← Product Code [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY) · K193143

# Stryker Facial iD Plating System (K193143)

_Stryker · JEY · Mar 26, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K193143

## Device Facts

- **Applicant:** Stryker
- **Product Code:** [JEY](/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY.md)
- **Decision Date:** Mar 26, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Intended Use

The Stryker Facial iD Plating System is intended for osteotomy, stabilization and rigid fixation of maxillofacial fractures and reconstruction in adults and adolescents (age 12 and higher). Specific Indication for Use: - Orthognathic Surgery - Reconstructive maxillofacial surgery - Mandible and maxillofacial trauma surgery

## Device Story

Stryker Facial iD Plating System consists of additively manufactured, patient-specific bone plates used for maxillofacial reconstruction, stabilization, and fixation. System includes design proposals, electronic IFUs, and optional anatomical models. Used by surgeons in clinical settings for orthognathic and trauma procedures. Software updates introduce automated bone thickness measurements and visualization tools to assist surgeons in plate design and screw hole placement. Device output (plates) provides rigid fixation to stabilize bone segments, facilitating healing and anatomical reconstruction. Compatibility with VSP System (K192192) allows integration with patient-specific surgical guides and plans.

## Clinical Evidence

No clinical testing performed. Substantial equivalence supported by bench testing, software verification/validation per IEC 62304, and cadaver lab validation confirming performance under specified use conditions.

## Technological Characteristics

Additively manufactured patient-specific bone plates. System includes software for automated bone thickness measurement and visualization. Software developed per IEC 62304. Compatible with VSP System (K192192).

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Predicate Devices

- Stryker Facial iD Plating System ([K182305](/device/K182305.md))

## Reference Devices

- VSP® System ([K192192](/device/K192192.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 26, 2020

Stryker Zainab Amini Senior Regulatory Affairs Specialist 750 Trade Centre Way- Suite 200 Portage, Michigan 49002

Re: K193143

Trade/Device Name: Stryker Facial iD Plating System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY Dated: February 27, 2020 Received: February 28, 2020

Dear Zainab Amini:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K193143

Device Name Stryker Facial iD Plating System

Indications for Use (Describe)

The Stryker Facial iD Plating System is intended for osteotomy, stabilization of maxillofacial fractures and reconstruction in adults and adolescents (age 12 and higher).

Specific Indications for Use: -Orthognathic surgery -Reconstructive maxillofacial surgery

-Mandible and maxillofacial trauma surgery.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### INTRODUCTION I.

This document provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.

#### SUBMITTER II.

| 510(k) Owner:                 | Stryker Leibinger GmbH & Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany                                                                                        |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/                    | Zainab Amini                                                                                                                                                                 |
| Submitter/<br>Contact Person: | Zainab Amini<br>Senior Regulatory Affairs Specialist<br>Stryker Craniomaxillofacial<br>750 Trade Centre Way<br>Portage, MI 49002<br>Phone: 269-389-8349<br>Fax: 877-648-7114 |

- Date prepared: March 26, 2020
#### DEVICE III.

- Trade Name: Stryker Facial iD Plating System
- Common or Bone Plating System Usual name:
- Classification Bone Plate; 21 CFR §872.4760 name:
- Regulatory Class: Class II

Product Code: JEY

#### IV. PREDICATE DEVICE

Primary Predicate: K182305, Stryker Facial iD Plating System Reference Device: K192192, VSP® System

{4}------------------------------------------------

# K193143

#### V. SUBJECT DEVICE DESCRIPTION

The Stryker Facial iD Plating System is intended for osteotomy, stabilization and rigid fixation of maxillofacial fractures and reconstruction in adults and adolescents (age 12 and higher), with the specific Indications for Use in orthognathic surgery, reconstructive maxillofacial surgery, and mandible and maxillofacial trauma surgery.

The Subject Device plate(s) are additively manufactured patient-specific plates, and the patientspecific design of the plates allows certain features to be configured to meet the individual needs of each patient. The Subject Device plate(s) are provided with a Design Proposal, an electronic Instruction for Use (IFU) and an optional Anatomical Model.

## INDICATIONS FOR USE

|                                       | Subject Device,                                                                                                                                                                                                                                                                                                                                                              | Primary Predicate Device,                                                                                                                                                                                                                                                                                                                                                    |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                       | K193143                                                                                                                                                                                                                                                                                                                                                                      | K182305                                                                                                                                                                                                                                                                                                                                                                      |
| Intended<br>Use/Indication<br>for Use | The Stryker Facial iD Plating<br>System is intended for osteotomy,<br>stabilization and rigid fixation of<br>maxillofacial fractures and<br>reconstruction in adults and<br>adolescents (age 12 and higher).<br><br>Specific Indication for Use:<br>- Orthognathic Surgery<br>- Reconstructive<br>maxillofacial surgery<br>- Mandible and<br>maxillofacial trauma<br>surgery | The Stryker Facial iD Plating<br>System is intended for osteotomy,<br>stabilization and rigid fixation of<br>maxillofacial fractures and<br>reconstruction in adults and<br>adolescents (age 12 and higher).<br><br>Specific indication for Use:<br>- Orthognathic surgery<br>- Reconstructive<br>maxillofacial surgery<br>- Mandible and<br>maxillofacial trauma<br>surgery |

| Table 5- 1: Comparison Of Intended Use/Indications For Use. |  |
|-------------------------------------------------------------|--|
|-------------------------------------------------------------|--|

The Intended Uses and the specific Indication for Use of the Subject Device and Primary Predicate Device are identical.

Reference Device: K192192, VSP® System - Patient specific maxillofacial anatomical models, templates, guides, and surgical plans. This is added for compatibility to the Subject Device. Information regarding this reference device is not necessary for the substantial equivalence comparison.

The purpose of this Special 510(k) submission is due to Stryker's software update, and additionally, to show compatibility of the Subject Device implants with the VSP System cleared in K192192.

{5}------------------------------------------------

#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PRIMARY PREDICATE DEVICE

The Subject Device is compared to the Primary Predicate Device for substantial equivalence of technological characteristics based on the following criteria:

- A. Principle of Operation
- B. Technological Characteristics

### A. Principle of Operation

The operating principle for the Subject Device is to reconstruct, stabilize and/or provide rigid fixation in the maxillofacial anatomy.

### B. Technological Characteristics

The fixation method, material, non-sterilization method, patient-specific offering and design are identical when comparing the Subject and Primary Predicate Device. Most importantly, performance test data shows the Subject Device is substantial equivalent to the Primary Predicate Device.

The bone plating system of the primary predicate device and the subject device remain identical. This 510(k) includes a software updates and shows compatibility to the VSP System (K192192). Updates include automated bone thickness measurements and visualization, which aids in visualization for plate design and screw hole placement; and plate design process improvements (i.e., naming of files, cosmetic display). Software Verification and Validation testing were performed for the software updates. As shown in the performance testing in the cadaver lab the interaction between the surgical guides and the Facial iD plates are not changing in any way compared to their previous clearance.

#### PERFORMANCE DATA 1.

The following performance testing to show substantial equivalency:

The Subject Device is identical to the Primary Predicate Device for cleaning and sterilization validation. Additionally, biocompatibility testing is not necessary for the Subject Device as there have been no changes in the Subject Device material and process, and therefore the Subject Device is identical to the Primary Predicate Device.

### Performance Bench Testing

Performance testing, cleaning and sterilization, shipping and handling done for the Primary Predicate Device in K182305 are valid for the Subject Device. The end-user test validation of the Subject Device in a cadaver lab showed that the subject device is performing as intended in the specified use conditions. Software verification and validation were performed according to internal procedures and IEC 62304. Therefore, the subject device met all predefined acceptance criteria as the primary predicate device, and the results of the tests support the substantial equivalence of the subject device to the primary predicate device.

{6}------------------------------------------------

# Animal Testing

Animal testing was not required as a basis for substantial equivalence.

### Clinical Testing

Clinical testing was not required as a basis for substantial equivalence.

#### CONCLUSIONS 2.

The results of the performance tests demonstrate that the Subject Device, Stryker Facial iD Plating System will perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence of the Subject Device to the Primary Predicate Device.

---

**Source:** [https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K193143](https://fda.innolitics.com/submissions/DE/subpart-e%E2%80%94surgical-devices/JEY/K193143)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
