons. Transformation: software-controlled motor operation with torque-limiting functionality to prevent screw stripping. Output: mechanical rotation of hex shank drills or driver stems. Benefits: precise screw placement, reduced risk of over-torquing. Context: sterile, single-use lithium battery; reusable driver stems; single-use drills. Clinician uses output to secure hardware in bone; device performance directly impacts surgical efficiency and hardware integrity.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by bench testing, including simulated use, sterilization/reprocessing validation, shipping validation, battery leak testing, and electrical safety/EMC testing (IEC 60601-1, IEC 60086-4, ES60601-1). Biocompatibility testing was performed per FDA Blue Book Memorandum #G95-1.
Indicated for driving screws and drilling into bone during dental, orthognathic, and mandibular surgical procedures.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 4, 2017
Osteomed Kathryn Jayne Senior Specialist, Regulatory Affairs 3885 Arapaho Road Addison, Texas 75001
Re: K162544
Trade/Device Name: Osteomed Pinnacle Driver Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument And Accessories Regulatory Class: Class II Product Code: KMW Dated: February 16, 2017 Received: March 9, 2017
Dear Kathryn Jayne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Susan Runno DDS, MA
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K162544
Device Name PINNACLE Driver
Indications for Use (Describe)
The PINNACLE Driver is intended for driving screws and for drilling into bone, in conjunction with dental, orthognathic, and mandibular surgical procedures.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Osteomed. The logo is in blue and gray, with the word "OSTEOMED" in blue, bold letters. The logo is surrounded by a blue and gray swoosh design. The blue swoosh is on top, and the gray swoosh is on the bottom.
# 510(k) SUMMARY
### I. SUBMITTER
OsteoMed 3885 Arapaho Rd. Addison, TX 75001 USA
Phone: 972-677-4766 Fax: 800-390-2620 Email: kjayne@osteomed.com
Contact Person: Kathryn A. Jayne Date Prepared: March 29, 2017
### II. DEVICE
Name of the Device: PINNACLE Driver Common or Usual Name: Battery-operated drill/driver Classification Name: Drill, Bone, Powered Regulation: 872.4120 Regulatory Class: II Product Code: KMW
### III. PREDICATE DEVICE
The OsteoMed "B" Power System, K933101, is the primary predicate. The OsteoMed OsteoPower System, K971692, is the reference predicate.
This primary predicate and reference predicate have not been subject to a designrelated recall.
### IV. DEVICE DESCRIPTION
The PINNACLE Driver is a handheld, cordless, software driven, battery operated driver for high speed insertion of screws and drilling of pilot holes. The device operates in both the forward and reverse directions. A sterile, single use battery powers the device and is intended for use solely with the PINNACLE Driver. Accessories for the PINNACLE Driver include hex shank drills and driver stems. The optional driver stems are reusable while the drills are single use. The PINNACLE Driver may be sterilized individually wrapped in two layers of 1-ply polypropylene wrap or unwrapped in a sterilization organizational tray inside a rigid sterilization container, which is available from OsteoMed.
The PINNACLE Driver is intended for use in a healthcare facility/hospital for use by a clinician. The PINNACLE is a prescription device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OSTEOMED" in blue letters. The logo has a blue swoosh above the word and a gray swoosh below the word. The logo is likely for a medical company that specializes in orthopedics.
The PINNACLE Driver and lithium battery is made of medical grade stainless steel, aluminum, silicone, PEEK, gold plated brass, and medical plastic. The driver stems and drills are made of medical grade stainless steel. The sterilization organizational trays and rigid sterilization containers are made of aluminum.
The PINNACLE Driver is a battery operated driver/drill that operates in the both the forward and reverse directions. The PINNACLE Driver contains torque limiting software to prevent over-torquing and stripping of screws in drive mode.
### V. INDICATIONS FOR USE
The PINNACLE Driver is intended for driving screws and for drilling into bone, in conjunction with dental, orthognathic, and mandibular surgical procedures.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
At a high level, the subject and predicate devices are based on the following same technological elements as demonstrated below.
| | Subject Device- | Primary Predicate- | Reference Predicate- |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | OsteoMed PINNACLE | OsteoMed "B" Power | OsteoMed OsteoPower |
| | Driver | System & Accessories | System & Accessories |
| 510(k) | Pending | K933101 | K971692 |
| Regulation No. | 872.4120 | 872.4120 | 872.4120 |
| Product Code | KMW | KIJ | KMW |
| Device Class | II | Unclassified | II |
| Regulatory Panel | Dental | Dental | Dental |
| Indications for<br>Use | The PINNACLE Driver is<br>intended for driving<br>screws and for drilling<br>into bone, in conjunction<br>with dental, orthognathic,<br>and mandibular surgical<br>procedures. | The OsteoMed "B" Power<br>Handpiece System is<br>intended for drilling and<br>cutting bone and teeth and<br>for driving pins, screws,<br>and/or wires into bone for<br>dental, craniofacial,<br>orthognathic, mandibular,<br>hand, foot, wrist, or<br>extremity reconstruction<br>surgical procedures. | The OsteoPower System<br>and Accessories are<br>indicated for drilling or<br>cutting bone or teeth, and<br>for driving screws and/or<br>pins and wires into bone,<br>in conjunction with<br>dental, craniofacial,<br>craniotomies,<br>orthognathic, spinal,<br>mandibular, hand, foot,<br>wrist, and extremity<br>reconstruction surgical<br>procedures. |
| Power Source | Battery operated<br>(Lithium, single patient<br>use) | Battery operated<br>(rechargeable NiCad<br>battery) | AC/DC External power<br>supply |
| Material | --Handpiece and<br>battery-Aluminum,<br>PEEK, brass, medical<br>grade stainless steel,<br>silicone | --Handpiece and<br>battery-stainless steel,<br>aluminum, brass, silver,<br>medical grade plastics,<br>fiber reinforced ceramics | -- Handpiece only-<br>stainless steel or carbide<br>--Cutting portion of the<br>drills or burrs—stainless<br>steel, carbide, diamond, or<br>fiber reinforced ceramic |
| | Subject Device—<br>OsteoMed PINNACLE<br>Driver | Primary Predicate—<br>OsteoMed “B” Power<br>System & Accessories | Reference Predicate—<br>OsteoMed OsteoPower<br>System & Accessories |
| | | --accessories—medical<br>grade stainless steel or<br>carbide | --Accessories—stainless<br>steel, aluminum, or<br>medical grade plastics |
| Speed | 2,000 RPM forward<br>275 RPM reverse | 120-1800 RPM | 608-70,000 RPM |
| Battery | Disposable, lithium,<br>single use, sterile battery<br>(gamma) | Rechargeable Nickel<br>Cadmium battery | NA |
| Modes | Drill<br>Drive | Drill<br>Drive | Speed ranges SAFETY,<br>100%, 75%, 50%, 25%,<br>and Di (12.5%) of<br>maximum speed |
| Activation<br>Buttons | 1 Forward button<br>1 Reverse button | 1 Forward button<br>1 Reverse button | 1 Forward button<br>1 Reverse button |
| Insertion<br>Material | Bone | Bone | Bone |
| Screw<br>sizes/insertion | 1.2mm, 1.6mm, 2.0mm | 1.6mm, 2.0mm | 1.2mm, 1.6mm, 2.0mm |
| Collet | Pull-back collet | Pull-back collet | Twist collet |
| Accessories | Hex Shank drills, Hex<br>Shank driver stems,<br>sterilization<br>organizational trays, rigid<br>sterilization container | Round burrs, oval burrs,<br>twist drills, wire pass<br>twist drills, side cutting<br>burs, and end-cutting<br>drills, saw blades, K-<br>wires and pins, screw<br>drivers, burr racks, blade<br>containers, sterilization<br>trays, battery charger | Round burrs, oval burrs,<br>twist drills, wire pass<br>twist drills, side cutting<br>burrs, end-cutting drills,<br>saws, motor units,<br>modules, footswitches,<br>screw drivers |
| Sterilization of<br>re-usable devices | Pre-vacuum steam<br>(wrapped)—Hand piece<br>only<br>--Min. Temp: 132° C<br>(270°F)<br>--Full Cycle Time: 4<br>minutes<br>--Min Dry Time: 30<br>minutes<br>Pre-vacuum steam (rigid<br>container)—<br>1-Driver Case & 2-Driver<br>Case<br> | Pre-vacuum Hi Vac<br>(wrapped or unwrapped)<br>Temp: 270-270 °F<br>Exposure Time: 4 min<br>Dry Time: min 8 min<br>270 °F Gravity<br>(wrapped on tray)<br>Temp: 270-270 °F<br>Exposure Time: 35 min<br>Dry Time: min 8 min | Pre-vacuum steam<br>(wrapped)<br>Temp: 132° C (270°F)<br>Dwell Time: 10 minutes<br>Dry Time: 70 minutes |
| | Subject Device—<br>OsteoMed PINNACLE<br>Driver | Primary Predicate—<br>OsteoMed “B” Power<br>System & Accessories | Reference Predicate—<br>OsteoMed OsteoPower<br>System & Accessories |
| Parameter<br>Control<br>Technology | Software | Non-programmable,<br>discrete digital and analog<br>logic | Non-programmable,<br>discrete digital and analog<br>logic |
| Software/safety<br>features | -Senses torque to prevent<br>over-torquing & stripping<br>of screws<br>-bit recognition | NA | Motor unit control switch<br>to temporarily disable the<br>Electric Drive |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for OsteoMed. The logo is in blue and gray, with the word "OsteoMed" in a bold, sans-serif font. The logo is surrounded by a gray and blue swoosh design. The blue swoosh is on top and the gray swoosh is on the bottom.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OsteoMed. The logo is in blue and white, with the word "OsteoMed" in large, bold letters. The logo is surrounded by a gray and white swoosh design. The logo is simple and modern, and it is likely used to represent the company's brand.
The primary predicate device contains non-programmable, discrete digital and analog logic to operate. The subject device contains software to sense torque to prevent over-torquing and stripping of screws. Software validation is located in section 16 of this submission.
The primary predicate and subject devices are similar in basic shape and structure. The main difference between the two devices is the addition of software and the sterilization/reprocessing. The evaluation performed and supporting documentation within this submission demonstrates that the subject is comparable to the predicate device. Whenever possible, testing, located in section 18, was executed using recognized standards to establish product performance.
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
The biocompatibility evaluation for the device was conducted in accordance with the FDA Blue Book Memorandum #G95-1.
- Cytotoxicity
- Sensitization ●
- Irritation ●
- . Systemic toxicity
The device and associated accessories are considered tissue/bone/dentin contacting for a duration of less than 24 hours. However, OsteoMed conducted a biocompatibility evaluation of the exposed material of the driver as worst case, based on an implant device due to the close proximity to the patient. The sterilization trays and rigid containers are considered non-patient contacting.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Osteomed. The logo is in blue and gray, with the word "OSTEOMED" in blue, sans-serif font. The logo is surrounded by a gray and blue swoosh design. The swoosh design is gray on the bottom and blue on the top.
## Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the device. The system complies with the IEC 60601-1 and IEC 60086-4 standards for safety and the ES60601-1 standard for EMC.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices." The software for this device was considered as "moderate" level of concern as documented in this submission.
## Bench Testing
The following bench testing was conducted:
- Simulated Use
- Sterilization Validation
- Reprocessing Validation
- Shipping Validation ●
- o Battery Leak Testing
- Electrical Safety and Electromagnetic Compatibility ●
## Animal Study
No animals studies were needed to demonstrate substantial equivalence to the predicates.
## Clinical Studies
No clinical studies were needed to demonstrate substantial equivalence to the predicates.
### CONCLUSIONS VIII.
The basis of substantial equivalence for the device is based on similarities in indications for use, material, function, performance, design, technology, sterilizations, and operating principles. The non-clinical data demonstrates substantial equivalence to the declared predicate and the verification and validation demonstrate that the device should perform as intended in the specified use conditions.
(End of Summary)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.