← Product Code [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY) · K974555

# SYNTHES 2.0 MM LOCKING PLATE SYSTEM (2.0 LPS) (K974555)

_Synthes (Usa) · JEY · Mar 3, 1998 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K974555

## Device Facts

- **Applicant:** Synthes (Usa)
- **Product Code:** [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY.md)
- **Decision Date:** Mar 3, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Intended Use

The Synthes 2.0 LPS is intended for oral, maxillofacial surgery: trauma; reconstructive surgery; and orthognathic surgery (surgical correction of dentofacial deformities).

## Device Story

The Synthes 2.0 Locking Plate System (2.0 LPS) is a bone fixation implant system used in oral and maxillofacial surgery. It consists of various plate configurations (straight, crescent, atrophic, reconstruction, and specialty Y- or L-plates) and 1.5 mm or 2.0 mm self-drilling locking screws. The system features threaded screw holes that accept screws with mating external threads on the head; this design locks the screw to the plate, enhancing stability and reducing the risk of screw loosening. The device is intended for use by surgeons in clinical settings to stabilize bone segments during trauma repair, reconstructive procedures, or orthognathic surgery. By providing rigid fixation, the system supports bone healing and structural correction.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Plates manufactured from CP Titanium and Ti-15Mo; screws manufactured from titanium alloy (Ti-6Al-7Nb). System utilizes a threaded locking interface between screw head and plate hole to provide rigid fixation. Components include various plate geometries and 1.5 mm/2.0 mm self-drilling screws.

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- Synthes Mandibular Modular Fixation System ([K954385](/device/K954385.md))
- 2.4 mm Universal Locking Plate System ([K961421](/device/K961421.md))
- SMF Ti Alloy Bone Screws ([K963546](/device/K963546.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Synthes. The logo consists of a circular symbol on the left, followed by the word "SYNTHES" in bold, black letters. There is a trademark symbol to the right of the word "SYNTHES".

974555

## Summary of Safety and Effectiveness Information [510(k) Summary]

MAR - 3 1998 Synthes (USA) SPONSOR 1690 Russell Road Paoli, PA 19301 (610) 647-9700 Contact: Angela Silvestri Bone Plate and Bone Screw COMMON OR USUAL NAME Class II, 21 CFR 872.4760 and 872.4880 DEVICE CLASSIFICATION Synthes Mandibular Modular Fixation System (K954385), Synthes PREDICATE DEVICE 2.4 mm Universal Locking Plate System (K961421), and SMF Ti Alloy Bone Screws (K963546) The 2.0 Locking Plate System consists of five different types of DESCRIPTION plates and two screws, as follows: 2.0 Straight Locking Plates, 2.0 Crescent Locking Plates, 2.0 Atrophic Locking Plates, 2.0 的 Reconstruction Locking Plates, 2.0 Specialty Plates (Y- and L-), and 1.5 mm and 2.0 mm Self-drilling Locking Screws. The plates feature a threaded screw hole which accepts a locking screw with a mating external thread on the screw head. Engaging these screw threads fixes the screw to the plate thereby reducing the likelihood of screw loosening and improving the screw to plate stability of the implant system. The Synthes 2.0 LPS is intended for oral, maxillofacial surgery: trauma; INTENDED USE reconstructive surgery; and orthognathic surgery (surgical correction of dentofacial deformities). MATERIAL The plates are manufactured from CP Titanium and Ti-15Mo, and the screws are manufactured from titanium alloy (Ti-6Al-7Nb).

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with its wings spread and head turned to the right. The image is in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 3 1998

Ms. Angela J. Silvesti Manager, Regulatory Affairs Synthes (USA) 1101 Synthes Avenue Monument, Colorado 80132

K974555 Re : Synthes 2.0mm Locking Plate System (2.0 LPS) Trade Name: Regulatory Class: II Product Code: JEY December 3, 1997 Dated: December 4, 1997 Received:

Dear Ms. Silvesti:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with ..... the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{2}------------------------------------------------

Page 2 - Ms. Silvesti

through 542 of the Act for devices under the Electronic enrough Siz or on Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA asbeing of substantial equivalence of your device to a legally marketed predicate device results in a classification for your mavice and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Patricia Cusinatt/hor

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized figure inside. The logo and the word "SYNTHES" appear to be the brand name of a company. The overall image has a clean and professional look.

SYNTHES (USA) 1690 Russell Road Post Office Box 1766 Paoli, Pennsylvania 19301 Telephone 610-647-9700

Page 1 of 1

510(k) Number (if known):

K974555

Synthes (USA) 2.0 Locking Plate System Device Name:

Indications for use:

The Synthes 2.0 LPS is intended for oral, maxillofacial surgery: trauma; reconstructive surgery; and orthognathic surgery (surgical correction of dentofacial deformities).

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Rumer
(Division Sign-Off)
Division of Dental, Infectioratrol,
and General Hospital Devices
510(k) Number. K974335

escription Use OR Over-The-Counter Use

Prescription Use V (Per 21 CFR 801.109)

OH

Over-The-Counter Use __

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K974555](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K974555)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
